K Number
K130434
Manufacturer
Date Cleared
2013-05-01

(69 days)

Product Code
Regulation Number
888.3080
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Octane Straight Intervertebral Fusion Device is intended for spinal fusion procedures at one or two contiguous levels in the lumbar spine from L2 to S1 in patients with Degenerative Disc Disease (DDD,) with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin, with degeneration of the disc confirmed by history and radiographic studies. Patients should be skeletally mature, and have had at least 6 months of non-operative treatment. The device is intended for use with autogenous graft, and with supplemental fixation systems cleared for use in the lumbosacral spine.

Device Description

Proposed Octane Straight Intervertebral Fusion Device components (Figure 1) are line extensions of the Octane interbody fusion device system cleared for US distribution per 510(k) #K082270.

AI/ML Overview

The provided text is a 510(k) summary for a medical device, the Exactech® Octane® Straight Intervertebral Fusion Device. It focuses on demonstrating substantial equivalence to a predicate device, primarily through non-clinical performance data and a comparison of technological characteristics. It is not an AI or diagnostic device, so many of the requested criteria are not applicable.

Here's an analysis of the provided information, noting where specific criteria cannot be met:

Acceptance Criteria and Reported Device Performance

The device is an intervertebral fusion device, and its performance is evaluated based on mechanical integrity and subsidence, not diagnostic accuracy or AI-driven metrics. The "acceptance criteria" in this context are the standards for mechanical testing, and the "reported device performance" is that it met these standards, demonstrating substantial equivalence to its predicate.

Acceptance Criteria (Evaluations)Reported Device Performance (Activities Performed)
Loss of mechanical integrity: fracture/breakageStatic compression testing per ASTM F2077
Dynamic compression testing per ASTM F2077
Static compressive shear testing per ASTM F2077
Implant subsidenceSubsidence Yield Force per ASTM F2267

Note: The summary states that "Comparison analysis and results of engineering studies referenced in this 510(k) submission demonstrate proposed Octane Straight Intervertebral Fusion Device implants are substantially equivalent to cited predicate devices cleared for US distribution." This implies that the device met or exceeded the performance of the predicate device in these mechanical tests. Specific numerical thresholds for "acceptance" and detailed numerical "performance" results are not provided in this summary but would be part of the full submission.

Study Information (Not applicable for this type of device)

This is a physical implant, not a diagnostic or AI device, thus many of the requested study parameters (sample size for test set, data provenance, expert-established ground truth, adjudication, MRMC studies, standalone performance, training set details) are not applicable. The "study" here refers to non-clinical, mechanical testing.

1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

  • Not Applicable. The "test set" for this device refers to physical prototypes or samples of the implant undergoing mechanical testing. The document does not specify the number of samples used for each test, nor is "data provenance" (country/retrospective/prospective) relevant in the same way it would be for clinical or AI data. These were laboratory tests.

2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

  • Not Applicable. Ground truth in this context refers to the defined mechanical standards (ASTM standards). The "experts" are implicitly the engineering and testing professionals who conducted and interpreted these standardized tests.

3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not Applicable. Adjudication methods are typically for clinical or diagnostic interpretations. For mechanical testing, adherence to established ASTM methodologies and passing predefined criteria are the "adjudication."

4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not Applicable. This device is an implant, not an AI or diagnostic tool, so MRMC studies are irrelevant.

5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not Applicable. This is a physical device, not an algorithm.

6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • Engineering Standards/Laboratory Testing Parameters: The "ground truth" for the device's performance is adherence to established ASTM (American Society for Testing and Materials) standards for orthopedic implants. These standards define the methods and acceptance criteria for evaluating mechanical properties like compression, shear, and subsidence.

7. The sample size for the training set:

  • Not Applicable. There is no "training set" for this type of physical implant device.

8. How the ground truth for the training set was established:

  • Not Applicable. There is no "training set" for this type of physical implant device.

{0}------------------------------------------------

K130434 Page 1 of 3

Exactech® Octane® Straight Intervertebral Fusion Device 510(k) Summary

Sponsor:Exactech ® Inc.
2320 N.W. 66 th Court
Gainesville, FL 32653

Phone: (352) 327-4762 Fax: (352) 378-2617

FDA Establishment Number 1038671

  • Contact: Patrick Hughes Senior Regulatory Affairs Specialist
    May 1, 2013 Date:

Trade or Proprietary or Model Name(s): Exactech® Octane® Straight Intervertebral Fusion Device

Common Name: Intervertebral body fusion device

Classification Name:

21 CFR 888.3080 - Intervertebral body fusion device

Product Code: MAX

FDA Classification: Class II

Predicate Device:

510(k) NumberTrade or Proprietary or Model NameManufacturer
K082270Octane® P Intervertebral Body Fusion DeviceExactech

Page 1 of 3

{1}------------------------------------------------

Exactech® Octane® Straight Intervertebral Fusion Device 510(k) Summary

Indications for Use:

The Octane Straight Intervertebral Fusion Device is intended for spinal fusion procedures at one or two contiguous levels in the lumbar spine from L2 to S1 in patients with Degenerative Disc Disease (DDD,) with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin, with degeneration of the disc confirmed by history and radiographic studies. Patients should be skeletally mature, and have had at least 6 months of non-operative treatment. The device is intended for use with autogenous graft, and with supplemental fixation systems cleared for use in the lumbosacral spine.

Device Description:

Image /page/1/Picture/5 description: The image shows two perspective views of a gray-colored medical implant. The implant has a rectangular base with a hollowed-out center. The top surface of the implant has a series of ridges or teeth-like structures. The implant also has a hole on the side.

Figure 1: Octane Straight Intervertebral Fusion Device implant options

Proposed Octane Straight Intervertebral Fusion Device components (Figure 1) are line extensions of the Octane interbody fusion device system cleared for US distribution per 510(k) #K082270.

Comparison of Technological Characteristics

Both predicate and proposed devices have the same intended use, basic fundamental scientific technology, and share the following similarities:

  • the same indications for use .
  • similar design features .
  • the same materials .
  • . the same 8-year shelf life
  • packaging and sterilization using the same materials and processes .

The proposed Octane Straight components are not being submitted as the result of a recall or any corrective action related to the Octane product lines.

Non-Clinical Performance Data

Table 1 shows non-clinical performance data provided, cited, or referenced in this submission to support a conclusion of substantial equivalence:

EvaluationActivities Performed
Loss of mechanical integrity: fracture/breakageStatic compression testing per ASTM F2077Dynamic compression testing per ASTM F2077Static compressive shear testing per ASTM F2077
Implant subsidenceSubsidence Yield Force per ASTM F2267

Table 1: Octane Straight Non-Clinical Performance Data

{2}------------------------------------------------

Exactech® Octane® Straight Intervertebral Fusion Device 510(k) Summary

ئ

Substantial Equivalence Conclusion:

Comparison analysis and results of engineering studies referenced in this 510(k) submission demonstrate proposed Octane Straight Intervertebral Fusion Device implants are substantially equivalent to cited predicate devices cleared for US distribution.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. The logo is presented in black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 1, 2013

Exactech®, Incorporated % Mr. Patrick Hughes Senior Regulatory Affairs Specialist 2320 Northwest 66th Court Gainesville, Florida 32653

Re: K130434

Trade/Device Name: Exactech® Octane® Straight Intervertebral Fusion Device Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: March 18, 2013 Received: March 19, 2013

Dear Mr. Hughes:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{4}------------------------------------------------

Page 2 - Mr. Patrick Hughes

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Image /page/4/Picture/8 description: The image shows the name "Erin Keith" in a stylized font. The letters "E", "r", "i", and "n" are in a bold, sans-serif font. The letters "K", "e", "i", "t", and "h" are in a similar font, but with a more decorative design.

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Exactech® Octane® Straight Intervertebral Fusion Device Special 510(k) - Indications for Use

· 510(k) Number: K130434

Device Name: Exactech® Octane® Straight Intervertebral Fusion Device

INDICATIONS

The Octane Straight Intervertebral Fusion Device is intended for spinal fusion procedures at one or two contiguous levels in the lumbar spine from L2 to S1 in patients with Degenerative Disc Disease (DDD,) with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin, with degeneration of the disc confirmed by history and radiographic studies. Patients should be skeletally mature, and have had at least 6 months of non-operative treatment. The device is intended for use with autogenous graft, and with supplemental fixation systems cleared for use in the lumbosacral spine.

Prescription Use X(Part 21 CFR 801 Subpart D)and/orOver-The-Counter Use(21 CFR 807 Subpart C)
-----------------------------------------------------------------------------------------------------------

Please do not write below this line - use another page if needed.

Concurrence of CDRH, Office of Device Evaluation (ODE)

RonaldiP: Jean -S

(Division Sign-Off)

Division of Orthopedic Devices

510(k) Number: K130434

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.