K Number
K131573
Device Name
CP150 ELECTROCARDIOGRAPH
Manufacturer
Date Cleared
2013-07-03

(34 days)

Product Code
Regulation Number
870.2340
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The Welch Allyn CP150TM Electrocardiograph is intended for use by trained operators in health facilities. The electrocardiograph provides the following diagnostic functions: Acquiring, viewing, storing and printing ECG waveforms using ECG Front-End modules (patient cables) and associated accessories that provide signal acquisition for up to twelve (12) leads of patient ECG waveforms through surface electrodes adhered to the body.Using optional algorithms to generate measurements, data presentations and graphical presentations, and interpretative statements on an advisory basis. These are presented for review and interpretation by the clinician based upon knowledge of the patient, the result of physical examination, the ECG tracings, and other clinical findings. The electrocardiograph is one of the tools that clinicians use to evaluate, diagnose, and monitor patient cardiac function. The 12-lead optional interpretation algorithm provides a computer-generated analysis of potential patient cardiac abnormalities, which must be confirmed by a physician with other relevant clinical information.
Device Description
The Welch Allyn CP150™ Electrocardiograph is an electrocardiograph used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart. Its features include a 7" color touch screen display for ECG preview and user-friendly interface, full-size user-programmable reports, and the ability to operate on either battery or AC power. The CP150™ Electrocardiograph is able to connect either via USB cable or via wired ethernet (RJ45 connector) across the ethernet network, which in turn can connect with other electronic patient-information systems, such as billing and medical records. The USB port can also be used to connect other accessory devices.
More Information

No
The document mentions "optional algorithms to generate measurements, data presentations and graphical presentations, and interpretative statements on an advisory basis" and a "12-lead optional interpretation algorithm". While these are algorithms, the description does not provide any indication that they utilize AI or ML techniques. The "Mentions AI, DNN, or ML" section explicitly states "Not Found".

No.
The device is used for diagnostic purposes (acquiring, viewing, storing, and printing ECG waveforms and providing interpretive statements on an advisory basis), not for therapy.

Yes

The "Intended Use / Indications for Use" section explicitly states that the device provides "diagnostic functions" and is "one of the tools that clinicians use to evaluate, diagnose, and monitor patient cardiac function." It also mentions "Using optional algorithms to generate measurements, data presentations and graphical presentations, and interpretative statements on an advisory basis" and a "12-lead optional interpretation algorithm provides a computer-generated analysis of potential patient cardiac abnormalities".

No

The device description explicitly mentions hardware components such as a 7" color touch screen display, battery or AC power operation, USB cable and wired ethernet connectivity, and the use of ECG Front-End modules (patient cables) and surface electrodes. This indicates it is a hardware device with integrated software, not a software-only medical device.

Based on the provided information, the Welch Allyn CP150™ Electrocardiograph is not an In Vitro Diagnostic (IVD) device.

Here's why:

  • IVD Definition: In Vitro Diagnostics are medical devices used to perform tests on samples taken from the human body, such as blood, urine, or tissue, to provide information about a person's health.
  • Device Function: The Welch Allyn CP150™ Electrocardiograph directly measures the electrical activity of the heart through surface electrodes placed on the body. It processes and displays this electrical signal.
  • Lack of Sample Analysis: The device does not analyze samples taken from the body. It is a non-invasive device that measures a physiological signal on the body.

Therefore, the Welch Allyn CP150™ Electrocardiograph falls under the category of a physiological monitoring device rather than an In Vitro Diagnostic device.

N/A

Intended Use / Indications for Use

The Welch Allyn CP150TM Electrocardiograph is intended for use by trained operators in health facilities. The electrocardiograph provides the following diagnostic functions: Acquiring, viewing, storing and printing ECG waveforms using ECG Front-End modules (patient cables) and associated accessories that provide signal acquisition for up to twelve (12) leads of patient ECG waveforms through surface electrodes adhered to the body. Using optional algorithms to generate measurements, data presentations, graphical presentations, and interpretative statements on an advisory basis. These are presented for review and interpretation by the clinician based upon knowledge of the patient, the result of physical examination, the ECG tracings, and other clinical findings.

The electrocardiograph is one of the tools that clinicians use to evaluate, diagnose, and monitor patient cardiac function. The 12-lead optional interpretation algorithm provides a computer-generated analysis of potential patient cardiac abnormalities, which must be confirmed by a physician with other relevant clinical information.

Product codes

DPS

Device Description

The Welch Allyn CP150™ Electrocardiograph is an electrocardiograph used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart. Its features include a 7" color touch screen display for ECG preview and user-friendly interface, full-size user-programmable reports, and the ability to operate on either battery or AC power.

The CP150™ Electrocardiograph is able to connect either via USB cable or via wired ethernet (RJ45 connector) across the ethernet network, which in turn can connect with other electronic patient-information systems, such as billing and medical records. The USB port can also be used to connect other accessory devices.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Adult and pediatric patients

Intended User / Care Setting

trained operators in health facilities

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

No clinical studies were utilized for the purpose of obtaining safety and effectiveness data.

Non-Clinical Tests: The Welch Allyn CP150™ Electrocardiograph was tested to evaluate its safety and effectiveness based on the following standards: EN/IEC 60601-1, EN/IEC 60601-1-2, EN/IEC 60601-1-4, AAMI EC-11, IEC/EN60601-2-25, IEC/EN60601-2-51, EN/ISO 14971.

Additional performance Bench Testing was conducted for:

  • CP150 Connectors Reliability: PASS: The connectors meet the requirements of TRS and reliability plan.
  • CP150 Battery Capacity Test: PASS: The battery meets the requirements of TRS.
  • CP150 Battery Door & Connector: PASS: The battery door and battery connector meet the requirements of TRS.
  • CP150 Power Button Test: PASS: The Power Button meets the requirements of reliability plan.
  • CP150 LCD Touchscreen Test: PASS: The test results shows the LCD touchscreen meets the requirements of Reliability plan.
  • CP150 Environment Test: PASS: The test results shows the CP150 meets the environmental test requirements of Reliability plan.
  • CP150 Barcode scanner verification: PASS: The test results shows the bar code scanner meets the requirements of Reliability plan.
  • 60601-1 Safety test: PASS Devices are compliant with 60601-1 standard.
  • 60601-1-2 Electromagnetic Compatibility test: PASS Devices are compliant with 60601-1-2 standard.
  • 60601-1-4 Medical Electrical Equipment - Part 1-4: General Requirements For Safety - Collateral Standard: General Requirements for Programmable Electrical Medical Systems: PASS Devices are compliant with 60601-1-4 standard.
  • AAMI EC-11 Diagnostic electrocardiographic devices: PASS Devices are compliant with AAMI EC-11.
  • 60601-2-25 Medical electrical equipment - Part 2-25: Particular requirements for the basic safety and essential performance of electrocardiographs: PASS Devices are compliant with 60601-2-25.
  • 60601-2-51 Particular Requirements For Safety, Including Essential Performance, Of Recording And Analysing Single Channel And Multichannel Electrocardiographs: PASS Devices are compliant with 60601-2-51.
  • 14971 Medical Devices - Application of Risk Management to Medical Devices: PASS Risk Analysis Summary.
  • 60051724-- Welch Allyn mobile stand - Large Platform Cart (LPC) Safety test: PASS the LPC is compliant with 60601-1 standard and can be used with the subject device.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K072449

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 870.2340 Electrocardiograph.

(a)
Identification. An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.(b)
Classification. Class II (performance standards).

0

Welch

K131573 pg 1 of 13

Special 510(k) Premarket Notification Welch Allyn CP150TM Electrocardiograph

510(k) Summary

[As described in 21 CFR 807.92]

JUL 0 3 2013

Submitted by:Welch Allyn Inc.
4341 State Street Road
Skaneateles Falls, NY 13153-0220
Contact Person:Kevin Crossen
Director Regulatory Affairs
Phone: (315) 685-2609
Fax: (315) 685-2532
E-mail: Kevin.Crossen@welchallyn.com
Date Prepared:May 30, 2013
Trade Name:Welch Allyn CP150 TM Electrocardiograph
901049 Electrocardiograph
Common Name:Electrocardiograph
Classification Reference:Class II, Electrocardiograph (21 CFR 870.2340, Product Code DPS)
Predicate Device:Welch Allyn CP100 TM and CP200 TM Electrocardiographs
510(k) Number: K072449
Electrocardiograph, 21 CFR 870.2340
Class II, DPS

Description of the Device:

The Welch Allyn CP150™ Electrocardiograph is an electrocardiograph used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart. Its features include a 7" color touch screen display for ECG preview and user-friendly interface, full-size user-programmable reports, and the ability to operate on either battery or AC power.

The CP150™ Electrocardiograph is able to connect either via USB cable or via wired ethernet (RJ45 connector) across the ethernet network, which in turn can connect with other electronic patient-information systems, such as billing and medical records. The USB port can also be used to connect other accessory devices,

1

Image /page/1/Picture/1 description: The image shows the logo for Welch Allyn. The text is in a bold, sans-serif font. The words "Welch" and "Allyn" are next to each other, with a small registered trademark symbol after "Allyn".

Special 510(k) Premarket Notification Welch Allyn CP150"M Electrocardiograph

The CP150™ Electrocardiograph is specifically intended for acquiring and printing ECG signals from adults and pediatric patients. It will be used in clinical settings by trained healthcare providers.

The optional interpretation algorithm analyzes these ECG signals to generate measurements and interpretive statements. The interpretive results are intended only as guidance for qualified physicians and must not be relied upon as diagnoses.

Indications for Use:

The electrocardiograph is one of the tools that clinicians use to evaluate, diagnose, and measure patient cardiac function.

The 12-lead optional interpretation algorithm provides a computer-generated analysis of potential patient cardiac abnormalities, which must be confirmed by a physician with other relevant clinical information.

Contraindications:

There are no known contraindications for use.

Technological Characteristics:

The subject device has the same technological characteristics and indications for use as the predicate; The hardware, software, and mechanical aspects of the CP150 have been updated to current technology equivalent to the cleared devices (CP100" and CP200™ Electrocardiographs, K072449, S.E. dated Nov. 29, 2007) as described below. The modification is to replace the display, i.e. 5.7 inches color LCD display, with a 7' color touch screen display for ECG preview. The physical QWERTY keyboard and other hard function keys used on CP100/CP200™ will be replaced by the touch screen interface. Additionally included are minor software and connectivity enhancements to improve performance and customer experience.

Non-Clinical Tests:

The Welch Allyn CP150" Electrocardiograph was tested to evaluate its safety and effectiveness based on the following standards:

StandardVersionTitle
EN/IEC 60601-12nd Edition
2000Medical Electrical Equipment - Part 1: General Requirements for Basic Safety
and Essential Performance
EN/IEC 60601-1-22007Medical Electrical Equipment - Part 1-2: General Requirements for Basic
Safety and Essential Performance - Collateral Standard: Electromagnetic
Compatibility - Requirements and Tests

2

Image /page/2/Picture/0 description: The image shows the logo for Welch Allyn. The text is in a bold, serif font. The words "Welch" and "Allyn" are next to each other, with a small dot after the word "Allyn".

K131573 pg 3 of 13

Special 510(k) Premarket Notification

Welch Allyn CP150™ Electrocardiograph

| EN/IEC 60601-1-4 | 2000 | Medical Electrical Equipment - Part 1-4: General Requirements for Safety -
Collateral Standard: General Requirements for Programmable Electrical
Medical Systems |
|------------------|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| AAMI EC-11 | 1991 (R2007) | Diagnostic electrocardiographic devices |
| IEC/EN60601-2-25 | 1993 AMD 1
1999 | Medical electrical equipment - Part 2-25: Particular requirements for the basic
safety and essential performance of electrocardiographs |
| IEC/EN60601-2-51 | 2003 | Particular Requirements For Safety, Including Essential Performance, Of
Recording And Analysing Single Channel And Multichannel
Electrocardiographs |
| EN/ISO 14971 | 2007/2007 | Medical Devices - Application of Risk Management to Medical Devices |

.

:

·

3

K131573 pg 4 of 13

VechAllyn.

Special 510(k) Premarket Notification
Welch Allyn CP150™ Electrocardiograph

Rench Testi Additional per

Additional performance Bench Testing:
Report DIR --
DescriptionObjective of the TestTest Method/ ProcedurePass/Fail CriteriaConclusions
60048998 --- CP150
Connectors
ReliabilityTo verify the reliability of
all connectors of CP150 as
per CP150 Technical
Requirements Specification
(TRS) (DIR 60047341) and
CP150 Reliability Plan
(DIR 60046890) including:
• AC power inlet
• ECG (DB15)
• Ethernet (RJ45)
• mini USB (USB B)
• USB1. Insert connector into the inlet
  1. Conduct an Interval Inspection after
    every 10 insert/remove cycles.
    3: Repeat steps 1-2 for greater than the
    minimum number of cycles listed in
    Pass/Fail criteria | • AC power inlet; 50
    connect/disconnect
    cycles.
    • ECG (DB15), 50
    connect/disconnect
    cycles.
    • Ethernet (RJ45), 100
    connect/disconnect
    cycles.
    • mini USB (USB B)
    100
    connect/disconnect
    cycles.
    • USB, 400
    connect/disconnect
    cycles. | PASS: The
    connectors
    meet the
    requirements of
    TRS and
    reliability plan |
    | 60048999 --- CP150
    Battery Capacity Test | To verify the battery
    capacity for number of
    standard use cycles
    available without re-
    charging the battery | 1. Using only the battery (fully charged),
    perform 25x ECG print tests
  2. ECG test duration should take 15
    minutes
  3. Check battery capacity after each ECG
    print test | The device shall have the
    capacity to complete 25
    ECGs over a 12 hour period
    with a single fully charged
    battery. Each ECG test is
    assumed taken in 15 minutes
    cycle with two copies of
    ECG prints | PASS: The
    battery meets
    the
    requirements
    of TRS |
    | 60049000 --- CP150
    Battery Door &
    Connector | To verify the reliability of
    Battery Door and Battery
    Connector for number of
    standard use
    (insert/remove) cycles | 1. Unscrew the Battery Door. Remove
    the battery fully. Insert back the battery
    and close the battery door and screw back
    to secure.
  4. Conduct an Interval Inspection after
    every 10 insert/remove cycles. | TRS 4.8 4.8: The battery
    connector shall have a
    useable life shall be at least
    10 connect/disconnect cycles
    and the door needs to pass
    230 cycles of
    opening/closing. | PASS: The
    battery door
    and battery
    connector meet
    the
    requirements
    of TRS |
    | 60049001 --- CP150
    Power Button Test | To verify the reliability of
    Power Button by number
    of presses as per CP150
    Reliability Plan (DIR
  1. requirements | • Press the Power Button with a nominal
    force as to power up the unit. Hold for
    one second then release.
    • Conduct an Interval Inspection after
    every 2,500 presses. | Reliability Plan 8.5: The test
    should meet zero failures for
    a total of 48,909 presses on 1
    unit | PASS: The
    Power Button
    meets the
    requirements
    of reliability
    plan |
    | 60049002 --- CP150
    LCD Touchscreen Test | To verify the reliability of
    LCD Touchscreen by
    applying mechanical
    touches on the screen as
    per CP150 Reliability Plan
    (DIR 60046890)
    requirements | • Install the 'touch finger' with silicone
    rubber region above the LCD
    Touchscreen. Adjust the input force to be
    350g.
    • Set the same for the other 5 'touch
    finger'
    • Setup test speed to 1 Hz and start the
    test.
    • Conduct an Interval Inspection after
    every 60,000 touches.
    • Shift the 'touch finger' to other positions | Reliability Plan 8.5: All the
    functional tests must pass for
    total of 1,687688 touches on
    3 unites (please see the
    rational on analysis column
    on the right) | PASS: The test
    results shows
    the LCD
    touchscreen
    meets the
    requirements
    of Reliability
    plan |
    | 60049003 --- CP150
    Environment Test | To verify the device shall
    operate at temperature
    between 10.0C and 40.0C
    and at a relative humidity
    of 15% to 95% (non
    condensing) For printing
    the humidity is limited to
    30% to 70% (non-
    condensing) | Device dwells for at least 6 hours at each
    operating conditions
    After every 3 hours dwell at each
    condition, perform an intermediate
    verification test | All the intermediate
    verification test was passed
    for following conditions:
    Acquiring ECG with
    Printing
    • +5°C / 30% RH for 6 hours
    • +5°C / 70% RH for 6 hours
    • +45°C / 30% RH for 6
    hours
    • +45°C / 30% RH for 6
    hours
    Acquiring ECG without
    Printing
    • +5°C / 10% RH for 6 hours
    • +5°C / 95% RH for 6 hours
    • +45°C / 10% RH for 6
    hours
    • +45°C / 95% RH for 6
    hours | PASS: The test
    results shows
    the CP150
    meets the
    environmental
    test
    requirements
    of Reliability
    plan |
    | --- CP150 Barcode
    scanner verification | To verify that the device
    will have the ability to
    input barcode
    alphanumeric inputs via a
    bar code scanner | 1. Connect barcode scanner to the device
  1. Scan the sample Serial number and
    verify the result
  2. Repeat step 2 for 58 times | Scanned results should
    match the sample serial
    number in the box | PASS: The test
    results shows
    the bar code
    scanner meets
    the
    requirements
    of Reliability
    plan |
    | 50052441 ---60601-
    1 Safety test | Test the device per 60601-
    1 to ensure that the device
    meets the safety
    requirements for medical
    devices | Units delivered to independently certified
    test lab for testing per compliance with
    standard. | Devices pass criteria of
    60601-1 | PASS Devices
    are compliant
    with 60601-1
    standard |
    | 50052442 --- 60601-
    1-2 Electromagnetic
    Compatibility test | Test the device per 60601-
    1-2 to ensure that the
    device meets the
    requirements for
    Electromagnetic
    Compatibility | Units delivered to independently certified
    test lab for testing per compliance with
    standard. | Devices pass criteria of
    60601-1-2 | PASS Devices
    are compliant
    with 60601-1-2
    standard |
    | 50052441 - 60601-1-4
    Medical Electrical
    Equipment - Part 1-4:
    General Requirements
    For Safety - Collateral
    Standard: General
    Requirements for
    Programmable
    Electrical Medical
    Systems | Test the device per 60601-
    1-4 to ensure that the
    device meets the
    requirements for
    Programmable Electrical
    Medical Systems | Units delivered to independently certified
    test lab for testing per compliance with
    standard. | Devices pass criteria of
    60601-1-4 | PASS Devices
    are compliant
    with 60601-1-4 |
    | AAMI EC-11
    Diagnostic
    electrocardiographic
    devices | Test the device per to
    ensure that the device
    meets the AAMI standard
    requirements for Diagnostic
    electrocardiographic devices | Units delivered to independently certified
    test lab for testing per compliance with
    standard. | Devices pass criteria of
    AAMI EC-11 | PASS Devices
    are compliant
    with AAMI EC-
    11 |
    | 60601-2-25 Medical
    electrical equipment -
    Part 2-25: Particular
    requirements for the
    basic safety and
    essential performance
    of electrocardiographs | 60601-2-51 Particular
    Requirements For
    Safety, Including
    Essential Performance,
    Of Recording And
    Analysing Single
    Channel And
    Multichannel
    Electrocardiographs | 14971 Medical Devices
  • Application of Risk
    Management to Medical
    Devices | 60051724-- Welch
    Allyn mobile stand
  • Large Platform Cart
    (LPC) Safety test | |
    | Test the device per to
    ensure that the device
    meets the 60601-2-25
    requirements for the basic
    safety and essential
    performance of
    electrocardiographs | Test the device per to
    ensure that the device
    meets 60601-2-51 the
    requirements for Safety,
    Including Essential
    Performance, Of Recording
    And Analysing Single
    Channel And Multichannel
    Electrocardiographs | Test the device per to
    ensure that the device meet
    the requirements for the
    Application of Risk
    Management to Medical
    Devices | To verify the LPC meet the
    safety requirements per
    60601-1 | |
    | Units delivered to independently certified
    test lab for testing per compliance with
    standard. | Units delivered to independently certified
    test lab for testing per compliance with
    standard. | Risk management review performed
    according to Welch Allyn procedures | The subject device is verifies to be
    assembled with the cart as per assembly
    instructions and the LPC is verified to
    meet the safety requirements per 60601-1
    by third party | |
    | Devices pass criteria of
    60601-2-25 | Devices pass criteria of
    60601-2-51 | Devices pass criteria of
    14971 | The device can be installed
    as per assembly instruction
    and the LPC pass criteria of
    60601-1 | |
    | PASS Devices
    are compliant
    with 60601-2-
    25 | PASS Devices
    are compliant
    with 60601-2-
    51 | PASS Risk
    Analysis
    Summary | PASS the LPC
    is compliant
    with 60601-1
    standard and
    can be used
    with the
    subject device | |

4

WelchAllyn

K131573 pg 5 of 13
Special Special 510(k) Premarket Notification
Welch Allyn CP150™ Electrocardiograph

.

Welch Allyn CP150™ Electrocardiograph

5

K131573 pg 6 of 13

VectorAly

Special 510(k) Premarket Notification
Welch Allyn CP150™ Electrocardiograph

Page 8-7

.

6

K131573 pg 7 of 13

Page 8-8

WelchAllyn

7

K131573 pg 8 of 13

V
ely
Allyn
.

Special 510(k) Premarket Notification
Welch Allyn CP150™ Electrocardiograph

.

8

K131573 pg 9 of 13

Veatchlyn.

Special 510(k) Premarket Notification
Welch Allyn CP150™ Electrocardiograph

ge
8-10

9

Image /page/9/Picture/1 description: The image shows the logo for Welch Allyn. The logo is in a bold, sans-serif font. The words "Welch Allyn" are written in black. There is a registered trademark symbol to the right of the word "Allyn".

Clinical Performance Data:

No clinical studies were utilized for the purpose of obtaining safety and effectiveness data.

Device Comparison Table:

The Welch Allyn CP150™ Electrocardiograph is substantially equivalent in operation and performance to the Welch Allyn CP100™ and CP200TM Electrocardiographs (K072449).

Subject Device and Predicate Device Comparison
CharacteristicSubject DevicePredicate DevicesDifferences
DeviceCP150TM
ElectrocardiographCP100TM and CP200TM
ElectrocardiographsModel number
ManufacturerWelch Allyn, Inc.Welch Allyn, Inc.Same
510(k) NumberN/AK072449N/A
Product CodeDPSDPSSame
Classification NameElectrocardiograph, Class
IIElectrocardiograph, Class
IISame
Regulation Number:21 CFR 870.234021 CFR 870.2340Same
Intended UseThe Welch Allyn CP150TM
Electrocardiograph is
intended for use by trained
operators in health facilities.
The electrocardiograph
provides the following
diagnostic functions:
Acquiring, viewing,
storing and printing ECG
waveforms using ECG
Front-End modules
(patient cables) and
associated accessories
that provide signal
acquisition for up to
twelve (12) leads of
patient ECG waveforms
through surface electrodes
adhered to the body.Using optional algorithms
to generate
measurements, data
presentations, graphical
presentations, andThe Welch Allyn
electrocardiographs are
intended for use by trained
operators in health facilities.
The electrocardiograph will
provide the following
diagnostic functions:
Acquiring, viewing,
storing, and printing
ECG waveforms using
ECG Front End modules
and associated
accessories that provide
signal acquisition for up
to twelve (12) leads of
patient ECG waveforms
through surface
electrodes adhered to the
body.Using optional
algorithms to generate
measurements, data
presentations, graphical
presentations andSame
Subject Device and Predicate Device Comparison
CharacteristicSubject DevicePredicate DevicesDifferences
interpretative statements
on an advisory basis.
These are presented for
review and interpretation
by the clinician based
upon knowledge of the
patient, the result of
physical examination, the
ECG tracings, and other
clinical findings.interpretative statements on
an advisory basis. These
are presented for review
and interpretation by the
clinician based upon
knowledge of the patient,
the result of physical
examination, the ECG
tracings, and other
clinical findings.
Indications for UseThe electrocardiograph is one
of the tools that clinicians use
to evaluate, diagnose, and
monitor patient cardiac
function.
The 12-lead interpretive
algorithm provides a
computer-generated analysis
of potential patient cardiac
abnormalities, which must be
confirmed by a physician
with other relevant clinical
information.The electrocardiograph is
one of the tools that
clinicians use to evaluate,
diagnose, and monitor
patient cardiac function.
The 12-lead interpretive
algorithm provides a
computer-generated analysis
of potential patient cardiac
abnormalities, which must
be confirmed by a physician
with other relevant clinical
information.Same
Target PopulationAdult and pediatric patientsAdult and pediatric patientsSame
Where UsedHealth care facilitiesHealth care facilitiesSame
ECG Storage100 Adults ECG50 Adults ECGUpgrading to
higher storage
capacity
PrinterThermal printer (internal)Thermal printer (internal)Same
Display typeLCD color touch screenColor LCDUpgrading to color
touch screen
Alphanumeric keyboardQWERTY keyboard - On
screenPhysical QWERTY
keyboardSame function
SterilityDevice not supplied sterileDevice not supplied sterileSame
PowerThe electrocardiograph can
run on AC or battery powerThe electrocardiograph can
run on AC or battery powerSame
Subject Device and Predicate Device ComparisonDifferences
CharacteristicSubject DevicePredicate DevicesDifferences
Battery OperationYesYesSame
12 hours of continuous use or
continuous printing of 250
ECG pages. The continuous
test is based on performing
25 ECG's in a period of 12
hours

Battery recharge - 6 hours to
full capacity | 4 hours of continuous use or
continuous printing of
100ECG pages. The
continuous test is based on
performing 5 ECG's in a
period of 4 hours

Battery recharge - 12 hours
to full capacity | Upgrading the
battery capacity |
| Standard Compliance | EC 11 (AAMI/ANSI)
IEC 60601-1
IEC 60601-1-2
IEC 60601-1-4
IEC 60601-2-25
IEC 60601-2-51
EN/ISO 14971 | EC 11 (AAMI/ANSI)
IEC 60601-1
IEC 60601-1-1
IEC 60601-1-2
IEC 60601-1-4
IEC 60601-2-25
IEC 60601-2-51 | Same - except IEC
60601-1-1 does not
apply to the
CP150. The
CP150 (like the
CP100) does not
have spirometry
like the CP200. It
no longer meets
the definition of
Medical System in
IEC 60601-1-1 |
| Filters | - 0.5 Hz high-performance
base line filter

  • 35 Hz muscle-tremor
    filter
  • AC interference filter | - 0.5 Hz high-
    performance base line
    filter
  • 35 Hz muscle-tremor
    filter
  • AC interference filter | Same |
    | ECG acquisition | ECG signal acquisition of up
    to 12 leads | ECG signal acquisition of up
    to 12 leads | Same |
    | ECG Interpretation | Optional algorithm for adult
    (MEANS) and pediatric
    (PEDMEANS) patients | Optional algorithm for adult
    (MEANS) and pediatric
    (PEDMEANS) patients | For the PEDMEANS
    analysis - QT
    interval corrected for
    heart rate according
    to Hodges' formula:
    QTc = QT + 1.75 ×
    (HR-60) was added
    as a user selectable
    option. This same
    analysis software is
    also used in Welch |
    | Subject Device and Predicate Device Comparison | | | |
    | Characteristic | Subject Device | Predicate Devices | Differences |
    | | | | Allyn's
    Cardioperfect
    Workstation, and
    this modification to
    the PEDMEANS
    was covered in the
    submission K082478
    for the Cardioperfect
    Workstation. |
    | Weight | 9.9 lb | 11.6 lb | Slightly lighter |
    | Dimensions | 15x14.1x5.4 in. | 16.2x15.6x6.2 in. | Slightly smaller |
    | Connectivity | - 1 USB client,
  • 4 USB host ports,
  • 1 Ethernet port. | - 1 USB client
  • 1 SD slot
  • Wireless network | - Add 4 USB host
    ports
  • Remove the SD
    slot
  • Add 1 Ethernet
    port to replace
    wireless network |
    | Accessories | - Patient Electrodes
  • Printer Paper
  • Patient Lead Cables
    DB15 connector to
    CP150 | - Patient Electrodes
  • Printer Paper
  • Patient Lead Cables
    DB9 connector to
    CP100/CP200 | All accessories are
    the same with the
    exception of a
    minor change to
    the connector type
    on the Unit
    connection end of
    patient lead cables |

10

K131573 pg 11 of 13

Welch Allyn

Special 510(k) Premarket Notification Welch Allyn CP150™ Electrocardiograph

.

11

K131573 pg 12 of 13

Image /page/11/Picture/1 description: The image shows the logo for Welch Allyn. The logo is in a bold, sans-serif font. The words "Welch Allyn" are written in black. There is a registered trademark symbol to the right of the word "Allyn".

Special 510(k) Premarket Notification Welch Allyn CP150™ Electrocardiograph

"

12

Image /page/12/Picture/0 description: The image shows the logo for Welch Allyn. The logo is in a bold, serif font. The words "Welch Allyn" are written in black. There is a small circle after the word Allyn.

K131573 pg 13 of 13

Conclusion

Based on the information presented in this 510(k) premarket notification the Welch Allyn CP150™ Electrocardiograph is considered substantially equivalent (as safe, as effective and performs as well as) the currently marketed devices (K072449) citied in this submission. The differences noted between the CP150 and the predicate devices do not impact safety or effectiveness based on the successfully conducted testing of the modified device.

13

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

Image /page/13/Picture/3 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three wing-like shapes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the emblem.

July 3, 2013

Welch Allyn, Inc. C/O Kevin Crossen 4341 State St. Rd. P.o. Box 220 Skaneateles Falls, NY 13153-0220 US

Re: K131573

Trade/Device Name: Cp150 electrocardiograph Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: May 30, 2013 Received: June 3, 2013

Dear Kevin Crossen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

14

Page 2 - Kevin Crossen

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

M.A. Lillehemen

for Bram D. Zuckerman, Ph.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

15

K131573

Indications for Use

510(k) Number (if known): K

Device Name: Welch Allyn CP150™ Electrocardiograph

Indications for Use:

The electrocardiograph is one of the tools that clinicians use to evaluate, diagnose, and monitor patient cardiac function.

The 12-lead optional interpretation algorithm provides a computer-generated analysis of potential patient cardiac abnormalities, which must be confirmed by a physician with other relevant clinical information.

Prescription Use × (Part 21 CFR 801 Subpart D) Over-The-Counter Use, (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

AND/OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

M.A. Hillemann