K Number
K131213
Date Cleared
2013-06-21

(53 days)

Product Code
Regulation Number
892.1550
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SonoScape S40 device is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic(neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (neonatal and adult), OB/Gyn and Urology.

Device Description

The SonoScape S40 Digital Color Doppler Ultrasound System is an integrated preprogrammed color ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications. The all digital architecture with progressive dynamic receive focusing allows the system to maximize the utility of all imaging transducers to enhance the diagnostic utility and confidence provided by the system. The exam dependent default setting allows the user to have minimum adjustment for imaging the patient, while the in-depth soft-menu control allows the advanced user to set the system for different situations. The architecture allows cost-effective system integration to a variety of upgrade-able options and features. This SonoScape system is a general purpose, software controlled, diagnostic ultrasound system. Its basic function is to acquire ultrasound data and display the image in B-Mode (including Tissue Harmonic Image), M-Mode, TDI, Color-Flow Doppler, Pulsed Doppler and Power Doppler, or a combination of these modes, 3D/4D.

AI/ML Overview

This document describes the SonoScape S40 Digital Color Doppler Ultrasound System and its substantial equivalence to a predicate device (SonoScape S6). The information provided is primarily for regulatory submission and does not detail a clinical study with an acceptance criterion for device performance in the way a clinical trial for an AI algorithm would. Instead, it focuses on technical specifications and compliance with safety standards to demonstrate substantial equivalence.

Based on the provided text, a "study" in the sense of a clinical performance study with human readers or standalone algorithm performance, as typically found in AI/ML device submissions, was not conducted. The study supporting this device's acceptance is a bench testing and comparison study against a predicate device and relevant safety standards.

Here's a breakdown of the requested information based on the provided 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are primarily defined by the performance characteristics of the predicate device (SonoScape S6) and compliance with international safety and performance standards. The "reported device performance" is the comparison of the SonoScape S40 against the S6.

Feature / ParameterAcceptance Criteria (Predicate Device S6 Performance)Reported Device Performance (SonoScape S40)Remark
Intended UseGeneral-purpose ultrasonic imaging for Fetal, Abdominal, Pediatric, Small Organ, Cephalic, Trans-rectal, Trans-vaginal, Peripheral Vascular, Musculo-skeletal, Cardiac, OB/Gyn, Urology.SameSame (Substantially Equivalent)
ClassificationSame as predicate: Ultrasonic Pulsed Doppler Imaging System, Ultrasonic Pulsed Echo Imaging System, Diagnostic Ultrasound Transducer; Class II; Product Code IYN, IYO, ITX.SameSame (Substantially Equivalent)
Supported ProbesSpecific list of Linear, Curved, Micro-curved, and Phased Array probes.Includes new L752, C353, C322, 5P2, 8P1 probes, and different configuration of Phased Array probes (2P2, 3P1). Handles from 2 to 15 MHz.Differences noted, but all comply with IEC 60601-2-37. Considered Substantially Equivalent in safety and effectiveness; no new risk. New probes potentially offer improved image quality.
Channel Beam Former64 channel full digital beam former.128 channel full digital beam former.S40 has better image quality due to 128 channels. Considered Substantially Equivalent in safety and effectiveness.
Image Sector Size32 lines to full B (256 lines).32 lines to full B (512 lines).S40 has better image quality due to 512 lines. Considered Substantially Equivalent in safety and effectiveness.
Display DepthMax 32.9 cm; (Probe depend); Error range: ±3%.Same (Max 32.9 cm; (Probe depend); Error range: ±3%).Same (Substantially Equivalent)
Distance Measurement0~31.0 cm; Error range: ±3%.Same (0~31.0 cm; Error range: ±3%).Same (Substantially Equivalent)
Area MeasurementMax. ≥855 cm²; Error range: ±7%.Same (Max. ≥855 cm²; Error range: ±7%).Same (Substantially Equivalent)
Angle Measurement10~193°; Error range: ±3%.Same (10~193°; Error range: ±3%).Same (Substantially Equivalent)
Circumference Measurement200 cm; Error range: ±3%.Same (200 cm; Error range: ±3%).Same (Substantially Equivalent)
Volume MeasurementMax. 25000 cm³; Error range: ±10%.Same (Max. 25000 cm³; Error range: ±10%).Same (Substantially Equivalent)
M-Mode Time2,4,6,8 S; Error range: ±1%.Same (2,4,6,8 S; Error range: ±1%).Same (Substantially Equivalent)
Heart Rate8 ~ 1000 beats/sec; Error range: ±3%.Same (8 ~ 1000 beats/sec; Error range: ±3%).Same (Substantially Equivalent)
Slope1300 cm/s; Error range: ±3%.Same (1300 cm/s; Error range: ±3%).Same (Substantially Equivalent)
Velocity (PW)0.04-2940 cm/s; Angle ≤60°, ≤5% error.Same (0.04-2940 cm/s; Angle ≤60°, ≤5% error).Same (Substantially Equivalent)
Velocity (CW)0.13-3529 cm/s; Angle ≤60°, ≤5% error.0.12-3795 cm/s; Angle ≤60°, ≤5% error.S40 has better/wider velocity range. Considered Substantially Equivalent in safety and effectiveness.
Velocity (Color)2-226 cm/s; Angle ≤60°, ≤5% error.1-298 cm/s; Angle ≤60°, ≤5% error.S40 has better/wider velocity range. Considered Substantially Equivalent in safety and effectiveness.
Acoustic OutputTrack 3:MI, TIS, TIC, TIB (Derated Ispta: 720Mw/cm2 maximum, TIS/TIB/TIC: 6.0 Maximum, Mechanical Index: 1.9 Maximum, or Derated Isppa: 190W/cm2 max).Same (Track 3:MI, TIS, TIC, TIB Derated ispta: 720Mw/cm2 maximum, TIS/TIB/TIC: 6.0 Maximum, Mechanical Index: 1.9 Maximum, or Derated Isppa: 190W/cm2 max).Same (Substantially Equivalent) - Measured and calculated per NEMA UD 2: 2004.
Electrical SafetyComplies with IEC 60601-1.Complies with IEC 60601-1.Same (Substantially Equivalent)
EMCComplies with IEC 60601-1-2.Complies with IEC 60601-1-2.Same (Substantially Equivalent)
Performance (Safety)Complies with IEC 60601-2-37.Complies with IEC 60601-2-37.Same (Substantially Equivalent)
BiocompatibilityComplies with ISO 10993-5, ISO 10993-10.Complies with ISO 10993-5, ISO 10993-10.Same (Substantially Equivalent)
Level of Concern of SoftwareModerate.Moderate.Same (Substantially Equivalent)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated as a "test set" in the context of a clinical performance study on patients. The testing involved laboratory testing to verify design specifications and compliance with standards. It's implied that various transducers were tested for their acoustic output and safety characteristics.
  • Data Provenance: The testing was conducted internally by SonoScape Company Limited ("Laboratory testing was conducted to verify that the S40 system with added transducer met all design specification..."). The country of origin for the data would be China, where SonoScape is based.
  • Retrospective or Prospective: This was likely prospective bench testing and engineering validation against predefined specifications and standards, rather than retrospective or prospective clinical data collection from patients.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • This type of information is not provided in the 510(k) summary. The submission does not describe a clinical study requiring human expert ground truth for interpretation of images. The ground truth for the technical measurements and safety compliance would be established by validated test equipment and calibration standards, assessed by qualified engineers and technicians, not clinical experts for diagnostic accuracy.

4. Adjudication Method for the Test Set

  • Not applicable. Given that no clinical test with human experts interpreting results was conducted, an adjudication method like 2+1 or 3+1 is not relevant to this submission.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

  • No, an MRMC comparative effectiveness study was not done. The summary explicitly states: "No clinical testing was required." (Section 8). The comparison was primarily technical specifications against a predicate device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

  • Not applicable / No. This device is a diagnostic ultrasound system, not an AI algorithm intended for standalone diagnostic interpretation. Its performance is evaluated based on its ability to acquire and display ultrasound data accurately and safely, matching or improving upon the predicate device's technical specifications.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

  • The "ground truth" for the acceptance of this device relies on:
    • Predicate Device Performance: The established performance characteristics and safety profile of the legally marketed SonoScape S6 ultrasound system.
    • International Safety and Performance Standards: Compliance with standards like IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, NEMA UD 2, NEMA UD3, ISO 10993-5, and ISO 10993-10.
    • Bench Test Results: Direct measurements of technical parameters (e.g., measurement accuracy for distance, area, velocity, acoustic output) using calibrated equipment.

8. The Sample Size for the Training Set

  • Not applicable. This device is a hardware and software system, not an AI/ML algorithm that is "trained" on a specific dataset. Its development involves standard engineering design, testing, and validation processes.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. As there is no "training set" in the AI/ML sense, there is no ground truth established for one. The device's functionality is based on established ultrasound physics and engineering principles.

{0}------------------------------------------------

510(k) Submission

510(k) Summary

JUN 2 1 2013

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(k) number: K131213

    1. Date of submission: April 26, 2013
    1. Submitter

SonoScape Company Limited Address: Yizhe Building, Yuquan Road,Nanshan, Shenzhen 518051, P.R.China Tel: (86) 755-26722890 Fax: (86) 755-26722850 Contact Person: Toki Wu E-mail:Faith@sonoscape.net/wusq@sonoscape.net

3. Proposed Device Identification

Trade/Proprietary Name: S40 Digital Color Doppler Ultrasound System Common Name: Diagnostic Ultrasound System and Transducers Classification:

21 FR 892.1550 Ultrasonic Pulsed Doppler Imaging System (90-IYN)

21 FR 892.1560 Ultrasonic Pulsed Echo Imaging System (90-IYO)

21 CFR 892.1570 Diagnostic Ultrasound Transducer (90-ITX)

Classification Panel: Radiology

Device Class: Il

4. Legally Marketed Predicate Device

SonoScape Company Limited, Diagnostic Ultrasound System, Model S6 has been cleared by FDA through 510(k) No.K112602 (Decision Date - November 07, 2011).

{1}------------------------------------------------

5. Device Description

The SonoScape S40 Digital Color Doppler Ultrasound System is an integrated preprogrammed color ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications.

The all digital architecture with progressive dynamic receive focusing allows the system to maximize the utility of all imaging transducers to enhance the diagnostic utility and confidence provided by the system. The exam dependent default setting allows the user to have minimum adjustment for imaging the patient, while the in-depth soft-menu control allows the advanced user to set the system for different situations. The architecture allows cost-effective system integration to a variety of upgrade-able options and features.

This SonoScape system is a general purpose, software controlled, diagnostic ultrasound system. Its basic function is to acquire ultrasound data and display the image in B-Mode (including Tissue Harmonic Image), M-Mode, TDI, Color-Flow Doppler, Pulsed Doppler and Power Doppler, or a combination of these modes, 3D/4D.

6. Intended Use Statement

The SonoScape S40 device is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic(neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (neonatal and adult), OB/Gyn and Urology.

Transducer Information is shown in the table 1.

No.ProbeTypeFrequencyRangeIntended Use
1C344curved Array2.0-5.0 MHzFetal / Abdominal/ Ob/GYN
2C353curved Array2.0-6.0 MHzFetal / Abdominal/ Ob/GYN
3C322curved Array2.0-6.0 MHzFetal / Abdominal/ Ob/GYN
4VC6-2curved Array2.0-6.0 MHzFetal / Abdominal/ Ob/GYN
56V1Micro-curvedArray4.0-8.0 MHzTrans-rectalTrans-vaginal
66V3Micro-curved5.0-9.0 MHzTrans-rectal

Table 1 Transducer Information

510(k) Summary

{2}------------------------------------------------

510(k) Submission

No.ProbeTypeFrequencyRangeIntended Use
ArrayTrans-vaginal
7L741Linear Array5.0-10.0 MHzSmall Organ (breast, thyroid, testes)Musculo-skeletal (Conventional)Peripheral vessel
8L742Linear Array5.0-12.0 MHzSmall Organ (breast, thyroid, testes)Musculo-skeletal (Conventional)Musculo-skeletal (Superficial)Peripheral vessel
9L752Linear Array5.0-12.0 MHzSmall Organ (breast, thyroid, testes)Musculo-skeletal (Conventional)Musculo-skeletal (Superficial)Peripheral vessel
102P2Phase Array1.0-5.0 MHzAbdominalCephalic(neonatal and adult)Cardiac (neonatal and adult)
113P1Phase Array1.0-5.0 MHzAbdominalCephalic(neonatal and adult)Cardiac (neonatal and adult)
125P2Phase Array3.0-8.0 MHzPediatricNeonatal CephalicCardiac Pediatric
138P1Phase Array4.0-12.0 MHzPediatricNeonatal CephalicCardiac Pediatric

7. Testing

Laboratory testing was conducted to verify that the S40 system with added transducer met all design specification and was substantially equivalent to the Predicate Device. The device has been found to conform to applicable medical device safety standards in regards to thermal, mechanical and electrical safety as well as biocompatibility. The acoustic output is measured and calculated per "NEMA UID 2: 2004 Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment".

510(k) Summary

{3}------------------------------------------------

StandardsNo.Standards TitleVersionDate
IEC 60601-1Medical Electrical Equipment - Part1.General Requirements for Safety1988+A1:1991+A2:199510/31/2005
IEC60601-1-2Medical Electrical Equipment, Part 1-2:General Requirements for Safety -Collateral Standard: ElectromagneticCompatibility - Requirements and Tests200703/01/2007
IEC60601-2-37Medical Electrical Equipment, Part 2-37:Particular Requirements for the Safety ofUltrasonic Medical Diagnostic andMonitoring Equipment200708/01/2007
NEMA UD 2Acoustic Output Measurement Standard forDiagnostic Ultrasound Equipment Version 3200401/01/2004(R 2009)
NEMA UD3Standard for Real-Time Display of Thermaland Mechanical Acoustic Output Indices onDiagnostic Ultrasound Equipment200401/01/2004(R 2009)
ISO 10993-5Biological evaluation of medical devices -Part 5: Tests for In Vitro cytotoxicity199905/15/1999
ISO10993-10Biological evaluation of medical devices -Part 10: Tests for irritation and delayed-typehypersensitivity200209/01/2002

Tab 2 Applicable Safety Standards

8. Clinical Test:

No clinical testing was required.

9. Comparison Table

The differences between the S40 and the predicate device S6 in almost every part are listed in the tables below.

{4}------------------------------------------------

IDItemsProposed DeviceSonoScape S40Predicate DeviceSonoScape S6Remark
1Intended UseThe device is ageneral-purpose ultrasonicimaging instrumentintended for use by aqualified physician forevaluation of Fetal,Abdominal, Pediatric, SmallOrgan (breast, testes,thyroid), Cephalic (neonataland adult), Trans-rectal,Trans-vaginal, PeripheralVascular, Musculo-skeletal(Conventional andSuperficial), Cardiac(neonatal and adult),Urology and OB/Gyn.The device is ageneral-purpose ultrasonicimaging instrumentintended for use by aqualified physician forevaluation of Fetal,Abdominal, Pediatric, SmallOrgan (breast, testes,thyroid), Cephalic(neonataland adult), Trans-rectal,Trans-vaginal, PeripheralVascular, Musculo-skeletal(Conventional andSuperficial), Cardiac(neonatal and adult),OB/Gyn and Urology.Same

Table 3 Intended Use Comparison

·

Table 4 General Comparison

IDComparison ItemsProposed DeviceSonoScape S40Predicate DeviceSonoScape S6Remark
2Classification NameUltrasonic Pulsed DopplerImaging SystemUltrasonic Pulsed EchoImaging SystemDiagnostic UltrasoundTransducerUltrasonic Pulsed DopplerImaging SystemUltrasonic Pulsed EchoImaging SystemDiagnostic UltrasoundTransducerSame
3Product Code90-IYN/90-IYO/90-ITX90-IYN/90-IYO/90-ITXSame
4Regulation Number892.1550/892.1560/892.1570892.1550/892.1560/892.1570Same
5PanelRadiologyRadiologySame
6ClassIIIISame
7Probe Type &ConnectorsL741 Linear Array, 5.0-10.0 MHzL742 Linear Array, 5.0-12.0 MHzL752 Linear Array, 5.0-12.0 MHzL741 Linear Array, 5.0-10.0 MHzL742 Linear Array, 5.0-12.0 MHzL743 Linear Array, 5.0-10.0 MHzSEAnalysis1

510(k) Summary

8-5

{5}------------------------------------------------

510(k) Submission

IDComparisonItemsProposed DeviceSonoScape S40Predicate DeviceSonoScape S6Remark
C344 Curved Array, 2.0-5.0MHzC322 Curved Array, 2.0-6.0MHzC353 Curved Array, 2.0-6.0MHzC344 Curved Array, 2.0-5.0MHzC362 Curved Array, 2.0-6.0MHz
C611 Micro-curved Array,4.0-8.0 MHz
VC6-2 Curved Array, 2.0-6.0MHz6V1 Micro-curved Array,4.0-8.0 MHz6V3 Micro-curved Array,5.0-9.0 MHzVC6-2 Curved Array, 2.0-6.0MHz6V1 Micro-curved Array,4.0-8.0 MHz6V3 Micro-curved Array,5.0-9.0 MHzEC9-5 Micro-curved Array,5.0-9.0 MHz
2P2 Phased Array, 1.0-5.0MHz3P1 Phased Array, 1.0-5.0MHz5P2 Phased Array, 3.0-8.0MHz8P1 Phased Array, 4.0-12.0MHzMulti-port connector connects2P1 Phased Array, 2.0-4.0MHz5P1 Phased Array, 4.0-7.0MHzMulti-port connector connects
4 transducers2 transducers
8AcousticTrackTRACK 3TRACK 3Same

Table 5 Functions Comparison

IDComparisonItemsProposed DeviceSonoScape S40Predicate DeviceSonoScape S6Remark
9Based on an embedded Linuxoperating system.Based on an embedded Linuxoperating system.Same
DesignBased on a 128 channel fulldigital beam former.Based on a 64 channel fulldigital beam former.SEAnalysis2
Autocorrelation for colorprocessing and FFT for pulseAutocorrelation for colorprocessing and FFT for pulseSame
IDComparisonItemsProposed DeviceSonoScape S40and CW Doppler processing.Predicate DeviceSonoScape S6and CW Doppler processing.Remark
Supporting Linear, Curve linearand Phase array probes from 2to 15 MHz.Supporting Linear, Curve linearand Phase array probes from 2to 15 MHz.Same
Cine play back capabilityCine play back capabilitySame
Image file archiveImage file archiveSame
Software upgrade with USBflash drive.Software upgrade with USBflash drive.Same
Digital Scan Converter800×600Digital Scan Converter800×600Same
With touch panelWith full keyboardSEAnalysis3
TGC 8 sliderTGC 8 sliderSame
Depth Range: 3 to 32 cmDepth Range: 3 to 32 cmSame
Image sector size: 32 lines tofull B (512 lines)Image sector size: 32 lines tofull B (256 lines)SEAnalysis4
Image Sector position: Steeringwithin full maximumImage Sector position: Steeringwithin full maximumSame
B orientation flip :L/R key withmarking on the screenB orientation flip :L/R key withmarking on the screenSame
B Dynamic range control:preset 14 curves over 140 dBB Dynamic range control:preset 14 curves over 140 dBSame
Gray Scale Control: 7 SettingsGray Scale Control: 7 SettingsSame
Focal Number: 12 focal zonesettingFocal Number: 12 focal zonesettingSame
10OperationControlsB persistence: 0-95%B persistence: 0-95%Same
Image Processing: Smoothing,edge enhancementImage Processing: Smoothing,edge enhancementSame
PW sweeping speed 2,4,6,8sec over displayPW sweeping speed 2,4,6,8sec over displaySame
PW Wall filter setting:16settings,25 to 750 HZPW Wall filter setting:16settings,25 to 750 HZSame
PW sample volume:0.5 to20mmPW sample volume:0.5 to20mmSame
PW/B update: with UPDATEkeyPW/B update: with UPDATEkeySame
PW cursor steering: Steer softkeyPW cursor steering: Steer softkeySame
PW angle correction:0 to 80degree user controlPW angle correction:0 to 80degree user controlSame
IDComparison ItemsProposed Device SonoScape S40Predicate Device SonoScape S6Remark
PW spectrum dynamic range:10 preset curve over 15-48 dBPW spectrum dynamic range:10 preset curve over 15-48 dBSame
Spectrum baseline shift and invertSpectrum baseline shift and invertSame
Color ROI setting: trackball and set key to control size and positionColor ROI setting: trackball and set key to control size and positionSame
Color steering on flat probe: $\pm$ 20, $\pm$ 16,0Color steering on flat probe: $\pm$ 20, $\pm$ 16,0Same
Color Wall Filter: Color wall filter with 16 selection, 25-750 of PRFColor Wall Filter: Color wall filter with 16 selection, 25-750 of PRFSame
Color priority-B priority soft menuColor priority-B priority soft menuSame
Color Packet size: preset per Exam, horizontal, vertical, offColor Packet size: preset per Exam, horizontal, vertical, offSame
Zoom adjustableZoom adjustableSame
Freeze control: Toggling freeze keyFreeze control: Toggling freeze keySame
Cine control: step, play backward, play continuouslyCine control: step, play backward, play continuouslySame
11Operation ModeB, M, PW, CW, CFM, DPI, TDI, Tissue Harmonic Image 3D/4D Mode Color M ModeB, M, PW, CW, CFM, DPI, TDI, Tissue Harmonic Image 3D/4D Mode Color M ModeSame
12Display ModesDual B, Quad Display, B and M, B and Doppler B + Color, Dual B(Flow) Triplex mode: B,CFM, and PW/CW; B,DPI, and PW/CW;B,THI and Color M, steer M Dual B and Color in real time Compound Imaging, Panoramic Imaging, Trapezoid Imaging.Dual B, Quad Display, B and M, B and Doppler B + Color, Dual B(Flow) Triplex mode: B,CFM, and PW/CW; B,DPI, and PW/CW;B,THI and Color M, steer M Dual B and Color in real time Compound Imaging, Panoramic Imaging, Trapezoid Imaging.Same
13Measurement ItemsDistance; area; circumference; calipers; volume, velocity, HR, PI, RI. Cardiac. OB/GYN, Urology, Vascular and smallDistance; area; circumference; calipers; volume, velocity, HR, PI, RI. Cardiac. OB/GYN, Urology, Vascular and smallSame
IDComparisonItemsProposed DeviceSonoScape S40Predicate DeviceSonoScape S6Remark
part packagepart package
14Cine LoopAutomatic review/ manualreviewAutomatic review/ manualreviewSame
Review speed can be adjustedReview speed can be adjustedSame

510(k) Summary

8-6

{6}------------------------------------------------

:

:

·

·


510(k) Submission

510(k) Summary

.

8-7

.

{7}------------------------------------------------

510(k) Submission .

510(k) Summary

:

.

8-8

.

{8}------------------------------------------------

510(k) Submission

Table 6 Specifications Comparison

.

IDComparisonItemsProposed DeviceSonoScape S40Predicate DeviceSonoScape S6Remark
15PowerSupplyVoltage: 110-127/220-240 VACFrequency: 50/60 HzPower Consumption: 450VAVoltage: 110-240 VACFrequency: 50/60 HzPower Consumption: 110-240V AC, 2.7-1.2ASEAnalysis5
16OperatingConditionTemperature: 1040°CRelative humidity: 3075%Air pressure: 700hPa ~1060hPaTemperature: 1040°CRelative humidity: 3075%Air pressure: 700hPa ~1060hPaSame
17StorageConditionTemperature: -2055°CRelative humidity: 2090%Air pressure: 700hPa ~1060hPaTemperature: -2055°CRelative humidity: 2090%Air pressure: 700hPa ~1060hPaSame
18Screen Size18.5 inch Widescreen LCDmonitor15 inch Widescreen LCD monitorSEAnalysis6
19MeasurementAccuracyParameterValue rangeErrorrangeParameterValue rangeErrorrange
DisplaydepthMax 32.9 cm; (Probedepend)±3%DisplaydepthMax 32.9 cm; (Probedepend)±3%Same
Distance0~31.0 cm±3%Distance0~31.0 cm±3%Same
AreaMax. ≥855 cm²±7%AreaMax. ≥855 cm²±7%Same
Angle10~193°±3%Angle10~193°±3%Same
Circumference200 cm±3%Circumference200 cm±3%Same
VolumeMax. 25000 cm³±10%VolumeMax. 25000 cm³±10%Same
M-Modetime2,4,6,8 S±1%M-Modetime2,4,6,8 S±1%Same
HeartRate8 ~ 1000beats/sec±3%HeartRate8 ~ 1000beats/sec±3%Same
Slope1300 cm/s±3%Slope1300 cm/s±3%Same

510(k) Summary

{9}------------------------------------------------

510(k) Submission

IDComparison ItemsProposed Device SonoScape S40Predicate Device SonoScape S6Remark
VelocityAngleVelocityAngle
Velocity (PW)0.04-2940 cm/s≤60°, ≤5%0.04-2940 cm/s≤60°, ≤5%Same
Velocity (CW)0.12-3795 cm/s≤60°, ≤5%0.13-3529 cm/s≤60°, ≤5%SE Analysis 7
Velocity (Color)1-298 cm/s≤60°, ≤5%2-226cm/s≤60°, ≤5%
20Acoustic OutputTrack 3:MI, TIS, TIC, TIBDerated ispta: 720Mw/cm2 maximum.TIS/TIB/TIC: 6.0 Maximum,Mechanical Index: 1.9 Maximum,or Derated Isppa: 190W/cm2 max
Same

Table 7 Safety Comparison

IDComparison ItemsProposed DeviceSonoScape S40Predicate DeviceSonoScape S6Remark
21Electrical Safety-IEC 60601-1-IEC 60601-1Same
22EMC-IEC 60601-1-2-IEC 60601-1-2Same
23Performance-IEC 60601-2-37-IEC 60601-2-37Same
24Biocompatibility-ISO 10993-5,-ISO 10993-10-ISO 10993-5,-ISO 10993-10Same
25Level of ConcernOf SoftwareModerate level ofconcern systemModerate level ofconcern systemSame

SE Analysis 1:

Probe Type & Connectors, Compare to the predicate device, the proposed device is with different probe type or frequency, such as L752, C353 etc. But no new intended use is added and all of them comply with IEC 60601-2-37; And the proposed device has 4 probe connection ports and the predicate device has 2 probe connection ports, but both of them comply with IEC 60601-1 and IEC 60601-1-2, therefore they can be considered Substantially Equivalent in safety and effectiveness, and no new risk is raised, so the SE is not affected.

{10}------------------------------------------------

SE Analysis 2: ·

The proposed device is based on a 128 channel full digital beam former and the predicate device is 64, but the 128 channel is better than 64 channel in terms of the image quality. Therefore, they can be considered Substantially Equivalent in safety and effectiveness. So the SE is not affected.

SE Analysis 3:

The predicate device is with full keyboard and the proposed device is not, but both of them comply with IEC 60601-1 and IEC 60601-1-2. Therefore, they can be considered Substantially Equivalent in safety and effectiveness. So the SE is not affected.

SE Analysis 4:

The proposed device is with 512 lines image sector size and the predicate device is 256, but the 512 lines is better than 256 lines in terms of the image quality. Therefore, they can be considered Substantially Equivalent in safety and effectiveness. So the SE is not affected.

SE Analysis 5

The Power Supply of the proposed device and the predicate device are 110-127/220-240 VAC, 450VA and 110-240VAC, 2.7-1.2A respectively, but both of them comply with IEC60601-1 and IEC 60601-1-2. Therefore, power supply can be considered Substantially Equivalent in safety and effectiveness.

SE Analysis 6

The screen size of the proposed is larger than that of the S6. This difference is considered to have no effect on effectiveness and safety.

SE Analysis 7:

The proposed device and the predicate device are with different measurement accuracy in Velocity (CW/ Color), but the proposed device is better. Therefore, they can be considered Substantially Equivalent in safety and effectiveness. So the SE is not affected.

Discussion of Substantially Equivalent

The subject device has same intended use, similar product design, and same performance effectiveness, performance safety as the predicate device. The differences above between the subject device and predicate device do not affect the basic design

8-11

{11}------------------------------------------------

K131213
Page 12 of 12

510(k) Submission

principle, usage, effectiveness and safety of the subject device. And no question is raised regarding to effectiveness and safety.

10. Substantially Equivalent Conclusion

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, SonoScape Company Limited concludes that S40 Digital Color Doppler Ultrasound System is substantially equivalent to predicate devices with regard to safety and effectiveness.

510(k) Summary

{12}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/12/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird-like figure with three wing-like shapes.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 21, 2013

SonoScape Company Limited % Ms. Toki Wu Yizhe Building, Yuquan Road, NanShan Shenzhen, Guangdong 518051 P.R. CHINA

Re: K131213

Trade/Device Name: S40 Digital Color Doppler Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: Class II Product Code: IYN, IYO, and ITX Dated: April 26, 2013 Received: May 2, 2013

Dear Ms. Wu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the S40 Digital Color Doppler Ultrasound System, as described in your premarket notification:

Transducer Model Number

2P2 Phase Array 8P1 Phase Array C344 Curved Array VC6-2 Curved Array 3PI Phase Array 6V1 Micro-curved Array C353 Curved Array L741 Linear Array L752 Linear Array

5P2 Phase Array 6V3 Micro-curved Array C322 Curved Array L742 Linear Array

{13}------------------------------------------------

Page 2 - Ms. Toki Wu

If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050;

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/About/DA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH) s) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

If you have any questions regarding the content of this letter, please contact Shing Chun Benny Lam, Ph.D. at (301) 796-9328.

Sincerely yours,

for

Janine M. Morris Director, Division of Radiological Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosures

{14}------------------------------------------------

Indications for Use

510(k) Number:K131213
Device Name:S40 Digital Color Doppler Ultrasound System
Indications for Use:The SonoScape S40 device is a general-purpose ultrasonicimaging instrument intended for use by a qualifiedphysician for evaluation of Fetal, Abdominal, Pediatric,Small Organ (breast, testes, thyroid), Cephalic(neonataland adult), Trans-rectal, Trans-vaginal, PeripheralVascular, Musculo-skeletal (Conventional and Superficial),Cardiac (neonatal and adult), OB/Gyn and Urology.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use _ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

$m(x-h)

(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health

K131213 510(k) _

Indications for Use

Page 1 of 15

{15}------------------------------------------------

System: SonoScape S40

Diagnostic Ultrasound Pulsed Echo System

Diagnostic Ultrasound Pulsed Doppler Imaging System

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalNNNNNNote 1Notes 2,4,5
AbdominalNNNNNNote 1Notes 2,4,5
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
PediatricNNNNNNote 1Notes 2,4
Small Organ (specify)NNNNNNote 1Notes 2,4,6
Neonatal CephalicNNNNNNNote 1Notes 2,3,4
FetalAdult CephalicNNNNNNNote 1Notes 2,3,4
Imaging&Trans-rectalNNNNNNote 1Notes 2,4
OtherTrans-vaginalNNNNNNote 1Notes 2,4
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)NNNNNNote 1Notes 2,4
Musculo-skeletal(Superficial)NNNNNNote 1Notes 2,4
Intravascular
Other (Ob/GYN)NNNNNNote 1Notes 2,4,5
Cardiac AdultNNNNNNNote 1Notes 2,3,4
Cardiac PediatricNNNNNNNote 1Notes 2,3,4
CardiacIntravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vesselNNNNNNote 1Notes 2,4
VesselOther (specify)

E = added under this appendix P = previously cleared by FDA; N = new indication; Note 1: Other Combined includes: BM; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health

510(k) _

Indications for Use

Page 2 of 15

{16}------------------------------------------------

Transducer: 2P2 Phase Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalImaging&OtherFetal
AbdominalNNNNNNote 1Notes 2,4
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal CephalicNNNNNNNote 1Notes 2,3,4
Adult CephalicNNNNNNNote 1Notes 2,3,4
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
CardiacCardiac AdultNNNNNNNote 1Notes 2,3,4
Cardiac PediatricNNNNNNNote 1Notes 2,3,4
Intravascular(Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Other (specify)

P = previously cleared by FDA; E = added under this appendix N = new indication; Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health

510(k) _

Indications for Use

Page 3 of 15

{17}------------------------------------------------

Transducer: 3P1 Phase Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude) DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalImaging&OtherFetal
AbdominalNNNNNNote 1Notes 2,4
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal CephalicNNNNNNNote 1Notes 2,3,4
Adult CephalicNNNNNNNote 1Notes 2,3,4
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
CardiacCardiac AdultNNNNNNNote 1Notes 2,3,4
Cardiac PediatricNNNNNNNote 1Notes 2,3,4
Intravascular(Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Other (specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health

510(k) __

Indications for Use

Page 4 of 15

{18}------------------------------------------------

Transducer: 5P2 Phase Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude) DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalImaging&OtherFetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
PediatricNNNNNNote 1Notes 2,4
Small Organ (specify)
Neonatal CephalicNNNNNNNote 1Notes 2,3,4
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
CardiacCardiac Adult
Cardiac PediatricNNNNNNNote 1Notes 2,3,4
Intravascular(Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Other (specify)

P = previously cleared by FDA; E = added under this appendix N = new indication; Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color

Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TD! Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health

510(k) __

Indications for Use

Page 5 of 15

{19}------------------------------------------------

Transducer: 8P1 Phase Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude) DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalImaging&OtherFetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
PediatricNNNNNNote 1Notes 2,4
Small Organ (specify)
Neonatal CephalicNNNNNNNote 1Notes 2,3,4
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
CardiacCardiac Adult
Cardiac PediatricNNNNNNNote 1Notes 2,3,4
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Other (specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppier; B/Color

Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health

510(k) __

Indications for Use

Page 6 of 15

{20}------------------------------------------------

Transducer: 6V1 Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther*
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalImaging&OtherFetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPNote 1Notes 2,4
Trans-vaginalPPPPPNote 1Notes 2,4
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Other (specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color

Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 5: 4D Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health

510(k) _

Indications for Use

Page 7 of 15

{21}------------------------------------------------

Transducer: 6V3 Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther*
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerCombinedSpecify
OphthalmicOphthalmic
FetalImaging&OtherFetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPNote 1Notes 2,4
Trans-vaginalPPPPPNote 1Notes 2,4
Trans-urethral
Trans-esoph. (non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Other (specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color

Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health

510(k) _

Indications for Use

Page 8 of 15

{22}------------------------------------------------

Transducer: C344 Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalImaging&OtherFetalPPPPPNote 1Notes 2,4
AbdominalPPPPPNote 1Notes 2,4
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)PPPPPNote 1Notes 2.4
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Other (specify)

P = previously cleared by FDA; E = added under this appendix N = new indication; Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD; B/Power Doppter/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health

510(k) _

Indications for Use

Page 9 of 15

{23}------------------------------------------------

Transducer: C353 Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther*
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerCombinedSpecify
OphthalmicOphthalmic
FetalImaging&OtherFetalNNNNNNote 1Notes 2,4
AbdominalNNNNNNote 1Notes 2,4
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)NNNNNNote 1Notes 2,4
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Other (specify)

N = new indication; Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD: B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D .

Note 6: Small Organ: breast, thyroid, testes

(Division Sign Off) Division of Radiological Health Office of in Vitro Diagnostic and Radiological Health

510(k) _

Indications for Use

Page 10 of 15

{24}------------------------------------------------

Transducer: C322 Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther*
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerCombinedSpecify
OphthalmicOphthalmic
FetalImaging&OtherFetalNNNNNNote 1Notes 2,4
AbdominalNNNNNNote 1Notes 2,4
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)NNNNNNote 1Notes 2,4
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Other (specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health

510(k) _

Page 11 of 15

{25}------------------------------------------------

Transducer: VC6-2 Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
GeneralSpecificColor· PowerOtherOther
(TRACK 1ONLY)(TRACKS 1 & 3)BMPWDCWDDoppler(Amplitude)DopplerCombinedSpecify
OphthalmicOphthalmic
FetalFetalPPPPPNote 1Notes 2,4,5
lmaging &AbdominalPPPPPNote 1Notes 2,4,5
OtherIntra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph (non-Card)
Musculo-skeletal
(Conventional)
Musculo-skeletal
(Superficial)
Intravascular
Other (Ob/GYN)PPPPPNote 1Notes 2,4,5
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselOther (specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color

Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 4: 3D Note 5: 4D Note 3: TDI

Note 6: Small Organ: breast, thyroid, testes

(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health

510(k) __

Indications for Use

Page 12 of 15

{26}------------------------------------------------

Transducer: L741 Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalImaging&OtherFetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,4,6
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)PPPPPPNote 1Notes 2,4
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vesselPPPPPPNote 1Notes 2,4
Other (specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color

Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 4: 3D Note 5: 4D Note 3: TDI

Note 6: Small Organ: breast, thyroid, testes

(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health

510(k) _

Indications for Use

Page 13 of 15

{27}------------------------------------------------

Transducer: L742 Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows;

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalFetal
Imaging&OtherAbdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,4,6
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2,4
Musculo-skeletal(Superficial)PPPPPNote 1Notes 2,4
Intravascular
Other (Ob/GYN)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vesselPPPPPNote 1Notes 2,4
Other (specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 4: 3D Note 5: 4D Note 3: TDI

Note 6: Small Organ: breast, thyroid, testes

(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health

510(k) .

Indications for Use

Page 14 of 15

{28}------------------------------------------------

Transducer: L752 Linear Array

Diagnostic Ultrasound Transducer

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalImaging&OtherFetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)NNNNNNote 1Notes 2,4,6
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)NNNNNNote 1Notes 2,4
Musculo-skeletal(Superficial)NNNNNNote 1Notes 2,4
Intravascular
Other (Ob/GYN)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vesselNNNNNNote 1Notes 2,4
Other (specify)

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

P = previously cleared by FDA; E = added under this appendix N = new indication; Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color

Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 5: 4D Note 3: TDI Note 4: 3D

  • Note 6: Small Organ: breast, thyroid, testes
    (Division Sign Off) . Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health

510(k) ___

Indications for Use

Page 15 of 15

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.