K Number
K130476
Manufacturer
Date Cleared
2014-01-08

(317 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Dental Morelli Orthodontic Dental Implant is a threaded titanium dental implant screw intended to serve as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. The device is used temporarily and is removed after orthodontic treatment.

Device Description

The Orthodontic Miniscrew for Absolute Anchorage is an implantable medical device used as a temporary skeletal anchorage point for orthodontic movements. The mini screw is intended to receive low intensity static clinical loads. Maximum dynamic forces may occur at low frequencies for short periods; which is inadequate to compromise the performance of the product through fatigue. The device is available in different lengths (6, 8, 10, mm), diameter 1.5 mm, and supplied in sealed sterile polyethylene bags.

AI/ML Overview

The provided document is a 510(k) summary for the "Orthodontic Miniscrew for Absolute Anchorage" and focuses on demonstrating substantial equivalence to a predicate device, the "Aarhus Anchorage System K041527." It does not describe a study involving a comparison of device performance against detailed acceptance criteria in the manner typically seen for novel device efficacy. Instead, it relies on demonstrating equivalence through material properties, dimensions, intended use, and general performance characteristics.

Therefore, many of the requested sections (e.g., sample size for test set, number of experts for ground truth, MRMC study, standalone performance) are not applicable or cannot be extracted from this document, as the submission primarily focuses on equivalence based on established standards and similarity to a predicate.

Here's an analysis based on the information available in the provided text:


Acceptance Criteria and Device Performance

The acceptance criteria are implicitly defined by compliance with established material and sterilization standards, and by demonstrating performance comparable to the predicate device.

Acceptance Criterion (Implicit)Reported Device Performance
Material Composition (conforms to surgical implant standards)Titanium Alloy ASTM F 136 (same as predicate). Adheres to chemical, mechanical, and metallurgical requirements for wrought annealed titanium-6aluminum-4vanadium ELI (extra low interstitial) alloy (R56401).
BiocompatibilityTitanium Alloy in medical grade according to ASTM-F 136 is accepted for endosseous implant (same as predicate).
SterilityGamma Radiation (same as predicate). Conforms to ISO 11137 (requirements for developing, validating, and routinely controlling sterilization process). Conforms to ASTM F1980-07 (addresses loss of sterile barrier system integrity).
Mechanical Safety/PerformanceTensile strength of material according to ASTM - F 136 is accepted for endosseous implants (same as predicate). Performance testing conducted to determine torque levels, forces, and insertion lengths. Measured performance is "comparable to the Aarhus Anchorage System K041527."
Dimensions (comparable to predicate)Diameter ranges from 1.5 mm, length ranges from 6-10 mm. (Predicate: Diameter ranges from 1.5-2 mm, length ranges from 9-11 mm). The submission states "similar dimensions."
Performance Characteristics (comparable to predicate)Self-trapping/Self-drilling (same as predicate).
Intended Use / Indications for UseThreaded titanium dental implant screw intended to serve as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. Used temporarily. (Identical to predicate).
Substantial Equivalence Conclusion"in all respects based on Intended Use, Indications of Use, Construction Material, Performance and Bench Tests that the Orthodontic Miniscrew for Absolute Anchorage and the Aarhus Anchorage System K041527 are substantially equivalent." Specifically, "Potential adverse effects are identical." and "made from the same material ... and has similar dimensions and characteristics."

2. Sample Size Used for the Test Set and Data Provenance

The document describes "Performance Testing" which included "testing was done to determine the torque levels, forces and insertion lengths while inserting the Orthodontic Miniscrew for Absolute Anchorage." However, no specific sample sizes (N) for these tests are mentioned.

The data provenance is implied to be from bench testing conducted by Dental Morelli Ltda. The country of origin for the device manufacturer is Brazil. The data itself would be from laboratory experiments, not patient data, in this context of a 510(k) submission for mechanical equivalence.


3. Number of Experts Used to Establish Ground Truth and Qualifications

Not Applicable. This submission is for a medical device (orthodontic miniscrew) and focuses on demonstrating substantial equivalence through material and mechanical properties, as well as adherence to standards. It does not involve a diagnostic algorithm or a scenario where expert-established ground truth is required in the context of a "test set" as one might find in an AI/ML medical device submission. The "ground truth" here is adherence to engineering standards and established performance of the predicate.


4. Adjudication Method for the Test Set

Not Applicable. See point 3. No clinical or diagnostic "test set" requiring expert adjudication is described.


5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

No. This document does not mention any MRMC comparative effectiveness study. This type of study is typically used for diagnostic devices to assess human reader performance with and without AI assistance. This device is a physical implant.


6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not Applicable. This is a physical medical device, not an algorithm or AI system.


7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the performance testing mentioned (torque levels, forces, insertion lengths), the ground truth is based on instrumentation measurements comparing the new device against the performance characteristics of the predicate device. For material and sterilization claims, the ground truth is demonstrated by conformance to established international standards (ASTM F136, ASTM F1980-07, ISO 11137). There is no "expert consensus" or "pathology" in the typical sense for this submission regarding performance.


8. The sample size for the training set

Not Applicable. This is not an AI/ML device, so there is no concept of a "training set."


9. How the ground truth for the training set was established

Not Applicable. See point 8.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows a logo for Morelli Ortodontia. The logo features a stylized symbol above the text. The symbol is an abstract design with intersecting lines and curved elements, resembling a stylized "X". Below the symbol, the word "Morelli" is written in a bold, stylized font, and below that, the word "Ortodontia" is written in a smaller, simpler font.

Dental Morelli Ltda.

SECTION 7

510(k) SUMMARY

Proprietary Name

Date Prepared

Orthodontic Miniscrew for Absolute Anchorage

December 1, 2013

Submitter

DENTAL MORELLI LTDA Alameda Jundiaí, 230 -Jardim Saira - Sorocaba CEP: 18085-090 Brazil Telephone: 55 (15) 3238-8200

Official Contact

Lilian Llull / Tara Conrad TechLink International Consulting 18851 NE 29th Avenue Suite 720 Aventura, FL 33180 TEL- (305) 377-0077

Common Name Classification Name Requlation Number & Product Codes Proposed Requilatory Class Predicate Device Identification

Orthodontic Mini Implants Endosseous Orthopedic Implants OAT - 21 CFR 872.3640 Class II Aarhus Anchorage System K041527

Description of Proposed Device

The Orthodontic Miniscrew for Absolute Anchorage is an implantable medical device used as a temporary skeletal anchorage point for orthodontic movements. The mini screw is intended to receive low intensity static clinical loads. Maximum dynamic forces may occur at low frequencies for short periods; which is inadequate to compromise the performance of the product through fatigue. The device is available in different lengths (6, 8, 10, mm), diameter 1.5 mm, and supplied in sealed sterile polyethylene bags.

Intended Use

The Orthodontic Miniscrew for Absolute Anchorage is threaded titanium dental imo ort screws intended to serve as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. The device is used temporarily and is removed after orthodontic treatment.

Page 12 of 115 K130476

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows a logo with a stylized "X" shape at the top. Below the "X" is the word "MORELLI" in a bold, stylized font. Underneath "MORELLI" is the word "ORTODONTIA" in a smaller, less bold font.

Device Comparison

.

.

Orthodontic Miniscrew for Absolute Anchorage by DentalMorelli K130476Aarhus Anchorage SystemK041527
Indications of UseThreaded titanium dental implant screw intended to serve as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. The device is used temporarilyThreaded titanium dental implant screw intended to serve as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. The device is used temporarily
Target PopulationPatients in need of teeth alignment correctionPatients in need of teeth alignment correction
Anatomical SitesJawJaw
Location of Use, hospital, home, ambulance, etcUse only by professional dentists / orthodontistsUse only by professional dentists / orthodontists
Diameter and LengthDiameter ranges from 1.5 mm, length ranges from 6-10 mmDiameter ranges from 1.5-2 mm, length ranges from 9-11 mm
PerformanceSelf-trapping/Self-drillingSelf-trapping/Self-drilling
MaterialsTitanium Alloy ASTM F 136Titanium Alloy ASTM F 136
BiocompatibilityTitanium Alloy in medical grade according to ASTM-F 136 is accepted for endosseous implantTitanium Alloy in medical grade according to ASTM-F 136 is accepted for endosseous implant
Compatibility with the environment and other devicesTitanium Alloy in medical grade according to ASTM-F 136 is accepted for endosseous implantTitanium Alloy in medical grade according to ASTM-F 136 is accepted for endosseous implant
SterilityGamma RadiationGamma Radiation
Mechanical SafetyTensile strength of material according to ASTM - F 136 is accepted for endosseous implantsTensile strength of material according to ASTM - F 136 is accepted for endosseous implants

·

.

{2}------------------------------------------------

Dental Morelli Ltda.

Image /page/2/Picture/1 description: The image shows a logo with the word "MORELLI" in a bold, stylized font above the word "ORTODONTIA" in a smaller, less bold font. Above the words is a geometric design that resembles an "X" shape, with each arm of the "X" having a block at the end and a curved section in the middle. The logo appears to be for a company or product related to orthodontics.

Substantial Equivalence

The Orthodontic Miniscrew for Absolute Anchorage is made from the same material as the predicate Aarhus Anchorage System K041527 and has similar dimensions and characteristics. Micro Titanium Plate System following ASTMF 136-98 which is generally used in this kind of bone screw therefore no further testing was conducted.

All reports show that the Orthodontic Miniscrew for Absolute Anchorage is substantially equivalent in design, material, intended use and function to Aarhus Anchorage System K041527. They are made of the same material and have similar dimensions and characteristics. Potential adverse effects are identical. Both devices are manufactured from titanium ASTMF 136-98 which is generally used in this kind of bone screw.

Performance Testing

The Orthodontic Miniscrew for Absolute Anchorage conforms to the following . standards:

  • ASTM F136 .
  • ASTM F1980-07 .
  • ISO 11137 �

ASTM F136: This specification covers the chemical, mechanical, and metallurgical reguirements for wrought annealed titanium-6aluminum-4vanadium ELI (extra low interstitial) alloy (R56401) to be used in the manufacture of surgical implants. The products are classified into: strip, sheet, plate, bar, forging bar, and wire. The heat analysis shall conform to the chemical composition requirements specified. Product analysis tolerances do not broaden the specified heat analysis requirements but cover variations between laboratories in the measurement of chemical content. Tension test and bend test shall be performed to meet the requirements specified.

ASTM F1980-07: The loss of sterile barrier system integrity may occur as a result of physical properties of the materials and adhesive or cohesive bonds degrading over time and by subsequent dynamic events during shipping and handling.

ISO 11137: Sterilization of Health Care Products Package provides the requirements for developing, validating and routinely controlling the sterilization process of medical devices. In addition to providing the requirements for the sterilization process of medical devices, this package also considers the

Page 14 of 115
K130476

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a logo for Morelli Ortodontia. The logo consists of a stylized "X" shape above the word "MORELLI" in a bold, sans-serif font. Below "MORELLI" is the word "ORTODONTIA" in a smaller, sans-serif font. The logo is black and white.

Dental Morelli Ltda.

Page 15 of 115 K130476

products' environment, personnel and their hygiene, packaging / storage, sterilization doses and more to inactivate microbiological contaminants on medical devices.

Conclusion

Testing was done to determine the torque levels, forces and insertion lengths while inserting the Orthodontic Miniscrew for Absolute Anchorage. The level of torque is comparable to the Aarhus Anchorage System K041527. Which were investigated with the measurement equipment under the same parameter. In conclusion it can be said that in all respects based on Intended Use, Indications of Use, Construction Material, Performance and Bench Tests that the Orthodontic Miniscrew for Absolute Anchorage and the Aarhus Anchorage System K041527 are substantially equivalent.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 8, 2014

DENTAL MORELLI LTDA C/O Ms. Lillian Llull Senior Partner TechLink International Consulting 18851 NE 29th Avenue Suite 720 Aventura, FL 33180

Re: K130476

Trade/Device Name: Orthodontic Miniscrew for Absolute Anchorage Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: OAT Dated: December 1, 2013 Received: December 6, 2013

Dear Ms. Llull:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register,

{5}------------------------------------------------

Page 2 - Ms. Llull

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Image /page/5/Picture/7 description: The image shows the name "Kwame O. Ulmer-S" in bold, black font. The text is arranged in two lines, with "Kwame O." on the first line and "Ulmer-S" on the second line. There is some noise and artifacts around the text, possibly due to the image quality or scanning process.

for

Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use Statement

510(k) Number: K130476

Device Name: Orthodontic Miniscrew for Absolute Anchorage

Indications for Use:

Dental Morelli Orthodontic Dental Implant is a threaded titanium dental implant screw intended to serve as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. The device is used temporarily and is removed after orthodontic treatment.

X Prescription Use _ (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEED

Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1

Runner -S 00 TO 14 07 09 13-54-51-05-00

Page 21 of 115 K130476

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.