K Number
K123801
Date Cleared
2013-05-28

(169 days)

Product Code
Regulation Number
870.2700
Panel
AN
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

PM601 Finger pulse oximeter is intended for spot-checking oxygen saturation in blood (SpO2) and pulse rate. The pulse oximeter is used on adults at hospital, clinics, and/or home. Not for continuously monitoring.

Device Description

The proposed device, Finger Pulse Oximeters PM601 is a fingertip device, which can display % SpO2. pulse rate value and pulse strength. It is based on digital blood oxygen. Power consumption of the proposed device is low and the two originally-equipped AAA alkaline batteries can be operated continuously for 24 hours. It will automatically turned off when no signal is detected for more than 8 seconds. And low voltage warning will be displayed and battery symbol flash when battery voltage is low.

AI/ML Overview

Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

MetricAcceptance Criteria (Predicate Device)Reported Device Performance (PM601)
SpO2 measurement range70%-100%70%-100%
SpO2 accuracy70%~100%, ±3%

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).