K Number
K123521
Date Cleared
2013-02-07

(84 days)

Product Code
Regulation Number
888.3070
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The RELIANCE Spinal Screw System is a pedicle screw systems intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic. Jumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis).

The RELIANCE Spinal Screw System is also indicated for pedicle screw fixation for the treatment of severe degenerative spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

When used posteriorly, the RELIANCE Spinal Screw System is also a hook and sacral/iliac screw fixation system of the noncervical spine indicated for, degenerative spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudarthrosis). Levels of fixation are for the thoracic, lumbar and sacral spine.

The RELIANCE Spinal Screw System when used with staples and two rods as an anterior thoracic/lumbar screw fixation system, is indicated for spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudarthrosis).

When used in a percutaneous, posterior approach with MIS Instrumentation, the RELIANCE Spinal Screw System components are intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curyatures (i.e., scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.

Device Description

The Posterior RELIANCE Spinal Screw System consists of longitudinal rods, monoaxial screws, polyaxial screws, reduction screws, cannulated poly axial screws, cannulated reduction screws, hooks, reduction hooks, set screws, transverse connectors.

The Anterior RELIANCE Spinal Screw System consists of two spinal rods, monoaxial screws, staples, and set screws. The Anterior RELIANCE staples and screws are intended to be attached to the lateral aspect of the vertebral bodies from T5 to L4, and SHOULD NOT be attached to the anterior aspect. Furthermore, only Titanium components should be used anteriorly. (See Precautions section.)

The RELIANCE Spinal Screw System components are available in titanium alloy conforming to ASTM F-136 specifications as well as stainless steel conforming to ASTM F-138 specifications. Furthermore, various rods of the RELIANCE Spinal Screw System are available in Cobalt Chrome conforming to ASTM F-75 and ASTM F-1537 specifications. Components of the differing diameter rod systems are NOT interchangeable. The components of one material should not be used with components of another material, with the exception that the Cobalt Chrome rods may be used with titanium alloy implants. The extension tabs on the reduction screw and hook components are intended to be removed intraoperatively.

AI/ML Overview

This appears to be a 510(k) summary for a medical device (Reliance Spinal Screw System), which focuses on establishing substantial equivalence to previously marketed devices rather than presenting a study proving the device meets specific performance acceptance criteria.

Therefore, the provided text does not contain the information requested regarding acceptance criteria and a study demonstrating the device meets those criteria, as it's not a performance study report. Instead, it's a regulatory document for device marketing clearance based on equivalence.

Specifically:

  • No acceptance criteria are mentioned for device performance. The document focuses on equivalence in terms of material, intended use, levels of attachment, size range, and strength compared to predicate devices.
  • No study proving the device meets acceptance criteria is presented. The document explicitly states: "No additional testing was performed on the subject RELIANCE Spinal Screw System as none of the modifications resulted in a new worst-case scenario." This indicates a lack of new performance studies for the modifications.

Given this, I cannot provide the requested table or answer the specific questions about sample size, expert ground truth, adjudication, MRMC studies, or standalone performance, as this information is not present in the provided 510(k) summary. The document's purpose is to argue for "substantial equivalence," not to detail a new performance study.

{0}------------------------------------------------

510(k) Summarv

Date Prepared: 5 February 2013 Reliance Medical Systems, LLC Bret M. Berry 545 South 500 West Suite 100 Bountiful, UT 84010 801-388-0700 Telephone: 801-294-0079 Fax:

Bret M. Berry Contact: Member-Manager

Common or Usual Name: Proposed Proprietary or Trade Name: Spinal Fixation Device Reliance Spinal Screw System Spinal Intervertebral Body Fixation Orthosis (per 21

CFR 888.3060)

888.3050)

Spondylolisthesis Spinal Fixation Device System

Spinal Interlaminal Fixation Orthosis (per 21 CFR

and Pedicle Screw System (per CFR 888.3070) Product Code:

KWP, KWQ, MNH, MNI

Substantial Equivalence

The Reliance Spinal Screw is substantially equivalent to the legally marketed Reliance Spinal Screw (K081978, K101112, and K110896) and the Nuvasive SpherX (K102514). The Reliance Spinal Screw is equivalent to these commercially available devices in terms of material, intended use, levels of attachment, size range, and strength.

Device Description

The Posterior RELIANCE Spinal Screw System consists of longitudinal rods, monoaxial screws, polyaxial screws, reduction screws, cannulated poly axial screws, cannulated reduction screws, hooks, reduction hooks, set screws, transverse connectors.

The Anterior RELIANCE Spinal Screw System consists of two spinal rods, monoaxial screws, staples, and set screws. The Anterior RELIANCE staples and screws are intended to be attached to the lateral aspect of the vertebral bodies from T5 to L4, and SHOULD NOT be attached to the anterior aspect. Furthermore, only Titanium components should be used anteriorly. (See Precautions section.)

The RELIANCE Spinal Screw System components are available in titanium alloy conforming to ASTM F-136 specifications as well as stainless steel conforming to ASTM F-138 specifications. Furthermore, various rods of the RELIANCE Spinal Screw System are available in Cobalt Chrome conforming to ASTM F-75 and ASTM F-1537 specifications. Components of the differing diameter rod systems are NOT interchangeable. The components of one material should not be used with components of another material, with the exception that the Cobalt Chrome rods

FEB 0 7 2013

{1}------------------------------------------------

may be used with titanium alloy implants. The extension tabs on the reduction screw and hook components are intended to be removed intraoperatively.

Intended Use/Indications for Use

The RELIANCE Spinal Screw System is a pedicle screw systems intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic. Jumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis).

The RELIANCE Spinal Screw System is also indicated for pedicle screw fixation for the treatment of severe degenerative spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

When used posteriorly, the RELIANCE Spinal Screw System is also a hook and sacral/iliac screw fixation system of the noncervical spine indicated for, degenerative spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudarthrosis). Levels of fixation are for the thoracic, lumbar and sacral spine.

The RELIANCE Spinal Screw System when used with staples and two rods as an anterior thoracic/lumbar screw fixation system, is indicated for spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudarthrosis).

When used in a percutaneous, posterior approach with MIS Instrumentation, the RELIANCE Spinal Screw System components are intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curyatures (i.e., scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.

Non-Clinical Testing

No additional testing was performed on the subject RELIANCE Spinal Screw System as none of the modifications resulted in a new worst-case scenario.

Technological Modifications

The subject RELIANCE Spinal Screw offers ball rods, tailed screws, and additional sizes of cobalt chrome rods. The subject RELIANCE Spinal Screw System is substantially equivalent to the predicate devices in terms of sterilization and biocompatibility.

Conclusions

The subject RELIANCE Spinal Screw is substantially equivalent to the predicate devices in terms of strength, size, indications for use, sterilization and biocompatibility.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three overlapping shapes that resemble a person embracing or supporting another, with flowing lines suggesting movement or support.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Letter dated: February 7, 2013

Reliance Medical Systems, LLC % Mr. Bret M. Berry 545 South 500 West, Suite 100 Bountiful, Utah 84010

Re: K123521

Trade/Device Name: Reliance Spinal Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNH, MNI, KWP, KWP, KWQ Dated: January 7, 2013 Received: January 15, 2013

Dear Mr. Berry:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. Bret M. Berry

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K123521

Device Name: Reliance Spinal Screw System

Indications for Use:

The Reliance Spinal Screw System is a pedicle screw systems intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis). The Reliance Spinal Screw System is also indicated for pedicle screw fixation for the treatment of severe degenerative spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

When used posteriorly, the Reliance Spinal Screw System is also a hook and sacral/iliac screw fixation system of the noncervical spine indicated for degenerative spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudarthrosis). Levels of fixation are for the thoracic, lumbar and sacral spine.

The Reliance Spinal Screw System when used with staples and two rods as an anterior thoracic/lumbar screw fixation system, is indicated for spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudarthrosis).

When used in a percutaneous, posterior approach with MIS Instrumentation, the Reliance Spinal Screw System components are intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Ronald P. Jean -S

Page 1 of 1

(Division Sign-off) Division of Orthopedic Devices 510(k) Number: K123521

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.