K Number
K130931

Validate with FDA (Live)

Date Cleared
2013-08-30

(148 days)

Product Code
Regulation Number
886.5925
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

EASY DAY (methafilcon A) 1-Day Soft ( Hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia) in not-aphakic persons with non-diseased eyes that may exhibit astigmatism up to 0.75 diopters that does not interfere with visual acuity. The lens may be prescribed in spherical powers ranging from 0 ~ - 12.00D. The lens is indicated for single use daily disposable wear and patients are instructed to discard the lens after each removal.

Device Description

EASY DAY (methafilcon A) 1-Day Soft (hydrophilic) contact lens is available as a single vision aspherical lens. The lens material, methafilcon A is a copolymer of 2-hydroxyethyl methacrylate (HEMA) and methacrylic acid (MAA) cross-linked with ethylene glycol dimethacrylate (EGDMA) . The lens with visible tint is tinted blue using reactive blue 4K to make the lens more visible for handling. Lenses are supplied sterile in sealed blister packers containing buffered saline solution.

AI/ML Overview

The provided 510(k) summary for the "EASY DAY (methafilcon A) 1-Day Soft (hydrophilic) Contact Lens for Daily Wear" focuses on demonstrating substantial equivalence to a predicate device rather than outright proving performance against specific acceptance criteria with detailed metrics in a table.

However, based on the nonclinical studies section, we can infer some "acceptance criteria" through the reported device properties and comparability to the predicate. The "study" proving these are the nonclinical performance tests.

Here's an attempt to structure the information as requested, though some sections will highlight the lack of detailed quantification regarding "performance" against explicit "acceptance criteria" as would be seen in an AI/software device submission.


1. Table of Acceptance Criteria and Reported Device Performance

Parameter/Acceptance Criteria (Inferred from Predicate Equivalence)Reported Device Performance (EASY DAY Contact Lens)
Technological Characteristics
Diameter14.20 ± 0.2mm
Power Range-0.0 D to -12.00 D
Center Thickness0.06 ~ 0.100mm (varies with power)
Refractive Index1.409 ± 0.005
Light Transmittance>92%
Water Content53 to 57 %
Oxygen Permeability (Dk/t)19.7 ± 20% {10-11(cm2/sec)(mlO2/ml-mmHg)}
Biocompatibility
Irritation (Rabbit Eye)Non-irritated
Systemic ToxicityNon-toxic
CytotoxicityNot cytotoxic
Microbiology
Sterility Assurance Level (SAL)Minimum 10⁻⁶ (validated steam sterilization)
Shelf-life SterilityStable through expiration date
Leachability
Leachable Monomers and Addictive ResiduesNone detected (at reported detection levels)
Clinical Performance (Inferred)
Vision CorrectionNearly 100% of participants' vision corrected
Patient Satisfaction (Wearing & Care)Nearly all participants satisfied
Overall Safety and EffectivenessGood and safe for customers

Note: For this type of device (contact lens), the "acceptance criteria" are largely derived from established standards for contact lens properties and biocompatibility, and by demonstrating equivalence to a legally marketed predicate device. Specific numerical acceptance thresholds beyond what is shown are often found in the detailed test protocols which are not included in this summary.


2. Sample Size Used for the Test Set and Data Provenance

  • Clinical Study: At least 30 evaluable subjects' eyes were tested separately for 1 month.
  • Data Provenance: The document does not explicitly state the country of origin for the clinical study data or whether it was retrospective or prospective. Given the submitter's location (Jiangsu, China), it is highly probable the clinical study was conducted in China. The "summary of clinical study" describes an evaluation over a month, suggesting a prospective study.
  • Nonclinical Studies: No explicit sample sizes are provided for individual nonclinical tests (e.g., irritation, cytotoxicity). For technological characteristics, it is implied that representative samples of the manufactured lenses were tested.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Clinical Study: Not applicable in the sense of expert consensus for ground truth as would be used for diagnostic imaging devices. The study assessed patient-reported outcomes (satisfaction) and vision correction, likely measured by optometrists or ophthalmologists, but their number or specific qualifications are not detailed.
  • Nonclinical Studies: Not applicable. These involve laboratory testing against established physical, chemical, and biological standards, not expert consensus on findings.

4. Adjudication Method for the Test Set

  • Clinical Study: Not applicable for adjudication in the context of diagnostic ground truth. Clinical outcomes (vision correction, satisfaction) were likely recorded directly.
  • Nonclinical Studies: Not applicable.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No, an MRMC comparative effectiveness study was not conducted. This type of study is typical for diagnostic algorithms (e.g., AI in radiology) where human readers interpret cases with and without AI assistance. This submission pertains to a medical device (contact lens), not a diagnostic algorithm.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • No, a standalone (algorithm only) performance study was not done. This category is not relevant for a contact lens.

7. The Type of Ground Truth Used

  • Clinical Study: The "ground truth" for the clinical study was based on:
    • Patient outcomes: Corrected vision (measured objectively).
    • Patient satisfaction: Subjective feedback on lens wearing and care.
    • General safety: Observation for adverse events or complications.
  • Nonclinical Studies: The "ground truth" (or reference) for these tests involved:
    • Industry standards and guidelines: For properties like oxygen permeability, light transmittance, water content, sterility, etc.
    • Reference materials/controls: In biocompatibility and leachability testing.
    • Predicate device characteristics: For comparison of technological characteristics.

8. The Sample Size for the Training Set

  • Not applicable. This device is a physical medical product (contact lens), not an AI algorithm, so there is no "training set" in the computational sense.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable, as there is no training set for this device.

{0}------------------------------------------------

K13093

510 (k) Summary SUMMARY OF SAFETY AND EFFECTIVENESS FOR EASY DAY (methafilcon A )1-Day Soft (hydrophilic) Contact Lens for Daily Wear

Prepared Date: Aug. 20, 2013

Prepared Date:Aug. 20, 2013
Submitter Information:
Company:Jiangsu Horien Contact Lens Co., Ltd.
No.88 Danfu RD, Situ Town, Danyang City,Jiangsu Province,
China.,212331
Contact Person:Jinhua Hu
Manager of Registration Division
Phone:86-511-86809314
FAX:86-511-86206366
AUG 3 0 2013
Identification of Device:
Classification Name:Lens, Contact, (Disposable)
Trade Name:EASY DAY (methafilcon A) 1-Day Soft (hydrophilic) ContactLens
Common or usual Name:Soft (hydrophilic) Contact lens (daily wear)
FDA Classification:Class II

Predicate Device:

FREQUENCY 55, FREQUENCY 55 ASPHERIC, ENCORE, ENCORE TORIC, ONEVUE, COOPERFLEX (methafilcon A ) Soft (hydrophilic) Contact Lens cleared via K000384 COOPERVISION, INC.

Indications for Use

EASY DAY (methafilcon A) 1-Day Soft ( Hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia) in not-aphakic persons with non-diseased eyes that may exhibit astigmatism up to 0.75 diopters that does not interfere with visual acuity. The lens may be prescribed in spherical powers ranging from 0 ~ - 12.00D. The lens is indicated for single use daily disposable wear and patients are instructed to discard the lens after each removal.

Description of Device

EASY DAY (methafilcon A) 1-Day Soft (hydrophilic) contact lens is available as a single vision aspherical lens. The lens material, methafilcon A is a copolymer of 2-hydroxyethyl

{1}------------------------------------------------

methacrylate (HEMA) and methacrylic acid (MAA) cross-linked with ethylene glycol dimethacrylate (EGDMA) . The lens with visible tint is tinted blue using reactive blue 4K to make the lens more visible for handling. Lenses are supplied sterile in sealed blister packers containing buffered saline solution.

Summary of Clinical Study:

The EASY DAY (methafilcon A) 1-Day Soft (hydrophilic) Contact Lenses for Daily Wear were tested at least 30 evaluable subjects' eyes separately for 1 month. Nearly a hundred percent of the participants' vision was corrected and nearly all were satisfied with the lens wearing and care of lenses. In general, these products are good and safe for customers.

Nonclinical Studies:

A series of nonclinical performance tests were performed to demonstrate the safety and effectiveness of the EASY DAY (methafilcon A) 1-Day Soft (hydrophilic) Contact Lens for Daily Wear, and establish substantial equivalence to predicate lenses - FREQUENCY 55, FREQUENCY 55 ASPHERIC, ENCORE, ENCORE TORIC, ONEVUE, COOPERFLEX (methafilcon A ) Soft (hydrophilic) Contact Lens (K000384). All testing was conducted in accordance with the May 1994 FDA guideline titled Premarket Notification [510(k)] Guidance Document for Class II Daily Wear Contact Lenses, and in conformance to applicable device regulations. The evidence of substantial equivalent to the predicate lens described as follow:

  • a) Technological characteristics studies
    EASY DAY (methafilcon A) 1-Day Soft (hydrophilic) Contact Lens for Daily Wear designs in the following parameter ranges:

Diameter range:14.20 ±0.2mm

Power range: -0.0 D to -12.00 D

Center Thickness: 0.06~0.100mm varies with power

Lenses have the following properties:

Refractive index: 1.409 ±0.005

Light transmittance: >92%

Water content: 53 to 57 %

Oxygen permeability (edged corrected): 19.7±20% { 10++1(cm2/sec)(mlO2/ml-mmHg) }

There characterizations of EASY DAY (methafilcon A) 1-Day Soft (hydrophilic) Contact Lenses are equivalent and comparable to those of predicate lenses.

b) Biocompatibility

"In accordance with the May 1994 FDA guideline titled Premarket Notification [510(k)] Guidance Document for Class II Daily Wear Contact Lenses, toxicity studies have been conducted on the model: EASY DAY (methafilcon A) 1-Day Soft (hydrophilic) Contact Lens

{2}------------------------------------------------

for Daily Wear. The Irritation test in the rabbit eye and Systemic toxicity studies indicate the extracts would be considered as non-toxic and nor irritated. The Cytotoxicity testing demonstrates the lens is not cytotoxic under the conditions of the study.

c) Microbiology

Steam sterilization process has been validated to deliver a minimum SAL of 10°, thereby complying with the requirement of FDA Group 4. There is shelf-life stability data supporting that the lens remains sterile through the expiration date claimed for the product.

d) Leachability

Studies were conducted to determine the leachable materials from the finished lens. The results show that, at the levels of the detection reported, there are no leachable monomers and addictive residues.

Substantial equivalence Statement:

Testing performed on the EASY DAY (methafilcon A) I-Day Soft (hydrophilic) Contact Lens for Daily Wear indicated that it can support the efficiency and security use as well as the predicate devices- FREQUENCY 55, FREQUENCY 55 ASPHERIC, ENCORE, ENCORE TORIC. ONEVUE, COOPERFLEX (methafilcon A ) Soft (hydrophilic) Contact Lens (K000384), when used in accordance with the instructions for use. It is due to the facts that the risks and benefits of the subject device are the same as soft contact lenses for to the daily wear.

In conclusion, it is Jiangsu Horien's EASY DAY (methafilcon A )1-Day Soft (hydrophilic) Contact Lens conviction that data submitted in this 510(k) to validate the claim of substantial equivalency, substantiates our ability to manufacture a soft contact lens, the EASY DAY (methafilcon A) 1-Day Soft (hydrophilic) Contact Lens for Daily Wear, with the same established safety profile and effectiveness as the predicate device-- FREQUENCY 55, FREQUENCY 55 ASPHERIC, ENCORE, ENCORE TORIC, ONEVUE, COOPERFLEX (methafilcon A ) Soft (hydrophilic) Contact Lens (K000384).

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features the department's emblem, which is a stylized representation of a human figure embracing a globe. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the emblem in a circular fashion.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

August 30, 2013

Jiangsu Horien Contact Lens Co., Ltd. % Ms. Jennifer Reich Harvest Consulting Corp 2904 N. Boldt Drive, Flagstaff, AZ 86001

Re: K130931

Trade/Device Name: EASY DAY(methafilcon A ) 1-Day Soft (Hydrophilic) Contact Lens Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) Contact lens (daily wear) Regulatory Class: Class II Product Code: LPL Dated: July 29, 2013 Received: July 19, 2013

Dear Ms. Reich:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{4}------------------------------------------------

Page 2 - Ms. Jennifer Reich

You must comply with all the Act's requirements, including, but not limited to: Registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kesia Y. Alexander -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number: K130931

Device Name: EASY DAY (methafilcon A) I-Day Soft ( Hydrophilic) Contact Lens

Indications for Use:

EASY DAY (methafilcon A) I-Day Soft ( Hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia) in not-aphakic persons with non-diseased eyes that may exhibit astigmatism up to 0.75 diopters that does not interfere with visual acuity. The lensmay be prescribed in spherical powers ranging from 0 ~ - 12.00D.

The lens is indicated for single use daily disposable wear and patients are instructed to discard the lens after each removal.

Prescription Use______________________________________________________________________________________________________________________________________________________________ X (part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Leonid Livshitz · -2 2013.08.23 : '00'04- 17:19:55 (Division Sign-Off) Division of Ophthalmic and Ear, Nose, and Throat Devices

510(k) Number:

page 1 of _ l ________________________________________________________________________________________________________________________________________________________________

§ 886.5925 Soft (hydrophilic) contact lens.

(a)
Identification. A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.