K Number
K130041

Validate with FDA (Live)

Date Cleared
2013-02-06

(29 days)

Product Code
Regulation Number
892.1570
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Sterile ultrasound couplant for use with medical diagnostic ultrasound. It is intended for non-invasive use in medical diagnostic ultrasound procedures to couple sound waves between a patient and the medical imaging electronics. The gel is intended for use in all diagnostic ultrasound procedures which require ultrasound coupling gel or fluid.

Device Description

Sheathing Technologies, Inc Ultrasound Gel is a water-based coupling agent for diagnostic ultrasonic procedures. This device is an accessory used on diagnostic ultrasound probes. The material is a water-based gel. Gel will be sold in sterile packets for single patient/procedure, disposable use.

AI/ML Overview

This sterile ultrasound gel 510(k) submission describes testing to support substantial equivalence to predicate devices. The acceptance criteria and supporting studies are outlined below:

1. Acceptance Criteria and Reported Device Performance

Test TypeAcceptance CriteriaReported Device Performance
BiocompatibilityMeets ISO 10993-1:2009 Biocompatibility StandardMeets ISO 10993-1:2009 for Irritation/Intracutaneous Toxicity and Sensitization. Cytotoxicity is equivalent to the predicate Sonotech's Natural Image Couplant.
    CytotoxicityEquivalent cytotoxicity to predicate deviceEquivalent to the cytotoxicity of Sonotech's Natural Image Couplant.
    Irritation/Intracutaneous ToxicityMeets ISO 10993-1:2009 standardMeets ISO 10993-1:2009 standard.
    SensitizationMeets ISO 10993-1:2009 standardMeets ISO 10993-1:2009 standard.
Bench Testing (Acoustic Performance)Equivalent acoustical performance to predicate devicesEquivalent acoustical performance to the predicate Sheathing Technologies gel and to the UltraBio™ predicate gel.
Sound VelocityEquivalent to predicate devicesEquivalent to predicate devices.
Acoustic ImpedanceEquivalent to predicate devicesEquivalent to predicate devices.
Sound AttenuationEquivalent to predicate devicesEquivalent to predicate devices.
Physical MeasurementsDensity and viscosity within the range measured in predicate devicesDensity and viscosity are within the range measured in the predicate devices.
DensityWithin range of predicate devicesWithin the range measured in the predicate devices.
ViscosityWithin range of predicate devicesWithin the range measured in the predicate devices.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify exact sample sizes for each test in the non-clinical studies (biocompatibility, bench testing, physical measurements). The data provenance is derived from laboratory testing of the device and comparison to predicate devices, rather than clinical patient data. Therefore, concepts like country of origin for patient data (as in medical image analysis) or retrospective/prospective clinical studies do not apply here.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This submission pertains to a sterile ultrasound gel, an accessory, and its substantial equivalence is demonstrated through non-clinical laboratory testing and comparison to predicate devices, not through human interpretation of medical data. Therefore, the concept of "experts" to establish ground truth in the context of diagnostic interpretation is not applicable here. The experts involved would be laboratory technicians, engineers, and quality assurance personnel performing and verifying the tests according to established international standards (like ISO 10993-1:2009). Their qualifications are typically in biomedical engineering, chemistry, biology, or related scientific fields, with expertise in laboratory testing and regulatory compliance.

4. Adjudication Method for the Test Set

The concept of an adjudication method (like 2+1, 3+1) is relevant for studies where multiple expert readers assess the same cases and a consensus is needed to establish ground truth. As this submission focuses on objective laboratory and physical tests of a medical device accessory, such an adjudication method is not applicable. The "ground truth" for each test is established by the validated measurement methods and comparison against the characteristics of the legally marketed predicate devices.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was conducted. This type of study is relevant for evaluating the performance of AI-powered diagnostic devices where human reader performance is a key metric, and improvement with AI assistance is assessed. This submission is for an ultrasound gel, which is not an AI-powered diagnostic device.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

No standalone study (in the context of an algorithm's performance without human interaction) was conducted. The device is a physical product (ultrasound gel), not an algorithm. Its performance is entirely standalone in the sense that it functions as a coupling agent for ultrasound, and its physical and acoustic properties are measured directly.

7. Type of Ground Truth Used

The ground truth used for this submission is based on:

  • Established International Standards: For biocompatibility, the ISO 10993-1:2009 standard serves as the benchmark.
  • Predicate Device Characteristics: For acoustic performance and physical measurements (density, viscosity), the characteristics of legally marketed predicate devices (Sonotech Clear Image™ Sterile Scanning Gel, K931909; Sheathing Technologies Ultrasound Gel, K112827; Sonotech UltraBio™ Sterile Ultrasound Imaging Couplant, K042619) serve as the "ground truth" or reference for substantial equivalence. The new device's performance is compared directly to these established products.

8. Sample Size for the Training Set

The concept of a "training set" is relevant for machine learning algorithms. This submission is for a physical medical device accessory and does not involve AI or machine learning. Therefore, there is no training set.

9. How the Ground Truth for the Training Set was Established

Since there is no training set for an AI/machine learning algorithm, this question is not applicable.

{0}------------------------------------------------

1130041

FEB 0 6 2013

Section 5: 510(K) Summary

(As required by 21 CFR 807.92)

Sterile Ultrasound Gel

January 30, 2013

Food and Drug Administration Center for Devices and Radiological Health Document Mail Center - WO66-G609 10903 New Hampshire Avenue Silver Spring, Maryland 20993-0002

To Whom it may Concern:

This letter, along with the attached materials is to notify your office of the intention of Sheathing Technologies to market the following device starting on or after (90) days from this date.

Classification:The FDA has classified ultrasound gel as aClass II device in the Code of FederalRegulations (CFR) 892.1570, Diagnosticultrasonic transducer.
Classification Panel:Radiology
Classification Procode:MUI
Device/Specification Developer:Sheathing Technologies, Inc.18431 Technology DriveMorgan Hill, CA 95037
Establishment Registration No.:2950776
Contact Persons:Jennifer Downing

Manager of Quality & Research

5-1

{1}------------------------------------------------

1-408-782-2720

Richard Stevens VP, Research & Development 1-408-782-2720

Trade Name:

Common Name:

Classification Name:

Equivalence:

Labeling and Usage:

Device Description:

Sheathes™ Sterile Ultrasound Gel

Sterile Ultrasound Gel

Ultrasonic pulsed echo imaging system accessorv

Sonotech Clear Image™ Sterile Scanning Gel, K931909 Sheathing Technologies Ultrasound Gel, K112827 Sonotech UltraBio™ Sterile Ultrasound Imaging Couplant, K042619.

The following information will be found on each box/bag:

    1. Proprietary name
  • Quantity of gel 2.
    1. Name and Location of Manufacturer
    1. Statement of Sterility
    1. Expiration date
    1. Prescription Statement: "Caution: Federal law restricts this device to sale by or on the order of a physician or a practitioner trained in its use."

Sheathing Technologies, Inc Ultrasound Gel is a water-based coupling agent for diagnostic ultrasonic procedures.

This device is an accessory used on diagnostic ultrasound probes.

The material is a water-based gel.

Gel will be sold in sterile packets for single patient/procedure, disposable use.

5-2

{2}------------------------------------------------

Substantial Equivalence:

The Sheathing Technologies, Inc. Ultrasound gel is identified as substantially equivalent to Sonomed/Sonotech's Clear Image Sterile Scanning Gel, K931909, to Sheathing Technologies's Ultrasound Gel, K112827, and to Sonotech/Sonomed's UltraBio Sterile Ultrasound Imaging Couplant, K042619.

Non-Clinical Tests:

    1. Biocompatibility
    • a. Cytotoxicity
    • b. Irritation/Intracutaneous Toxicity
    • c. Sensitization
    1. Bench testing

. :

  • a. Sound Velocity
  • b. Acoustic Impedance
  • c. Sound Attenuation
    1. Physical measurements
    • a. Viscosity
    • b. Density

Conclusions from Non-Clinical Tests:

Sheathing Technologies's ultrasound gel meets the ISO 10993-1:2009 biocompatibility standard for both Irritation/intracutaneous toxicity and sensitization. The cytotoxicity of the ultrasound gel is equivalent to the cvtotoxicity Sonotech's Natural Image Couplant.

Sheathing Technologies's ultrasound gel has equivalent acoustical performance to the predicate Sheathing Technologies gel and to the UltraBio™ predicate gel, and the density and viscosity are within the range measured in the predicate devices.

5-3

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three curved lines representing its wings and tail. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 6, 2013

Sheathing Technologies, Inc c/o Jennifer Downing Senior Manager of Quality & Research 18431 Technologies Inc. MORAN HILL CA 95037

Re: K130041.

Trade/Device Name: Sterile Ultrasound Gel Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: MUI Dated: January 5, 2013 Received: January 8, 2013

Dear Ms. Downing:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If vour device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2- Ms. Downing

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Sean M. Boyd -S for

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name: Ultrasound Gel

Indications for Use: Sterile ultrasound couplant for use with medical diagnostic ultrasound. It is intended for non-invasive use in medical diagnostic ultrasound procedures to couple sound waves between a patient and the medical imaging electronics. The gel is intended for use in all diagnostic ultrasound procedures which require ultrasound coupling gel or fluid.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Sean M. Boyd -S

(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health

K130041 510(k)

Page 1 of __1

§ 892.1570 Diagnostic ultrasonic transducer.

(a)
Identification. A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.(b)
Classification. Class II.