(484 days)
The EasyRA Creatine Kinase-MB (CK-MB) Reagent is intended for the quantitative determination of CK-MB activity in human serum and plasma, using the MEDICA "EasyRA Chemistry Analyzer" in clinical laboratories. Measurements of CK-MB activity are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy. For in vitro diagnostic use only. For prescription use only.
Medica's EasyRA C-Reactive Protein (CRP) Reagent is intended for use in the quantitative in-vitro diagnostic determination of C-reactive protein in human serum or plasma using the EasyRA clinical chemistry analyzer. Measurements of C-reactive protein aids in evaluation of the amount of injury to body tissues. For in-vitro diagnostic use only. For prescription use only.
The EasyCAL C-Reactive Protein (CRP) Calibrator Kit is used for calibrating the CRP assay on the EasyRA clinical chemistry analyzer when used in conjunction with EasyRA CRP Reagent. The CRP calibrators are used to establish points of reference that are used in the determination of values in the measurement of CRP in human serum and plasma. For in-vitro diagnostic use only. For prescription use only.
The EasyQC C-Reactive Protein (CRP) Quality Control Materials are intended for use as quality control material for the CRP turbidimetric assay, using the EasyRA CRP Reagent and calibrator kit on the EasyRA clinical chemistry analyzer. For in-vitro diagnostic use only. For prescription use only.
Not Found
This document is a letter from the FDA regarding the 510(k) premarket notification for several diagnostic reagents and related calibrator/control materials. It does not contain information about the acceptance criteria or a study proving the device meets those criteria.
Therefore, I cannot provide the requested information. The document focuses on regulatory approval and indications for use, not performance studies.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. The text is in all caps and appears to be in a sans-serif font.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
MEDICA CORPORATION PHOTIOS MAKRIS DIRECTOR, REGULATORY AFFAIRS 5 OAK PARK DRIVE BEDFORD MA 01730
May 13, 2014
Re: K130080
Trade/Device Name: EasyRA Creatine Kinase-MB (CK-MB) Reagent EasyRA C-Reactive Protein (CRP) Reagent EasyCAL C-Reactive Protein (CRP) Calibrator Kit EasyQC C-Reactive Protein (CRP) Quality Control Material Regulation Number: 21 CFR 862.1215 Regulation Name: Creatine phosphokinase/creatine kinase or isoenzymes test system Regulatory Class: II Product Code: JHY, DCK, JIT, JJX Dated: May 06, 2014 Received: May 07, 2014
Dear Dr. Photios Makris:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the
{1}------------------------------------------------
Page 2-Dr. Photios Makris
clectronic product radiation control provisions (Scctions 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm. Also, picase note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincercly yours,
Yung W. Chan -S for
Courtney H. Lias. Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K130080
Device Name
EasyRA Creatine Kinase-MB (CK-MB) Reagent
Indications for Use (Describe)
The EasyRA Creatine Kinase-MB (CK-MB) Reagent is intended for the quantitative determination of CK-MB activity in human serum and plasma, using the MEDICA "EasyRA Chemistry Analyzer" in clinical laboratories. Measurements of CK-MB activity are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
For in vitro diagnostic use only. For prescription use only,
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Ruth A. Chesler -S
This section applies only to requirements of the Paperwork Reduction Act of 1995.
"DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) k 130080
Device Name
EasyRA C-Reactive Protein (CRP) Reagent EasyCAL C-Reactive Protein (CRP) Calibrator Kit EasyQC C-Reactive Protein (CRP) Quality Control Material
Indications for Use (Describe)
Medica's EasyRA C-Reactive Protein (CRP) Reagent is intended for use in the quantitative in-vitro diagnostic determination of C-reactive protein in human serum or plasma using the EasyRA clinical chemistry analyzer. Measurements of C-reactive protein aids in evaluation of the amount of injury to body tissues. For in-vitro diagnostic use only. For prescription use only.
The EasyCAL C-Reactive Protein (CRP) Calibrator Kit is used for calibrating the CRP assay on the EasyRA clinical chemistry analyzer when used in conjunction with EasyRA CRP Reagent. The CRP calibrators are used to establish points of reference that are used in the determination of values in the measurement of CRP in human serum and plasma. For in-vitro diagnostic use only. For prescription use only.
The EasyQC C-Reactive Protein (CRP) Quality Control Materials are intended for use as quality control material for the CRP turbidimetric assay, using the EasyRA CRP Reagent and calibrator kit on the EasyRA clinical chemistry analyzer. For in-vitro diagnostic use only. For prescription use only.
Type of Use (Select one or both, as applicable)
2 Prescription Use {Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
,对于比
FOR FDA USE ONLY . . . .
.
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Ruth A. Chesler -S
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
§ 862.1215 Creatine phosphokinase/creatine kinase or isoenzymes test system.
(a)
Identification. A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.(b)
Classification. Class II.