K Number
K123519

Validate with FDA (Live)

Device Name
INVIVO DENTAL
Manufacturer
Date Cleared
2013-02-01

(78 days)

Product Code
Regulation Number
892.2050
Reference & Predicate Devices
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

In VivoDental is a software application used for the display and 3D visualization of medical image files from scanning devices, such as CT, MRI, or 3D Ultrasound. It is intended for use by radiologists, clinicians, referring physicians and other qualified individuals to retrieve, process, render, review, store, print, assist in diagnosis and distribute images, utilizing standard PC hardware. Additionally, In VivoDental is a preoperative software application used for the simulation of dental implants, orthodontic planning and surgical treatments.

This device is not indicated for mammography use.

Device Description

InVivoDental is a volumetric imaging software designed specifically for clinicians, doctors, physicians, and other qualified medical professionals. The software runs in Windows operating systems and visualizes medical imaging data on the computer screen. The software is downloaded over the internet and installed on the customer's computer. Users are able to examine anatomy on a computer screen and use software tools to move and manipulate rendered images by turning, zooming, flipping, adjusting contrast and brightness, cutting, and slicing. There are several rendering settings to emphasize bone and/or soft tissue. The software also has the ability to create panoramic images, superimpose images, and create measurements of volume, angle, and length. There are multiple tools to annotate and otherwise mark areas of interest on the images. The software has specialized tools to simulate surgical treatments using models rendered by Antomage, and any view or user modified area can be saved to a gallery. These simulations and galleries can then be reviewed by medical personnel or used for consultation between the doctor and patient.

AI/ML Overview

This section describes the acceptance criteria and the study that proves the device meets the acceptance criteria.

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaReported Device Performance
Software stability and operation as designed."Testing confirmed that the software is stable and operating as designed."
Evaluation of hazards and reduction of risk to acceptable levels."Testing also confirmed that the software evaluated for hazards and that the risk has been reduced to acceptable levels."
Effectiveness of measurement and rendering functions compared to predicate software."Bench testing of the software with predicate software was performed by evaluation of images rendered by InVivoDental and predicate software. This testing and evaluation included testing of measurement tools in both predicate and subject software and was performed by an expert in the field of radiology. This testing confirms that InVivoDental is as effective as its predicates in its ability to perform its essential functions of measurement and rendering of DICOM data."

2. Sample Size Used for the Test Set and Data Provenance:

The document does not explicitly state the sample size (number of cases or images) used for the test set during the bench testing. The data provenance (country of origin, retrospective or prospective) is also not specified.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

The bench testing "was performed by an expert in the field of radiology." The specific qualifications (e.g., years of experience, subspecialty) of this expert are not detailed beyond "an expert in the field of radiology."

4. Adjudication Method for the Test Set:

The document does not describe an adjudication method (such as 2+1, 3+1, or none) for the test set. The evaluation was performed by a single expert.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

A multi-reader multi-case (MRMC) comparative effectiveness study was not conducted according to the provided information. The testing involved comparing InVivoDental's performance to predicate devices by a single expert.

6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance:

The primary evaluation described is a standalone performance assessment where the software's rendering and measurement capabilities were directly compared against predicate devices. The "human-in-the-loop" aspect was the expert radiologist evaluating the outputs of both the subject and predicate software, rather than evaluating the AI assisting a human. The software itself is designed to be used by clinicians independently.

7. Type of Ground Truth Used:

The ground truth for the bench testing was established by comparing the performance (specifically measurement and rendering accuracy) of InVivoDental against that of predicate software. The "truth" was based on the accepted functionality and output of already-marketed equivalent devices.

8. Sample Size for the Training Set:

The submission does not provide any information regarding a training set size. This suggests that the device's development and validation relied on comparison to predicate devices and established software engineering practices rather than a machine learning model that would require a distinct training set.

9. How the Ground Truth for the Training Set Was Established:

Since no information regarding a training set or machine learning model is provided, the method for establishing ground truth for a training set is not applicable in this context. The validation focuses on verification through quality assurance measures and performance testing against predicate devices.

{0}------------------------------------------------

5. 510K Summary

FEB 0 1 2013

Submitter Anatomage Inc. 111 N. Market Street #800 San Jose, CA 95113 Phone: (408) 885-7471 Fax: (408) 295-9786 Contact Person Mike Mendez Phone Number: (408) 885-7471 ext 119 Device Name InVivoDental 510(k) Preparation Date 11/02/12 System, Image Processing, Radiological Common Name Classification Class II Imaging Processing System, LLZ, 21 CFR 892.2050 Classification Name Product Code LLZ, 21 CFR 892.2050 Device Description InVivoDental is a volumetric imaging software designed specifically for clinicians, doctors, physicians, and other qualified medical professionals. The software runs in Windows operating systems and visualizes medical imaging data on the computer screen. The software is downloaded over the internet and installed on the customer's computer. Users are able to examine anatomy on a computer screen and use software tools to move and manipulate rendered images by turning, zooming, flipping, adjusting contrast and brightness, cutting, and slicing. There are several rendering settings to emphasize bone and/or soft tissue. The software also has the ability to create panoramic images, superimpose images, and create measurements of volume, angle, and length. There are multiple tools to annotate and otherwise mark areas of interest on the images. The software has specialized tools to simulate surgical treatments using models rendered by Antomage, and any view or user modified area can be saved to a gallery. These simulations and galleries can then be reviewed by medical personnel or used for consultation between the doctor and

K 123519

{1}------------------------------------------------

Intended Use InVivoDental is a software application used for the display and 3D visualization of medical image files from scanning devices. such as CT, MRI, or 3D Ultrasound. It is intended for use by radiologists, clinicians, referring physicians and other qualified individuals to retrieve, process, render, review, store, print, assist in diagnosis, and distribute images utilizing standard PC hardware. Additionally, InVivoDental is a preoperative software application used for the simulation and evaluation of dental implants, orthodontic planning and surgical treatments. This device is not indicated for mammography use. Equivalent Devices K070803 Anatomage, InVivoDental IMTEC ILUMA Vision v2.0 K081347 Technological Characteristic InVivoDental and the predicate devices are standalone software. All of these software devices are designed to be installed on standard off-the-shelf x86 processor based computers running the Windows operating system. DICOM data from medical imaging devices are visualized as volume or 2D images. A difference between InVivoDental and the predicate software is the output of DICOM files. DICOM files can be distributed and read by a device that reads DICOM files while the format that InVivoDental exports is intended to be read by the InVivoDental reader. This does not impact safety or effectiveness of the device. Limiting the rendering of images to Anatomage software allows for consistent rendering of images and information by eliminating the use of substandard software that is freely distributed on the internet. Non-Clinical Test Results Testing confirmed that the software is stable and operating as designed. The following quality assurance measures were applied to the development of the system: · Risk Analysis · Requirements Reviews · Design Reviews · Performance testing (Verification) · Safety testing (Verification) · Final acceptance testing (Validation) · Bench testing to compare with predicate software

i

patient.

{2}------------------------------------------------

Testing confirmed that the software is stable and operating as designed. Testing also confirmed that the software evaluated for hazards and that the risk has been reduced to acceptable levels.

Bench testing of the software with predicate software was performed by evaluation of images rendered by InVivoDental and predicate software. This testing and evaluation included testing of measurement tools in both predicate and subject software and was performed by an expert in the field of radiology. This testing confirms that InVivoDental is as effective as its predicates in its ability to perform its essential functions of measurement and rendering of DICOM data.

Based on the intended use, product, performance, and testing information provided in this notification, the subject device has been shown to be substantially equivalent in technology. functionality, and indicated use to the currently marketed predicate devices.

Summary

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized graphic of what appears to be an abstract representation of the human form.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 1, 2013

Anatomage, Inc. c/o Mr. Mike Mendez Official Correspondent 111 N. Market Street, Suite 800 SAN JOSE CA 95113

Re: K123519

Trade/Device Name: InVivoDental Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 2, 2012 Received: November 15, 2012

Dear Mr. Mendez:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/Reportal.roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Sean M. Boyd -S for

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

4. Indications For Use Statement

510(k) Number (if known): _

Device Name: __ InVivoDental

Indications for Use:

In VivoDental is a software application used for the display and 3D visualization of medical image files from scanning devices, such as CT, MRI, or 3D Ultrasound. It is intended for use by radiologists, clinicians, referring physicians and other qualified individuals to retrieve, process, render, review, store, print, assist in diagnosis and distribute images, utilizing standard PC hardware. Additionally, In VivoDental is a preoperative software application used for the simulation of dental implants, orthodontic planning and surgical treatments.

This device is not indicated for mammography use.

Prescription Use
(Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use
(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Sean M. Boyd -S

Division Sign-Off Office of In Vitro Diagnostics and Radiological Health (OIR) Evaluation and Safety K123519 510(k)

Page 1 of

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).