K Number
K123910
Date Cleared
2013-10-29

(314 days)

Product Code
Regulation Number
N/A
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

To cleanse minor wounds and scrapes.

The product is not to be used for body cavities, canals, eyes, or mouth.

Device Description

Dr. Mehta's Wound Wash Saline SprayTM consists of drug and preservative-free sterile isotonic saline solution that is delivered in atomized spray form by means of activating the actuator. Dr. Mehta's Wound Wash Saline SprayTM distributed as a single can includes 0.9% saline (sodium chloride) solution. The product utilizes bag-on-valve technology (bag in can). The propellant, compressed air, is charged into the container between bag and the can creating a means to dispense the contents of the bag, i.e., sterile isotonic salution. The product consists of specially designed actuator to deliver an effective consistent spray of the solution on the wound area. The filled device is sterilized by gamma radiation utilizing parameters those are validated according to ISO/AAMI 11137 requirements (Sterilization of health care products -Requirements for validation and routine control - Radiation sterilization). The product passes USP<71> Sterility Test. Dr. Mehta's Wound Wash Saline Spray™ is available as an over-thecounter product, and will be labeled with the following indication: "To cleanse minor wounds and scrapes.".

AI/ML Overview

This document describes Dr. Mehta's Wound Wash Saline SprayTM, which is a medical device. The document is primarily a 510(k) premarket notification summary, which focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting a detailed study of its performance against specific acceptance criteria.

Therefore, the information required to fully answer your request regarding acceptance criteria and a study proving the device meets them is not fully available in the provided text. The document does not describe a performance study as you might expect for a novel or high-risk device. Instead, it relies on demonstrating that the device is substantially equivalent to an existing, legally marketed predicate device.

Here's an analysis based on the available information:

1. A table of acceptance criteria and the reported device performance:

Since this is a submission for substantial equivalence rather than a de novo or PMA application, explicit "acceptance criteria" are not framed in terms of clinical performance metrics like sensitivity, specificity, or reader improvement. Instead, the "acceptance criteria" are implicit in demonstrating substantial equivalence to the predicate device.

Acceptance Criteria (Implicit for Substantial Equivalence)Reported Device Performance (from Provided Text)
Safety:- "Dr. Mehta's Wound Wash Saline Spray™ is as safe as the predicate device."
Technological Characteristics:- "It has the same technological characteristics and basic principles of operation as its predicate devices."
Sterility:- "The filled device is sterilized by gamma radiation utilizing parameters those are validated according to ISO/AAMI 11137 requirements... The product passes USP<71> Sterility Test." - "The subject device passes USP <71> Sterility Test, like the predicate device."
Biocompatibility (LAL Test / Cytotoxicity):- "The subject device also passes LAL Test or Bacterial Endotoxins Test, & MEM Elution Test or Cytotoxicity Test."
Composition:- "consists of drug and preservative-free sterile isotonic saline solution... includes 0.9% saline (sodium chloride) solution." - "0.9% sterile saline solution is colorless and consists of USP grade sodium chloride and USP grade purified water." (Matches description of a wound wash product)
Spray Delivery Mechanism:- "product utilizes bag-on-valve technology (bag in can)... propellant, compressed air, is charged into the container between bag and the can creating a means to dispense the contents of the bag..." - "consists of an aerosol bag on valve technology which includes a bag-in-can system attached to actuator for spray dispensing in all positions... The technological characteristics with respect to the spray including pressurization method, blog technology and sterilization are the same as the predicate device."
Intended Use:- "To cleanse minor wounds and scrapes." (Matches predicate device's implied use for wound cleansing)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

No separate "test set" in the context of clinical performance data (e.g., patient images, clinical cases) is described. The "tests" mentioned are non-clinical bench tests:

  • USP <71> Sterility Test
  • LAL Test (Bacterial Endotoxins Test)
  • MEM Elution Test (Cytotoxicity Test)

The sample sizes for these bench tests are not specified in the document, nor is the country of origin for the data or whether they were retrospective or prospective. It is standard practice that these are prospective tests conducted on manufactured batches of the device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not applicable as there is no human-interpreted "test set" like imaging data requiring expert ground truth establishment. The ground truth for the non-clinical tests (sterility, endotoxins, cytotoxicity) is established by laboratory parameters and standards (e.g., USP <71> criteria, LAL assay thresholds).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

This is not applicable for the non-clinical bench tests described.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This is not applicable. The device is a wound wash saline spray, not an AI-powered diagnostic or assistive tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

This is not applicable. The device is not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

The "ground truth" for this device's evaluation (for its 510(k) submission) is based on established laboratory standards and regulatory requirements for sterility, biocompatibility, and chemical composition. For instance:

  • Sterility: USP <71> criteria.
  • Biocompatibility: LAL test and MEM Elution test standards.
  • Composition: USP grade specifications for sodium chloride and purified water, and 0.9% isotonic saline concentration.

8. The sample size for the training set:

This is not applicable as the device does not employ an AI algorithm requiring a training set.

9. How the ground truth for the training set was established:

This is not applicable as the device does not employ an AI algorithm requiring a training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for NeilMed. The logo is in black and features a bold, sans-serif font. A registered trademark symbol is present to the right of the word "Med".

K123910

Attachment 5

510(k) SUMMARY

(As prepared in accordance with 21 CFR 807.87(h) and 21 CFR 807.92)

Date Prepared: December 4th 2012

1. Submitter's Name, Address, Telephone Number, Contact PersonOCT 2 9 2013
NEILMED PHARMACEUTICALS, INC.
Address: 601 Aviation Blvd, Santa Rosa, CA 95403
Contact: Phone: +1 707 525 3784Fax: +1 707 525 3785
Contact Person: Dr Ketan MehtaKetan@neilmed.com
2. Name of Device and Name/Address of Sponsor
Device: Dr. Mehta's Wound Wash Saline SprayTM
Name of Sponsor: NEILMED PHARMACEUTICALS, INC.
Address: 601 Aviation Blvd, Santa Rosa, CA 95403
3. Common or Usual Name
Wound Cleanser
4. Device Classification Name and Classification Code
Name: Dressing, Wound, Drug
Code: FRO
Review Panel: General & Plastic Surgery
Device Class: Unclassified
5. Description
Dr. Mehta's Wound Wash Saline SprayTM consists of drug and preservative-free sterile isotonicsaline solution that is delivered in atomized spray form by means of activating the actuator. Dr.Mehta's Wound Wash Saline SprayTM distributed as a single can includes 0.9% saline (sodiumchloride) solution. The product utilizes bag-on-valve technology (bag in can). The propellant,compressed air, is charged into the container between bag and the can creating a means to

dispense the contents of the bag, i.e., sterile isotonic salution. The product consists of specially designed actuator to deliver an effective consistent spray of the solution on the wound area. The filled device is sterilized by gamma radiation utilizing parameters those are validated according to ISO/AAMI 11137 requirements (Sterilization of health care products -Requirements for validation and routine control - Radiation sterilization). The product passes USP<71> Sterility Test. Dr. Mehta's Wound Wash Saline Spray™ is available as an over-thecounter product, and will be labeled with the following indication: "To cleanse minor wounds

  1. Predicate Devices

and scrapes.".

Wound Wash Saline® (K083355) (Blairex Laboratories, Inc.)

  1. Intended Use / Indications for Use

{1}------------------------------------------------

Indications for OTC Use :

"To cleanse minor wounds and scrapes".

The product is not to be used for body cavities, canals, eyes, or mouth.

8. Technological Characteristics

Dr. Mehta's Wound Wash Saline Spray™ consists of an aerosol bag on valve technology which includes a bag-in-can system attached to actuator for spray dispensing in all positions. The is a laminate system including sandwiched aluminum layer providing an impermeable barrier between the propellant (compressed air) and the sterile saline solution within the bag. The solution comes in contact with the inner layer (polypropylene) of the bag. 0.9% sterile saline solution is colorless and consists of USP grade sodium chloride and USP grade purified unter. The propellant, compressed air, is charged into the container between bag and the can creating a means to dispense the contents of the bag, sterile isotonic saline solution. The device manufacturing process includes crimping, pressure testing and filling and packaging for sterilization which is followed by gamma sterilization. The product is tested against established specifications and meets USP<71> sterility requirements. The technological characteristionshow subject device are substantially equivalent to those of the predicate device.

9. Substantial Equivalence

Dr. Mehta's Wound Wash Saline Spray™ is as safe as the predicate device. It has the same technological characteristics and basic principles of operation as its predicate devices The technological characteristics with respect to the spray including pressurization method, blog technology and sterilization are the same as the predicate device.

The non-clinical tests performed on the subject device include (i) USP <71> Sterility Test, (ii) LAL Test, & (iii) MEM Elution Test. The subject device passes USP <71> Sterility Test, like the predict to Aevice. The subject device also passes LAL Test or Bacterial Endotoxins Test, & MEM Elution Test or Cytotoxicity Test.

Thus, Dr. Mehta's Wound Wash Saline Spray™ is substantially equivalent to the predicate device.

This concludes the 510(k) Summary.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 29, 2013

NeilMed Pharmaceuticals, Incorporated % Mr. Dinesh Patel Vice President of Quality Assurance and Regulatory Affairs 601 Aviation Boulevard Santa Rosa, California 95403

Re: K123910

Trade/Device Name: Dr. Mehta's Wound Wash Saline Spray Regulation Name: Dressing, Wound, Drug. Regulatory Class: Unclassified Product Code: FRO Dated: August 26, 2013 Received: August 28, 2013

Dear Mr. Patel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. Dinesh Patel

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please

contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

David Krause -S

for Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K123910

Device Name

Dr. Mehta's WoundWash Saline Spray(TM)

Indications for Use (Describe)

To cleanse minor wounds and scrapes.

The product is not to be used for body cavities, canals, eyes, or mouth.

Type of Use (Select one or both, as applicable)

_ Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 " 时时彩玩 .. . . . . Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Image /page/4/Picture/13 description: The image shows the name "Jiyoung Dang -S" in a bold, sans-serif font. The name is written in black ink on a white background. The letters are slightly distorted, giving the image a vintage or distressed look. The word "Dang" has a unique design, with the letters appearing to be constructed from geometric shapes.

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

N/A