(61 days)
Not Found
No
The description focuses on electromagnetic and optical technology for tracking transponders and does not mention AI or ML.
No.
The device is used to assist in treatment planning and radiation therapy by aligning and monitoring the patient's position. It does not directly treat a disease or condition.
No
The Calypso System is for treatment planning and radiation therapy, and it monitors patient position during therapy. It does not diagnose diseases or conditions.
No
The device description explicitly mentions the use of "non-ionizing electromagnetic and optical technology" and "Beacon® transponders" which are physical components used for detection and localization. It also mentions "hardware and system verification and validations testing," indicating the presence of hardware.
Based on the provided information, the Calypso System is not an In Vitro Diagnostic (IVD) device.
Here's why:
- IVD Definition: In Vitro Diagnostics are devices intended for use in the collection, preparation, and examination of specimens taken from the human body (such as blood, urine, or tissue) to provide information for diagnostic, monitoring, or compatibility purposes.
- Calypso System's Function: The Calypso System's intended use and device description clearly state that it is used for patient alignment and monitoring during radiation therapy. It utilizes electromagnetic and optical technology to track the position of implanted or external markers on the patient's body.
- No Specimen Analysis: The system does not involve the collection, preparation, or analysis of any biological specimens from the patient. Its function is entirely focused on the physical positioning and movement of the patient or their internal anatomy.
Therefore, the Calypso System falls under the category of a medical device used for treatment planning and delivery, not an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
The Calypso System is intended for use as an adjunct in treatment planning and radiation therapy, to align and/or monitor the patient's position relative to the isocenter of a radiation therapy system. The Calypso System provides accurate, precise and continuous localization of a treatment isocenter by using two or more Beacon transponders.
The optional Dynamic Edge™ Gating component may signal a radiation therapy system to impose a beam-hold when the treatment target position has moved outside the defined tracking limits and to signal the radiation therapy system to remove the beam-hold upon the target's return to a position inside the defined tracking limits. The Dynamic Edge Gating feature has been shown to be compatible with Varian and Siemens radiation therapy treatment systems with external gating interfaces.
Implanted Beacon® transponders are indicated for use to radiographically and electromagnetically mark soft tissue for future therapeutic procedures.
Permanent Beacon transponders are indicated for implantation in the prostate and the peri-prostatic tissue (i.e., prostatic bed) to align and monitor the treatment isocenter in real time during radiation therapy.
Surface Beacon® transponders are indicated for temporary external placement on the skin, to monitor respiratory motion and other patient motion in real time during radiation therapy.
Product codes (comma separated list FDA assigned to the subject device)
IYE
Device Description
The Calypso System utilizes non-ionizing electromagnetic and optical technology to provide accurate, objective, and continuous localization of a treatment target for patient alignment and target position monitoring during radiation therapy. Use of the Calypso System for target localization is based on the system's detection of non-ionizing electromagnetic signals from passive markers, called Beacon® transponders. The Beacon transponders are implanted in or near the treatment target or placed externally on the surface of a patient.
Implanted Beacon transponders when used with the Calypso System enable objective measurement of the location of the target in 3 dimensions. The system operator uses this information to align the patient's treatment target to the isocenter of the therapy system prior to treatment. This information can also be used to monitor (track) the position of the treatment target during radiation therapy treatment.
When using Surface Beacon transponders this information can be used to monitor respiratory motion and other patient motion in real time during radiotherapy. When correlated to the patient's treatment target motion the system provides objective measurement of the location of the target in 3 dimensions.
The Dynamic Edge Gating feature enables the Calypso System to connect with radiation therapy systems configured with gating capabilities via an interface to external systems. With this feature added to the Calypso System, a beam hold is signaled to the radiation therapy system when the treatment target position has moved outside the defined tracking limits. The radiation therapy system is signaled to remove the beam-hold upon the target's return to a position inside the defined tracking limits.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
prostate and the peri-prostatic tissue (i.e., prostatic bed)
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Results of software, hardware and system verification and validations testing demonstrate that the Calypso System satisfies the intended use as described above.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 892.5050 Medical charged-particle radiation therapy system.
(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
0
K123137: 510(K) SUMMARY
DEC 0 5 2012
This 510(k) Summary is provided in accordance with 21 CFR 807.92.
Date of preparation: | 04 October 2012 |
---|---|
Revised Date: | 05 November 2012 |
Submitter information: | Varian Medical Systems |
2101 Fourth Avenue, Suite 500 | |
Seattle, WA 98121 | |
Phone: 206-254-0600 | |
Fax: 206-254-0606 | |
Contact: | Marcia A Page |
Director Quality Assurance and Regulatory Affairs | |
Device trade name: | Calypso® System |
Calypso® System with Dynamic Edge™ Gating | |
Common name: | Patient localization system |
Classification name: | Medical charged-particle radiation therapy system |
Classification: | CFR 892.5050 |
Class II | |
Product code: IYE | |
Predicate devices: | Calypso System (K060906, K102373, K112768) |
Device Description:
The Calypso System utilizes non-ionizing electromagnetic and optical technology to provide accurate, objective, and continuous localization of a treatment target for patient alignment and target position monitoring during radiation therapy. Use of the Calypso System for target localization is based on the system's detection of non-ionizing electromagnetic signals from passive markers, called Beacon® transponders. The Beacon transponders are implanted in or near the treatment target or placed externally on the surface of a patient.
1
Implanted Beacon transponders when used with the Calypso System enable objective measurement of the location of the target in 3 dimensions. The system operator uses this information to align the patient's treatment target to the isocenter of the therapy system prior to treatment. This information can also be used to monitor (track) the position of the treatment target during radiation therapy treatment.
When using Surface Beacon transponders this information can be used to monitor respiratory motion and other patient motion in real time during radiotherapy. When correlated to the patient's treatment target motion the system provides objective measurement of the location of the target in 3 dimensions.
The Dynamic Edge Gating feature enables the Calypso System to connect with radiation therapy systems configured with gating capabilities via an interface to external systems. With this feature added to the Calypso System, a beam hold is signaled to the radiation therapy system when the treatment target position has moved outside the defined tracking limits. The radiation therapy system is signaled to remove the beam-hold upon the target's return to a position inside the defined tracking limits.
Indications for Use:
The Calypso System is intended for use as an adjunct in treatment planning and radiation therapy, to align and/or monitor the patient's position relative to the isocenter of a radiation therapy system. The Calypso System provides accurate, precise and continuous localization of a treatment isocenter by using two or more Beacon transponders.
The optional Dynamic Edge™ Gating component may signal a radiation therapy system to impose a beam-hold when the treatment target position has moved outside the defined tracking limits and to signal the radiation therapy system to remove the beam-hold upon the target's return to a position inside the defined tracking limits. The Dynamic Edge Gating feature has been shown to be compatible with Varian and Siemens radiation therapy treatment systems with external gating interfaces.
Implanted Beacon® transponders are indicated for use to radiographically and electromagnetically mark soft tissue for future therapeutic procedures.
Permanent Beacon transponders are indicated for implantation in the prostate and the peri-prostatic tissue (i.e., prostatic bed) to align and monitor the treatment isocenter in real time during radiation therapy.
Surface Beacon® transponders are indicated for temporary external placement on the skin, to monitor respiratory motion and other patient motion in real time during radiation therapy.
2
Technological Characteristics – See device comparison table below
| Feature and/or
Specification of the
Calypso System | Cleared Calypso System
Feature/Specification | Modified Calypso
System Feature/
Specification |
|------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| Infrared optical system | Cameras and infrared targets
on array and Beacon
Transponders | No change |
| Monitoring | Real Time Monitor patient
motion | No change |
| Tracking | Real Time Tracking Patient
and Respiratory motion | No change |
| Gating | Interface to external systems | No change |
| Records | Display and Record motion | No change |
| Computer hardware and
software | Computer systems (console
and tracking station) to
control device functions and
provide for user interface | No change |
| Electromagnetic
technology (array) | Used to track motion during
therapeutic procedures | No change |
| Compatibility with the
environment and other
devices | Operates in a radiation
therapy system environment | No change |
| Electrical safety
mechanical safety | EN60601-1: 1990 / A1:1993 /
A2:1995
EN 60601-1-1: 2001 | EN60601-1 EN
60601-1: 2005 +
CORR. 1 (2006) +
CORR. 2 (2007) |
| EMC Safety | EN60601-1-2 ed. 2.0 | No Change |
Summary of Performance Testing:
Results of software, hardware and system verification and validations testing demonstrate that the Calypso System satisfies the intended use as described above.
.
3
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three wavy lines extending from its body, representing the department's commitment to health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle symbol. The logo is simple, yet recognizable, and represents the department's role in providing essential services to the American people.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
December 5. 2012
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Ms. Marcia Page Director Quality Assurance and Regulatory Affairs Varian Medical Systems 2101 Fourth Avenue South, Suite 500 SEATTLE, WA 98121
Re: K123137
Trade/Device Name: Calypso® System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: October 4, 2012 Received: October 5, 2012
Dear Ms. Page:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
4
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostics and Radiological Health at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Michael D. O'Hara
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
5
Indications for Use
510(k) Number (if known): K123137
Device Name: Calypso® System
Indications for Use:
The Calypso System is intended for use as an adjunct in treatment planning and radiation therapy, to align and/or monitor the patient's position relative to the isocenter of a radiation therapy system. The Calypso System provides accurate, precise and continuous localization of a treatment isocenter by using two or more Beacon transponders.
The optional Dynamic Edge™ Gating component may signal a radiation therapy system to impose a beam-hold when the treatment target position has moved outside the defined tracking limits and to signal the radiation therapy system to remove the beam-hold upon the target's return to a position inside the defined tracking limits. The Dynamic Edge Gating feature has been shown to be compatible with Varian and Siemens radiation therapy treatment systems with external gating interfaces.
Implanted Beacon® transponders are indicated for use to radiographically and electromagnetically mark soft tissue for future therapeutic procedures.
Permanent Beacon transponders are indicated for implantation in the prostate and the peri-prostatic tissue (i.e., prostatic bed) to align and monitor the treatment isocenter in real time during radiation therapy.
Surface Beacon® transponders are indicated for temporary external placement on the skin. to monitor respiratory motion and other patient motion in real time during radiation therapy.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR -
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Michael D. O'Hara
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health
510(k)