K Number
K130606
Date Cleared
2014-01-03

(302 days)

Product Code
Regulation Number
884.4120
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The FMwand Ferromagnetic Surgical System is indicated for cutting and coagulation of soft tissue during surgical procedures, including Gynecologic procedures (open, transabdominal only).

Device Description

The FMwand Ferromagnetic Surgical System is a soft tissue cutting and coagulation device that consists of the FMwand Generator that is electrically connected to the sterile, single-use FMwand Handpiece by the FMwand Power Module (a resterilizable, reusable accessory cable) and a single-use, sterile air line. Like the predicate device, the handpiece includes actuation buttons that can be used to activate the handpiece tip. Also like the predicate device, an optional footpedal can be used to activate the handpiece. The connectors at both ends of the FMwand Power Module are designed to fit only the FMwand Generator and FMwand sterile, disposable accessories. The FMwand Power Module is provided non-sterile. It is intended to be cleaned and sterilized prior to the initial use and before each subsequent reuse.

AI/ML Overview

The provided 510(k) summary for the FMwand Ferromagnetic Surgical System (K130606) does not describe a study involving an AI/Machine Learning device or a Multi-Reader Multi-Case (MRMC) comparative effectiveness study. This submission is for a traditional medical device (an electrosurgical system) and focuses on demonstrating substantial equivalence to a predicate device through bench testing and compliance with existing standards.

Therefore, many of the requested criteria related to AI/ML device testing (such as sample size for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and training set information) are not applicable to this document.

The document primarily discusses that the new device is "substantially equivalent" to an existing predicate device (Domain Surgical System, K121881) by demonstrating that differences do not raise new questions of safety and effectiveness, and that the device functions as intended based on bench testing.

Here's an analysis based on the information available, noting where specific AI/ML related criteria are not met:

1. Table of acceptance criteria and reported device performance:

The document doesn't explicitly state "acceptance criteria" in a quantitative, measurable sense for clinical performance. Instead, it relies on demonstrating that the new device performs identically or equivalently to the predicate device through bench testing and compliance with standards. The "performance" is implicitly deemed acceptable if it matches the predicate and meets relevant safety standards.

Feature / CriterionPredicate Device Performance (Domain Surgical System K121881)Subject Device Performance (FMwand System K130606)Discussion / Equivalence
Intended Use (Cutting and coagulation of soft tissue)YesYesIdentical
Indications for Use (Specific for Gyn procedures)YesYesIdentical
Heat Generation Method (Ferromagnetic induction)YesYesIdentical
Power Requirements (100-240 VAC, 50-60 Hz)YesYesIdentical
Maximum Output Energy (5-60W)YesYesIdentical
Maximum Operational Temperature (450°C)YesYesIdentical
Operational Control Method (Controlled power to tip)YesYesIdentical
Mode of Operation (Intermittent)YesYesIdentical
Active Cooling (Air cooled handpiece)YesYesIdentical
Operating Frequency (Pure Sinusoidal 40.68 MHz)YesYesIdentical
Delivery System Configuration (Various tip lengths)YesYesIdentical
Bench TestingPassed allPassed all (details in Section 15 of submission)Identical performance
IEC 60601-2-2 ComplianceYesYes (details in Section 14 of submission)Identical compliance
Biocompatibility (ISO 10993-5, -10, -11)Passed allPassed all (details in Section 12 of submission)Identical compliance
Handpiece/Tip Sterilization (Ethylene Oxide: SAL 10-6)ValidatedValidated for reusable cable (new component)Equivalent sterilization methods and SAL 10-6 used
Output TypeType BF minimumType CF (more stringent leakage current restrictions)Subject device meets a more stringent standard (CF vs BF)
Handpiece ConfigurationOne-piece disposableDisposable handpiece section with reusable cableDifferences in form and sterilization methods, but equivalent tip function. Validated for new sterilization process.
Handpiece Actuation ButtonOne button (single power level)Two buttons (dual user-selected high/low power)Added user convenience; single button mode still supported.
Footpedal ConfigurationOne pedal (single power level)Two pedals (dual user-selected high/low power)Added user convenience; single pedal mode still supported.
Generator User InterfaceSupports single power levelAdded capability to support dual user-selected power levelsAccommodates new dual power feature; single power still supported.

Study Proving Device Meets Acceptance Criteria:

The document refers to "bench testing" as the primary study type to demonstrate safety and effectiveness and substantial equivalence.

  • "All applicable bench testing was performed with the FMwand Ferromagnetic Surgical System to assure that it functions as intended. Bench tests verified that the generator output was within specification and that the thermal output of the handpiece and tip corresponded to the target values."
  • "The FMwand Ferromagnetic Surgical System passed all bench tests performed. Details are found in Section 15: of this submission."
  • Biocompatibility testing was performed according to ISO 10993-1 and 510(k) Memorandum G95-1.
  • Compliance with electrical safety standards included IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, and IEC 60601-2-2.

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not specified in the provided text. Bench testing typically involves a number of units or measurements, but specific quantities are not given here.
  • Data Provenance: Not specified. Bench tests are usually conducted in laboratories by the manufacturer or third-party testing facilities. The country of origin of testing data is not mentioned, and the testing is inherent to the device's design and manufacturing rather than involving patient data. This is a retrospective analysis of the device's physical and electrical properties, not clinical data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not Applicable. This is a traditional medical device (electrosurgical system), not an AI/ML device requiring expert ground truth for interpretation of outputs like images. Bench testing involves objective measurements against predefined specifications and standards.

4. Adjudication method for the test set:

  • Not Applicable. See point 3. Testing involves objective measurements, not subjective expert reviews that would require adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. This is not an AI/ML device. No MRMC study was conducted or is relevant for this type of device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not Applicable. This is not an AI/ML device. Standalone performance for an algorithm is not relevant. The device's "standalone" performance refers to its ability to meet specifications when operated.

7. The type of ground truth used:

  • For the technical performance of the device (generator output, thermal output, electrical safety): Predefined specifications and recognized international or national safety and performance standards (e.g., IEC 60601 series). These are objective criteria.
  • For biocompatibility: ISO 10993 series standards define the "ground truth" for material safety.

8. The sample size for the training set:

  • Not Applicable. This is not an AI/ML device. There is no "training set" in the context of machine learning.

9. How the ground truth for the training set was established:

  • Not Applicable. See point 8.

{0}------------------------------------------------

JAN - 3 2014

K130606 pg 1 of 4

510(k) SUMMARY

FMwand Ferromagnetic Surgical System Date of Summary: March 6, 2013

General Provisions

510(k) Owner's Name:Domain Surgical, Inc.
Address:1370 South 2100 East
Salt Lake City, Utah 84108
Contact Person:Curtis Jensen, Vice President of Quality and Regulatory Affairs
Phone Number:(801) 924-4958
Fax Number:(801) 924-4951
Classification Name:Electrosurgical cutting and coagulation device and accessories (21 CFR §878.4400, Product Code GEI, HGI)
Proprietary Name:FMwand Ferromagnetic Surgical System
Common Name:Electrosurgical cutting and coagulation device and accessories

Name of Predicate Device(s)

Domain Surgical System, (Product Code GEI, HGI) 510(k) #K121881 .

Device Description

The FMwand Ferromagnetic Surgical System is a soft tissue cutting and coagulation device that consists of the FMwand Generator that is electrically connected to the sterile, single-use FMwand Handpiece by the FMwand Power Module (a resterilizable, reusable accessory cable) and a single-use, sterile air line. Like the predicate device, the handpiece includes actuation buttons that can be used to activate the handpiece tip. Also like the predicate device, an optional footpedal can be used to activate the handpiece. The connectors at both ends of the FMwand Power Module are designed to fit only the FMwand Generator and FMwand sterile, disposable accessories. The FMwand Power Module is provided non-sterile. It is intended to be cleaned and sterilized prior to the initial use and before each subsequent reuse.

Technological Comparison

Both the FMwand Ferromagnetic Surgical System and the Domain Surgical System employ ferromannetic induction to generate heat in the handpiece tip for the purpose of cutting and coagulation of soft tissue. Both devices offer active cooling of the handpiece by circulating air through the handpiece and air line.

The FMwand Ferromagnetic Surgical System handpiece consists of a disposable, single-patient use portion (the FMwand Handpiece) and a reusable, user-resterilizable connection cable (FMwand Power Module), whereas the Domain Surgical System Handpiece is a one piece disposable accessory.

The FMwand Ferromagnetic Surgical System differs from the predicate device in that it offers a second button on the handpiece as well as a second footpedal to allow the ability to quickly switch two user-selected power levels when desired.

{1}------------------------------------------------

K130606

Device Comparison Tablepg 2 of 4
PerformanceFeatureFMwand FerromagneticSurgical SystemDomain Surgical SystemDiscussion
ManufacturerDomain Surgical Inc.Domain Surgical Inc.Identical
510(k) NumberTo be assignedK121881N/A
Prescription/ OTCPrescription OnlyPrescription OnlyIdentical
Product CodeGELIGIGELIGIIdentical
Classification21 CFR 878.440021 CFR 878.4400Identical
Regulation
Basis IntendedUseCutting and coagulation of softtissueCutting and coagulation of softtissueIdentical
Indications forUseThe FMwand FerromagneticSurgical System is indicated forcutting and coagulation of softtissue during surgicalprocedures, includingGynecologio procedures (opeлtransabdominal only).The Domain Surgical System isindicated for cutting andcoagulation of soft tissue duringsurgical procedures includingGynecologic procedures (opentransabdominal only).Identical
Heat GenerationMethodFerromagnetic induction heatingprovides an elevatedtemperature blade which willcauterize tissue as it cutsFerromagnetic induction heatingprovides an elevatedtemperature blade which willcauterize tissue as it cuts.Identical
PowerRequirements100-240 VAC, 50-60 Hz100-240 VAC, 50-60 HzIdentical
Generator SizeApproximately 4" x 14" x 11"Approximately 4" x 14" x 11"Identical
Generator WeightApproximately 10 lbs. (4.5 kg)Approximately 10 lbs. (4.5 kg)Identical
Maximum OutputEnergy5 to 60 W155 Volts maximum5 to 60 W166 Volts maximumIdentical
MaximumOperationalTemperature450°C450°CIdentical
OperationalControl MethodControlled power delivered to thehandpiece tipControlled power delivered tothe handpiece tipIdentical
Mode ofOperationIntermittent OperationIntermittent OperationIdentical
Active CoolingAir cooled handpieceAir cooled handpieceIdentical
Output TypeType CFType BF minimumLeakage current restrictions for aType CF rating are morestringent than for a Type BFrating.
OperatingFrequencyPure Sinusoidal Waveform(40.68 MHz)Pure Sinusoidal Waveform(40.68 MHz)Identical
Delivery systemconfigurationLengthDiameterTips of various lengths.Tips of various lengths.Identical
HandpieceConfigurationHandpiece consists of adisposable, single-patient usehandpiece section connected toa reusable cable (FMwandPower Module) section.Handpiece consists of adisposable, single-patient usehandle with a disposable,single-patient use tip.The internal circuitry as well asthe tip function is equivalent inthese two configurations. Thechange is to the form of thehandpiece and to the sterilizationmethods.
HandpieceActuation ButtonConfigurationTwo actuation buttonscorresponding to dual user-selected high/low power levels.One actuation buttoncorresponding to a single user-selected power level.Having two power settingsavailable for actuation at thehandpiece is for userconvenience. Single buttonhandpiece is still supported.
Tip configurationsVarious handpiece tip shapes.Various handpiece tip shapes.Identical
FootpedalConfigurationTwo actuation pedalscorresponding to dual user-selected high/low power levels.One actuation pedalcorresponding to a single user-selected power level.With two power levels available,a dual footpedal is available forthose users who prefer footpedalactuation of the device. Theinterface still supports a singlefootpedal if desired.
PerformanceFeatureFMwand FerromagneticSurgical SystemDomain Surgical SystemDiscussion
Generator UserInterfaceUser interface has addedcapability to support the dualuser-selected power levels.User interface can only supporta single user-selected powerlevel.With two power levels available, the generator needs the ability toallow the user to select a low and high power setting(FMmin/FMmax). The interface still supports single-buttonhandpieces and singlefootswitches.
TreatmentModalityCutting and/or coagulation usingconducted heat from an elevatedtemperature handpiece tip.Cutting and/or coagulation usingconducted heat from anelevated temperature handpiecetip.Identical
Bench TestingThe FMwand FerromagneticSurgical System passed allbench tests performed. Detailsare found in Section 15: of thissubmission.The Domain Surgical Systempassed all bench testsperformed. Details are found inthe original 510(k) submission,K121881Identical
Meets applicableportions ofIEC 60601-2-2Yes, Details are found in Section14: of this submission.Yes, Details are found in theoriginal 510(k) submission,K121881Identical
BiocompatibilityTestsMaterials used in the patient-contacting portions of theFMwand Ferromagnetic SurgicalSystem are either known to bebiocompatible or have passedtesting performed according toISO 10993-5 (Cytotoxicity),10993-10 (Acute SystemicToxicity) and 10993-11(Sensitization and Irritation).Details are found in Section 12:of this submission.Materials used in the patient-contacting portions of theDomain Surgical System areeither known to bebiocompatible or have passedtesting performed according toISO 10993-5 (Cytotoxicity),10993-10 (Acute SystemicToxicity) and 10993-11(Sensitization and Irritation).Details are found in the original510(k) submission, K121881Identical
GeneratorSterilizationMethodGenerator is not intended to beprovided sterile and is notintended to be sterilized by theuser.Generator is not intended to beprovided sterile and is notintended to be sterilized by theuser.Identical
Handpiece/TipSterilizationHandpiece/Tip are single-patientuse and provided sterile.Sterilization method is EthyleneOxide: SAL 10-6.Connection cable is not providedsterile and is intended to becleaned and sterilized by theuser prior to initial use and eachsubsequent use up to amaximum of 100 uses.Handpiece/Tip, including cableand connector are single-patientuse and provided sterile.Sterilization method is EthyleneOxide: SAL 10-6.The sterilization method of thepredicate device has beenvalidated to an SAL of 10-6. Thesubject device cleaning andsterilization methods will bevalidated to an SAL of 10-6.These are equivalent.

{2}------------------------------------------------

K130606

pg 3 of 4

Figure 5-1: Device Comparison Table

Indications for Use

Indications for Use: The FMwand Ferromagnetic Surgical System is indicated for cutting and coagulation of soft tissue during surgical procedures, including Gynecologic procedures (open, transabdominal only).

The indications statement for the FMwand Ferromagnetic Surgical System is the identical to that of the predicate device.

Performance Testing Data Summary

Questions of safety and effectiveness are the same for this device as they are for the predicate devices and other similar devices on the market. All applicable bench testing was performed with the FMwand Ferromagnetic Surgical System to assure that it functions as intended. Bench tests verified that the generator output was within specification and that the thermal output of the handpiece and tip corresponded to the target values.

{3}------------------------------------------------

K130606 pg 4 of 4

Safety and Biocompatibility Summary

The patient contacting materials used in the FMwand Reusable Power Module were chosen for their biocompatibility, function and suitability for the intended use of this device. All necessary biocompatibility testing of applicable parts of the system was successfully completed by accredited independent testing laboratories according to ISO 10993-1 and 510(k) Memorandum G95-1.

It was also tested by an accredited independent testing laboratory to assure that it complies with the applicable electrical safety standards for medical electrosurgical devices, including the applicable sections of IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, and IEC 60601-2-2.

Conclusion

The FMwand Ferromagnetic System is substantially equivalent to the Domain Surgical System (510(k)# K121881). The intended use of the FMwand Ferromagnetic Surgical System is the same as the predicate device. The FMwand Ferromagnetic Surgical System differs from the predicate device; however the differences do not raise different types of questions of safety and effectiveness. The information presented in this premarket submission, including the bench testing, demonstrates that the FMwand Ferromagnetic System is as safe and effective as the predicate device for its intended use and functions as well as or better than the predicate device.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular fashion around the caduceus symbol. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 3, 2014

DOMAINTM Surgical Curtis Jensen Vice President of Quality and Regulatory Affairs 1370 South 2100 East Salt Lake City, UT 84108

Re: K130606

Trade/Device Name: FMwand Ferromagnetic Surgical System Regulation Number: 21 CFR§ 884.4120 Regulation Name: Gynecologic electrocautery and accessories Regulatory Class: II Product Code: HGI Dated: December 3, 2013 Received: December 4, 2013

Dear Curtis Jensen,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{5}------------------------------------------------

Page 2 - Curtis Jensen

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/5/Picture/6 description: The image shows the text "Benjamin Fisher -S". The text is in a bold, sans-serif font. The word "Benjamin" is on the left, followed by "Fisher" and then "-S". The letters "FDA" are superimposed over the word "Benjamin".

Benjamin R. Fisher. Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known): K130606

Device Name: FMwand Ferromagnetic Surgical System

ﻌ ﻴﺔ

ﺮ ﺩ

Indications for Use: The FMwand Ferromagnetic Surgical System is indicated for cutting and coagulation of soft tissue during surgical procedures, including Gynecologic procedures (open, transabdominal only).

Prescription Use __ × AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use: (21 CFR 801 Subpart C)

(Please do not write below this line - Continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Benjamin R. Fisher -S 2014.01.03 13:19:31 -05'00'

Section 4:- Indications for Use Statement

Page 4-1

§ 884.4120 Gynecologic electrocautery and accessories.

(a)
Identification. A gynecologic electrocautery is a device designed to destroy tissue with high temperatures by tissue contact with an electrically heated probe. It is used to excise cervical lesions, perform biopsies, or treat chronic cervicitis under direct visual observation. This generic type of device may include the following accessories: an electrical generator, a probe, and electrical cables.(b)
Classification. Class II (performance standards).