K Number
K131588
Device Name
VIGILEO ARTERIAL PRESSURE CARDIAC OUTPUT/OXIMETRY MONITOR; VIGILEO APCO/OXIMETRY MONITOR
Date Cleared
2014-05-22

(356 days)

Product Code
Regulation Number
870.1435
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Vigileo™ APCO/Oximetry Monitor is indicated for continuously measuring hemodynamic parameters such as cardiac output and oximetry to assess oxygen delivery and consumption. When connected to an Edwards oximetry catheter, the monitor measures oximetry in adults and pediatrics. The monitor also displays parameters, such as stroke volume and stroke volume variation, used to assess fluid status and vascular resistance. The Vigileo™ APCO/Oximetry Monitor may be used in all settings in which critical care is provided.
Device Description
The Vigileo™ Arterial Pressure Cardiac Output (APCO)/Oximetry Monitor (Vigileo™ Monitor) is a microprocessor-based instrument. When used with an Edwards FloTrac sensor, the Vigileo™ Monitor is capable of continuously monitoring the following parameters: Cardiac Output (CO); Cardiac Index (CI): Stroke Volume (SV): Stroke Volume Variation (SVV); Systemic Vascular Resistance (SVR); Systemic Vascular Resistance Index (SVRI); Oxygen Delivery (DO2); and Oxygen Delivery Index (DO2I). When used with Edwards Oximetry catheters, the Vigileo™ Monitor is capable of continuously monitoring the following parameters: Central venous oxygen saturation (ScvO2); and, Mixed venous oxygen saturation (SvO2).
More Information

Not Found

No
The document describes a microprocessor-based monitor that calculates hemodynamic parameters based on sensor inputs, without mentioning any AI or ML algorithms for data processing or interpretation.

No
The device is described as a monitor that measures and displays hemodynamic parameters to assess oxygen delivery and consumption, fluid status, and vascular resistance. It is used for diagnostic and monitoring purposes, not for treating a condition or restoring function.

Yes
The device is described as "continuously measuring hemodynamic parameters" and "displaying parameters... used to assess fluid status and vascular resistance," which indicates its role in providing objective measurements for diagnosis and monitoring of physiological conditions.

No

The device description explicitly states it is a "microprocessor-based instrument" and is used "with an Edwards FloTrac sensor" and "with Edwards Oximetry catheters," indicating it is a hardware device that processes data from sensors and catheters.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use describes the device as continuously measuring hemodynamic parameters and oximetry to assess oxygen delivery and consumption, fluid status, and vascular resistance. These are physiological measurements taken directly from the patient's body, not tests performed on samples in vitro (outside the body).
  • Device Description: The description details a microprocessor-based instrument that connects to sensors and catheters to monitor various physiological parameters. This aligns with a patient monitoring device, not an IVD.
  • Lack of IVD Characteristics: There is no mention of analyzing biological samples (blood, urine, tissue, etc.) or performing tests on these samples to diagnose or monitor a condition.

Therefore, the Vigileo™ APCO/Oximetry Monitor is a patient monitoring device, not an In Vitro Diagnostic device.

N/A

Intended Use / Indications for Use

The Vigileo™ APCO/Oximetry Monitor is indicated for continuously measuring hemodynamic parameters such as cardiac output and oximetry to assess oxygen delivery and consumption. When connected to an Edwards oximetry catheter, the monitor measures oximetry in adults and pediatrics. The monitor also displays parameters, such as stroke volume and stroke volume variation, used to assess fluid status and vascular resistance. The Vigileo™ APCO/Oximetry Monitor may be used in all settings in which critical care is provided.

Product codes (comma separated list FDA assigned to the subject device)

DXG, DQE

Device Description

The Vigileo™ Arterial Pressure Cardiac Output (APCO)/Oximetry Monitor (Vigileo™ Monitor) is a microprocessor-based instrument. When used with an Edwards FloTrac sensor, the Vigileo™ Monitor is capable of continuously monitoring the following parameters: Cardiac Output (CO); Cardiac Index (CI): Stroke Volume (SV): Stroke Volume Variation (SVV); Systemic Vascular Resistance (SVR); Systemic Vascular Resistance Index (SVRI); Oxygen Delivery (DO2); and Oxygen Delivery Index (DO2I). When used with Edwards Oximetry catheters, the Vigileo™ Monitor is capable of continuously monitoring the following parameters: Central venous oxygen saturation (ScvO2); and, Mixed venous oxygen saturation (SvO2).

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

adults and pediatrics

Intended User / Care Setting

all settings in which critical care is provided.
OR and ICU environments.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Verification and validation testing was conducted to compare the performance and functionality of the subject and predicate devices. This testing regimen included side-by-side bench, pre-clinical studies, and comparative analysis of clinical data. The Vigileo™ Monitor has been shown to be safe, effective, and substantially equivalent to the cited predicate device for its intended use critical care environments.

The Vigileo™ Monitor has successfully passed functional and performance testing, including software verification and validation, mechanical and electrical testing, bench studies, pre-clinical animal studies, comparison testing of clinical cases, and clinical utility.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K103094

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 870.1435 Single-function, preprogrammed diagnostic computer.

(a)
Identification. A single-function, preprogrammed diagnostic computer is a hard-wired computer that calculates a specific physiological or blood-flow parameter based on information obtained from one or more electrodes, transducers, or measuring devices.(b)
Classification. Class II (performance standards).

0

Traditional 510(k) - Vigileo™ Monitor, ver 3.08

KAY 2 2 2014

SECTION 5 - 510(K) SUMMARY

.

VIGILEO™ MONITOR 510(k) SUMMARY
510(k) SubmitterEdwards Lifesciences
Contact PersonSally L. Maher
One Edwards Way
Irvine, CA 92614-5686
Date PreparedMay 6, 2014
Trade NameVigileo™ Arterial Pressure Cardiac Output (APCO)/Oximetry Monitor
Common NameCardiac Output/Oximetry Computer
Classification
NameSingle-Function, Preprogrammed Diagnostic Computer
(21 CFR 870.1435, product codes DXG, DQE )
Regulation Class/
Product CodeClass II/
DXG, DQE
Predicate DeviceK103094 (cleared 17 March 2011), Edwards Lifesciences, Vigileo™
Arterial Pressure Cardiac Output/Oximetry Monitor
Device
DescriptionThe Vigileo™ Arterial Pressure Cardiac Output (APCO)/Oximetry
Monitor (Vigileo™ Monitor) is a microprocessor-based instrument.
When used with an Edwards FloTrac sensor, the Vigileo™ Monitor is
capable of continuously monitoring the following parameters:
Cardiac Output (CO);
Cardiac Index (CI):
Stroke Volume (SV):
Stroke Volume Variation (SVV);
Systemic Vascular Resistance (SVR);
Systemic Vascular Resistance Index (SVRI);
Oxygen Delivery (DO2); and
Oxygen Delivery Index (DO2I).
When used with Edwards Oximetry catheters, the Vigileo™ Monitor is
capable of continuously monitoring the following parameters:
Central venous oxygen saturation (ScvO2); and,
Mixed venous oxygen saturation (SvO2).
VIGILEO™ MONITOR 510(k) SUMMARY, con't.
Indications for
Use/Intended UseThe Vigileo™ APCO/Oximetry Monitor is indicated for continuously
measuring hemodynamic parameters such as cardiac output and oximetry
to assess oxygen delivery and consumption. When connected to an
Edwards oximetry catheter, the monitor measures oximetry in adults and
pediatrics. The monitor also displays parameters, such as stroke volume
and stroke volume variation, used to assess fluid status and vascular
resistance. The Vigileo™ APCO/Oximetry Monitor may be used in all
setting in which critical care is provided.
Comparative
AnalysisVerification and validation testing was conducted to compare the
performance and functionality of the subject and predicate
devices. This testing regimen included side-by-side bench, pre-
clinical studies, and comparative analysis of clinical data. The
Vigileo™ Monitor has been shown to be safe, effective, and
substantially equivalent to the cited predicate device for its
intended use critical care environments.
Functional/ Safety
TestingThe Vigileo™ Monitor has successfully passed functional and
performance testing, including software verification and validation,
mechanical and electrical testing, bench studies, pre-clinical animal
studies, comparison testing of clinical cases, and clinical utility.
ConclusionThe Vigileo™ Monitor has been shown to be safe, effective, and is
substantially equivalent to the cited predicate devices for their intended
use in the OR and ICU environments.

1

K131588

.

.

2

Image /page/2/Picture/1 description: The image shows a black and white seal for the Department of Health & Human Services - USA. The seal is circular with text around the perimeter. The right side of the seal contains a symbol of three curved lines.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-00002

May 22, 2014

Edwards Lifesciences LLC Mr. Stephen M. Enos Director, Regulatory Affairs One Edwards Way Irvine, CA 92614

Re: K131588

Edwards Lifesciences™ Vigileo™ Arterial Pressure Cardiac Output/Oximetry Monitor Regulation Number: 21 CFR 870.1435 Regulation Name: Single-function, Preprogrammed Diagnostic Computer Regulatory Class: Class II (two) Product Code: DXG, DQE Dated: May 6, 2014 Received: May 7, 2014

Dear Mr. Enos:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

3

Page 2 - Mr. Stephen M. Enos

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm

Sincerely yours,

Image /page/3/Picture/7 description: The image shows a logo with the letters FDA. The letters are in a bold, outlined font. The logo appears to be crossed out with a thick, dark line. The image is in black and white.

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

4

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K131588

Device Name

Edwards Lifesciences™ Vigileo™ Arterial Pressure Cardiac Output/Oximetry Monitor

Indications for Use (Describe)

The Vigileo™ APCO/Oximerry Monitor is indicated for continuously measuring hemodynamic parameters such as cardiac output and oximetry to assess oxygen delivery and connected to an Edwards oximetry catheter, the monitor measures oximetry in adults and pediatrics.

The monitor also displays parameters, such as stroke volume variation, used to assess fluid status and vascular resistance. The Vigileo™ APCO/Oximetry Monitor may be used in all settings in which critical care is provided.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

2014.05.22 11:18:12

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY -- -- -- T Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

J. Russell Blein

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of Information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."