(30 days)
Not Found
No
The summary describes a quality control material for laboratory testing, with no mention of AI or ML in the intended use, device description, or performance studies.
No.
This device is an assayed quality control material used to monitor the precision of laboratory testing procedures, not to treat or diagnose a disease.
No
This device is a quality control material used to monitor the precision of laboratory testing procedures, not to diagnose a condition in a patient.
No
The device is a quality control material prepared from human whole blood, which is a physical substance, not software.
Based on the provided information, yes, this device is an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use explicitly states it's for "monitoring the precision of laboratory testing procedures for the analyte listed in the package insert." This is a classic function of a quality control material used in laboratory testing, which falls under the scope of IVDs.
- Device Description: The description mentions it's prepared from human whole blood and used in a laboratory setting.
- Intended User / Care Setting: It specifies "laboratory testing procedures," further reinforcing its use in a clinical laboratory environment.
Quality control materials like this are essential components of IVD testing, ensuring the accuracy and reliability of the diagnostic results.
N/A
Intended Use / Indications for Use
Liquichek Diabetes Control is intended for use as an assayed quality control material to monitor the precision of laboratory testing procedures for the analyte listed in the package insert.
Product codes
JJX
Device Description
This product is prepared from human whole blood and contains preservatives and stabilizers. The control is provided in liquid form for convenience.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Real-time stability studies were conducted to establish the thawed stability claims (openvial and unopened). Accelerated stability studies were conducted to establish the shelflife claims at -20 to -50 ℃. Based on the available data, product claims are as follows:
- Thawed Open Vial Stability: 14 days at 2°C to 8 °C .
- Thawed Unopened Stability: 180 days at 2°C to 8 °C .
- Shelf Life Stability: 3 years at -10°C to -50°C .
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 862.1660 Quality control material (assayed and unassayed).
(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
0
JAN 0 9 2013
Page 1 of 2
Bio-Rad Laboratories Premarket Notification Section 510 (k) Liquichek Dlabetes Control
Page 1 of 2
510(k) Summary Liquichek Diabetes Control
Submitter 1.0
Bio-Rad Laboratories 9500 Jeronimo Road, Irvine, Callfornia 92618-2017 (949) 598-1200 Telephone: (949) 598-1557 Fax:
Contact Person
Suzanne Parsons Regulatory Affairs Manager Telephone: (949) 598-1467
Date of Summary Preparation
December 21, 2012
Device Identification 2.0
Product Trade Name: | Liquichek Diabetes Control |
---|---|
Classification Panel: | Clinical Chemistry |
Common Name: | Single (Specified) Analyte Controls (Assayed and Unassayed) |
Classification: | Class I, Reserved |
Product Code: | JJX |
Regulation Number: | 21 CFR 862.1660 |
Device to Which Substantial Equlvalence is Claimed 3.0
Liquichek Diabetes Control Bio-Rad Laboratories Irvine, California
510 (k) Number: K052838
Description of Device 4.0
This product is prepared from human whole blood and contains preservatives and stabilizers. The control is provided in liquid form for convenience.
Value Assignment 5.0
The mean values and the corresponding ±3SD ranges printed in this insert were derived The mean values and the confessionality as on this lot of product. The tests listed were performed by the manufacturer and/or independent laboratories using manufacturer performed by the manufacturer andror incopenabiling of this lot of product. It is supported Teagents and a Toprocomative camon acceptable ranges and use those recommended that back laboratory established ranges may vary from those listed during provided only as guides. Edboratory over time and between laboratories may be caused by the the of this control. Vanutions over annualion and reagents, or by manufacturer test method modifications.
1
6.0 Intended Use
Liquichek Dlabetes Control is intended for use as an assayed quality control material to monitor the precision of laboratory testing procedures for the analyte listed in the package insert.
Comparison of the new device with the Predicate Device 7.0
Liquichek Diabetes Control claims substantial equivalence to the Liquichek Diabetes Control currently in commercial distribution (K052838). Table 1 (below) contains comparison information of similarities and differences between the new and predicate device to which substantial equivalence is claimed.
| Characteristics | Liquichek Diabetes Control
(New Device) | Liquichek Diabetes Control
(Predicate Device, K052838) |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Similarities | | |
| Intended Use | Liquichek Diabetes Control is intended for
use as an assayed quality control material
to monitor the precision of laboratory
testing procedures for the analyte listed in
this package Insert. | Liquichek Diabetes Control Is intended
for use as an assayed quality control
material to monitor the precision of
laboratory testing procedures for the
analyte listed in this package insert. |
| Matrix | Human Whole Blood | Human Whole Blood |
| Form | Liquid | Liquid |
| Thawed Opened
Stability | 14 days at 2 to 8°C | 14 days at 2°C to 8 °C |
| Thawed Unopened
Stability | 180 days at 2 to 8°C | 180 days at 2 to 8°C |
| Differences | | |
| Fill Volume | 2 mL | 1 mL |
| Storage unopened
(Shelf life) | -10°C to -50°C until expiration date | -10°C to -70°C until expiration date |
| Analytes | Hemoglobin A1C | Hemoglobin A1C
Hemoglobin |
. Statement of Supporting Data 8.0
Real-time stability studies were conducted to establish the thawed stability claims (openvial and unopened). Accelerated stability studies were conducted to establish the shelflife claims at -20 to -50 ℃. Based on the available data, product claims are as follows:
- Thawed Open Vial Stability: 14 days at 2°C to 8 °C .
- Thawed Unopened Stability: 180 days at 2°C to 8 °C .
- Shelf Life Stability: 3 years at -10°C to -50°C .
9.0 Conclusion
Based on the performance characteristics Indicated above, Bio-Rad's Liquichek Diabetes control is substantially equivalent to the predicate device K052838.
All supporting data is retained on file at Bio-Rad Laboratories.
2
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three curved lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
DEPARTMENT OF HEALTH & HUM AN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
January 9, 2013
Bio-Rad Laboratories c/o Suzanne Parsons 9500 Jeronimo Road Irvine, CA 92618-2017
Re: K123798
Trade/Device Name: Liquichek Diabetes control Regulation Number: 21 CFR §862.1660 Regulation Name: Ouality Control Material Regulatory Class: Class I, Reserved Product Code: JJX Dated: December 7, 2012 Received: December 10, 2012
Dear Ms. Parsons:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
3
Page 2 -- Suzanne Parsons
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostics and Radiological Health at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
for Carol C. Benson
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
4
Indications for Use
510(k) Number (if known): K123798
Device Name: Liquichek Diabetes control
Indications for Use: Liquichek Diabetes Control is intended for use as an assayed quality control material to monitor the precision of laboratory testing procedures for the analyte listed in the package insert.
Prescription Use _ X (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use __ (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Katherine Serrano 2013.01.09 15:11:51 -05'00'
Division Sign-Off Office of In Vitro Diagnostics and Radiological Health
510(k) K123798