(150 days)
The Pioneer Elite Diode Laser is indicated for dentistry and oral soft tissue procedures of:
- The removal of lesions, excision, incision, vaporization, ablation, hemostasis, and photocoagulation on soft tissue including abscess treatment, contouring, curettage, sulcular debridement, pulpotomy, frenectomy, gingivectomy, troughing, and removal of inflamed edematous tissue.
- Temporary relief of minor muscle and joint pain, stiffness, minor arthritis pain, muscle spasm, temporary increase in local blood circulation, and temporary relaxation of muscles by means of topical elevated tissue temperature from infrared spectral emissions;
- Light activation of bleaching materials for teeth whitening and laser-assisted whitening/bleaching of teeth.
The Pioneer Elite Diode Laser is a device for delivering laser energy to human soft tissue for a variety of surgical procedures and treatments. This energy is generated by solid-state diodes, which provide a consistent and reliable generation of laser energy at 810 ± 20mm for a maximum of 3 watts of energy output. The laser energy is delivered to surgical site by means of an optical fiber system, which allows for the safe transmission of laser energy to the site without creating undue risk to the patient or operatory staff by errant or collateral laser emissions. The device features some user definable settings, including a switchable 630nm aiming beam, adjustable power output for both the working beam and aiming beam, and continuous delivery or pulse delivery options. The terminal end of the delivery fiber is contained within a metal handpiece with a disposable single-use tip. This handpiece system is provided with the device. The activation of the working beam diode is completed by use of a foot-actuated switch.
Here's an analysis of the provided text regarding the acceptance criteria and study for the Pioneer Elite Diode Laser:
Acceptance Criteria and Device Performance for Pioneer Elite Diode Laser
This 510(k) summary focuses on demonstrating substantial equivalence to predicate devices for the Pioneer Elite Diode Laser, rather than presenting a traditional clinical study with detailed acceptance criteria and performance metrics against a specific clinical outcome. Therefore, the "acceptance criteria" here are primarily conformity to standards and meeting design specifications as demonstrated through bench testing.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Criteria / Requirement | Reported Device Performance |
|---|---|---|
| Safety Standards | Compliance with 21 CFR 1040.10 and 1040.11 (Laser Product Performance Standards) and Laser Notice 50 | "The Pioneer Elite Diode Laser is designed to comply with the performance requirements of 21 CFR 1040.10 and 1040.11, with permissible deviations relative to Laser Notice 50, dated June 24, 2007." |
| Medical Electrical Equipment Safety | Compliance with IEC 60601-2-22 Edition 2 | "The device also complies with the recognized standards of IEC 60601-2-22 Edition 2" |
| Laser Safety Management | Compliance with IEC 60825-1 Edition 1.2 | "and IEC 60825-1 Edition 1.2." |
| General Medical Electrical Equipment Safety | Compliance with IEC 60601-1: 2nd Edition, IEC 60601-1-2, and IEC 60601-1-4 | "The device is designed in compliance to the entirety of IEC 60601-1: 2nd Edition, IEC 60601-1-2, and IEC 60601-1-4." |
| Design Criteria & Performance Requirements | Meet design criteria for essential performance and satisfy performance requirements in 21 CFR 1010 and 21 CFR 1040 | "Bench testing on an evaluation sample of the current device revealed that the device met the design criteria for essential performance, and satisfied the performance requirements indicated in 21 CFR 1010 and 21 CFR 1040." "Device outputs were within performance requirements and all safety features and functions were operating correctly." |
| Substantial Equivalence | Similar intended uses, operating principles, design features, functional and performance characteristics to predicate devices. | "The Pioneer Elite Diode Laser is substantially equivalent to the listed predicate devices without raising any new issues of safety or effectiveness. This device shares similar intended uses, operating principles, design features, and similar functional and performance characteristics." (This is the overarching conclusion of the submission, based on the comparison of features and the performance data summarized above). |
2. Sample Size for Test Set and Data Provenance
- Sample Size for Test Set: The document states "Bench testing on an evaluation sample of the current device." A specific number for the sample size is not provided.
- Data Provenance: The data is generated from bench testing performed by the manufacturer (CAO Group, Inc.). There is no information about the country of origin of this testing other than the manufacturer's address in West Jordan, UT, USA. The testing is implicitly prospective in the sense that it was conducted on the manufactured device to verify its performance prior to submission.
3. Number of Experts and Qualifications for Ground Truth
- Not Applicable. This submission relies on technical standards compliance and bench testing against engineering specifications, not expert-derived ground truth for clinical outcomes or diagnostic accuracy.
4. Adjudication Method
- Not Applicable. There is no mention of an adjudication method as the primary evaluation is technical compliance and bench testing.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No. A MRMC comparative effectiveness study was not conducted or described in this 510(k) summary. The submission focuses on demonstrating substantial equivalence to predicate devices through technical specifications and compliance with safety standards, not a clinical comparison of AI-assisted human performance.
6. Standalone Performance Study (Algorithm Only)
- Yes (implicitly, for the device's technical functions). The "Performance Data" section describes "Bench testing on an evaluation sample of the current device" where "device outputs were within performance requirements and all safety features and functions were operating correctly." This describes the standalone performance of the device's hardware and software (if any control software is involved) in meeting its design specifications. It's not an "algorithm-only" study in the modern sense of AI, but rather a verification of the entire device's functional performance.
7. Type of Ground Truth Used
- The "ground truth" for the bench testing was the design criteria and performance requirements specified by the manufacturer, and the criteria outlined in the cited federal regulations (21 CFR 1010, 21 CFR 1040) and international standards (IEC 60601 series, IEC 60825-1).
8. Sample Size for Training Set
- Not Applicable. This device is a laser surgical instrument, and its performance evaluation does not involve machine learning algorithms that require a "training set" of data.
9. How Ground Truth for Training Set Was Established
- Not Applicable. As no training set is relevant for this type of device and performance evaluation, no ground truth for a training set was established.
{0}------------------------------------------------
510(k) Summary of Safety and Effectiveness
CAO Group, Inc. 4628 West Skyhawk Drive West Jordan, UT 84084 Tel: 801-256-9282 Fax: 801-256-9287
Prepared By: Robert K. Larsen. Preparation Date: September 11, 2013
Device Name:
| Trade Name: | Pioneer Elite Diode Laser |
|---|---|
| Common Name: | Soft Tissue Diode Laser |
| Product Classification: | Powered Laser Surgical Instrument |
Legally Marketed Predicate Devices for Substantial Equivalence:
Odyssey Navigator Diode Laser, manufactured for Ivoclar Vivadent, Inc. (K062258)
Pilot Diode Laser, manufactured by CAO Group, Inc. (K100143)
ezLase, manufactured by Biolase, Inc. (K082938)
Rationale for Substantial Equivalence:
The present device shares similar specific indications with the aforementioned devices as now specified:
- . Excision, incision, ablation, and photocoagulation (K062258)
- Temporary relief of minor muscle and joint pain, stiffness, minor arthritis pain, muscle o spasm, temporary increase in local blood circulation, and temporary relaxation of muscles by means of topical elevated tissue temperature from infrared spectral emissions (K100143)
- A light source to activate tooth whitening materials and assist in tooth whitening . procedures (K082938)
The predicate devices and submitted device share similar design features including wavelength. operating controls, and laser delivery method. The devices share similar methods of control systems, safety features, and performance monitoring. The devices share similar performance specifications including power output and energy type.
SEP 1-3-2013
{1}------------------------------------------------
Description of Submitted Device:
The Pioneer Elite Diode Laser is a device for delivering laser energy to human soft tissue for a variety of surgical procedures and treatments. This energy is generated by solid-state diodes, which provide a consistent and reliable generation of laser energy at 810 ± 20mm for a maximum of 3 watts of energy output. The laser energy is delivered to surgical site by means of an optical fiber system, which allows for the safe transmission of laser energy to the site without creating undue risk to the patient or operatory staff by errant or collateral laser emissions. The device features some user definable settings, including a switchable 630nm aiming beam, adjustable power output for both the working beam and aiming beam, and continuous delivery or pulse delivery options.
The terminal end of the delivery fiber is contained within a metal handpiece with a disposable single-use tip. This handpiece system is provided with the device. The activation of the working beam diode is completed by use of a foot-actuated switch.
Intended Uses of the Submitted Device:
The Pioneer Elite Diode Laser is indicated for dentistry and oral soft tissue procedures of:
- l) The removal of lesions, excision, incision, vaporization, ablation, hemostasis, and photocoagulation on soft tissue including abscess treatment, contouring, curettage, sulcular debridement, pulpotomy, frenectomy, gingivectomy, troughing, and removal of inflamed edematous tissue.
-
- Temporary relief of minor muscle and joint pain, stiffness, minor arthritis pain, muscle spasm, temporary increase in local blood circulation, and temporary relaxation of muscles by means of topical elevated tissue temperature from infrared spectral emissions;
-
- Light activation of bleaching materials for teeth whitening and laser-assisted whitening/bleaching of teeth.
Technological Characteristics and Substantial Equivalence:
The Odyssey Navigator Diode Laser is a device for delivering laser energy to human oral soft tissue for a variety of surgical procedures and treatments. This energy is generated by solidstate diodes, which provide a consistent and reliable generation of laser energy at 810 ± 20nm for a maximum of 3 watts of energy output. The laser energy is delivered to the treatment site by means of an optical fiber system, which allows for the safe transmission of laser energy to the site without creating undue risk to the patient or operatory staff by errant or collateral laser emissions. The device features some user definable settings, including a switchable 630nm aiming beam, adjustable power output for both the working beam and aiming beam, and continuous delivery or pulse delivery options.
{2}------------------------------------------------
K131059
The working end of the delivery fiber is contained within a metal handpiece with a disposable single-use tip. This handpiece system is incorporated into the device. The activation of the working beam diodes is completed by use of a foot-actuated switch.
The Pilot Diode Laser uses solid state diodes to generate laser energy in the 810nm range. This system uses a fiber delivery system to transmit laser energy to the treatment site. The system also features a 630nm aiming beam and features controls that allow for adjusting the output of the working beam, and switching between a continuous or pulsed-mode laser emissions. The maximum output of the working beam is 9 watts.
The ezlase uses solid state diodes to generate laser energy in the 940nm range. This system uses a fiber delivery system to transmit laser energy to the treatment site. The system also features a 630nm aiming beam and features controls that allow for adjusting the output of the working beam, and switching between a continuous or pulsed-mode laser emissions. The maximum output of the working beam is 7 watts.
Conformity to Standards:
The Pioneer Elite Diode Laser is designed to comply with the performance requirements of 21 CFR 1040.10 and 1040.11, with permissible deviations relative to Laser Notice 50, dated June 24, 2007. The device also complies with the recognized standards of IEC 60601-2-22 Edition 2 and IEC 60825-1 Edition 1.2. The device is designed in compliance to the entirety of IEC 60601-1: 2nd Edition, IEC 60601-1-2, and IEC 60601-1-4.
Performance Data
Bench testing on an evaluation sample of the current device revealed that the device met the design criteria for essential performance, and satisfied the performance requirements indicated in 21 CFR 1010 and 21 CFR 1040. Device outputs were within performance requirements and all safety features and functions were operating correctly.
Conclusion
The Pioneer Elite Diode Laser is substantially equivalent to the listed predicate devices without raising any new issues of safety or effectiveness. This device shares similar intended uses, operating principles, design features, and similar functional and performance characteristics. The device is designed to comply with relevant federal and international safety and performance standards.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle-like bird with three curved lines representing its body and wings. The bird is oriented towards the upper right of the image. Encircling the bird is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Ducument Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 13, 2013
CAO Group, Incorporated Mr. Robert K. Larsen Regulatory Alfairs Manager 4628 West Skyhawk Drive West Jordan. Utah 84084
Re: K131059
Trade/Device Name: Pioneer Elite Diode Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX. ILY Dated: August 22, 2013 Received: August 23. 2013
Dear Mr. Larsen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Mr. Robert K. Larsen
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
For
Peter D. Rumm -A
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
K131059 510(k) Number (if known):
Device Name: Pioneer Elite Diode Laser
The Pioneer Elite Diode Laser is indicated for dentistry and oral soft tissue procedures of:
-
- The removal of lesions, excision, incision, vaporization, ablation, hemostasis, and photocoagulation on soft tissue including abscess treatment, contouring, curettage, sulcular debridement, pulpotomy, frenectomy, gingivectomy, troughing, and removal of inflamed edematous tissue.
-
- Temporary relief of minor muscle and joint pain, stiffness, minor arthritis pain, muscle spasm, temporary increase in local blood circulation, and temporary relaxation of muscles by means of topical elevated tissue temperature from infrared spectral emissions;
-
- Light activation of bleaching materials for teeth whitening and laser-assisted whitening/bleaching of teeth.
Prescription Use AND/OR Over-The-Counter Use (Per 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
.............................................................................................................................................................................. Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R Ogden: Ja 2013.09.13 09:26:51 -04'00'
(Division Sign-Off) for MXM
Division of Surgical Devices
KI31050 510(k) Number _
Page 1 of of 1 _
4-1
§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.
(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.