(385 days)
Indicated for use in patients with a grossly rotator cuff deficient shoulder joint with severe arthropathy or a previously failed joint replacement with a grossly rotator cuff deficient shoulder joint. The patient's joint must be anatomically and structurally suited to receive the selected implant(s), and a functional deltoid muscle is necessary to use the device. The glenoid baseplate is intended for cementless application with the addition of screws for fixation. The humeral stem is intended for cemented use only.
The RSP Modular Glenoid Baseplate is a multiple piece construct as opposed to a single piece construct as the predicate Reverse Shoulder Prosthesis Baseplate (K041066). The RSP Modular Glenoid Baseplate is fabricated from wrought/forged Ti-6A 1-4V that conforms to ASTM F136. The bottom surface of the baseplate is porous coated with commercially pure titanium (ASTM F67 grade 2) beads. This is the same coating previously cleared under K974294. The baseplate is Hydroxyapatite coated that conforms to ASTM F1185. The distal surface of the glenoid baseplate is porous coated with an incorporated 6.5mm cancellous screw and is intended to be used with four peripheral screws (3.5mm non-locking and/or 5mm locking and non-locking) for additional fixation. The RSP Modular Glenoid Baseplate female taper of the adapter have geometry identical to those used in the predicate Turon Shoulder (K080402).
Here's an analysis of the provided text regarding the acceptance criteria and study for the RSP Modular Glenoid Baseplate, organized to address your specific points:
Acceptance Criteria and Device Performance
| Acceptance Criteria Category | Specific Criteria (Implicitly from "as safe, as effective, and performs as well as the predicate device") | Reported Device Performance |
|---|---|---|
| Mechanical Fatigue | Comparable fatigue life to predicate devices (K041066 and K080402) | "Results of non-clinical testing demonstrate that the device is as safe, as effective, and performs as well as the predicate device." |
| Dynamic Performance | Comparable dynamic performance to predicate devices (K041066 and K080402) | "Results of non-clinical testing demonstrate that the device is as safe, as effective, and performs as well as the predicate device." |
| Biocompatibility | Materials conform to established ASTM standards (Ti-6A1-4V to ASTM F136; pure titanium to ASTM F67 grade 2; Hydroxyapatite to ASTM F1185). | Explicitly stated: "fabricated from wrought/forged Ti-6A 1-4V that conforms to ASTM F136. ... porous coated with commercially pure titanium (ASTM F67 grade 2) beads. ... Hydroxyapatite coated that conforms to ASTM F1185." |
| Fixation Strength | Adequate fixation comparable to predicate devices. | The design includes "an incorporated 6.5mm cancellous screw and is intended to be used with four peripheral screws (3.5mm non-locking and/or 5mm locking and non-locking) for additional fixation," implying a design to meet fixation needs. |
| Sterilization | Sterilization method is equivalent to predicate devices. | "Features comparable to predicate devices include the same indications, sterilization, packaging and intended use." |
| Packaging | Packaging is equivalent to predicate devices. | "Features comparable to predicate devices include the same indications, sterilization, packaging and intended use." |
Study Information:
Based on the provided text, the device clearance was based on non-clinical testing only. There were no clinical studies performed. The "study" mentioned below refers to the non-clinical testing described.
-
Sample Size Used for the Test Set and Data Provenance:
- Sample Size: Not explicitly stated. The text only mentions "Mechanical fatigue testing, Dynamic testing." The specific number of devices tested for fatigue and dynamic performance is not provided.
- Data Provenance: Not applicable as this was non-clinical (laboratory) testing, not human data.
-
Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
- Not applicable. As this was non-clinical mechanical testing, "experts" in the sense of clinical reviewers establishing ground truth for patient data are not relevant. The ground truth would be established by engineering standards and measurements.
-
Adjudication Method for the Test Set:
- Not applicable. This was non-clinical mechanical testing, not a clinical study requiring adjudication of expert opinions.
-
If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:
- No, an MRMC comparative effectiveness study was not done. The document explicitly states "Clinical Testing: None provided." This type of study is for comparing human reader performance with and without AI assistance, which is not relevant for a mechanical device clearance based on non-clinical testing.
-
If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done:
- No, a standalone algorithm performance study was not done. This device is a physical implant, not an AI algorithm.
-
The Type of Ground Truth Used:
- For the non-clinical testing, the "ground truth" would be established by engineering standards and measurements. This includes material specifications (ASTM F136, F67 grade 2, F1185), as well as the measured fatigue life and dynamic performance data compared against the performance of predicate devices.
-
The Sample Size for the Training Set:
- Not applicable. There was no "training set" in the context of an AI/machine learning model, as this device uses non-clinical mechanical testing for clearance.
-
How the Ground Truth for the Training Set Was Established:
- Not applicable. There was no training set.
{0}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized human figure with three lines representing the head, body, and legs. The figure is positioned within a circular border that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a sans-serif font. The text is arranged around the top half of the circle, with the figure centered below.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 2, 2016
Encore Medical, L.P. Mr. William Garzon Regulatory Affairs Specialist 9800 Metric Boulevard Austin, Texas, 78758
Re: K130550
Trade/Device Name: RSP Modular Glenoid Baseplate Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PHX, KWS Dated: February 13, 2014 Received: February 18, 2014
Dear Mr. Garzon:
This letter corrects our substantially equivalent letter of March 24, 2014.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
K130550 510(k) Number (if known):
Device Name: RSP Modular Glenoid Baseplate
Indications for Use:
Indicated for use in patients with a grossly rotator cuff deficient shoulder joint with severe arthropathy or a previously failed joint replacement with a grossly rotator cuff deficient shoulder joint. The patient's joint must be anatomically and structurally suited to receive the selected implant(s), and a functional deltoid muscle is necessury to use the device. The glenoid baseplate is intended for comentless application with the addition of screws for fixation. The humeral stem is intended for cemented use only.
Prescription Usc X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE.CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Casey L. Hanley, Ph.D.
Division of Orthopedic Devices
Page 1 of 1
{2}------------------------------------------------
Page 2 - Mr. William Garzon
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Lori A. Wiggins -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
MAR 2 4 2014
510(k) Summary
Date: March 1, 2013
Manufacturer: Encore Medical, L.P. Trade Name: DJO Surgical 9800 Metric Blvd Austin, TX 78758
Contact Person: William Garzon Regulatory Affairs Specialist Phone: (512) 834-6391 Fax: (512) 834-6313 Email: william.garzon@djoglobal.com
| Product | Classification | Product Code |
|---|---|---|
| RSP Modular Glenoid Baseplate | Class II | KWS |
| Product Code | Regulation and Classification Name |
|---|---|
| KWS | Shoulder joint metal/polymer semi-constrained prosthesis per 21 CFR 888.3660 |
Description:
The purpose of this application is to include a new modularity option for the existing RSP baseplate. The RSP Modular Glenoid Baseblate is a multiple piece construct as opposed to a single piece construct as the predicate Reverse Shoulder Prosthesis Baseplate (K041066).
The RSP Modular Glenoid Baseplate is fabricated from wrought/forged Ti-6A 1-4V that conforms to ASTM F136. The bottom surface of the baseplate is porous coated with commercially pure titanium (ASTM F67 grade 2) beads. This is the same coating previously cleared under K974294. The baseplate is Hydroxyapatite coated that conforms to ASTM F1185. The distal surface of the glenoid baseplate is porous coated with an incorporated 6.5mm cancellous screw and is intended to be used with four peripheral screws (3.5mm non-locking and/or 5mm locking and non-locking) for additional fixation.
The RSP Modular Glenoid Baseplate female taper of the adapter have geometry identical to those used in the predicate Turon Shoulder (K080402).
There are no changes to sterilization, packaging, indications or intended use.
Indications for Use:
Indicated for use in patients with a grossly rotator cuff deficient shoulder joint with severe arthropathy or a previously failed joint replacement with a grossly rotator cuff deficient shoulder joint. The patient's joint must be anatomically and structurally suited to receive the selected implant(s), and a functional deltoid muscle is necessary to use the device. The glenoid baseplate is intended for cementless application with the addition of screws for fixation. The humeral stem is intended for cemented use only.
Predicate Device:
Comparable Features to Predicate Device(s):
{4}------------------------------------------------
Features comparable to predicate devices include the same indications, sterilization, packaging and intended use.
Non-Clinical Testing:
Mechanical fatigue testing, Dynamic testing. Results of non-clinical testing demonstrate that the device is as safe, as effective, and performs as well as the predicate device.
Clinical Testing: None provided
N/A