K Number
K130366
Manufacturer
Date Cleared
2013-05-15

(91 days)

Product Code
Regulation Number
870.3375
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DENALI® Filter System - Femoral and Jugular/Subclavian Delivery Kits are indicated for use in the prevention of recurrent pulmonary embolism via placement in the vena cava in the following situations:

  • Pulmonary thromboembolism when anticoagulants are contraindicated. .
  • . Failure of anticoagulant therapy for thromboembolic disease.
  • . Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced.
  • . Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated.
    DENALI® Filter may be removed according to the instructions supplied under the section labeled: Optional Procedure for Filter Removal.
Device Description

The DENALI® Filter consists of twelve Nitinol appendages emanating from a central snareable tip. These twelve appendages (six legs and six arms) form two levels of filtration for emboli: the legs provide the lower level of filtration and the arms provide the upper level of filtration. Four out of the six legs have cranial anchors and the remaining two legs have caudal anchors. In addition, all of the legs have penetration limiters. The anchors have been designed to resist cranial and caudal migration, while allowing the filter to be percutaneously removed. The DENALI® Filter is intended to be used in the inferior vena cava with diameters less than or equal to 28 mm.
The DENALI® Delivery Systems consist of an introducer sheath and dilator, and a preloaded DENALI® Filter in a storage tube with a pusher. The dilator accepts a 0.035" guidewire and allows for an 800 psi maximum pressure contrast power injection. Radiopaque marker bands are on the end of the dilator to aid in measuring the maximum indicated IVC diameter. They are spaced at a distance of 28 mm (outer-to-outer). The 55 cm, 8.4 French I.D. introducer sheath contains a radiopaque marker band at the distal tip and hemostasis valve with a side port. The pusher advances the filter through the introducer sheath to the pre-deployment mark and is then used to fix the filter in place while the filter is unsheathed.

AI/ML Overview

The DENALI® Filter System's acceptance criteria and performance are detailed across various tests, including in-vitro, in-vivo animal studies, and a clinical study.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria CategorySpecific Criteria/EndpointAcceptance GoalReported Device Performance
Clinical Efficacy (CSP)Clinical Success Placement (CSP)Lower bound of the 95% confidence interval for CSP was greater than 80%.CSP was 96.1%, and the lower bound of the 95% confidence interval was 91.2%. Goal Met.
Clinical Efficacy (TSP)Technical Success of Placement (TSP)Not explicitly quantified as a percentage goal, but satisfactory deployment.TSP was 100%. Goal Met.
Filter RetrievalRetrieval Success RateAll filters successfully removed with acceptable retrieval force (in animal study).Animal Study: All 12 filters successfully removed with acceptable retrieval force. Clinical Study: 97.7% (86/88) retrieval success rate. Goal Met.
Safety (Animal Study)Caval occlusion/thrombosisNo observed.No observed caval occlusion/thrombosis. Goal Met.
IVC penetrations/perforationsNo definitive.No definitive IVC penetrations/perforations. Goal Met.
Contrast extravasationNo extravasation from VC after filter removal.No contrast extravasation from the VC after filter removal. Goal Met.
Significant filter tiltingNo significant.No significant filter tilting. Goal Met.
Hemodynamically significant caval stenosisNo.No hemodynamically significant caval stenosis. Goal Met.
Safety (Clinical Study)Recurrent PELow incidence, non-fatal.2 / 139 (1.4%) symptomatic PEs, neither caused patient death. Goal Met.
Caval Occlusion0%0 / 137 (0%). Goal Met.
Filter Fracture0%0 / 137 (0%). Also, no findings of filter fracture in the clinical study. Goal Met.
Cranial Migration0%0 / 137 (0%). Also, no findings of cranial migration in the clinical study. Goal Met.
Caudal Migration0%0 / 137 (0%).
Filter Tilt at Placement0%0 / 175 (0%). Also, no findings of filter tilt at placement in the clinical study. Goal Met.
Filter Tilt at Retrieval0%0 / 88 (0%). Also, no findings of filter tilt at retrieval in the clinical study. Goal Met.
Filter PenetrationLow incidence, no clinical sequelae (asymptomatic).3 / 175 (1.7%) at placement, 2 / 88 (2.3%) at retrieval. Five total cases of asymptomatic penetration; none had clinical sequelae. Goal Met.
New/Worsening DVTMonitored, incidence reported.New DVT: 10 / 137 (7.3%). Worsening DVT: 3 / 137 (2.2%). All new DVTs were in patients with active disease, hypercoagulable states, or orthopedic procedures.
Delivery System AttributesVarious (Visibility, Trackability, Pushability, Deployment, Centering, Entanglement)Acceptable performance as assessed by clinical evaluator (in animal study).All assessed attributes (dilator visibility, marker band visibility, introducer sheath/marker band visibility, dilator/introducer trackability, dilator/introducer pushability, aspiration, delivery system trackability, delivery system pushability, ease of deployment, deployment accuracy, filter centering, arm/leg entanglement, filter visibility under fluoroscopy, snare tip visibility, and pusher assembly visibility) were found acceptable. Goal Met.
In-Vitro Tests(Fatigue, Corrosion, Migration, Penetration, Tensile, Removal Force, Clot Trapping, Radiopacity, MRI Compatibility, Deployment Force/Accuracy, etc.)Successful completion of testing as per risk assessment.All acceptance criteria were met. Goal Met.
Biocompatibility(Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Subacute Toxicity, Genotoxicity, Hemocompatibility)Compliant with ISO 10993.All acceptance criteria were met. Goal Met.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • Clinical Study (Test Set):

    • Sample Size: 175 patients (107 males, 68 females).
    • Data Provenance: Prospective, multi-center clinical study conducted at 20 investigational sites across the United States.
  • Animal Studies (Test Set for In-Vivo Performance):

    • Filter Assessment: 12 filter retrievals (ovine animal model).
    • Delivery System Assessment: 12 Femoral and 12 Jugular/Subclavian DENALI® Filter Systems (acute ovine animal model).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Clinical Study:

    • Independent Clinical Events Committee (CEC): Responsible for adjudicating patient deaths, all site-reported adverse events, and abnormalities from venacavograms. Specific number of experts and their detailed qualifications (e.g., years of experience) are not explicitly provided, but they are described as an "independent Clinical Events Committee."
    • Clinical Evaluators: Assessed "ease of removal" in the animal study and deemed various delivery system attributes acceptable. Qualifications are not further specified.
    • Pathologist: Confirmed findings in the animal study (e.g., no caval occlusion/thrombosis) and determined the cause of a death. Qualifications are not further specified.
  • Animal Studies:

    • Clinical Evaluator: Assessed ease of removal, delivery system attributes.
    • Pathologist: Confirmed findings for cava wall damage, caval occlusion/thrombosis, etc.
    • Attending Veterinarian: Determined the cause of a death in the animal study.
    • Specific number and detailed qualifications of these individuals are not explicitly provided.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Clinical Study: An Independent Clinical Events Committee (CEC) was used to adjudicate:

    • Patient deaths.
    • All site-reported adverse events.
    • Abnormalities from venacavograms related to the device.
      The specific method (e.g., 2+1, 3+1 for consensus) is not detailed, but the involvement of a committee implies a form of expert adjudication.
  • Animal Study: Adjudication involved clinical evaluators, a pathologist, and an attending veterinarian for various aspects. The specific method for dispute resolution or consensus is not detailed.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No MRMC comparative effectiveness study was done, and no AI component is mentioned. This submission is for a medical device (vena cava filter) and its associated delivery system, not an AI-based diagnostic or assistive software. Therefore, the questions related to human reader improvement with/without AI assistance are not applicable.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This submission is for a physical medical device (vena cava filter system), not an algorithm or AI.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Clinical Study:
    • Outcomes Data: Primary ground truth for clinical efficacy (CSP, TSP), recurrent PE, DVT, migration, fracture, tilt, and penetration were derived from patient follow-up visits, imaging (venacavograms), and clinical assessments.
    • Expert Adjudication (CEC): Played a crucial role in establishing ground truth for adverse events, device-relatedness, and complex clinical outcomes.
  • Animal Studies:
    • Gross Evaluation and Histopathology: Used to assess cava wall damage, intimal irregularities, penetration, perforation, and thrombus.
    • Venography: Used to assess cava wall damage.
    • Clinical Evaluation: Used to assess ease of removal and various delivery system attributes.

8. The sample size for the training set

  • Not explicitly mentioned as a distinct "training set" in the context of machine learning. For this physical device, the "training" equivalent would be the extensive in-vitro testing and prior experience with predicate devices that informed the DENALI® design.
    • The "training" for the device's design and manufacturing processes would be based on years of experience and data from previous iterations like the ECLIPSE® Filter System.
    • Performance testing is conducted on the final device to confirm its safety and efficacy, not to "train" an algorithm.

9. How the ground truth for the training set was established

  • Not applicable in the AI/machine learning sense. The ground truth for the development of the physical device design (if one were to consider it "training") would be based on:
    • Engineering principles and design specifications.
    • Pre-clinical (in-vitro and animal) test results for predicate devices and earlier prototypes.
    • Regulatory standards and guidance documents.
    • Clinical experience and literature for the medical need and prior devices.

{0}------------------------------------------------

Page 2

DENALI® Filter System 510(k) Summary 21 CFR 807.92

MAY 1 5 2013

As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:

Submitter Information

Applicant:Bard Peripheral Vascular, Inc1625 West 3rd StreetTempe, Arizona 85281
Phone:480-638-2906
Fax:480-449-2546
Contact:Joni Creal, Sr. Regulatory Affairs Specialist
Date:May 14, 2013

Subject Device Name

Device Trade Name:DENALI® Filter System –Jugular/Subclavian Delivery Kit
Common or Usual Name:Filter, Intravascular, Cardiovascular
Classification:Class II
Classification Panel:Cardiovascular Devices
Product Code:DTK
Predicate Devices:ECLIPSE® Filter System – Femoral and Jugular/SubclavianDelivery Kit (K101431; Clearance June 25, 2010)

Summary of Change

The ECLIPSE® Filter received FDA clearance under K101431 on June 25, 2010. As part of the product improvement life cycle, Bard Peripheral Vascular, Inc. (BPV) has chosen to re-design its vena cava filter platform. The new fitter design is named the DENALL® Filter and incorporates cranial anchors, caudal anchors, penetration limiters and will be terminally electropolished. In addition, minor changes have been made to the IFU.

Bard Peripheral Vascular, Inc.

Image /page/0/Picture/13 description: The image contains the word "BAIRD" in a bold, sans-serif font. The letters are all capitalized and evenly spaced. The word appears to be a logo or brand name.

{1}------------------------------------------------

Device Description

The DENALI® Filter consists of twelve Nitinol appendages emanating from a central snareable tip. These twelve appendages (six legs and six arms) form two levels of filtration for emboli: the legs provide the lower level of filtration and the arms provide the upper level of filtration. Four out of the six legs have cranial anchors and the remaining two legs have caudal anchors. In addition, all of the legs have penetration limiters. The anchors have been designed to resist cranial and caudal migration, while allowing the filter to be percutaneously removed. The DENALI® Filter is intended to be used in the inferior vena cava with diameters less than or equal to 28 mm.

The DENALI® Delivery Systems consist of an introducer sheath and dilator, and a preloaded DENALI® Filter in a storage tube with a pusher. The dilator accepts a 0.035" guidewire and allows for an 800 psi maximum pressure contrast power injection. Radiopaque marker bands are on the end of the dilator to aid in measuring the maximum indicated IVC diameter. They are spaced at a distance of 28 mm (outer-to-outer). The 55 cm, 8.4 French I.D. introducer sheath contains a radiopaque marker band at the distal tip and hemostasis valve with a side port. The pusher advances the filter through the introducer sheath to the pre-deployment mark and is then used to fix the filter in place while the filter is unsheathed.

Indications for Use of Device

The subject device, the DENAL® Filter System – Femoral and Jugular/Subclavian Delivery Kits, is indicated for use in the prevention of recurrent pulmonary embolism via placement in the vena cava in the following situations:

  • Pulmonary thromboembolism when anticoagulants are contraindicated. .
  • . Failure of anticoagulant therapy for thromboembolic disease.
  • . Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced.
  • Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is . contraindicated.

DENALI® Filter may be removed according to the instructions supplied under the section labeled: Optional Procedure for Filter Removal.

Image /page/1/Picture/13 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are outlined in black, giving them a distinct and prominent appearance. The overall design is simple and eye-catching, making the word easily readable.

{2}------------------------------------------------

Technological Comparison to Predicate Devices

The DENALI® Filter System – Femoral and Jugular/Subclavian Delivery Kit has the following similarities to its predicate device, the ECLIPSE® Filter System ~ Femoral and Jugular/Subclavian Delivery Kit (clearance to market via K101431 on June 25, 2010):

  • . Same intended use
  • Same indications for use .
  • . Similar filter and delivery system design
  • . Same target population
  • . Same operating principle
  • . Same fundamental scientific technology
  • . Similar packaging configuration and materials
  • . Same sterility assurance level and method of sterilization

The differences between the subject device and its predicate device are as follows:

  • Laser-cut filter from a Nitinol tube (single piece with welded snare tip) .
  • . Caudal anchors
  • . Penetration limiters (one on each filter leg)
  • Dimensional modifications .
  • . Delivery System Modifications
  • . Updated IFU

The successful completion of the testing required per the risk assessment demonstrates that the technological characteristics and performance criteria of the DENAL® Filter System - Femoral and Jugular/Subclavian Delivery Kits are comparable to the predicate device and that the subject device can perform in a manner substantially equivalent to devices currently on the market for the same intended use.

Performance Testing – In-Vitro Testing

To demonstrate substantial equivalence of the subject device to the predicate device. the technological characteristics and performance criteria were evaluated using the in-vitro testing as outlined below:

  • . In-Vitro - Filter
    • Fatigue Resistance

Image /page/2/Picture/25 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are outlined in black, giving them a three-dimensional appearance. The font is uniform in size and style, and the letters are evenly spaced.

{3}------------------------------------------------

  • । Corrosion Resistance
  • Cranial Migration Resistance l
  • -Caudal Migration Resistance
  • l Penetration Resistance (Radial force)
  • -Tensile
  • -Removal Force
  • Clot Trapping -
  • Filter Tip Visibility (Radiopacity) ।
  • । MRI Compatibility
  • In-Vitro Delivery System .
    • Deployment Force -
    • Deployment Accuracy ।
    • -Arm/Leg Entanglement (Configuration)
    • Filter Centering (Tilt)
    • -Simulated Use
    • Delivery System Tensile Strength -
    • | Delivery System Torque
    • -Visual Inspection (Freedom from surface defects)
    • Delivery System Visibility (Radiopacity) l
    • Dimensional Verification ー
    • Burst Pressure ।
  • Biocompatibility, per ISO 10993 Filter .
    • Cytotoxicity ।
    • Sensitization ।
    • -Irritation - Intracutaneous Reactivity
    • Acute Systemic Toxicity -
    • l Subacute Toxicity and Implantation
    • l Genotoxicity
    • । Hemocompatibility
  • Biocompatibility, per ISO 10993 Delivery System .
    • Cytotoxicity ।
    • Sensitization ၊
    • Irritation Irracutaneous Reactivity |

Image /page/3/Picture/36 description: The image shows the word "BARD" in a stylized, bold font. The letters are outlined in black, giving them a distinct and prominent appearance. The font style is blocky and somewhat geometric, with a unique design for the letter "A" that resembles a triangle.

{4}------------------------------------------------

  • । Acute systemic toxicity
  • । Hemocompatibility

Performance Testing - In-Vivo Testing

Two GLP compliant animal studies were performed in support of the DENALL® Filter System. one to assess the filter and one to assess the delivery systems.

To assess the fitter, an animal study was performed with the primary objective of the study being to evaluate the retrievability of the DENALI® Filter in an ovine animal model following implantation periods of 4 and 12 weeks. Twelve (12) filter retrievals were evaluated for: 1) ease of removal (as assessed by a clinical evaluator) and 2) cava wall damage as assessed by venography, gross evaluation, and histopathology (in two separate subgroups - immediately post-euthanasia (0 week healing) and after an 8 week healing period). The secondary objectives of this study were to assess caval narrowing/stenosis, caval patency, extravasation, filter strut configuration, filter visibility under fluoroscopy, fracture, intimal irregularities, migration, penetration, perforation, thrombus and tilt.

To assess the delivery systems, an acute animal study in an ovine animal model was performed. The acute study validated the following attributes of 12 Femoral and 12 Jugular/Subclavian DENALI® Filter Systems:

  • . Dilator Visibility
  • Dilator Marker Band Visibility .
  • Introducer Sheath Tip (Jugular) and Introducer Marker Band (Femoral) . Visibility
  • Dilator/Introducer Trackability .
  • Dilator/Introducer Pushability .
  • . Aspiration
  • Delivery System Trackability .
  • . Delivery System Pushability
  • . Ease of Deployment (Deployment Force)
  • . Deployment Accuracy
  • Filter Centering (Tilt) .
  • Arm/Leg Entanglement .

4

Image /page/4/Picture/21 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are outlined in black, giving them a three-dimensional appearance. The overall design is simple and eye-catching.

{5}------------------------------------------------

  • Filter Visibility Under Fluoroscopy .
  • Snare Tip Visibility under Fluoroscopy .
  • Pusher Assembly Visibility Under Fluoroscopy .

Based upon the performance assessment, all acceptance criteria were met and the DENALL® Filter System was deemed acceptable by clinical evaluators. Specifically to the retreivability animal study, all 12 filters were successfully removed with an acceptable retrieval force. In addition, based on clinical evaluation and confirmed by the pathologist there was no observed caval occlusion/thrombosis, no definitive IVC penetrations/perforations, no contrast extravasation from the VC after filter removal, no significant filter tilting, and no hemodynamically significant caval stenosis. One death occurred following filter placement which the pathologist and attending veterinarian determined not to be device related.

The delivery systems were assessed by a clinical evaluator and the following delivery system attributes were found to be acceptable: dilator visibility, dilator marker band visibility, introducer sheath tip (Jugular), introducer markerband (Femoral) visibility, dilator/introducer trackability, dilator/introducer pushability, aspiration, delivery system (with Filter) trackability, delivery system (with Filter) pushability, ease of deployment (Deployment Force), deployment accuracy, filter centering (Tilt), arm/leg entanglement, filter visibility under fluoroscopy, snare tip visibility under fluoroscopy and pusher assembly visibility under fluoroscopy.

Performance Testing - Clinical Testing

A single-arm, prospective, multi-center clinical study was conducted to assess the safety of the DENALI® Filter as both a permanent and retrievable device. Clinical Success Placement (CSP) was defined as freedom from subsequent PE, filter embolization, caval occlusion, filter or procedure related death, insertion adverse events, and technical failure of placement. The preestablished performance goal was that the lower bound of the 95% confidence interval for the CSP was greater than 80%. Technical success of placement (TSP) was defined as deployment of the filter such that the physician judged the location to be suitable to provide sufficient mechanical protection against PE. Additionally, the secondary endpoints of recurrent PE, new or worsening DVT. filter migration. filter fracture and tilt were assessed at the six month visit or the one month post-retrieval visit.

Image /page/5/Picture/11 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are outlined in black, with a white fill. The word is horizontally oriented and appears to be a logo or brand name.

{6}------------------------------------------------

Traditional 510/k) DENALI® Filter System - Femoral and Jugular/Subclavian Delivery Kit

One hundred seventy five (175) patients (107 males, 68 females) were enrolled at 20 investigational sites across the United States. The mean age was 56.7 ± 15.8 years (range 18 – 89 years). Eighty six (86) patients had their filter successfully retrieved.

Of the 175 patients who underwent DENAL® Filter placement, 95 had active thromboembolic disease (the presence of DVT or PE at the time of filter placement). Of these 95 patients, 63 had a contraindication to anticoagulation, 7 had a complication related to the use of anticoagulant medication, 8 had a failure of anticoagulation, and 17 had a filter placed without contraindication, complication or failure related to anticoagulant medication. Eighty (80) patients without active thromboembolic disease (neither DVT nor PE at the time of filter placement) were enrolled in the study.

Reasons for filter placement were as follows: Surgery (n=86, 49%), Trauma (n=41, 23%), Hypercoagulopathy (n=33, 19%), Cancer (n=7, 4%), Stroke (n=3, 2%) and Other (n=5, 3%).

Sixty nine (69) patients completed a six month visit. Longer term data was available in 21 patients at 12-month follow up and in one patient at 18-month follow up. The study will continue to follow all patients to 24 month post retrieval, whichever comes first. Three (3) patients withdrew their consent, 3 were lost to follow up and 7 died from pre-existing conditions. An independent Clinical Events Committee (CEC) determined that no patient deaths were attributed to the filter device, or implant or retrieval procedures. This clinical experience will be updated in the IFU once the study is complete.

{7}------------------------------------------------

EligibleforVisit(N)VisitCompleted(N, %)RetrievedReason Visit Not CompletedEvents OccurringBefore Next VisitPendingVisit
Death1Lost toFollow-UpWithdrawnMissedVisitMigrationFracture
Baseline/Implant175175N/A000000N/A
3 Months128114 (89%)286105002
6 Months8869 (78%)2321135009
12 Months2521 (84%)3230001003
18 Months41 (25%)30000003
24 Months00N/AN/AN/AN/AN/AN/AN/A0
Retrieval888886N/AN/AN/AN/AN/AN/AN/A
30 Days Post-Retrieval8668 (79%)4N/AN/A1N/AN/AN/AN/A8

Table 1: Patient Accountability

1 CEC adjudicated, not device related

2 69 patients completed the 6-month visit, 9 patients were retrieved in the 6-month visit window (6-months plus or minus 30 days)

1 patient was retrieved in the 12-month visit window (12-months plus or minus 30 days)

477 patients completed the 30 Day Post-Retrieval Visit, 9 of which occurred during the six month visit window. 68 patients completing the one month post-retrieval visit are reported in the table above to prevent assessment at the 6-month visit and one month post-retrieval visit.

TSP for the DENALI® Filter was 100%. CSP for the DENALI® Filter was 96.1% and the lower bound of the 95% confidence interval was 91.2%. It was concluded that the performance goal was successfully met. Mean placement procedure time was 17.5 minutes.

There were no findings of caval occlusion, filter fracture, cranial migration, filter tilt at placement, or filter tilt at retrieval. There were two (2) cases of symptomatic PE; neither of which caused patient death. There were five (5) cases of asymptomatic penetration; none of which had clinical sequelae. Three (3) cases of penetration were noted at implant and two (2) cases of penetration were noted at retrieval. Twelve (12) patients reported thirteen (13) cases of new or worsening DVT. There were ten (10) cases of new DVT and three (3) cases of worsening DVT. All 10 new DVTs reported were in those patients that had active disease at the time of implant, were considered to be hypercoagulable, or those that had orthopedic procedures on their lower extremities. All site-reported adverse events were adjudicated by the CEC.

Image /page/7/Picture/11 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are all capitalized and appear to be filled with a solid color, likely black or a dark shade. The font style gives the word a strong and prominent appearance.

{8}------------------------------------------------

Table 2: Complication Rates
Recurrent PE2 / 1391 (1.4%)
Caval Occlusion0 / 1372 (0%)
New DVT10 / 137 (7.3%)
Worsening DVT3 / 137 (2.2%)
Filter Fracture0 / 137 (0%)
Cranial Migration0 / 137 (0%)
Caudal Migration0 / 137 (0%)
Filter Penetration at Placement3 / 175 (1.7%)
Filter Penetration at Retrieval2 / 883 (2.3%)
Filter Tilt at Placement0 / 175 (0%)
Filter Tilt at Retrieval0 / 883 (0%)

The denominator of 139 includes 69 patients who completed the 6-month visit. 68 patients who reached the one month post-retrieval visit and 2 patients who had a reported PE outside of the 6month visit or the one month post-retrieval visit

1All complication rates with a denominator of 137 include 69 patients who completed the 6-month visit and 68 patients who reached the one month post-retrieval visit

388 patients had a retrieval visit with 86 successful retrievals

DENALI® Filter retrieval was attempted in 88 patients and successful in 86 patients (97.7%). In the two (2) unsuccessful retrieval cases, the snare was unable to engage the filter retrieval hook due to anatomical curvature. Mean filter indwell time was 136.2 ± 90.6 days (median 120.0 days, range 5 – 454 days). The right internal jugular vein was used in all retrieval procedures and mean procedure time was 21.9 minutes.

Venacavograms taken before and after the retrieval procedures of the IVC implant site revealed abnormalities that the CEC determined to be related to the device in two patients. One patient had minimal, self limited contrast extravasation post retrieval and another patient experienced intimal injury and caval narrowing of the IVC post retrieval. No clinical sequelae were reported for either patient.

Seventy seven (77) of the 86 patients who had their filter retrieved completed one month followup, one (1) subject was lost to follow-up, and eight (8) were pending. No instances of recurrent PE or new or worsening DVT were reported for any patient completing the one month postretrieval visit.

Bard Peripheral Vascular, Inc.

Image /page/8/Picture/9 description: The image shows the text "Page 10" in a bold, sans-serif font. The text is black against a white background. The words are stacked on top of each other, with "Page" on the top line and "10" on the bottom line.

Image /page/8/Picture/10 description: The image shows the word "Baird" in a bold, sans-serif font. The letters are all capitalized and outlined in black. The word is slightly distorted, with the letters appearing to be stretched horizontally.

{9}------------------------------------------------

Table 3: DENALI® Filter Retrieval Details
Number of Filter Retrieval Attempts88
Number of Successful Retrievals86
Retrieval Success Rate97.7%
Mean Indwell Time136.2 days
Maximum Indwell Time454 days

Image /page/9/Figure/3 description: The image is a histogram titled "Figure 3: Time from Implantation to Retrieval (N=86)". The x-axis is labeled "Time Interval (In Days)" and ranges from 30 to 480 in increments of 30. The y-axis is labeled "Number of Subjects" and ranges from 0 to 15 in increments of 3. The histogram shows the distribution of time intervals, with the highest frequency at 90 and 180 days.

Conclusion

The DENALI® Filter System – Femoral and Jugular/Subclavian Delivery Kits are substantially equivalent to the legally marketed predicate device, the ECLIPSE® Filter System – Femoral and Jugular/Subclavian Delivery Systems (K101431).

Image /page/9/Picture/8 description: The image shows the word "BARD" in a stylized, outlined font. The letters are bold and sans-serif, with a unique design where the inner parts of the letters are not filled in, creating a hollow effect. The overall appearance is clean and modern.

{10}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/10/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a caduceus, which is a symbol of medicine. The caduceus is made up of three intertwined snakes and a staff.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 15, 2013

Bard Peripheral Vascular, Inc. c/o Ms. Joni Creal Regulatory Affairs Specialist II 1625 West Third Street Tempe, AZ 85281

Re: K130366

Trade/Device Name: Denali Filter System – Femoral Delivery Kit and Jugular Delivery Kit Regulation Number: 21 CFR 870.3375 Regulation Name: Cardiovascular intravascular filter Regulatory Class: Class II Product Code: DTK Dated: February 14, 2013 Received: February 15, 2013

Dear Ms. Creal:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability

warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{11}------------------------------------------------

Page 2 - Ms. Joni Creal

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Bram D. Zự chế man -S

Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{12}------------------------------------------------

Indications for Use

510(k) Number (if known): K130366

Device Name: DENALI® Filter System - Femoral and Jugular/Subclavian Delivery Kits

Indications for Use:

The DENALI® Filter System - Femoral and Jugular/Subclavian Delivery Kits are indicated for use in the prevention of recurrent pulmonary embolism via placement in the vena cava in the following situations:

  • Pulmonary thromboembolism when anticoagulants are contraindicated. .
  • . Failure of anticoagulant therapy for thromboembolic disease.
  • . Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced.
  • . Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated.

DENALI® Filter may be removed according to the instructions supplied under the section labeled: Optional Procedure for Filter Removal.

Prescription Use X (Part21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use_ (21CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Bram D. Zut 2013.05.1

Bard Peripheral Vascular, Inc.

§ 870.3375 Cardiovascular intravascular filter.

(a)
Identification. A cardiovascular intravascular filter is an implant that is placed in the inferior vena cava for the purpose of preventing pulmonary thromboemboli (blood clots generated in the lower limbs and broken loose into the blood stream) from flowing into the right side of the heart and the pulmonary circulation.(b)
Classification. Class II. The special controls for this device are:(1) “Use of International Standards Organization's ISO 10993 ‘Biological Evaluation of Medical Devices Part I: Evaluation and Testing,’ ” and
(2) FDA's:
(i) “510(k) Sterility Review Guidance and Revision of 2/12/90 (K90-1)” and
(ii) “Guidance for Cardiovascular Intravascular Filter 510(k) Submissions.”