(97 days)
The InterFuse L - Intervertebral Body Fusion Device is indicated for intervertebral spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The InterFuse L device is to be used in patients who have had at least six (6) months of non-operative treatment. These patients may have had a previous non-fusion surgery at the involved spinal level(s). The InterFuse L device is indicated for use with autogenous bone graft, and to be used with supplemental internal spinal fixation systems that have been cleared for use by the FDA in the lumbosacral spine.
VTI's InterFuse L - Intervertebral Body Fusion device is made of implant grade Polyetheretherketone (PEEK- OPTIMA®, Grade LT1), a polymer with a history of use in interbody fusion device designs, and which has a compressive modulus similar to bone. The device has two modules. Module A has a rail and Module B has a slot. The modules are installed by using VTI's proprietary slot riding the rail technology. Both modules have tantalum beads that aid in visualizing the implanted device under X-ravs. Each module incorporates a ramp lock to help ensure that it is properly aligned and engaged with the other module. Each module has two vertical slots through the device for the surgeon to fill with autogenous bone that will provide a path for solid bone growth during the fusion process. The device is produced in six heights (7 mm, 8 mm, 9 mm, 10 mm, 12 mm and 14 mm) each for parallel and 8 lordotic configurations to fit the angular geometry of the disc at each disc level. Four anterior - posterior (AP) lengths (16 mm, 18 mm, 21 mm and 24 mm) of the device are being offered. The device will be available in 7 Medial - Lateral (ML) dimensions (35 mm, 40 mm, 45 mm, 50 mm, 55 mm 60 mm and 65 mm). VTI will supply a set of Instruments (all non-sterile, re-usable manual instruments) for implantation of InterFuse L by a Direct-Lateral Lumbar Interbody Fusion technique.
The provided text describes the 510(k) summary for the "InterFuse L - Intervertebral Body Fusion Device." This document focuses on demonstrating substantial equivalence to previously cleared predicate devices through non-clinical bench testing. It does not contain information about a clinical study involving human subjects, AI performance, or ground truth establishment in a medical imaging context.
Therefore, many of the requested elements for describing "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of an AI/medical imaging device cannot be extracted from this document.
However, I can provide what is available, specifically related to the non-clinical testing performed for this device.
Acceptance Criteria and Device Performance (Non-Clinical Bench Testing)
| Acceptance Criteria Category | Reported Device Performance (Summary) |
|---|---|
| Durability (Static) | Verified by static testing per ASTM F2077-11 for axial load deflection and shear deflection. |
| Durability (Dynamic) | Verified by dynamic testing per ASTM F2077-11 for axial load cyclic fatigue and shear load cyclic fatigue. |
| Subsidence | Tested per ASTM F2267-04. |
| Biocompatibility | Tested for cytotoxicity per ISO 10993-5:2009 and found to be non-cytotoxic. |
| Sterility | Sterilized by gamma irradiation to a sterility assurance level of 10-6. |
Study Information (Based on Non-Clinical Testing):
The provided document describes non-clinical bench testing rather than a clinical study for an AI device. As such, the following requested information is not applicable or not available within this submission:
- Sample size used for the test set and the data provenance: Not applicable. The testing was bench testing on physical devices, not a dataset of patient images.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for mechanical testing is based on engineering standards and measurements, not expert clinician consensus.
- Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No MRMC study was done, as this is a physical medical device (intervertebral body fusion device), not an AI/imaging diagnostic device.
- If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable in the context of clinical AI/imaging ground truth. For the bench tests, the "ground truth" would be the measured physical properties and performance against established ASTM standards for intervertebral body fusion devices.
- The sample size for the training set: Not applicable. There is no training set mentioned for this type of device submission.
- How the ground truth for the training set was established: Not applicable.
Summary of the Study (Bench Testing):
The study described is a series of non-clinical bench tests designed to demonstrate the mechanical and material performance of the InterFuse L - Intervertebral Body Fusion Device. The tests were conducted according to recognized ASTM standards (ASTM F2077-11 for static and dynamic durability, and ASTM F2267-04 for subsidence) and ISO standards (ISO 10993-5:2009 for cytotoxicity). The purpose was to show substantial equivalence to predicate devices (K091988 and K102277) by demonstrating similar performance characteristics under these standardized test conditions. The device also underwent sterilization testing to ensure a sterility assurance level of 10-6.
{0}------------------------------------------------
4. 510(k) SUMMARY
510(k) Owner's Name:
Phone/Fax/Email:
Address:
ﻣﺮ
EP 0 3 2013
510(k) SUMMARY [as required by section 807.92(c)]
Vertebral Technologies, Inc.
5909 Baker Road, Suite 550 Minnetonka, MN 55345
Suresh Ghai, Ph.D.
AUG 08, 2013
Office - Direct Tel: (952) 979-9357 Company Main Tel: (952) 912-5400 Fax: (952) 912-5410 email: sghai@vti-spine.com
VP, Quality and Regulatory Affairs
Intervertebral Body Fusion Device
Name of Contact Person:
Date prepared:
Trade or Proprietary Name:
Common or Usual Name:
Classification Name:
Intervertebral Fusion Device with Bone Graft, Lumbar 21 CFR § 888.3080 Product code: MAX Device Class: II
InterFuse L - Intervertebral Body Fusion Device
4.1 LEGALLY MARKETED DEVICE TO WHICH SUBSTANTIAL EQUIVALENCE IS CLAIMED
The InterFuse L - Intervertebral Body Fusion Device is substantially equivalent in performance, indication, design and material to:
-
- InterFuse S Intervertebral Body Fusion Device (cleared under Premarket notification # K091988)
-
- InterFuse T Intervertebral Body Fusion Device (cleared under Premarket notification # K102277)
-
- PrimaLIF LLIF Unitary PEEK Lateral Interbody Fusion System (cleared under Premarket notification # K123207)
4.2 DEVICE DESCRIPTION
-31-
{1}------------------------------------------------
VTI's InterFuse L - Intervertebral Body Fusion device is made of implant grade Polyetheretherketone (PEEK- OPTIMA®, Grade LT1), a polymer with a history of use in interbody fusion device designs, and which has a compressive modulus similar to bone. The device has two modules. Module A has a rail and Module B has a slot. The modules are installed by using VTI's proprietary slot riding the rail technology. Both modules have tantalum beads that aid in visualizing the implanted device under X-ravs. Each module incorporates a ramp lock to help ensure that it is properly aligned and engaged with the other module. Each module has two vertical slots through the device for the surgeon to fill with autogenous bone that will provide a path for solid bone growth during the fusion process. The device is produced in six heights (7 mm, 8 mm, 9 mm, 10 mm, 12 mm and 14 mm) each for parallel and 8 lordotic configurations to fit the angular geometry of the disc at each disc level. Four anterior - posterior (AP) lengths (16 mm, 18 mm, 21 mm and 24 mm) of the device are being offered. The device will be available in 7 Medial - Lateral (ML) dimensions (35 mm, 40 mm, 45 mm, 50 mm, 55 mm 60 mm and 65 mm).
VTI will supply a set of Instruments (all non-sterile, re-usable manual instruments) for implantation of InterFuse L by a Direct-Lateral Lumbar Interbody Fusion technique.
4.3 INTENDED USE OF THE DEVICE
The InterFuse L - Intervertebral Body Fusion Device is indicated for intervertebral spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These patients may also have up to Grade I spondylolisthesis at the involved level(s). The InterFuse L device is to be used in patients who have had at least six (6) months of non-operative treatment. These patients may have had a previous non-fusion surgery at the involved spinal level(s). The InterFuse L device is indicated for use with autogenous bone graft and to be used with supplemental internal spinal fixation systems that have been cleared for use by the FDA in the lumbosacral spine.
4.4 TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE PREDICATE DEVICE
InterFuse L Intervertebral Body Fusion Device is substantially equivalent in performance, indication, design and materials to InterFuse S Intervertebral Body Fusion Device (cleared under premarket notification # K091988) and InterFuse T Intervertebral Body Fusion Device (cleared under premarket notification # K102277) both products of Vertebral Technologies. Inc.
4.5 SUMMARY AND CONCLUSIONS FROM THE NONCLINICAL TESTS SUBMITTED
The substantial equivalence is supported by bench testing comparing the InterFuse L Intervertebral Body Fusion Device to the predicate devices (K091988 and K102277).
The basic design of the device is similar to the 510(k) cleared predicate devices (K091988 and K102277). The device will be assembled (intra-operatively) via Direct-Lateral Lumbar Interbody Fusion (DLIF) technique as compared to the predicate devices which are
{2}------------------------------------------------
K131540
assembled (intra-operatively) via Posterior Lumbar Interbody Fusion and Transforaminal Lumbar Interbody Fusion techniques respectively.
The InterFuse L Intervertebral Body Fusion Device is assembled (intra-operatively) using the same rail and slot system as done for the predicate devices (K091988 and K102277).
The performance of the InterFuse L was tested, post gamma irradiation sterilization, in accordance with ASTM F2077-11 and ASTM F2267 - 04 as detailed below:
- 4.5.1 Device durability was verified by static testing per ASTM F2077-11 for axial load deflection and shear deflection.
- Device durability was also verified by dynamic testing per ASTM F2077-11 for 4.5.2 axial load cyclic fatigue and shear load cyclic fatigue.
- Subsidence testing was conducted per ASTM F2267-04. 4.5.3
Substantial equivalence comparison between InterFuse L and predicate devices performance results is given in section 10.
The detailed performance results are given in section 12.
The packaged and sterilized InterFuse L device was also tested for cytotoxicity (per ISO 10993-5:2009) and was found to be non-cytotoxic.
The InterFuse L device is sterilized by gamma irradiation by the same process as used for the predicate devices (K091988 and K102277). This process provides a sterility assurance level of 10-6.
On the basis of performance data it is concluded that the InterFuse L – Intervertebral Body Fusion Device is substantially equivalent to the predicate devices (K091988 and K102277).
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized human figure with three arms reaching upwards, symbolizing health, services, and people. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA", which is arranged in a circular fashion around the left side of the logo.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
September 3, 2013
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002
Vertebral Technologies, Incorporated % Suresh Ghai, Ph.D. Vice President, Quality and Regulatory Affairs 5909 Baker Road, Suite 550 Minnetonka, Minnesota 55345 US
Re: K131540
Trade/Device Name: Interfuse L- Intervertebral Body Fusion Device Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: August 14, 2013 Received: August 15, 2013
Dear Dr. Ghai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 - Suresh Ghai. Ph.D.
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also. please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Mark N: Melkerson -S
Mark N. Melkerson Director Division of Orthonedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
3. STATEMENT OF INDICATION FOR USE
Indication for Use
510(k) Number (if known): ____K131540
Device Name: InterFuse L - Intervertebral Body Fusion Device
Indications for Use:
The InterFuse L - Intervertebral Body Fusion Device is intervertebral spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The InterFuse L device is to be used in patients who have had at least six (6) months of non-operative treatment. These patients may have had a previous non-fusion surgery at the involved spinal level(s). The InterFuse L device is indicated for use with autogenous bone graft, and to be used with supplemental internal spinal fixation systems that have been cleared for use by the FDA in the lumbosacral spine.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
Anton E. Dmitriey, PhD Division of Orthopedic Devices
§ 888.3080 Intervertebral body fusion device.
(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.