K Number
K130088
Manufacturer
Date Cleared
2013-07-18

(185 days)

Product Code
Regulation Number
872.1800
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

PortView is indicated for use in acquiring images of patient teeth. It is intended for intra oral x-ray examinations and produces instant, digital, intra-oral x-ray images of a patient's teeth for diagnosis of diseases of the teeth and oral structures.

Device Description

The digital x-ray sensor system, Port View produces instant, digital, intra-oral x-ray images of a patient's mouth for diagnosis of the tecth and oral structures. It consists of digital intraoral sensors and image processing software as followings.

Digital Intraoral sensor: is an indirect light converting digital X-ray detector. Incident X-rays are converted to visible light by a scintillating device. (material) such as Csl(Cesium lodide), the light is coupled optically to a light detection imager which is based on CMOS technology. The sensor supports USB 2.0 Direct connectivity to personal computers and or laptops.

Image processing software; provides various functions such as Image acquisition. Image scarch, Image Processing including Enhancement. 21) analysis (Zoom. Rotate, and Measurement) and Image Printing & Saving and DICOM 3.0 application.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the PortView Digital X-ray sensor system, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The provided 510(k) summary primarily focuses on demonstrating substantial equivalence to predicate devices rather than setting explicit, quantifiable acceptance criteria with a performance table for a new, unique claim. Instead, the "acceptance criteria" are implied by meeting established safety standards and performing comparably to the predicate devices.

Metric/StandardAcceptance Criteria (Implied/Predicate Performance)Reported Device Performance (PortView)
SafetyCompliance with IEC 60601-1Performed to IEC 60601-1 standards; satisfactory results
EMCCompliance with IEC 60601-1-2Performed to IEC 60601-1-2 standards; satisfactory results
Effectiveness (Implied by Substantial Equivalence)As safe and effective as predicate devices (Schick CDR, Gemini DUSB) for acquiring intraoral x-ray images for diagnosis of dental diseases and oral structures.Bench and clinical evaluation indicates the new device is as safe and effective as the predicate devices.
Technological Characteristics Comparison to PredicatesMatching or superior technical specifications (e.g., sensor size, active area, technology, interface, dynamic range, cable length, CPU, memory, HDD, OS, DICOM compliance, monitor settings)See Comparison Table below for detailed matching.

Comparison Table (from the document, showing how PortView meets technological characteristics of predicates):

ManufacturerGENORAY Co., Ltd. (PortView)Schick Technologies. Inc. (CDR)Edlen Imaging LLC (Gemini DUSB)
ImplementationSoftware and SensorSoftware and SensorSensor Only
Sensor size (mm)Size1: 36.73 X 24.35 Size2: 42.8 X 30.49Size 1: 31 X 22 Size 2: 37 X 24 Size 2: 43 X 30Size1: 36.73 X 24.35 Size2: 42.8 X 30.49
Sensor Active Area (mm²)Size1: 600 Size2: 900Size1: 432 Size2: 600 Size3: 921Size 1: 600 Size2: 900
TechnologyCMOSCMOSCMOS
Interface to PCUSB 2.0USB 2.0USB 2.0
Dynamic Range4096:14096:14096:1
Cable length(m)323
CPUIntel Pentium Dual-Core 2.5GHz(E5200) or higherIntel Pentium IV or AMD 3GHz or higher2.0 GHz Pentium 4
Memory1GB or higher1 GB1GB
HDD10GB or higher (Client), 100GB or higher (Server)80 GB HD80GB
OSWindows XP service pack 2 or higherWindows XP Professional or Windows Vista Home PremiumWindows XP Pro / 7
DICOM ComplianceDICOM 3.0DICOM 4.5DICOM 3.0
Monitor Settings (Min.)1024 X 7681024 X 7681024 X 768

2. Sample size used for the test set and the data provenance

The document states "the result of bench and clinical evaluation indicates that the new device is as safe and effective as the predicate devices." However, it does not specify the sample size for a distinct "test set" (such as a specific number of patients or images).

Regarding data provenance:

  • The manufacturer is GENORAY Co., Ltd. from Korea.
  • The study is described as "clinical evaluation" which implies prospective data collection, but no further details are given to confirm this. The document does not specify the country of origin of the clinical data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document does not provide any information on the number of experts used or their qualifications for establishing ground truth.

4. Adjudication method for the test set

The document does not specify any adjudication method.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study, nor does it discuss AI assistance improving human readers. The device is a digital X-ray sensor system and image processing software, not an AI-driven interpretive tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This question is not directly applicable in the context of this 510(k) for a digital X-ray sensor system. The device's primary function is image acquisition and basic processing, not autonomous algorithm-based diagnosis. The performance is inherently tied to human interpretation of the images produced.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The document does not specify the type of ground truth used. Given the "clinical evaluation," it would likely involve clinical assessment or radiographic interpretation, but the method for establishing a definitive "ground truth" for comparison is not detailed.

8. The sample size for the training set

The document does not specify any training set size. This is consistent with the device being an imaging system, not typically a machine learning algorithm requiring a separate training set for its core functionality.

9. How the ground truth for the training set was established

Since no training set is mentioned in the context of an algorithm or AI model, there is no information on how ground truth for a training set would have been established.

{0}------------------------------------------------

K130088
Page 1 of 4

510(k) Summary Exhibit 5

Date of Summary Preparation: Dec 21, 2012

JUL 1 8 2013

1.Submitter and US Official Correspondent
Submitter:GENORAY Co., Ltd.
Address:#512, Byucksan Technopia
434-6. Sangdaewon 1-dong.
Jungwon-gu, Seongnam-city
Gyeonggi-do, 462-716, Korea
Telephone No.:+82-31-740-4100
Fax:+82-31-737-8018

Official Correspondent (U.S): Jae Kim - Business Manager

Correspondent :GENORAY America Inc.
Address:1073 N, Batavia St.Orange, CA 92867, USA
Telephone No.:855-436-6729
Fax:714-786-8919
Email:jac@genorayamerica.com
    1. Establishment Registration Number 3005843418

ਤੇ . Device Information

Proprietary/Trade Name: Digital x-ray sensor system / PortView Common/Usual Name: Digital x-ray sensor system / Intraoral X-ray sensor system Classification Name: Digital x-ray sensor(Intraoral sensor) system to capture Extraoral X-ray source Product Codc: MUH

Device Class: Class II per regulation 21 CFR 872.1800

    1. Equivalent Legally Marketed Device
CompanyDevice NameK-Number
Schick Technologies. Inc.Schick Computed Oral Radiology System.CDRK072134
Edlen Imaging LLCGemini DUSBK103290

{1}------------------------------------------------

5. Description of the Device

The digital x-ray sensor system, Port View produces instant, digital, intra-oral x-ray images of a patient's mouth for diagnosis of the tecth and oral structures. It consists of digital intraoral sensors and image processing software as followings.

Digital Intraoral sensor: is an indirect light converting digital X-ray detector. Incident X-rays are converted to visible light by a scintillating device. (material) such as Csl(Cesium lodide), the light is coupled optically to a light detection imager which is based on CMOS technology. The sensor supports USB 2.0 Direct connectivity to personal computers and or laptops.

Image processing software; provides various functions such as Image acquisition. Image scarch, Image Processing including Enhancement. 21) analysis (Zoom. Rotate, and Measurement) and Image Printing & Saving and DICOM 3.0 application.

It can operate on any hardware platform which mects the minimum requirements that Intel Pentium Dual-Core 2.5GHz with Windows XP operating system.

ItemsProductPortView
ManufacturerGENORAY Co., Ltd.
Product NamePortView
510(k) No.
ImplementationSoftware and Sensor
Indications for usePortView is indicated for use in acquiringimages of patient teethes. It is intended forintra oral x-ray examinations andproduces instant, digital, intra-oral x-rayimages of a patient's mouth for diagnosisof diseases of the teeth and oral structures.
Size(mm)Size1: 36.73 X 24.35Size2: 42.8 X 30.49
Intra oral sensorActive Arca(mm²)Size1: 600Size2: 900
TechnologyCMOS
Interface to PCUSB 2.0
Dynamic Range4096:1
Cable length(m)3
Specification1. Software name : PortView2. Software version : 1.03. File size : 100MB4. Language : C++5. DICOM 3.0 Support
CPUIntel Pentium Dual-Core 2.5GHz(E5200) orhigher
Memory1GB or higher
SoftwareHDD10GB or higher (Client)100GB or higher
VGAOnboard chipset or higher
OSWindows XP or higher
DICOMComplianceDICOM 3.0
IMAGE ARCHIVEDCM, BMP, JPG, PNG
WINDOWINGAdjust the Window width and level forcontrast windowing parameters.

{2}------------------------------------------------

Indications for use/ Rationale for Substantial Equivalence Q.

The PortView is to be used as a digital intra oral sensor system of X-rays in Dental radiography.

The PortView shares the same indication for use, materials, design, operational and functional features and is therefore substantially equivalent to the predicate devices listed above No.4.

There are many independent manufacturers of intraoral s-ray dental radiography svstems in the United States today. One is Schick Computed Oral Radiology System. CDR by Schick Technologies. Inc. (K0752134). The other currently marketed device is the Gemini DUSB by Edlen Imaging LLC. (K103290)

A comparison table for the different systems is available in the section on Substantial Equivalence and No.7 below.

The device has been tested by a third party and found to meet the international safety standards established by the IEC 60601-1, further details available in the section on Test report.

ManufacturerGENORAY Co., Ltd.Schick Technologies. Inc.Edlen Imaging LLC
Product NamePortViewSchick Computed OralRadiology System.CDRGemini DUSB
ImplementationSoftware and SensorSoftware and SensorSensor Only
Sensor size(mm)Size1: 36.73 X 24.35Size2: 42.8 X 30.49Size 1: 31 X 22Size 2: 37 X 24Size 2: 43 X 30Size1: 36.73 X 24.35Size2: 42.8 X 30.49
Sensor Active Area(mm²)Size1: 600Size2: 900Size1: 432Size2: 600Size3: 921Size 1: 600Size2: 900
TechnologyCMOSCMOSCMOS
Interface to PCUSB 2.0USB 2.0USB 2.0
Dynamic Range4096:14096:14096:1
Cable length(m)323
CPUIntel Pentium Dual-Core2.5GHz(E5200) or higherIntel Pentium IV or AMD3GHzor higher2.0 GHz Pentium 4
Memory1GB or higher1 GB1GB
HDD10GB or higher (Client)100GB or higher80 GB HD80GB
OSWindows XP service pack2 or higherWindows XP ProfessionalorWindows Vista HomePremiumWindows XP Pro / 7
DICOM ComplianceDICOM 3.0DICOM 4.5DICOM 3.0
Monitor Settings(Min.)1024 X 7681024 X 7681024 X 768

7. Substantial equivalence chart (Technological Characteristics)

{3}------------------------------------------------

K130088
Page 4 of 4

8. Safety, EMC and Performance data

The basic safety and essential performance testing according to standard IEC 60601-1 were performed. EMC testing was conducted in accordance with standard IEC 60601-1-2. All test results were satisfactory.

And the result of bench and clinical evaluation indicates that the new device is as safe and effective as the predicate devices.

9. Conclusion

The PortView is determined to be substantially equivalent to other legally marketed devices in the United States. They are the Schick Computed Oral Radiology System, CDR by Schick Technologies. Inc. and the Gemini DUSB by Edlen Imaging LLC.

We believe that the PortView is safe, effective and substantially equivalent to the predicate devices.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 18, 2013

GENORAY Co., Ltd. % Mr. Jae Kim Business Development Manager GENORAY America, Inc. 1073 N. Batavia Street ORANGE CA 92867

Re: K130088

Trade/Device Name: PortView, digital x-ray sensor system Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: Class II Product Code: MUH Dated: April 23, 2013 Received: April 26, 2013

Dear Mr. Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

Page 2 - Mr. Jae Kim

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.lda.gov/McdicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

for

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known): K130088

Device Name: Digital x-ray sensor system, PortView

Indications for Use:

PortView is indicated for use in acquiring images of patient teeth. It is intended for intra oral x-ray examinations and produces instant, digital, intra-oral x-ray images of a patient's teeth for diagnosis of diseases of the teeth and oral structures.

The digital x-ray sensor system, PortView consists of Image acquisition part as an Intraoral Sensor and Image processing Software which provides functions such as Image acquisition, store, and search and connects to Network. And the software can also display images from intra oral video cameras.

V Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Smh.7)

(Division Sign-Off) Division of Radiological I Icalth Office of In Vitro Diagnostics and Radiological Health

510(k) _______________________________________________________________________________________________________________________________________________________________________ K130088_______________________________________________________________________________________________________________________________________________________________________

Page 1 of 1

§ 872.1800 Extraoral source x-ray system.

(a)
Identification. An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.(b)
Classification. Class II.