K Number
K123487
Device Name
GSO GS1 CANNULATED SCREW SYSTEM
Date Cleared
2013-03-11

(118 days)

Product Code
Regulation Number
888.3070
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The GSO GS1 Cannulated Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion. It is intended to be used as a temporary construct that assists normal healing and is not intended to replace normal body structures. The GSO GS1 Cannulated Screw System should be removed after fusion. As a pedicle screw system, the GSO GS1 Cannulated Screw System is intended for patients: (a) having severe spondylolisthesis (Grade 3 and 4) of the fifth lumbar-first sacral (L5-S1) vertebral joint; (b) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); (c) who are receiving fusions using autogenous bone graft only; and (d) who are having the device removed after the development of a solid fusion mass. In addition, when used as a pedicle screw system, the GSO GS1 Cannulated Screw System is indicated for the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). When used as a posterior, non-cervical, non-pedicle screw and/or hook fixation system, the GSO GS1 Cannulated Screw System is indicated for: 1. Degenerative disc disease (as defined by chronic back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) 2. Idiopathic scoliosis 3. Kyphotic deformities of the spine 4. Paralytic scoliosis and/or pelvis obliquity 5. Vertebral fracture or dislocation 6. Neuromuscular scoliosis associated with pelvic obliquity 7. Vertebral fracture or dislocation 8. Tumors 9. Spondylolisthesis 10. Stenosis 11. Pseudarthrosis 12. Unsuccessful previous attempts at spinal fusion For posterior, non-pedicle, screw use, the GSO GS1 screws are intended for sacral/iliac attachment only and the GSO GS1 hooks and crosslinks are intended for thoracic and/or lumbar use only.
Device Description
The GSO GS1 Spinal System consists of rods, screws, hooks, and connecting components that can be locked rigidly into various configurations to build a spinal construct specific to the needs of each individual patient. The implants are attached to the spine posteriorly by means of screws and/or hooks joined with rods. The current submission, the GSO GS1 Cannulated Screw System, adds the following components to the GSO GS1 System: 5.5mm and 1/4" straight longitudinal rods in lengths of 50 - 480mm; 5.5mm and 1/4" pre-bent rods in lengths of 30 - 120mm; and multi-directional cannulated screws with standard or reduction heads in diameters of 5.5mm - 8.5 and lengths of 25mm - 100mm (depending on the screw diameter) to fit either 5.5mm or 1/2" rods. These additional components are compatible with the previously cleared GSO GS1 rods, screws, hooks and connectors. The GSO GS1 Spinal System can be installed with any suitable instrumentation. The GSO GS1 Spinal System components are manufactured from CP Titanium conforming to ASTM F67 and Ti-6Al-4V Titanium alloy conforming to ASTM F136.
More Information

Not Found

No
The 510(k) summary describes a mechanical spinal implant system and does not mention any software, algorithms, or data processing capabilities that would indicate the use of AI or ML.

Yes
The device is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for a variety of conditions, which are therapeutic goals.

No

The device is described as a "Spinal System" composed of rods, screws, hooks, and connecting components intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion. It is a temporary construct used in surgical procedures, not for diagnosing medical conditions.

No

The device description explicitly states it consists of physical components like rods, screws, hooks, and connecting components made from titanium.

Based on the provided text, the GSO GS1 Cannulated Screw System is not an IVD (In Vitro Diagnostic) device.

Here's why:

  • Intended Use: The intended use clearly states that the device is for "immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion." This describes a surgical implant used in vivo (within the body) to support the spine.
  • Device Description: The device is described as consisting of "rods, screws, hooks, and connecting components" made of titanium. These are physical implants designed to be surgically placed.
  • Lack of IVD Characteristics: IVD devices are used to examine specimens (like blood, urine, or tissue) in vitro (outside the body) to provide information about a patient's health. The description of the GSO GS1 system does not involve any such testing or analysis of biological samples.

The device is a surgical implant used for spinal fusion, which is a completely different category of medical device than an IVD.

N/A

Intended Use / Indications for Use

The GSO GS1 Cannulated Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion. It is intended to be used as a temporary construct that assists normal healing and is not intended to replace normal body structures. The GSO GS1 Cannulated Screw System should be removed after fusion.

As a pedicle screw system, the GSO GS1 Cannulated Screw System is intended for patients: (a) having severe spondylolisthesis (Grade 3 and 4) of the fifth lumbar-first sacral (L5-S1) vertebral joint; (b) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); (c) who are receiving fusions using autogenous bone graft only; and (d) who are having the device removed after the development of a solid fusion mass.

In addition, when used as a pedicle screw system, the GSO GS1 Cannulated Screw System is indicated for the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).

When used as a posterior, non-cervical, non-pedicle screw and/or hook fixation system, the GSO GS1 Cannulated Screw System is indicated for:

    1. Degenerative disc disease (as defined by chronic back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies)
    1. Idiopathic scoliosis
    1. Kyphotic deformities of the spine
    1. Paralytic scoliosis and/or pelvis obliquity
    1. Vertebral fracture or dislocation
    1. Neuromuscular scoliosis associated with pelvic obliquity
    1. Vertebral fracture or dislocation
    1. Tumors
    1. Spondylolisthesis
    1. Stenosis
    1. Pseudarthrosis
    1. Unsuccessful previous attempts at spinal fusion

For posterior, non-pedicle, screw use, the GSO GS1 screws are intended for sacral/iliac attachment only and the GSO GS1 hooks and crosslinks are intended for thoracic and/or lumbar use only.

Product codes (comma separated list FDA assigned to the subject device)

MNH, MNI, KWP

Device Description

The GSO GS1 Spinal System consists of rods, screws, hooks, and connecting components that can be locked rigidly into various configurations to build a spinal construct specific to the needs of each individual patient. The implants are attached to the spine posteriorly by means of screws and/or hooks joined with rods.

The current submission, the GSO GS1 Cannulated Screw System, adds the following components to the GSO GS1 System: 5.5mm and ¼" straight longitudinal rods in lengths of 50 - 480mm; 5.5mm and 1/4" pre-bent rods in lengths of 30 - 120mm; and multi-directional cannulated screws with standard or reduction heads in diameters of 5.5mm - 8.5 and lengths of 25mm - 100mm (depending on the screw diameter) to fit either 5.5mm or 1/2" rods. These additional components are compatible with the previously cleared GSO GS1 rods, screws, hooks and The GSO GS1 Spinal System can be installed with any suitable connectors. instrumentation.

The GSO GS1 Spinal System components are manufactured from CP Titanium conforming to ASTM F67 and Ti-6Al-4V Titanium alloy conforming to ASTM F136.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Spinal segments, fifth lumbar-first sacral (L5-S1) vertebral joint, lumbar and sacral spine (L3 and below), thoracic, lumbar, and sacral spine, posterior, non-cervical, sacral/iliac, thoracic and/or lumbar.

Indicated Patient Age Range

Skeletally mature patients

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Non-Clinical Testing:
Testing of the GSO GS1 components was conducted according to ASTM F1717-12. This testing demonstrated that the failure loads for the GSO GS1 Cannulated Screw System were substantially equivalent to the failure loads for the predicate GSO GS1 Spinal System.

Clinical Testing:
Clinical testing was not necessary to demonstrate the substantial equivalence of the GSO GS1 Cannulated Screw System to the predicate devices.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K070966, K121020, K032265

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

0

510(k) SUMMARYMAR 11 2013
Date:January 16, 2013
Applicant/Sponsor:Gold Standard Orthopaedics, LLC.
1226 Rowan St.
Louisville, KY 40203
Contact Person:David Baughman
President
David06@Baughmangroup.com
Phone (502) 581-8770
Proprietary Name:GSO GS1 Cannulated Screw System
Common Name:Spinal Implants
Classification Name:Spinal interlaminal fixation orthosis, 21 CFR 888.3050-KWP
Pedicle screw spinal system, 21 CFR 888.3070-MNH, MNI

Legally Marketed Devices to which Substantial Equivalence is claimed:

GSO GS1 Spinal System (K070966)

DePuy Spine VIPER System (K121020)

Medtronic Sofamor Danek CD Horizon Spinal System (K032265)

Device Description:

The GSO GS1 Spinal System consists of rods, screws, hooks, and connecting components that can be locked rigidly into various configurations to build a spinal construct specific to the needs of each individual patient. The implants are attached to the spine posteriorly by means of screws and/or hooks joined with rods.

The current submission, the GSO GS1 Cannulated Screw System, adds the following components to the GSO GS1 System: 5.5mm and ¼" straight longitudinal rods in lengths of 50 - 480mm; 5.5mm and 1/4" pre-bent rods in lengths of 30 - 120mm; and multi-directional cannulated screws with standard or reduction heads in diameters of 5.5mm - 8.5 and lengths of 25mm - 100mm (depending on the screw diameter) to fit either 5.5mm or 1/2" rods. These additional components are compatible with the previously cleared GSO GS1 rods, screws, hooks and The GSO GS1 Spinal System can be installed with any suitable connectors. instrumentation.

1

The GSO GS1 Spinal System components are manufactured from CP Titanium conforming to ASTM F67 and Ti-6Al-4V Titanium alloy conforming to ASTM F136.

Intended Use / Indications:

The GSO GS1 Cannulated Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion. It is intended to be used as a temporary construct that assists normal healing and is not intended to replace normal body structures. The GSO GS1 Cannulated Screw System should be removed after fusion.

As a pedicle screw system, the GSO GS1 Cannulated Screw System is intended for patients: (a) having severe spondylolisthesis (Grade 3 and 4) of the fifth lumbar-first sacral (L5-S1) vertebral joint; (b) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); (c) who are receiving fusions using autogenous bone graft only; and (d) who are having the device removed after the development of a solid fusion mass.

In addition, when used as a pedicle screw system, the GSO GS1 Cannulated Screw System is indicated for the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative . spondylolisthesis with objective .evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).

When used as a posterior, non-cervical, non-pedicle screw and/or hook fixation system, the GSO GS1 Cannulated Screw System is indicated for:

    1. Degenerative disc disease (as defined by chronic back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies)
  • Idiopathic scoliosis 2.
  • Kyphotic deformities of the spine 3.
  • Paralytic scoliosis and/or pelvis obliquity 4.
    1. Vertebral fracture or dislocation
  • Neuromuscular scoliosis associated with pelvic obliquity 6.
  • Vertebral fracture or dislocation 7.
  • Tumors 8.
  • Spondylolisthesis ರು.
    1. Stenosis
    1. Pseudarthrosis
    1. Unsuccessful previous attempts at spinal fusion

For posterior, non-pedicle, screw use, the GSO GS1 screws are intended for sacral/iliac attachment only and the GSO GS1 hooks and crosslinks are intended for thoracic and/or lumbar use only.

2

Summary of Technologies/Substantial Equivalence:

The GSO GS1 Cannulated Screw System has the same indications, a similar design, and is manufactured from the same materials as the GSO GS1 Spinal System cleared in K070966. The cannulated screw components of the GSO GS1 Cannulated Screw System are similar in design to the cannulated screw components of the DePuy Spine VIPER System and the Medtronic Sofamor Danek CD Horizon Spinal System.

Non-Clinical Testing:

Testing of the GSO GS1 components was conducted according to ASTM F1717-12. This testing demonstrated that the failure loads for the GSO GS1 Cannulated Screw System were substantially equivalent to the failure loads for the predicate GSO GS1 Spinal System.

Clinical Testing:

Clinical testing was not necessary to demonstrate the substantial equivalence of the GSO GS1 Cannulated Screw System to the predicate devices.

3

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name arranged in a circular pattern around a stylized emblem. The emblem consists of a caduceus-like symbol with three branches, representing health and human services.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 11, 2013

Gold Standard Orthopaedics, LLC % Mr. David Baughman President 1226 Rowan Street Louisville, Kentucky 40203

Re: K123487

Trade/Device Name: GSO GS1 Cannulated Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNH, MNI, K WP Dated: January 18, 2013 Received: January 22, 2013

Dear Mr. Baughman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

4

Page 2 – Mr. David Baughman

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Erin DKeith

Mark N. Melkerson Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

5

INDICATIONS FOR USE

510(k) Number (if known): K123487

Device Name: GSO GS1 Cannulated Screw System

Indications for Use:

The GSO GS1 Cannulated Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion. It is intended to be used as a temporary construct that assists normal healing and is not intended to replace normal body structures. The GSO GS1 Cannulated Screw System should be removed after fusion.

As a pedicle screw system, the GSO GS1 Cannulated Screw System is intended for patients: (a) having severe spondylolisthesis (Grade 3 and 4) of the fifth lumbar-first sacral (L5-S1) vertebral joint; (b) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); (c) who are receiving fusions using autogenous bone graft only; and (d) who are having the device removed after the development of a solid fusion mass.

In addition, when used as a pedicle screw system, the GSO GS1 Cannulated Screw System is indicated for the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal turnor, and failed previous fusion (pseudarthrosis).

Prescription Use __ X AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Ronald P. Jean -S

(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K123487

Page 1 of 2

6

When used as a posterior, non-cervical, non-pedicle screw and/or hook fixation system, the GSO GS1 Cannulated Screw System is indicated for:

    1. Degenerative disc disease (as defined by chronic back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies)
    1. Idiopathic scoliosis
  • Kyphotic deformities of the spine 3.
  • Paralytic scoliosis and/or pelvis obliquity 4.
  • Vertebral fracture or dislocation ഗ്
  • Neuromuscular scoliosis associated with pelvic obliquity 6.
    1. Vertebral fracture or dislocation
    1. Tumors
    1. Spondvlolisthesis
    1. Stenosis
    1. Pseudarthrosis
    1. Unsuccessful previous attempts at spinal fusion

For posterior, non-pedicle, screw use, the GSO GS1 screws are intended for sacral/iliac attachment only and the GSO GS1 hooks and crosslinks are intended for thoracic and/or lumbar use only.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Ronald P. Jean -S

(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K123487

Page 2 of 2

23