K Number
K131063
Manufacturer
Date Cleared
2013-08-15

(121 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AMPLATZER TorqVue Low Profile Delivery System is indicated to provide a pathway through which devices are introduced within the chambers and coronary vasculature of the heart or in the peripheral vasculature.

Device Description

The AMPLATZER TorqVue Low Profile Delivery System is a sterile, single-use device designed to facilitate the introduction of devices to chambers and coronary vasculature of the heart and for introducing therapeutic devices to a location within the peripheral vasculature. The catheter has a single lumen for passage of devices with maximum outer diameters of 4 and 5 French. The catheters will be provided in 60 cm and 80 cm usable lengths. The system includes the following components: Delivery Catheter, Loader, Delivery Wire (optional), Plastic Vise (included with delivery wire), Hemostasis Valve.

AI/ML Overview

The provided text describes the PMA for the AMPLATZER TorqVue Low Profile Delivery System. However, it does not include information about acceptance criteria, reported device performance in a table, or study details such as sample sizes, ground truth establishment, expert qualifications, or comparative effectiveness studies (MRMC). The submission is for a medical device (delivery system), not an AI/ML-based device, hence, many of the requested fields are not applicable.

Here's the information that can be extracted from the provided text, along with notes on what is not available due to the nature of the submission (a medical device 510(k) for a delivery system, not an AI/ML product):

1. Table of acceptance criteria and the reported device performance:

The document does not explicitly state acceptance criteria in a table format with corresponding reported performance values. Instead, it states that "Verification and validation testing confirms that the functional characteristics of the AMPLATZER® TorqVue® Low Profile Delivery Systems are substantially equivalent to the predicate devices cited."

The testing included:

  • Catheter integrity
  • Catheter kink resistance
  • Leak resistance
  • Hub strength
  • "The ability to deliver various implantable devices."

The implication is that the performance met the standards of the predicate devices for these functional characteristics.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

This information is not provided in the document. The submission references "laboratory verification tests and in vitro performance information," but offers no details on sample sizes or data provenance for these tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not applicable/not provided. For a medical device like a delivery system, "ground truth" as typically defined for AI/ML algorithms (e.g., expert consensus on image interpretation, pathology reports) is generally not relevant. The verification and validation testing would have involved engineering and functional tests to ensure the physical device met its design specifications and performance claims.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

This information is not applicable/not provided. Adjudication methods are typically used in studies involving human interpretation or consensus for subjective assessments, which is not the primary focus of the engineering tests performed on a delivery system.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This information is not applicable. This is a submission for a physical medical device (delivery system), not an AI software. Therefore, an MRMC study related to AI assistance would not be relevant.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

This information is not applicable. This is not an algorithm-only device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The "ground truth" for this medical device would be based on engineering specifications and established performance standards for medical catheters and delivery systems, demonstrating functional characteristics like integrity, kink resistance, leak resistance, hub strength, and delivery capability. It's not based on expert consensus, pathology, or outcomes data in the way an AI diagnostic device would be.

8. The sample size for the training set:

This information is not applicable/not provided. This is not an AI/ML device that requires a training set. The device's performance is based on its physical design and manufacturing.

9. How the ground truth for the training set was established:

This information is not applicable/not provided. As above, no training set for an AI/ML algorithm is involved.

{0}------------------------------------------------

.

4 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

4.1 Applicant Information

Date Prepared:March XX, 2013
Submitter:AGA Medical Corporation
Address:5050 Nathan Lane North
Plymouth, MN 55442AUG 1 5 2013
EstablishmentRegistration No:2135147
Contact Person:Heather M Taylor
Sr. Regulatory Affairs Specialist
Telephone Number:(651) 756-5765
Fax Number:(763) 647-5932
4.2Device Information
Trade Name:AMPLATZER® TorqVue® Low Profile Delivery System
Common Name:Delivery System
Classification Name:Catheter, Percutaneous
Classification:Class II, 21 CFR 870.1250
Product Code:DQY
Predicate Devices:AMPLATZER® TorqVue® Delivery System510(k) K072313, Reg. No. 870.1250; Product Code: DQY
Predicate Device Intended Use: The AMPLATZER® TorqVue®Delivery System is intended to provide a pathway throughwhich devices are introduced within the chambers andcoronary vasculature of the heart or in the peripheralvasculature.

:

{1}------------------------------------------------

AMPLATZER® TorqVue® Low Profile Delivery System

510(k) K080757, Reg. No. 870.1250; Product Code: DQY

Predicate Device Intended Use: The AMPLATZER TorqVue LP delivery system is intended to provide a pathway through which devices are introduced into the peripheral vasculature.

Device Description: The AMPLATZER TorqVue Low Profile Delivery System is a sterile, single-use device designed to facilitate the introduction of devices to chambers and coronary vasculature of the heart and for introducing therapeutic devices to a location within the peripheral vasculature.

The catheter has a single lumen for passage of devices with maximum outer diameters of 4 and 5 French. The catheters will be provided in 60 cm and 80 cm usable lengths. The system includes the following components:

  • . Delivery Catheter - used to deliver devices
  • . Loader – used to help introduce the selected implantable device into the delivery catheter
  • . Delivery Wire (optional) - attaches to the implantable device and facilitates advancement through the catheter, placement and, if desired, recapture of the specified implantable device. [If this surgical accessory (Product Code = DWS) is included, the labeling will specify compatibility with the appropriate implantable devices.]
  • . Plastic Vise (included with delivery wire) – a handle that is attached to the Delivery Wire by means of a set screw
  • . Hemostasis Valve- used on the proximal end of the Loader to minimize bleeding from the Delivery Catheter and for flushing air from the system

The AMPLATZER® TorgVue Low Profile Delivery System is Intended Use: intended to provide a pathway through which devices are

AMPLATZER® TorqVue® Low Profile Delivery System 510(k) Premarket Notification

{2}------------------------------------------------

introduced within the chambers and coronary vasculature of the heart or in the peripheral vasculature.

Comparison to

The AMPLATZER TorqVue Low Profile Delivery System is Predicate Devices: substantially equivalent to the predicate devices cleared by K072313 and K080757. This submission seeks to add a cardiac indication to the already cleared K080757. All systems are handheld catheter systems designed to facilitate access and placement of specified implantable devices within the chambers and coronary vasculature of the heart. All three delivery catheters are single lumen design with a hemostasis valve. The device included in this submission is identical in design to the already cleared K080757, and is using the already cleared K072313 to support the cardiac indication for use. Test Data: Verification and validation testing confirms that the functional characteristics of the AMPLATZER® TorqVue® Low Profile Delivery Systems are substantially equivalent

to the predicate devices cited. This included catheter integrity, catheter kink resistance, leak resistance, hub strength and the ability to deliver various implantable devices.

Summary: Based on the technical information, intended use, laboratory verification tests and in vitro performance information provided, the AMPLATZER® TorqVue® Low Profile Delivery System is substantially equivalent to the currently marketed predicate devices.

{3}------------------------------------------------

Public Health Service

Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of two main elements: a circular seal with text and a stylized symbol. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the circumference of the circle. Inside the circle is a symbol that resembles a stylized caduceus, with three wavy lines representing snakes intertwined around a staff.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

August 15, 2013

St. Jude Medical C/O Heather Taylor 5050 Nathan Lane North Plymouth, MN 55442 US

Re: K131063

Trade/Device Name: AMPLATZER TorqVue Low Profile Delivery System Regulation Number: 21 CFR 870.1250 Regulation Name: Catheter, Percutaneous Regulatory Class: Class II Product Code: DQY Dated: July 25, 2013 Received: July 26, 2013

Dear Ms. Taylor:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

M.S. Hillman

for

for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

3 STATEMENT OF INDICATIONS FOR USE

K131063 510(k) Number: _______________________________________________________________________________________________________________________________________________________________

Device Name: AMPLATZER TorqVue Low Profile Delivery System

Indications for use:

The AMPLATZER TorqVue Low Profile Delivery System is indicated to provide a pathway through which devices are introduced within the chambers and coronary vasculature of the heart or in the peripheral vasculature.

Prescription Use ______x (Part 21 CFR 801 Subpart D) OR Over-The-Counter-Use (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

M. A. Hilleman

(Division Sign-Off) AMPLATZER® TorqVue® Low Profile Delivery System Division Sign-Off)
510(k) Premarket Notification 510(k) Premarket Notification K131063 510(k) Number:

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).