(116 days)
ROMEO posterior osteosynthesis system is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; spinal tumor; and failed previous fusion (pseudarthrosis).
When used as a posterior, non-cervical, non-pedicle screw fixation system, ROMEO posterior osteosynthesis system is intended for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.
The modifications to ROMEO posterior osteosynthesis system (K081165, K093170, K101678, K111127 and K112108) manufactured by SPINEART consist of:
- Addition of Romeo® 25D & 25T Screws made of Titanium alloy Ti6AI4V ELI . conforming to ISO 5832.3 and ASTM F136
- Addition of Romeo® Straight and Multiaxial Cross Connectors made of Titanium . alloy Ti6AI4V ELI conforming to ISO 5832-3 and ASTM F136
- Additional lengths of Percutaneous Pre-bent and Straight Rods made of Titanium . alloy (Ti6AI4V ELI conforming to ISO 5832-3 and ASTM F136).
These components are supplied either sterile or not sterile.
- Addition of a Cross connector caliper and a 3.5 mm tightener (not sterile). .
The medical device described in K130267 is a spinal osteosynthesis system, not an AI device. Therefore, the acceptance criteria and study information typically sought for AI/ML devices, such as performance metrics like accuracy, sensitivity, specificity, and related study designs (e.g., MRMC, standalone studies, ground truth establishment by experts, sample sizes for training and test sets), are not applicable in this context.
This 510(k) summary focuses on demonstrating substantial equivalence for an extension to an existing mechanical spinal implant system. The "performance data" section explicitly states: "Mechanical testing including static compression bending, static torsion, dynamic axial compression tests have been performed according to ASTM F1717-12. Results demonstrate that additional components perform as safely and effectively as their predicate devices. No clinical data has been presented."
Therefore, I cannot provide a table of acceptance criteria and device performance as requested in the format for an AI device. Instead, I can summarize the mechanical testing performed:
1. Table of Acceptance Criteria and Reported Device Performance (for mechanical testing):
This information is not explicitly detailed in a pass/fail table format within the provided 510(k) summary. However, the summary states:
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Conformity to special controls for Pedicle screw spinal system | Conforms to special controls |
| Conformity to "Spinal System 510(k)s - Guidance for Industry and FDA Staff Document" (May 3, 2004) | Conforms to guidance document |
| Performance as safely and effectively as predicate devices | Demonstrates performance as safely and effectively as predicate devices for additional components |
| Withstand mechanical testing according to ASTM F1717-12 | Mechanical testing (static compression bending, static torsion, dynamic axial compression) performed according to ASTM F1717-12. Results deemed adequate to demonstrate substantial equivalence. |
| Substantial equivalence in terms of intended use, material, design, mechanical properties, and function | Additional components are substantially equivalent to predicate devices |
2. Sample size used for the test set and the data provenance:
- Test Set Sample Size: Not applicable. For mechanical testing, this typically refers to the number of devices or components tested. The specific number of components and repetitions for each test (static compression bending, static torsion, dynamic axial compression) according to ASTM F1717-12 is not specified in the summary but would have been part of the full testing report.
- Data Provenance: Not applicable in the context of clinical data. For mechanical testing, the data originates from laboratory testing conducted by the manufacturer according to specified ASTM standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. Ground truth as typically understood for AI/ML refers to clinical labels or expert annotations on patient data. For mechanical testing, the "ground truth" is compliance with engineering standards and performance relative to predicate devices, verified by engineering and regulatory experts within the testing facility and the FDA.
4. Adjudication method for the test set:
- Not applicable. Adjudication methods are relevant for human expert review of clinical data. For mechanical testing, results are objectively measured and compared against predefined engineering specifications and predicate device performance.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is a mechanical device, not an AI/ML device that assists human readers.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is a mechanical device.
7. The type of ground truth used:
- For mechanical testing, the "ground truth" is adherence to validated engineering test standards (ASTM F1717-12) and comparison against the performance of legally marketed predicate devices. The objective is to demonstrate that the new components meet the same safety and effectiveness profiles as the existing, approved devices.
8. The sample size for the training set:
- Not applicable. There is no AI/ML algorithm involved to train.
9. How the ground truth for the training set was established:
- Not applicable. There is no training set for an AI/ML algorithm.
{0}------------------------------------------------
510(k) SUMMARY
As required by section 807.92
MAY 3 1 2013
| Submitter | SPINEART | |
|---|---|---|
| International Center Cointrin 20 route de pré-bois CP1813 | ||
| 1215 GENEVA 15 | ||
| SWITZERLAND | ||
| Contacts | Franck PENNESI Director of Industry & Quality | |
| Phone : +41 22 799 40 25 | ||
| Fax : +41 22 799 40 26 | ||
| Mail : fpennesi@spineart.ch | ||
| Regulatory contact : Dr Isabelle DRUBAIX (Idée Consulting) | ||
| idrubaix@nordnet.fr | ||
| Trade Name | ROMEO posterior osteosynthesis system | |
| SPECIAL 510k | ROMEO posterior osteosynthesis system - Line extension | |
| CFR section | 888.3070 and 888.3050 | |
| Classification Name | Pedicle screw spinal system / Spinal interlaminal fixation orthosis | |
| Class | II | |
| Product Codes | KWP Spinal interlaminal fixation orthosisMNH orthosis, spondylolisthesis spinal fixationMNI orthosis, spinal pedicle fixation | |
| Device panel | ORTHOPEDIC | |
| Legally marketedpredicate devices | ELLIPSE posterior osteosynthesis system (K081165) andROMEO posterior osteosynthesis system (K093170, K101678, K111127 and K112108) manufactured bySPINEART | |
| e-copy Statement | The eCopy is an exact duplicate of the paper copy |
Description: The modifications to ROMEO posterior osteosynthesis system (K081165, K093170, K101678, K111127 and K112108) manufactured by SPINEART consist of:
- Addition of Romeo® 25D & 25T Screws made of Titanium alloy Ti6AI4V ELI . conforming to ISO 5832.3 and ASTM F136
- Addition of Romeo® Straight and Multiaxial Cross Connectors made of Titanium . alloy Ti6AI4V ELI conforming to ISO 5832-3 and ASTM F136
- Additional lengths of Percutaneous Pre-bent and Straight Rods made of Titanium . alloy (Ti6AI4V ELI conforming to ISO 5832-3 and ASTM F136).
These components are supplied either sterile or not sterile.
- Addition of a Cross connector caliper and a 3.5 mm tightener (not sterile). .
{1}------------------------------------------------
Intended Use ROMEO posterior osteosynthesis system is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; spinal tumor; and failed previous fusion (pseudarthrosis). When used as a posterior, non-cervical, nonpedicle screw fixation system, ROMEO posterior osteosynthesis system is intended for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.
Performance data ROMEO posterior osteosynthesis additional components conforms to special control established for Pedicle screw spinal system and to « Spinal System 510(k)s - Guidance for Industry and FDA Staff Document » issued on: May 3, 2004.
Mechanical testing including static compression bending, static torsion, dynamic axial compression tests have been performed according to ASTM F1717-12. Results demonstrate that additional components perform as safely and effectively as their predicate devices. No clinical data has been presented.
ROMEO posterior osteosynthesis system additional Substantial equivalence components are substantially equivalent to their predicate devices in terms of intended use, material, design, mechanical properties and function.
Non clinical performance testing according to special control and Verification Activity and Validation Activity demonstrate that additional components are as safe, as effective, and performs as safely and effectively as their predicate devices.
Preparation date, January 31th, 2013
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Letter dated: May 31, 2013
SPINEART % Mr. Franck Pennesi Director of Industry & Quality International Center Cointrin 20 route de pré-bois CP1813 1215 Geneva 15 Switzerland
Re: K130267
Trade/Device Name: ROMEO posterior osteosynthesis system Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNH, MNI, KWP Dated: February 22, 2013 Received: March 15, 2013
Dear Mr. Pennesi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). -- You may, therefore, market the device, subject-to the general-controls provisions of-the-Act. -The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Mr. Franck Pennesi
forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Mark Nigelkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological. Health_
Enclosure
{4}------------------------------------------------
INDICATIONS FOR USE
K130267 510(k) Number (if known): _
Device Name: ROMEO Posterior osteosynthesis system Indications for Use:
ROMEO posterior osteosynthesis system is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; spinal tumor; and failed previous fusion (pseudarthrosis).
When used as a posterior, non-cervical, non-pedicle screw fixation system, ROMEO posterior osteosynthesis system is intended for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.
Prescription Use Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ronald P. Jean -S
(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K130267
N/A