K Number
K123197
Manufacturer
Date Cleared
2012-12-05

(55 days)

Product Code
Regulation Number
866.3130
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Verigene® Clostridium difficile Nucleic Acid Test (CDF) is a qualitative multiplexed in vitro diagnostic test for the rapid detection of toxin A (tcdA), toxin B (tcdB), and tcdC gene sequences of toxigenic Clostridium difficile and for presumptive identification of PCR ribotype 027 strains from unformed (liquid or soft) stool specimens collected from patients suspected of having C. difficile infection (CDI). Presumptive identification of the PCR ribotype 027 strain of C. difficile is by detection of the binary toxin (cdt) gene sequence and the single base pair deletion at nucleotide 117 in the tcdC gene encodes for a negative regulator in C. difficile toxin production. The test is performed on the Verigene System and utilizes automated specimen preparation and polymerase chain reaction (PCR) amplification, combined with a nanoparticle-based array hybridization assay to detect the toxin gene sequences associated with toxin-producing C. difficile.

The CDF Test is indicated for use as an aid in the diagnosis of CDI. Detection of PCR ribotype 027 strains of C. difficile by the CDF Test is solely for epidemiological purposes and is not intended to guide or monitor treatment for C. difficile infections. Concomitant culture is necessary only if further typing or organism recovery is required.

Device Description

The Verigene C. difficile Nucleic Acid Test (CDF) is a molecular assay which relies on detection of specific nucleic acid targets in a microarray format. For each of the bacterial nucleic acid sequences detected by CDF, unique Capture and Mediator oligonucleotides are utilized, with gold nanoparticle probe-based endpoint detection. The Capture oligonucleotides are covalently bound to the microarray substrate and hybridize to a specific portion of the nucleic acid targets. The Mediator oligonucleotides have a region which bind to a different portion of the same nucleic acid targets and also have a sequence which allows binding of a gold nanoparticle probe. Specific silver enhancement of the bound gold nanoparticle probes at the capture sites results in gold-silver aggregates that scatter light with high efficiency and provide accurate detection of target capture.

The CDF Test is performed on the Verigene System, a "sample-to-result", fully automated, bench-top molecular diagnostics workstation. The System consists of two components: the Verigene Reader and the Verigene Processor SP. The Reader is the user interface, serving as the central control unit for test processing, automated imaging, and result generation. The Processor SP automates sample preparation, target amplification, and hybridization-detection.

AI/ML Overview

The medical device being described is the Verigene® Clostridium difficile Nucleic Acid Test (CDF), which is a qualitative multiplexed in vitro diagnostic test for the rapid detection of toxin B (tcdB), and tcdC gene sequences of toxigenic Clostridium difficile and for presumptive identification of PCR ribotype 027 strains from unformed (liquid or soft) stool specimens collected from patients suspected of having C. difficile infection (CDI).


1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly stated as distinct predefined numerical targets in the provided text. Instead, the study's findings demonstrate the performance, which then implies the met "acceptance criteria." The main performance metrics provided are for sensitivity, specificity, and accuracy against reference methods.

Performance MetricAcceptance Criteria (Implied)Reported Device PerformanceStudy Section
Toxigenic C. difficile Detection
Sensitivity (Direct Culture & PCR Ribotyping)High sensitivity (e.g., >90%)98.7% (154/156) (95.5-99.8% CI)Method Comparison
Specificity (Direct Culture & PCR Ribotyping)High specificity (e.g., >85%)87.6% (1500/1713) (85.9-89.1% CI)Method Comparison
Sensitivity (Enriched Culture & PCR Ribotyping)High sensitivity (e.g., >90%)91.8% (247/269) (87.9-94.8% CI)Method Comparison
Specificity (Enriched Culture & PCR Ribotyping)High specificity (e.g., >90%)92.5% (1480/1600) (91.1-93.7% CI)Method Comparison
Sensitivity (Direct Culture & BDS)High sensitivity (e.g., >90%)98.7% (156/158) (95.5-99.9% CI)Method Comparison
Specificity (Direct Culture & BDS)High specificity (e.g., >85%)87.5% (1500/1715) (85.8-89.0% CI)Method Comparison
Sensitivity (Enriched Culture & BDS)High sensitivity (e.g., >90%)91.9% (251/273) (88.1-94.9% CI)Method Comparison
Specificity (Enriched Culture & BDS)High specificity (e.g., >90%)92.5% (1480/1600) (91.1-93.7% CI)Method Comparison
Toxigenic C. difficile / 027 Strain Detection
Positive Agreement (Direct Culture & PCR Ribotyping)High agreement (e.g., >90%)97.5% (39/40) (86.8-99.9% CI)Method Comparison
Negative Agreement (Direct Culture & PCR Ribotyping)High agreement (e.g., >90%)97.8% (1787/1828) (97.0-98.4% CI)Method Comparison
Positive Agreement (Enriched Culture & PCR Ribotyping)High agreement (e.g., >90%)91.4% (53/58) (81.0-97.1% CI)Method Comparison
Negative Agreement (Enriched Culture & PCR Ribotyping)High agreement (e.g., >90%)98.5% (1783/1811) (97.8-99.0% CI)Method Comparison
Positive Agreement (Direct Culture & BDS)High agreement (e.g., >90%)97.7% (42/43) (87.7-99.9% CI)Method Comparison
Negative Agreement (Direct Culture & BDS)High agreement (e.g., >90%)97.8% (1790/1830) (97.0-98.4% CI)Method Comparison
Positive Agreement (Enriched Culture & BDS)High agreement (e.g., >90%)93.7% (59/63) (84.5-98.2% CI)Method Comparison
Negative Agreement (Enriched Culture & BDS)High agreement (e.g., >90%)98.7% (1787/1810) (98.1-99.2% CI)Method Comparison
Analytical Sensitivity (LoD)LoD ≤ 1250 CFU/mL of stool63 to 1250 CFU/ml of stoolAnalytical Testing
Analytical Specificity (Cross-reactivity)100% specificity100%Analytical Testing
Microbial/InterferenceNo interference observedNo interference observedAnalytical Testing
Carry-over / Cross-contaminationNo carry-over or cross-contaminationNo carry-over or cross-contaminationAnalytical Testing
PrecisionExpected agreement rates (e.g., ~95% for LP, ~20-80% for HN)Met expected agreement ratesClinical Testing
ReproducibilityHigh inter-site agreementMet expected agreement ratesClinical Testing

Note: The acceptance criteria were not explicitly stated with specific numerical thresholds but are inferred from the reported performance which were considered sufficient for substantial equivalence.

2. Sample size used for the test set and the data provenance

  • Sample Size for Clinical (Method Comparison) Test Set: 1,875 specimens were evaluated in the clinical study (n=1875). For specific comparisons, some samples were excluded due to inconclusive or missing reference results:
    • 6 specimens were not PCR-ribotyped.
    • 2 specimens were not sequenced for tcdC Bi-Directional Sequencing (BDS).
  • Data Provenance: The data was collected from a multi-site prospective investigation study at five U.S. institutions.
    • This indicates the data is prospective and collected within the United States.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The text does not explicitly state the "number of experts" or their "qualifications" used to establish the ground truth. The ground truth for the clinical test set was established through a combination of reference culture followed by cell cytotoxicity testing on isolates, and subsequent strain typing by PCR Ribotyping and Bi-Directional Sequencing (BDS) at a central laboratory and an external third-party site. This suggests laboratory professionals and/or researchers were involved, but their specific expert qualifications (e.g., years of experience) are not detailed.

4. Adjudication method for the test set

The text describes using a "central laboratory" for initial culture and cytotoxin B testing, and an "external third-party site" for PCR Ribotyping. Bi-Directional Sequencing was done either on culture-confirmed isolates or DNA extracted during PCR Ribotyping. There isn't a described "adjudication method" in the sense of multiple experts reviewing and reaching a consensus on cases where initial results varied. Instead, the ground truth was established hierarchically and sequentially using definitive laboratory methods. For instance, if C. difficile was isolated from direct culture and was cytotoxin positive, it was deemed "toxigenic C. difficile positive." If not, enriched culture was used.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is an automated molecular diagnostic test, not an AI-assisted diagnostic device that human readers interpret. Therefore, the concept of "human readers improving with AI vs. without AI assistance" is not applicable.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, a standalone study was done. The Verigene CDF Test operates on the Verigene System, which is described as a "sample-to-result, fully automated, bench-top molecular diagnostics workstation." The results are generated by a "software-based decision algorithm resident in the Reader." The performance data presented (analytical and clinical) represent the standalone performance of this automated device and its algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The ground truth used for the clinical (method comparison) study was a combination of:

  • Reference Culture: Inoculation onto CCFA-D and CCMB-Tal media.
  • Cell Cytotoxicity Testing: Performed on isolates from the cultures.
  • Molecular Strain Typing:
    • PCR Ribotyping (at an external third-party site, with Agilent 2100 Bioanalyzer for added discrimination).
    • Bi-Directional Sequencing (BDS) of the tcdC gene.

This constitutes a robust, multi-faceted laboratory-based "reference standard" or "gold standard."

8. The sample size for the training set

The document does not provide information on the sample size for a training set. As this device is a molecular diagnostic test based on PCR and hybridization, it is likely designed and validated using analytical samples and clinical samples, but the term "training set" in the context of machine learning (where algorithms learn from data) is not used. The "cutoff verification" mentioned in analytical testing involved 59 C. difficile strains to verify cut-off values for the two-tiered filter algorithm, which might be considered an internal validation step, but not a "training set" in the common AI sense.

9. How the ground truth for the training set was established

As no "training set" is explicitly mentioned in the context of the device's development or a machine learning approach, the method for establishing its ground truth is not described. The analytical studies (LoD, reactivity, specificity, etc.) utilized independently-confirmed C. difficile strains and other microorganisms with known characteristics.

{0}------------------------------------------------

510(K) Summary

510(k) Number:

K123197: Verigene® Clostridium difficile Nucleic Acid Test (CDF)

Summary Preparation Date:

November 27, 2012

Submitted by:

Nanosphere, Inc. 4088 Commercial Avenue Northbrook, IL 60062 Phone: 847-400-9000 Fax: 847-400-9176

Contact:

Mark A. Del Vecchio Vice President, Regulatory Affairs

Proprietary Names:

For the instrument: Verigene® System For the assay: Verigene® Clostridium difficile Nucleic Acid Test (CDF)

Common Names:

For the instrument:

Bench-top molecular diagnostics workstation

For the assay:

Clostridium difficile Nucleic Acid Test Clostridium difficile identification and differentiation system C. difficile assay C. diff test

Nanosphere, Inc.

Page 1 of 10

November 27, 2012

DEC. 5 2012

{1}------------------------------------------------

Regulatory Information:

Regulation section:

866.3130

Classification:

Class II

Panel:

Microbiology (83)

Product Code(s):

OZN Clostridium difficile toxin gene amplification assay

Other codes used by predicate devices:

OMN C. difficile nucleic acid test assay

Reagents, Clostridium Difficile toxin LLH

Respiratory Virus Panel Nucleic Acid System OCC

Instrumentation for clinical multiplex test systems NSU

Predicate Devices:

Portrait Toxigenic C. Difficile Assay (K113358) (Great Basin Scientific)

Xpert C. difficile/Epi Assay (K110203) (Cepheid)

Verigene RVNATSP Test (K092566) (Nanosphere)

Indications for Use:

The Verigene® Clostridium difficile Nucleic Acid Test (CDF) is a qualitative multiplexed in vitro diagnostic test for the rapid detection of toxin B (tcdB), and tcdC gene sequences of toxigenic Clostridium difficile and for presumptive identification of PCR ribotype 027 strains from unformed (liquid or soft) stool specimens collected from patients suspected of having C. difficile infection (CDI). Presumptive identification of the PCR ribotype 027 strain of C. difficile is by detection of the binary toxin (cdt) gene sequence and the single base pair deletion at nucleotide 117 in the tedC gene encodes for a negative regulator in C. difficile toxin production. The test is performed on the Verigene System and utilizes automated specimen preparation and polymerase chain reaction (PCR) amplification, combined with a nanoparticle-based array hybridization assay to detect the toxin gene sequences associated with toxin-producing C. difficile.

Nanosphere, Inc.

Page 2 of 10

{2}------------------------------------------------

The CDF Test is indicated for use as an aid in the diagnosis of CDI. Detection of PCR ribotype 027 strains of C. difficile by the CDF Test is solely for epidemiological purposes and is not intended to guide or monitor treatment for C. difficile infections. Concomitant culture is necessary only if further typing or organism recovery is required.

Technological Characteristics:

The Verigene C. difficile Nucleic Acid Test (CDF) is a molecular assay which relies on detection of specific nucleic acid targets in a microarray format. For each of the bacterial nucleic acid sequences detected by CDF, unique Capture and Mediator oligonucleotides are utilized, with gold nanoparticle probe-based endpoint detection. The Capture oligonucleotides are covalently bound to the microarray substrate and hybridize to a specific portion of the nucleic acid targets. The Mediator oligonucleotides have a region which bind to a different portion of the same nucleic acid targets and also have a sequence which allows binding of a gold nanoparticle probe. Specific silver enhancement of the bound gold nanoparticle probes at the capture sites results in gold-silver aggregates that scatter light with high efficiency and provide accurate detection of target capture.

The CDF Test is performed on the Verigene System, a "sample-to-result", fully automated, bench-top molecular diagnostics workstation. The System enables automated nucleic acid extraction from unformed stool specimens (liquid or soft) and detection of bacterial-specific target DNA. The Verigene System consists of two components: the Verigene Reader and the Verigene Processor SP.

The Reader is the Verigene System's user interface, which serves as the central control unit for all aspects of test processing, automated imaging, and result generation using a touch-screen control panel and a barcode scanner. The Verigene Processor SP executes the test procedure, automating the steps of (1) Sample Preparation and Target Amplification - cell lysis and magnetic bead-based bacterial DNA isolation and amplification, and (2) Hybridizationdetection and identification of bacterial-specific DNA in a microarray format by using gold nanoparticle probe-based technology. Once the specimen is loaded by the operator, all other fluid transfer steps are performed by an automated pipette that transfers reagents between wells of the travs and finally loads the specimen into the Test Cartridge for hybridization. Single-use disposable test consumables and a self-contained Verigene Test Cartridge are utilized for each sample tested with the CDF assay.

To obtain the test results after test processing is complete, the user removes the Test Cartridge from the Processor SP, and inserts the substrate holder into the Reader for analysis. Light scatter from the capture spots is imaged by the Reader and intensities from the microarray spots are used to make a determination regarding the presence (Detected) or absence (Not Detected) of a bacterial nucleic acid sequence/analyte. This determination is made by means of software-based decision algorithm resident in the Reader.

Nanosphere, Inc.

Page 3 of 10

{3}------------------------------------------------

Performance Data - Analytical Testing

Analytical Sensitivity / Limit of Detection (LoD)

Analytical sensitivity (LoD) of the CDF Test was determined for seven strains of C. difficile, representing all major toxinotypes found in North America and including two PCR Ribotype 027 strains. The LoD was defined as the concentration at which the test produces a positive result greater than 95% of the time. Serial dilutions of the strains were tested and the putative LoD confirmed with 20 replicates. The LoDs for the seven strains are shown in the table below and ranged from 63 to 1250 CFU/ml of stool. Thus, the study established the overall limit of detection of the CDF Test to be 1250 CFU/ml of organism present in stool.

Strain Designation(Source ID)ToxinotypeCalculatedCFU/mL Stoolat LoDCFU perCDF Test atLoDLoDConfirmationResults
ATCC BAA-1805III250520/20
ATCC 43255 (VPI 10463)0631.2520/20
ATCC BAA-1875 (5325)V5001020/20
CDC 2007858IX/XXIII12502520/20
CDC 2009087012502520/20
CDC 2009292III12502520/20
ATCC 43598 (1470)VIII250520/20

Analytical Reactivity (Inclusivity)

Analytical reactivity of the CDF Test was demonstrated with a comprehensive panel of sixtythree (63) independently-confirmed C. difficile strains, tested in triplicate at three times the LoD (i.e. 3.750 CFU/mL). The panel was comprised of a wide range of toxinotypes, including toxinotypes 0, 1, IV, V, VIII, IX, X, XI, XII, XXI, XXII, IX/XXIII, XIV/XV, and six (6) PCR ribotype 027 strains (toxinotype III).

All tests correctly reported the expected results for the detection of gene sequences for toxigenic C. difficile and for presumptive PCR ribotype 027, with one exception. Strain CDC 2009048 strain, classified by the CDC as Toxinotype XIV/XV, is associated with non-027 strains (Ribotypes 111/122). However, the Verigene CDF Test reported detection of the tcdA, tedB, binary and tcdC-MUT targets as would be expected for a PCR ribotype 027 strain. Subsequent sequencing of the tcdC gene verified the presence of the A117 deletion.

Nanosphere, Inc.

Page 4 of 10

{4}------------------------------------------------

Analytical Specificity (Cross-reactivity)

Ninety-four (94) microorganisms, including two (2) non-toxigenic C. difficile strains and fourteen (14) non C. difficile Clostridium species, along with one human cell line, were tested with the CDF Test to determine analytical specificity. In addition, the cross-reactivity of Clostridium botulinum was evaluated by in silico analysis.

Each bacterial strain was prepared in Negative Stool Matrix and tested in triplicate in concentrations of 5×106 CFU/mL stool. Two (2) organisms. Cryptosporidium parvum and Giardia lamblia, were tested using genomic DNA at a concentration of 1×10° copies of gDNA. For the viruses. Echovirus 11 and Coxsackievirus were tested at 5×10° PFU/mL stool. Adenovirus, Enterovirus, Cytomegalovirus and Rotavirus were also tested using genomic DNA or RNA at a concentration of 1×10° copies per reaction. Noroviruses were tested as clinical samples.

Analytical specificity was observed to be 100%. including that determined by in silico analysis.

Microbial Interference

Microorganisms that may be encountered in clinical stool samples, but not detected by the CDF Test, were tested in this study to evaluate the potential for microbial interference. The CDF Test was tested against the same ninety-five (95) organisms/cell line that were used for analytical specificity, at the same medically relevant concentrations, using two strains of toxigenic C. difficile (ATCC BAA-1805 [toxinotype III] and ATCC 43255 [toxinotype 0]) at 1.5x LoD and 3x LoD, respectively.

No interference was observed with the CDF Test for any of the samples tested.

Interference

Thirty-four (34) products/exogenous substances (shown in the following table) that are possibly encountered in stool samples were evaluated for potential inhibitory effects with the CDF Test. Each interfering substance was evaluated at its "worst case" concentration, against two C. difficile strains (ATCC BAA-1805, ATCC 43255). Additionally, Carv-Blair media was tested, None of the thirty four (34) substances or the Cary-Blair media tested in this study showed any inhibitory effect on the detection of C. difficile using the CDF Test.

Nanosphere, Inc.

Page 5 of 10

{5}------------------------------------------------

Stearic AcidPreparation H® Hemorrhoidal OintmentAluminum Hydroxide, Reagent Grade
Palmitic AcidWalgreens Enema Mineral Oil LaxativeMesalazine
Whole BloodOptions Conceptrol® Vaginal Contraceptive GelImmodium® AD Anti-Diarrheal
Nasopharyngeal Swab Sample in UniversalTransport Media (UTM)Dulcolax® Laxative SuppositoriesPepto-Bismol Max Strength
Nystatin SuspensionDimenhydrinateEx-lax® Maximum Strength StimulantLaxative
Monistat® 3Neosporin® First Aid Antibiotic OintmentVancomycin
Preparation H® Medicated WipesWet Ones® Antibacterial Hand WipesMetronidazole Topical Cream (0.75%)
Vagisil Anti-Itch Crème Maximum Strength .K-Y® Personal Lubricant JellyNaproxen Sodium
Preparation H® Anti-Itch Hydrocortisone 1%Vaseline Original 100% Pure Petroleum JellyMucin from bovine submaxillaryglands, Type I-S (Dehydrated)
Desitin Maximum Strength Original PasteSarna Anti-Itch Lotion, SensitiveBarium Sulfate
Gaviscon® Extra Strength Liquid AntacidBile, bovine, dried, unfractionedCary-Blair Medium
Phillips® Genuine Milk of Magnesia SalineLaxativeTums® Antacid with Calcium Extra Strength 750

Carry-over / Cross-contamination

The potential for carry-over and cross-contamination of the CDF Test on the Verigene system was assessed by alternately testing a high positive C. difficile sample (toxigenic & ribotype 027 Clostridium difficile strain BAA-1805) at 5x108 CFU/mL, followed by testing a negative sample, comprising only of CDF-Negative Stool Matrix. The high-titer sample was alternated with the negative sample three times on three unique Verigene SP Processors, for a total of eighteen individual tests. No carry-over or cross-contamination was observed.

Cutoff Verification

Analytical testing of 59 strains of C. difficile, comprising a range of toxinotypes and nontoxinogenic strains, were performed in duplicate with the CDF Test to verify the cut off values of the two-tiered filter algorithm. Using the established cut-off levels for the assay, the CDF Test correctly detected the expected analytes for all the verification samples.

Performance Data - Clinical Testing

Precision

The precision study was conducted in-house by Nanosphere, during which a seven-member Precision Study Sample Panel was tested in duplicate twice daily by two operators for twelve non-consecutive days. This testing regime generated a total of 48 replicates per specimen and an overall total of 336 data points.

The seven sample panel was comprised of two different strains (ATCC 43255 and ATCC BAA-1805) at three different concentrations (six positive samples) and one negative sample. This panel included for each strain, a "High Negative" sample, which would be expected to produce a negative result approximately 20% to 80% of the time, a "Low Positive" sample, which would

Nanosphere, Inc.

Page 6 of 10

{6}------------------------------------------------

be expected to produce a positive result approximately 95% of the time, and a "Moderate Positive" sample, which would be expected to yield a positive result approximately 100% of the time. Results are summarized below.

PanelMemberStrainLevelExpectedOccurrenceTotal Agreementwith ExpectedResult*
1CDF Negative StoolMatrixNegative~100%Negative100%48/48(92.6-100%)
2Toxigenic Wild TypeC. difficileModerate Positive (MP)~100% Positive100%48/48(92.6-100%)
3Toxigenic Wild TypeC. difficileLow Positive (LP)~95% Positive97.9%47/48(88.9-100%)
4Toxigenic Wild TypeC. difficileHigh Negative (HN)~20-80%Negative12.5%6/48(4.7-25.3%)
5ToxigenicHypervirulentC. difficileModerate Positive (MP)~100% Positive100%48/48(92.6-100%)
6ToxigenicHypervirulentC. difficile 'Low Positive (LP)~95% Positive95.8%46/48(85.6-99.5%)
7ToxigenicHypervirulentC. difficileHigh Negative (HN)~20-80%Negative20.8%10/48(10.5-35.0%)

Reproducibility

The inter-laboratory reproducibility of the CDF was determined by conducting a reproducibility study at three external sites. Seven samples were tested daily in triplicate by two (2) operators for five (5) non-consecutive days at three (3) sites for a total of ninety (90) tests per sample (3 sites x 2 operators / site x 3 replicates / operator x 5 days = 90 tests per sample). The study tested a total of 630 samples.

The seven (7) sample panel for the reproducibility study was the same panel described previously for the precision study, comprised of two (2) different strains at three (3) different concentrations (six positive samples) and one (1) negative sample. The results of the Reproducibility Study are provided in the table below.

Nanosphere, Inc.

{7}------------------------------------------------

·

:

.

.

. . .

.

.

·

:

· :

.

:

PanelMemberStrainLevelExpectedOccurrenceTotal Agreement with Expected Result*(95% CI)
Site 1Site 2Site 3Total
1CDFNegativeStoolMatrixNegative~100%Negative100%30/30(88.4-100)100%30/30(88.4-100)100%30/30(88.4-100)100%90/90(96.0-100)
2ToxigenicWild TypeC. difficileModeratePositive(MP)~100%Positive100%30/30(88.4-100)100%30/30(88.4-100)100%30/30(88.4-100)100%90/90(96.0-100)
3ToxigenicWild TypeC. difficileLowPositive(LP)~95%Positive100%30/30(88.4-100)100%30/30(88.4-100)100%30/30(88.4-100)100%90/90(96.0-100)
4ToxigenicWild TypeC. difficileHighNegative(HN)~20-80%Negative30%9/30(14.7-49.4)33.30%10/30(17.3-52.8)16.70%5/30(5.6-34.7)26.70%24/90(17.9-37.0)
5ToxigenicHypervirulentC. difficileModeratePositive(MP)~100%Positive100%30/30(88.4-100)100%30/30(88.4-100)100%30/30(88.4-100)100%90/90(96.0-100)
6ToxigenicHypervirulentC. difficileLowPositive(LP)~95%Positive96.70%29/30(82.8-99.9)96.70%29/30(82.8-99.9)100%30/30(88.4-100)97.80%88/90(92.2-99.7)
7ToxigenicHypervirulentC. difficileHighNegative(HN)~20-80%Negative36.70%11/30(19.9-56.1)40.00%12/30(22.7-59.4)30.00%9/30(14.7-49.4)35.60%32/90(25.7-46.4)

:

.

Nanosphere, Inc.

:

.

.

.

Page 8 of 10

:

{8}------------------------------------------------

Method Comparison

The performance characteristics of the CDF Test were determined in a multi-site prospective investigation study at five U.S. institutions by comparing the Verigene CDF Test results to reference culture followed by cell cytotoxicity testing on the isolates and strain typing on the toxigenic strains by PCR Ribotyping and Bi-Directional Sequencing methods.

Subjects included individuals whose routine care called for C. difficile testing. A portion of each leftover unformed stool specimen was obtained for testing with the CDF Test. In parallel to Verigene CDF Testing, an aliquot of the same specimen was sent to a central laboratory for reference culture and cytotoxin B isolate testing. Each stool specimen was inoculated onto prereduced CCFA-D (cycloserine-cefoxitin-fructose agar-direct plate) and CCMB-Tal (cycloserine cefoxitin mannitol broth with taurocholate lysozyme cysteine). After 24 hours the CCMB-TAL was subcultured onto a second CCFA-E plate (CCFA-Enriched). The direct culture method is referred to hereafter as the "direct culture" and the enriched culture method is referred to hereafter as the "enriched culture".

If C. difficile was isolated from the CCFA-D plate and the isolate was positive by the cell cytotoxicity assay, the specimen was classified as "toxigenic C. difficile positive" and the CCFA-E plate was not further analyzed. If no C. difficile was isolated from the CCFA-D plate or if the isolate was negative by the cell cytoxicity assay, the CCFA-E plate was further analyzed.

If CCFA-E was positive for C. difficile and the isolate was positive for cell cytotoxicity assay, the specimen was classified as "toxigenic C. difficile positive". The specimen was reported as "negative" if CCFA-E was negative for C. difficile or the isolate was tested negative by the cell cytotoxicity assay.

Following central site culture testing, the toxigenic C. difficile positive isolates were sent for strain identification by PCR Ribotyping at an external third-party site. In addition to conventional agarose-gel electrophoresis, for added discrimination, PCR products were analyzed using the Agilent 2100 Bioanalyzer with the strain type assignment based on a comparison of isolate sizing profiles with known C. difficile reference strains were designated into two PCR Ribotyping categories, 027 and non-027.

In parallel, following central culture testing, CDF Test extracted DNA from the cultureconfirmed C. difficile positive isolates or the DNA extracted during the PCR Ribotyping were sent for tcdC Bi-Directional Sequencing. Sequencing templates were prepared by PCR amplification of the stored DNA using sequencing primers. The same set of primers was also used for sequencing the amplified material. The PCR primers were designed from conserved regions based on multiple sequence alignments of all available sequence entries for each target in GenBank at the time of this study.

The clinical performance of the CDF Test was evaluated against four basic metrics, consisting of the combination of two reference methods for the detection of toxigenic C. difficile and two reference methods for the detection of the 027 (hypervirulent) strain of C. difficile: 1) Direct Culture and PCR Ribotyping, 2) Enriched Culture and PCR Ribotyping, and 3) Direct Culture

Nanosphere, Inc.

{9}------------------------------------------------

and Bidirectional Sequencing, and 4) Enriched Culture and Bidirectional Sequencing. The overall performance of the CDF Test evaluated against these metrics is summarized in Table 1.

Toxigenic C. difficileToxigenic C. difficile / 027 Strain
Sensitivity(95% CI)Specificity(95% CI)Accuracy(95% CI)PPV(95% CI)NPV(95% CI)AgreementPPV(95% CI)NPV(95% CI)
POS(95% CI)NEG(95% CI)Total(95% CI)
DirectCulture &PCRRibotyping(a)98.7%154/156(95.5-99.8)87.6%1500/1713(85.9-89.1)88.5%1654/1869(87.0-89.9)42.0%154/367(36.9-47.2)99.9%1500/1502(99.5-99.9)97.5%39/40(86.8-99.9)97.8%1787/1828(97.0-98.4)97.7%1826/1869(96.9-98.3)48.2%39/81(36.9-59.5)99.9%1787/1788(99.7-100)
EnrichedCulture &PCRRibotyping(a)91.8%247/269(87.9-94.8)92.5%1480/1600(91.1-93.7)92.4%1727/1869(91.1-93.6)67.3%247/367(62.2-72.1)98.5%1480/1502(97.8-99.1)91.4%53/58(81.0-97.1)98.5%1783/1811(97.8-99.0)98.2%1836/1869(97.5-98.8)65.4%53/81(54.0-75.7)99.7%1783/1788(99.4-99.9)
DirectCulture &BDS (b)98.7%156/158(95.5-99.9)87.5%1500/1715(85.8-89.0)88.4%1656/1873(86.9-89.8)42.1%156/371(37.0-47.3)99.9%1500/1502(99.5-100)97.7%42/43(87.7-99.9)97.8%1790/1830(97.0-98.4)97.8%1832/1873(97.0-98.4)51.2%42/82(39.9-62.4)99.9%1790/1791(99.7-100)
EnrichedCulture &BDS (b)91.9%251/273(88.1-94.9)92.5%1480/1600(91.1-93.7)92.4%1731/1873(91.1-93.6)67.7%251/371(62.6-72.4)98.5%1480/1502(97.8-99.1)93.7%59/63(84.5-98.2)98.7%1787/1810(98.1-99.2)98.6%1846/1873(97.9-99.1)72.0%59/82(60.9-81.3)99.8%1787/1791(99.4-99.9)
Table 1:Summary of Overall CDF Test Clinical Performance n=1875
-------------------------------------------------------------------------

(ਕ) Of the 1,875 specimens evaluated, six specimens were not PCR-ribotyped because the isolate was either not sent or the result was inconclusive. These six specimens were not included in the performance characteristics above.

(b) Of the 1,875 specimens evaluated, two specifive but were not sequenced because the isolate was either not sent or the result was inconclusive. These two specimens were not included in the performance characteristics above.

Substantial Equivalence

The Verigene C. difficile Nucleic Acid Test (CDF) is as safe and effective as the combination predicate consisting of the Cepheid Xpert C. difficile/Epi Assay, the Great Basin Scientific Portrait C. difficile assay and the Nanosphere Verigene RVNATsp. CDF has similar intended use and indications, performance characteristics and principles of operation as Cepheid's Xpert C. difficile/Epi Assay and Great Basin Scientific's Portrait C. difficile assay. CDF has similar technological characteristics as Nanosphere's Verigene RVNATsp. The minor differences between CDF and its predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that the CDF is as safe and effective as the predicate devices. Thus, CDF is substantially equivalent.

Nanosphere, Inc.

{10}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/10/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

DEC 5 2012

Nanosphere, Inc. Mr. Mark A. Del Vecchio 4088 Commercial Avenue Northbrook. IL 60062

Re: K123197

Trade/Device Name: Regulation Number: Regulation Name: Regulatory Class: Product Code: Dated: Received:

Verigene® Clostridium difficile Nucleic Acid Test (CDF) 21 CFR 866.3130 C. Difficile Nucleic Acid Amplification Test Assay Class II OZN October 10, 2012 October 11, 2012

Dear Mr. Del Vecchio:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becement of the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosure) to regars actment date of the Medical Device Amendments, or to comments from to May 20, 1978, the encordance with the provisions of the Federal Food, Drug, devices that have been recuire approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. I ou may, increfore, manot the Act include requirements for annual registration, listing of I he general controls provisions and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (See acove) into the major regulations affecting your device can be may be subject to additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I DA has made a determination administered by other Federal agencies. You must or any I edelar sunates and regaraments, including, but not limited to: registration and listing (21 Comply with an the Act 3 requirements on the more reporting (reporting (reporting of medical CrK Fatt 607); labornig (21 OFR 803); good manufacturing practice requirements as set device-related adverse events) (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.

{11}------------------------------------------------

Page 2 - Mark A. Del Vecchio

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and n you desire specific arriver of In Vitro Diagnostics and Radiological Health at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

under the MDR regulation (21 Of NY are ook) poolem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Tou may obtain other general memational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Sally A. Hojvat

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{12}------------------------------------------------

4. Indications for Use Statement

K123197 510(k) Number (if known): _

Device Name: Verigene® Clostridium difficile Nucleic Acid Test (CDF) on the Verigene® System

The Verigene® Clostridium difficile Nucleic Acid Test (CDF) is a qualitative multiplexed in vitro diagnostic test for the rapid detection of toxin A (tcdA), toxin B (tcdB), and tedC gene sequences of toxigenic Clostridium difficile and for presumptive identification of PCR ribotype 027 strains from unformed (liquid or soft) stool specimens collected from patients suspected of having C. difficile infection (CDI). Presumptive identification of the PCR ribotype 027 strain of C. difficile is by detection of the binary toxin (edf) gene sequence and the single base pair deletion at nucleotide 117 in the tcdC gene. The tcdC gene encodes for a negative regulator in C. difficile toxin production. The test is performed on the Verigene System and utilizes automated specimen preparation and polymerase chain reaction (PCR) amplification, combined with a nanoparticle-based array hybridization assay to detect the toxin gene sequences associated with toxin-producing C. difficile.

The CDF Test is indicated for use as an aid in the diagnosis of CDI. Detection of PCR ribotype 027 strains of C. difficile by the CDF Test is solely for epidemiological purposes and is not intended to guide or monitor treatment for C. difficile infections. Concomitant culture is necessary only if further typing or organism recovery is required.

Prescription Use x (Part 21 CFR 801 Subpart D) and/or

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Nac cot
Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K123197

§ 866.3130 Clostridium difficile toxin gene amplification assay.

(a)
Identification. AClostridium difficile toxin gene amplification assay is a device that consists of reagents for the amplification and detection of target sequences inClostridium difficile toxin genes in fecal specimens from patients suspected of havingClostridium difficile infection (CDI). The detection of clostridial toxin genes, in conjunction with other laboratory tests, aids in the clinical laboratory diagnosis of CDI caused byClostridium difficile. (b)
Classification. Class II (special controls). The special controls are set forth in FDA's guideline document entitled: “Class II Special Controls Guideline: Toxin Gene Amplification Assays for the Detection ofClostridium difficile; Guideline for Industry and Food and Drug Administration Staff.” See § 866.1(e) for information on obtaining this document.