K Number
K131161

Validate with FDA (Live)

Date Cleared
2013-09-17

(146 days)

Product Code
Regulation Number
892.1570
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This device is indicated for use in adult males with known or suspected prostate cancer.

Device Description

The Stepping Unit holds an ultrasound imaging probe and facilitates manual linear and rotational positioning of the probe. The Stepping Unit consists of a Cradle designed to hold a specific ultrasound probe and a Carriage that moves longitudinally along the Stepping Unit. A Grid Platform connected at one end of the Stepper provides support for the Disposable Template Grid. The Stepping Unit connects to the Stabilizer.

The Stabilizer provides a base for the Stepping Unit and offers fixation and support during insertion and final placement of the ultrasound imaging probe. On certain models, a fine tune mechanism is used to micro-adjust the probe or instrument to the ideal orientation. The Stabilizer mounts to an operating room table or is supported by a floor stand.

The Disposable Template Grid is a single-use, sterile grid that consists of rows and columns of holes (channels) spaced 5mm apart. These channels are labeled and provide placement of needles in predefined areas of the prostate.

AI/ML Overview

Here's a summary of the acceptance criteria and study information for the CIVCO Medical Instruments Co., Inc. devices, based on the provided text:

Acceptance Criteria and Device Performance

Acceptance CriteriaReported Device Performance
New EX3 Stepper Specific Criteria:
Stepping Unit is able to provide continuous (free) longitudinal movement.Verified.
Encoders correctly report longitudinal and rotational motion of Stepping Unit.Verified.
Stepping Unit meets electrical safety requirements of IEC 60601-1 and IEC 60601-1-1.Verified.
Device remains safe and effective for its intended use, including with treatment planning software (Validation).Validation was conducted to confirm the device remains safe and effective for its intended use and included testing with treatment planning software. All testing confirmed that the EX3 Stepper is substantially equivalent to the predicate device in regards to safety and effectiveness and the new design features have not diminished the safety and effectiveness of the device.
Existing Devices (Classic Stepper, Multi-Purpose Workstation Stepper, Micro-Touch, Micro-Touch LP, Multi-Purpose Workstation, Multi-Purpose Workstation LP, Disposable Template Grid) with Expanded Indications for Use Criteria:
Safely and effectively used for cryotherapy, transperineal template-guided biopsy, and fiducial marker placement without new safety/effectiveness issues arising from expanded indications.A literature review was conducted to support the modification of the intended use. The literature demonstrates that these devices have been safely and effectively used in cryotherapy, transperineal template-guided biopsy, and fiducial marker placement. "These changes do not raise new issues of safety and effectiveness because the system continues to be used for positioning in males with suspected or diagnosed prostate cancer. These changes do not impact or modify the therapy and CIVCO did not modify the design of these devices to enable use in these additional applications." (Page 1) "No design changes were made to enable use of the devices in these additional applications." (Page 5)
Disp. Template Grid: Sterilization method change (EtO Gas vs. Gamma Irradiation) does not impact safety/effectiveness; Material change (ABS vs. Polycarbonate) does not impact safety/effectiveness; 3-year expiration date is acceptable.Implied by substantial equivalence claim; no specific test results for these changes are provided, but the overall conclusion of substantial equivalence covers these modifications.
Stabilizer: Ability to mount to an operating table (for some models) vs. only a floor stand for predicate (for some Stabilizers). Fine-tune adjustment mechanism on Micro-Touch Stabilizers.Implied by substantial equivalence claim; no specific test results for these changes are provided, but the overall conclusion of substantial equivalence covers these modifications.

Since this is a 510(k) submission for substantial equivalence for a medical device (specifically a stepping unit, stabilizer, and disposable template grid used in prostate procedures), the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are generally focused on demonstrating that the new/modified device is as safe and effective as a legally marketed predicate device. The evidence provided is primarily through:

  • Comparison to Predicate Devices: Showing similarity in technological characteristics and intended use.
  • Literature Review: For expanding indications of use for existing devices.
  • Verification and Validation Testing: Specifically for the new EX3 Stepper, confirming its new features meet specifications and do not compromise overall safety/effectiveness.

Here's the breakdown of the requested information based on the provided text:


2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not explicitly stated. The verification and validation testing for the EX3 Stepper (longitudinal movement, encoder reporting, electrical safety, and compatibility with treatment planning software) would have involved specific test units, but the number of units tested is not detailed in this summary.
  • Data Provenance (e.g., country of origin of the data, retrospective or prospective): Not explicitly stated. The verification and validation testing would have been conducted internally or by a contracted lab, likely in the US (where CIVCO is headquartered). This would be prospective testing on the physical device. The literature review for expanded indications of use would involve published, likely retrospective or prospective, clinical data (of various origins) that has already been peer-reviewed and published.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • This type of information is generally not applicable or explicitly stated for a 510(k) submission concerning a mechanical device like a stepping unit. The "ground truth" for mechanical function, electrical safety, or general utility for specific surgical procedures isn't typically established by expert consensus in the same way as an AI diagnostic algorithm's output would be. Instead, the "ground truth" here is the established safety and efficacy of the predicate device, engineering specifications, and universally accepted standards (e.g., IEC 60601-1 for electrical safety).
  • For the validation with "treatment planning software," it implies functionality was checked against expected outputs, likely by engineers or possibly clinical users, but no explicit "expert" review process is described for establishing a ground truth.

4. Adjudication Method for the Test Set

  • Not applicable/mentioned. Adjudication methods (like 2+1 or 3+1) are typically used in clinical studies, especially for evaluating diagnostic accuracy where there might be disagreements among human readers or between AI and human readers. This submission focuses on the mechanical and electrical performance of hardware, not diagnostic interpretation.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. This study type is highly specific to evaluating the clinical impact of AI-assisted diagnostic tools on human performance, which is not the subject of this 510(k) submission. This device is a mechanical accessory (stepping unit, stabilizer, template grid) used during procedures, not an AI diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This device is a mechanical accessory, not an algorithm or AI system. The EX3 Stepper has "firmware in the encoders report position," which is a functional component, but not a standalone algorithm in the typical sense of AI. Its performance (accuracy of position reporting) was verified.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

  • For EX3 Stepper's new features (encoders, continuous movement, electrical safety): The "ground truth" would be established engineering specifications, measurable physical properties, and adherence to recognized electrical safety standards (IEC 60601-1 and IEC 60601-1-1). For compatibility with treatment planning software, the ground truth would be the expected data output/interaction defined by the software's specifications.
  • For expanded indications of use of existing devices: The "ground truth" for safety and effectiveness in cryotherapy, transperineal template-guided biopsy, and fiducial marker placement was established through a literature review. This means previously published clinical studies and outcomes data effectively served as a form of "ground truth" demonstration of safe and effective use.

8. The Sample Size for the Training Set

  • Not applicable. This is not an AI/machine learning device that requires a training set. The firmware on the EX3 Stepper's encoders is for reporting position, not for learning or prediction from data.

9. How the Ground Truth for the Training Set was Established

  • Not applicable, as there is no training set for this device.

{0}------------------------------------------------

K131161
Page 1 of 6

Image /page/0/Picture/1 description: The image shows the logo for CIVCO Medical Solutions. The logo features a stylized globe on the left, followed by the word "CIVCO" in bold, sans-serif font. Below "CIVCO" are the words "MEDICAL SOLUTIONS" in a smaller, sans-serif font.

WWW.CIVCO.COM

Section 5 – 510(k) Summary

SEP 17 2013

A. Submitter Information

Submitter Name & Address: CIVCO Medical Instruments Co., Inc. d/b/a CIVCO MedicaSolutions102 First Street SouthKalona, lowa 52247
Contact Person:Amanda Stahle, Regulatory Affairs SpecialistTelephone: 319-248-6628, Fax: 877-218-0324amanda.stahle@civco.com
Date Summary Prepared:April 22, 2013
Trade Names:EX3 Stepper™, Classic Stepper, Multi-PurposeWorkstation™ Stepper;Micro-Touch™, Micro-Touch™ LP, Multi-PurposeWorkstation™, Multi-Purpose Workstation™ LP;Disposable Template Grid(These devices are marketed as part of the AccuCARE™product line)
Common Names:Stepping UnitStabilizerDisposable Template Grid
Classification Names:Transducer, Ultrasonic, DiagnosticSystem, Applicator, Radionuclide, ManualSystem, Applicator, Radionuclide, Remote-controlled
Classification Numbers:Class It under 21 CFR 892.1570Class I under 21 CFR 892.5650Class II under 21 CFR 892.5700
Review Panels:Radiology
Product Codes:ITX, IWJ & JAQ

B. Predicate Devices

The subject devices are substantially equivalent to the predicate devices included in the following 510(k)s:

Predicate DevicesMfg.
K972672: Brachystepper Stepping Unit, Brachystand Support andManual Adjustment AccessoryNote: The Brachystepper Needle Guide Template component in this510(k) is not being used as a predicate.Barzell-Whitmore MaroonBells, Inc. (acquired byCIVCO MedicalInstruments Co., Inc.)
Corporate Headquarters102 First Street South, Kalona, IA 52247 USAP: 319.248.6757 F: 319.248.6660
Coralville Office2301 Jones Blvd., Coralville, IA 52241 USAP: 319.248.6757 F: 319.248.6660
Europe OfficePasteurstraat 6, 2811 DX Reeuwijk, The NetherlandsP: +31(0) 182.394495 F: +31(0) 182.395014
Orange City Office1401 8th Street SE, Orange City, IA 51041 USAP: 712.737.8688 F: 712.737.8654

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for CIVCO Medical Solutions. The logo features a stylized globe on the left, followed by the word "CIVCO" in bold, sans-serif font. Below "CIVCO" are the words "MEDICAL SOLUTIONS" in a smaller, sans-serif font with wide spacing between the letters. The overall design is clean and professional.

WWW.CIVCO.COM

Predicate DevicesMfg.
K981873: Brachystepper™ Disposable Template GridBarzell-Whitmore MaroonBells, Inc. (acquired byCIVCO MedicalInstruments Co., Inc.)

The subject devices, with the exception of the EX3 Stepper, have been marketed for many years under K972672 and K981873. The purpose of this 510(k) is to modify the intended use of these devices to include cryotherapy, transperineal template-guided biopsy, and fiducial marker placement. These additional applications fall within the indications for use and the literature demonstrates that these devices have been safely and effectively used for many years in these applications. These changes do not raise new issues of safety and effectiveness because the system continues to be used for positioning in males with suspected or diagnosed prostate cancer. These changes do not impact or modify the therapy and CIVCO did not modify the design of these devices to enable use in these additional applications.

The second purpose of this 510(k) is to provide a premarket notification for the EX3 Stepper. The EX3 Stepper is similar to the predicate in that it holds an ultrasound imaging probe and facilitates manual linear and rotational positioning of the probe. The EX3 Stepper differs from the predicate in that it offers continuous (free) longitudinal movement and contains encoder components that are powered via a USB-hub. Firmware in the encoders report position and this position is read by treatment planning software marketed by other companies. This position reporting function has been added to the EX3 intended use. Verification and validation testing confirmed that these changes in design do not raise new questions of safety and effectiveness.

C. Device Description

Stepping Unit

The Stepping Unit holds an ultrasound imaging probe and facilitates manual linear and rotational positioning of the probe. The Stepping Unit consists of a Cradle designed to hold a specific ultrasound probe and a Carriage that moves longitudinally along the Stepping Unit. A Grid Platform connected at one end of the Stepper provides support for the Disposable Template Grid. The Stepping Unit connects to the Stabilizer.

Stabilizer

The Stabilizer provides a base for the Stepping Unit and offers fixation and support during insertion and final placement of the ultrasound imaging probe. On certain models, a fine tune mechanism is used to micro-adjust the probe or instrument to the ideal orientation. The Stabilizer mounts to an operating room table or is supported by a floor stand.

Disposable Template Grid

The Disposable Template Grid is a single-use, sterile grid that consists of rows and columns of holes (channels) spaced 5mm apart. These channels are labeled and provide placement of needles in predefined areas of the prostate.

Corporate Headquarters102 First Street SouthKalona, IA 52247USA319.248.6757319.248.6660
Coralville Office2301 Jones Blvd.Coralville, IA 52241USA319.248.6757319.248.6660
Europe OfficePasteurstraat 62811 DX ReeuwijkThe Netherlands+31(0) 182.394495+31(0) 182.395014
Orange City Office1401 8th Street SEOrange City, IA 51041USA712.737.8688712.737.8654

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for CIVCO Medical Solutions. The logo features a stylized globe-like graphic on the left, followed by the word "CIVCO" in large, bold letters. Below "CIVCO" are the words "MEDICAL SOLUTIONS" in smaller letters.

www.civco.com

.

The following part numbers are included in this 510(k):

. 1

CommonNameTrade NamePart No.Device Description
SteppingUnitEX3Stepper™609-004EX3 Stepper for use with Best Sonalistransducers
612-228EX3 Stepper for use with Acuson ER7Btransducers
614-119EX3 Stepper for use with Aloka UST-672-5/7.5transducers
620-119EX3 Stepper for use with B-K Medical 8658(8558), 8848 transducers
642-443EX3 Stepper for use with GE Healthcare ERBtransducers
644-081EX3 Stepper for use with Hitachi EUP-U533transducers
676-177EX3 Stepper for use with Siemens Endo P-IItransducers
683-003EX3 Stepper for use with Terason 8B4Stransducers
ClassicStepper614-092Classic Stepper for use with Aloka UST-672-5/7.5transducers
644-064Classic Stepper for use with Hitachi EUP-U533transducers
612-225Classic Stepper for use with Acuson ER7Btransducers
642-316Classic Stepper for use with GE Healthcare ERBtransducers
620-089Classic Stepper for use with B-K Medical 8658(8558) transducers
676-114Classic Stepper for use with Siemens Endo-P IItransducers
609-001Classic Stepper for use with Best Sonalistransducers
Multi-PurposeWorkstation™Stepper614-098Multi-Purpose Workstation Stepper for use withAloka UST-672-5/7.5
642-334Multi-Purpose Workstation Stepper for use withGE Healthcare ERB
644-066Multi-Purpose Workstation Stepper for use withHitachi EUP-U533
676-121Multi-Purpose Workstation Stepper for use withSiemens Endo-P II
620-094Multi-Purpose Workstation Stepper for use with B-K Medical 8658 (8558), 8808
620-113Multi-Purpose Workstation Stepper for use withBK Medical 8808e, 8818
620-117Multi-Purpose Workstation Stepper for use withBK Medical 8848
StabilizerMicro-Touch ™610-911610-911SMicro-Touch (Dual-Sided Table Mount)Micro-Touch Transportation Stand
Corporate Headquarters102 First Street SouthKalona, IA 52247USA319.248.6757319.248.6660
Coralville Office2301 Jones Blvd.Coralville, IA 52241USA319.248.6757319.248.6660
Europe OfficePasteurstraat 62811 DX ReeuwijkThe Netherlands+31(0) 182.394495+31(0) 182.395014
Orange City Office1401 8th Street SEOrange City, IA 51041USA712.737.8688712.737.8654

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for CIVCO Medical Solutions. The logo features a stylized globe on the left, followed by the word "CIVCO" in bold, sans-serif font. Below "CIVCO" are the words "MEDICAL SOLUTIONS" in a smaller font size. The logo is black and white.

WWW.CIVCO.COM
Micro-TouchTMLP610-912Micro-Touch LP (Single-Sided Table Mount-rightside)
610-912SMicro-Touch Transportation Stand
610-913Micro-Touch LP (Single-Sided Table Mount-leftside)
610-913SMicro-Touch Transportation Stand
Multi-PurposeWorkstationTM610-974Multi-Purpose Workstation with adjustable floorstand
Multi-PurposeWorkstationTMLP610-975Multi-Purpose Workstation LP with rail mount
DisposableTemplateGridDisposableTemplate Grid610-905Sterile 17GA Grid for use with AccuCAREPositioning and Stabilizing equipment
610-906Sterile 18GA Grid for use with AccuCAREPositioning and Stabilizing equipment
610-915Sterile 17GA Grid for use with B-K Medical UA-1084 Stepper
610-916Sterile 18GA Grid for use with B-K Medical UA-1084 Stepper

D. Indications for Use

This device is indicated for use in adult males with known or suspected prostate cancer.

E. Intended Use

System Intended Use:

Intended for use in ultrasonic procedures related to brachytherapy, transperineal template-guided biopsy, and/or fiducial marker placement (including volume determination of the prostate gland), and/or the application of radionuclide source(s) into the body during brachytherapy.

Component-specific Intended Uses:

Stepping Unit (Multi-Purpose Workstation Stepper, Classic Stepper): Holding and manipulating ultrasound imaging probes during prostate brachytherapy. transperineal template-guided biopsy, and/or fiducial marker placement procedures (including volume determination of the prostate gland), and/or the application of radionuclide source(s) into the body during brachytherapy.

Stepping Unit (EX3 Stepper): Holding and manipulating ultrasound imaging probes, and reporting position, during prostate brachytherapy, transperineal templatequided biopsy, and/or fiducial marker placement procedures (including volume determination of the prostate gland), and/or the application of radionuclide source(s) into the body during brachytherapy.

Stabilizer: Provides fixation, support and manipulation of transrectal ultrasound imaging probes during insertion and final placement.

Corporate Headquarters102 First Street SouthKalona, IA 52247USA319.248.6757319.248.6660
Coralville Office2301 Jones Blvd.Coralville, IA 52241USA319.248.6757319.248.6660
Europe OfficePasteurstraat 62811 DX ReeuwijkThe Netherlands+31(0) 182.394495+31(0) 182.395014
Orange City Office1401 8th Street SEOrange City, IA 51041USA712.737.8688712.737.8654

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for CIVCO Medical Solutions. The logo features a stylized globe on the left, followed by the word "CIVCO" in a bold, sans-serif font. Below the word "CIVCO" are the words "MEDICAL SOLUTIONS" in a smaller, sans-serif font. The logo is black and white.

WWW.CIVCO.COM

Disposable Template Grid: Accepting and guiding needles up to 1.3 mm (18 gauge) in diameter and providing coordinates as an aid to needle loading and positioning during radioactive seed implantation, cryotherapy, transperineal template-guided biopsy, and/or fiducial marker placement.

F. Technological Characteristics

Stepping Unit

Both the subject and predicate Stepping Units are designed to hold and manipulate ultrasound imaging probes during prostate procedures and enable manual linear and rotational positioning of the probe. The subject and predicate Stepping Units consist of a Cradle designed to hold and rotate an ultrasound probe, a Carriage that moves longitudinally along the Stepping Unit, and a Grid Platform connected at one end of the Stepper to attach the Disposable Template Grid. The proposed EX3 Stepper differs from the predicate in that it offers continuous (free) longitudinal movement of the Carriage and contains encoder components that are powered via a USB-hub. Firmware in the encoders report position and this position is read by treatment planning software marketed by other companies.

Stabilizer

Both the subject and predicate Stabilizers are designed to serve as a base for the Stepping Unit. The subject and predicate Stabilizers offer fixation and support during insertion and final placement of the ultrasound imaging probe. Whereas the predicate device only mounted on a floor stand, several of the subject Stabilizers can be mounted to an operating table. The fine tune adjustment mechanism on the predicate device is incorporated in the Micro-Touch Stabilizers, but not the Multi-Purpose Workstation Stabilizers.

Disposable Template Grid

Both the subject and predicate Disposable Template Grids are single-use, sterile grids that consist of rows and columns of holes (channels) spaced 5mm apart. These channels are labeled and provide placement of needles in predefined areas of the prostate. The subject device is sterilized via EtO gas, whereas the predicate device was sterilized by gamma irradiation. Also, the subject device is made of ABS while the predicate device was constructed from polycarbonate. The subject device is labeled with a 3 year expiration date.

G. Literature Review , Non-Clinical Performance Testing, and Conclusions

A literature review was conducted to support the modification of the intended use to include cryotherapy, transperineal template-guided biopsy, and fiducial marker placement. The literature demonstrates that these devices have been safely and effectively used in cryotherapy, transperineal template-guided biopsy, and fiducial marker placement. No design changes were made to enable use of the devices in these additional applications.

Corporate Headquarters102 First Street SouthKalona, IA 52247USAⓅ 319.248.6757Ⓕ 319.248.6660
Coralville Office2301 Jones Blvd.Coralville, IA 52241USAⓅ 319.248.6757Ⓕ 319.248.6660
Europe OfficePasteurstraat 62811 DX ReeuwijkThe NetherlandsⓅ +31(0) 182.394495Ⓕ +31(0) 182.395014
Orange City Office1401 8th Street SEOrange City, IA 51041USAⓅ 712.737.8688Ⓕ 712.737.8654

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for CIVCO Medical Solutions. The logo features a stylized globe on the left, with curved lines forming a grid pattern. To the right of the globe is the word "CIVCO" in bold, sans-serif font. Below "CIVCO" are the words "MEDICAL SOLUTIONS" in a smaller, sans-serif font.

K13 1161
Page 6 of 6

www.civco.com

Additionally, non-clinical performance testing was conducted on the EX3 Stepper including verification testing on the following characteristics:

  • Stepping Unit is able to provide continuous (free) longitudinal movement ●
  • Encoders correctly report longitudinal and rotational motion of Stepping Unit .
  • Stepping Unit meets electrical safety requirements of IEC 60601-1 and IEC . 60601-1-1

Validation was also conducted to confirm the device remains safe and effective for its intended use and included testing with treatment planning software.

All testing confirmed that the EX3 Stepper is substantially equivalent to the predicate device in regards to safety and effectiveness and the new design features have not diminished the safety and effectiveness of the device.

Corporate Headquarters102 First Street SouthKalona, IA 52247USA319.248.6757319.248.6660
Coralville Office2301 Jones Blvd.Coralville, IA 52241USA319.248.6757319.248.6660
Europe OfficePasteurstraat 62811 DX ReeuwijkThe Netherlands+31(0) 182.394495+31(0) 182.395014
Orange City Office1401 8th Street SEOrange City, IA 51041USA712.737.8688712.737.8654

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol consists of three abstract human figures connected by flowing lines, representing the department's mission to protect the health of all Americans.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 17, 2013

CIVCO Medical Instruments Co., Inc. d/b/a CIVCO Medical Solutions % Ms. Amanda Stahle Regulatory Affairs Specialist 102 First Street South KALONA IA 52247

Re: K131161

Trade/Device Name: Ex3 Stepper" Classic Stepper, Multi-Purpose Workstation " Stepper, Micro-Touch", Micro-Touch" LP, Multi-Purpose Workstation ", Multi-Purpose Workstation™ LP; and Disposable Template Grid

Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: ITX, IWJ, JAQ Dated: August 28, 2013 Received: August 29, 2013

Dear Ms. Stahle:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{7}------------------------------------------------

Page 2-Ms. Stahle

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

for

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Indications for Use

510(k) Number (if known): K131161

Device Name: Stepping Unit Stabilizer Disposable Template Grid

Indications for Use: This device is indicated for use in adult males with known or suspected prostate cancer.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Sm27)

(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health

K131161 510(k)_

Page 1 of

§ 892.1570 Diagnostic ultrasonic transducer.

(a)
Identification. A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.(b)
Classification. Class II.