(29 days)
It is intended that this device, in any system configuration, be removed after development of solid fusion mass. Hook components are indicated for use at C1-C7. Polyaxial pedicle screws and offset connectors are limited to the thoracic spine (T1-T3) in treating thoracic conditions only. These screws and offset connectors are not intended to be placed in the cervical spine. The components in the SOLANAS Posterior stabilization system can be linked to the components in the Zodiac Polyaxial spinal fixation system offered by Alphatec Spine using the axial rod connectors, parallel rod connectors or transitional rods
- Degenerative disc disease (DDD), defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
- Spondylolisthesis
- Spinal stenosis
- Fracture/Dislocation
- Atlanto/Axial fracture with instability
- Revision of previous cervical spine surgery
- Tumors
The Solanas Posterior Stabilization System facilitates the surgical correction of spinal deformities by providing temporary internal fixation and stabilization during bone graff healing and/or fusion mass development. Device implants include a range of sizes of bone screws, hooks, rods and bridge assemblies to provide the versatility required for the specific conditions listed in the indications. The implants are manufactured from surgical grade titanium alloy (Ti-6AI-4V, ASTM F 136) with an electrolytic conversion coating. All hook components are intended for fixation/attachment to the cervical spine only (CI-C7). It is intended that the implants be removed after successful fusion.
The Solanas Posterior Stabilization System is a rod system designed to improve the stability of the cervical, cervico-thoracic and upper thoracic spine from C1-T3. This titanium alloy (Ti6Al-4V ELI, ASTM F136) system is simple in design and in its application of posterior rod fixation. It contains integrated functioning screws, rods, hooks, and connectors. The laminar hooks are an integral part of this system providing stabilization to the cervical (C1-C7) portion of the spine. These hooks attach to the lamina via the blade portion of the hook on one end and to a rod via a set screw and collar connection on the other end.
The provided text is related to a 510(k) premarket notification for a medical device called the "Solanas Posterior Stabilization System." This document is a regulatory filing for a spinal implant system and does not describe a study involving device performance metrics, human readers, or AI.
Therefore, I cannot extract information related to:
- A table of acceptance criteria and reported device performance (in terms of clinical or algorithmic metrics)
- Sample sizes for test sets or data provenance in the context of device performance
- Number and qualifications of experts for ground truth
- Adjudication methods
- Multi-reader multi-case (MRMC) comparative effectiveness studies
- Standalone algorithm performance studies
- Type of ground truth used (expert consensus, pathology, outcomes data)
- Sample size for the training set
- How ground truth for the training set was established
However, I can provide information about the non-clinical testing performed to establish substantial equivalence for this spinal implant system.
Based on the provided text, here is the relevant information:
Acceptance Criteria and Study for the Solanas Posterior Stabilization System (Mechanical Testing)
The Solanas Posterior Stabilization System is a spinal implant. The "acceptance criteria" and "study" mentioned in the document refer to mechanical testing performed to demonstrate that the modified device (with larger hook blades of 7mm and 8mm) is substantially equivalent to a previously cleared predicate device.
Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|
| Static anterior/posterior mechanical property measurements as outlined in ASTM F1798 | "produced acceptable results" |
| Performance per recognized consensus standards and guidance document: "Guidance for Industry and FDA Staff: Spinal System 510(k)s (2004)" | "demonstrates that the Solanas Posterior Stabilization System is substantially equivalent to the predicate device." |
2. Sample size used for the test set and the data provenance:
- Sample size: The document does not specify the exact number of units or subassemblies tested. It refers to "a subassembly" and "the subject and predicate devices."
- Data provenance: Not applicable in the context of mechanical testing of physical devices. The testing was performed by Alphatec Spine, Inc.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):
- Not applicable. This was mechanical testing of a medical device, not a study involving ground truth established by medical experts. The "ground truth" here is the adherence to mechanical testing standards.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable. This refers to consensus among human readers, which is not relevant for mechanical testing.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is a 510(k) for a physical spinal implant, not an AI or imaging device.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Not applicable. This is for a physical spinal implant, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- For this mechanical test, the "ground truth" or reference standard would be the established mechanical testing standards and parameters defined in ASTM F1798 and FDA guidance documents. The performance of the modified device was compared against these standards and the predicate device's performance.
8. The sample size for the training set:
- Not applicable. This refers to mechanical testing, not a dataset used for training an algorithm.
9. How the ground truth for the training set was established:
- Not applicable.
In summary: The provided document describes a regulatory submission for a physical medical device. The "study" mentioned is a mechanical performance comparison to demonstrate substantial equivalence, rather than a clinical trial or an AI/imaging device performance study with expert-adjudicated ground truth.
{0}------------------------------------------------
K131119 :: page 1 of 2
MAY 2 1 2013
510(k) Summary
Solanas® Posterior Stabilization System Date Prepared: April, 19th, 2013
| I | Company: | Alphatec Spine, Inc.5818 El Camino RealCarlsbad, CA 92008.USA |
|---|---|---|
| II. | Contact: | Trevor J. DenboRegulatory Affairs SpecialistTelephone: 760-494-6951Fax: 760-431-0289 |
| III. | Product Trade Name: | Solanas Posterior Stabilization System |
| IV. | Common Name: | Pedicle screw spinal system |
| V. | Classification Name: | Pedicle screw spinal system |
| VI. | Regulation Number: | 888.3050, 888.3070 |
| VII. | Classification Product Code: | KWP, MNI, NKB |
| VIII. | Predicate Devices: | Solanas Posterior Stabilization System, K052201and Modification to Solanas PosteriorStabilization System, K071380 |
IX. Description:
The Solanas Posterior Stabilization System facilitates the surgical correction of spinal deformities by providing temporary internal fixation and stabilization during bone graff healing and/or fusion mass development. Device implants include a range of sizes of bone screws, hooks, rods and bridge assemblies to provide the versatility required for the specific conditions listed in the indications.
The implants are manufactured from surgical grade titanium alloy (Ti-6AI-4V, ASTM F 136) with an electrolytic conversion coating. All hook components are intended for fixation/attachment to the cervical spine only (CI-C7). It is intended that the implants be removed after successful fusion.
Special 510(k)
{1}------------------------------------------------
X. Indications for Use:
It is intended that this device, in any system configuration, be removed after development of solid fusion mass. Hook components are indicated for use at C1-C7. Polyaxial pedicle screws and offset connectors are limited to the thoracic spine (T1-T3) in treating thoracic conditions only. These screws and offset connectors are not intended to be placed in the cervical spine. The components in the Solanas Posterior stabilization system can be linked to the components in the Zodiac Polyaxial spinal fixation system offered by Alphatec Spine using the axial rod connectors, parallel rod connectors or transitional rods
-
- Degenerative disc disease (DDD), defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
-
- Spondylolisthesis
-
- Spinal stenosis
-
- Fracture/Dislocation
-
- Atlanto/Axial fracture with instability
-
- Revision of previous cervical spine surgery
-
- Tumors
XI. Summary of the Technological Characteristics:
The technological characteristics, materials, and indications for use of Solanas® Posterior Stabilization System are substantially equivalent to the previously cleared predicate. The subject device differs from the predicate in that the modified hooks contain 7mm and 8mm blades, which is larger than the predicate.
The Solanas Posterior Stabilization System is a rod system designed to improve the stability of the cervical, cervico-thoracic and upper thoracic spine from C1-T3. This titanium alloy (Ti6Al-4V ELI, ASTM F136) system is simple in design and in its application of posterior rod fixation. It contains integrated functioning screws, rods, hooks, and connectors. The laminar hooks are an integral part of this system providing stabilization to the cervical (C1-C7) portion of the spine. These hooks attach to the lamina via the blade portion of the hook on one end and to a rod via a set screw and collar connection on the other end.
XII. Discussion of the Non-clinical Testing:
Mechanical characteristics were examined in accordance to the static anterior/posterior mechanical property measurements of a subassembly as outlined in ASTM F1798. This performance testing was performed comparing the subject and predicate devices per the recognized consensus standards and per the guidance document. Guidance for Industry and FDA Staff: Spinal System 510/k)s (2004). This test produced acceptable results and demonstrates that the Solanas Posterior Stabilization System is substantially equivalent to the predicate device.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G6 Silver Spring, MD 20993-0002
Letter Dated: May 21, 2013
Alphatec Spine, Incorporated % Mr. Trevor J. Denbo Regulatory Affairs Specialist 5818 El Camino Real Carlsbad, California 92008
Re: K13119
Trade/Device Name: SOLANAS® Posterior Stabilization System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNI. KWP Dated: April 19, 2013 Received: April 22, 2013
Dear Mr. Denbo:
. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 – Mr. Trevor J. Denbo
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Laurence Coyne -A
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
K131119 510(k) Number (if known):
Device Name: SOLANAS® Posterior Stabilization System
Indications For Use:
It is intended that this device, in any system configuration, be removed after development of solid fusion mass. Hook components are indicated for use at C1-C7. Polyaxial pedicle screws and offset connectors are limited to the thoracic spine (T1-T3) in treating thoracic conditions only. These screws and offset connectors are not intended to be placed in the cervical spine. The components in the SOLANAS Posterior stabilization system can be linked to the components in the Zodiac Polyaxial spinal fixation system offered by Alphatec Spine using the axial rod connectors, parallel rod connectors or transitional rods
-
- Degenerative disc disease (DDD), defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
-
- Spondylolisthesis
-
- Spinal stenosis
-
- Fracture/Dislocation
-
- Atlanto/Axial fracture with instability
-
- Revision of previous cervical spine surgery
-
- Tumors
ਮ Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ronald P. Jean -S
(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K131119
Solanas Posterior Stabilization System
Special 510(k)
l Page 1 of
N/A