K Number
K123120
Manufacturer
Date Cleared
2012-12-12

(70 days)

Product Code
Regulation Number
886.4390
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LenSx Laser System is indicated for use:

  • In the creation of corneal cuts/incisions, anterior capsulotomy and laser phacofragmentation during cataract surgery. Each of these procedures may be performed either individually or consecutively during the same surgery.
  • In the creation of a lamellar cut/resection for lamellar keratoplasty, and in the creation of a penetrating cut/incision for penetrating keratoplasty.
  • In the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea.
Device Description

The LenSx Laser System uses focused femtosecond laser pulses to create incisions and separate tissue in the lens capsule, crystalline lens, and the cornea. Individual photodisruption locations are controlled by repeatedly repositioning the laser focus. The light pulse is focused into a sufficiently small spot in order to achieve photodisruption of the tissue inside the focus. A tiny volume of tissue, a few microns in diameter, is thereby photodisrupted at the laser focus. The surgical effect is produced by scanning thousands of individual pulses per second to produce a continuous incision or tissue separation. The location of the tissue photodisruption is controlled by moving the focus of the laser beam to the desired surgical target location. A computer-controlled scanning system directs the laser beam throughout a three-dimensional pattern to produce an incision. The laser pulses are delivered through a Patient Interface accessory that is placed on the surface of the cornea and fixes the eye with respect to the delivery system. This 510(k) premarket notification is being submitted to seek clearance for an additional patient interface accessory to be offered for use with cataract procedures in addition to the original patient interface accessory to the LenSx Laser System (cleared under K120732 for cataract and lamellar resection procedures). The SoftFit Patient Interface differs from the original Patient Interface (K120732) in that it has an extended suction ring skirt that enables positioning of a soft contact lens against the internal surface of the patient interface glass. The soft contact lens is similar to a standard daily wear contact lens indicated for the correction of ammetropia, with a slightly modified edge to enable fitting into the SoftFit Patient Interface. The materials and manufacturing processes used for the soft contact lens component of the proposed SoftFit Patient Interface accessory are identical to the processes used for the predicate daily wear soft contact lens (K100244). With the soft contact lens insert in place, the SoftFit Patient Interface is then mounted onto the LenSx laser system for docking onto the eye in a manner that is identical to the original LenSx Laser Patient Interface. As a result, use of the soft contact lens substantially reduces intraocular pressure (IOP) during the laser procedure, enhancing comfort and addressing potential concerns in patients with a history of glaucoma. For cataract procedures, either a LenSx Laser Patient Interface or a SoftFit Patient Interface may be used. The original LenSx Laser Patient Interface is used for keratoplasty and corneal flaps.

AI/ML Overview

The provided text describes a 510(k) premarket notification for an additional patient interface accessory (SoftFit Patient Interface) for the LenSx Laser System. The primary purpose of this submission is to demonstrate substantial equivalence to a predicate device and to show that the new accessory does not raise new questions of safety or effectiveness.

Here's an analysis of the acceptance criteria and study information based only on the provided text:

1. Table of acceptance criteria and reported device performance:

The document doesn't explicitly define formal "acceptance criteria" in a quantitative manner (e.g., "accuracy must be >X%"). Instead, it describes performance evaluations conducted to demonstrate substantial equivalence to the predicate device, particularly focusing on the differences introduced by the SoftFit Patient Interface.

Performance MetricReported Device Performance (SoftFit Patient Interface)
Holding Force and Intraocular Pressure (IOP) IncreaseEvaluated. The SoftFit Patient Interface results in a significantly lower intraocular pressure elevation during the laser procedure than the original Patient Interface.
Accuracy and Reproducibility of Cataract Treatment PatternsEvaluated for depths and geometry when using the SoftFit Patient Interface. (No specific values provided, but the context implies it meets requirements for substantial equivalence).

2. Sample size used for the test set and the data provenance:

The document does not explicitly state the sample sizes used for the performance evaluations (e.g., number of eyes, number of procedures).

The data provenance is not explicitly mentioned as retrospective or prospective, nor is the country of origin.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not provided in the document. The type of performance evaluation described (holding force, IOP, accuracy of laser patterns) suggests objective measurements rather than expert consensus on a diagnostic outcome.

4. Adjudication method for the test set:

Not applicable, as the performance tests described involve objective measurements rather than subjective assessments requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This device is a surgical laser system with a patient interface accessory, not an AI-assisted diagnostic or imaging device that would typically involve a multi-reader multi-case study to assess human reader improvement.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. The "device" in question is a surgical laser system designed for in-vivo use with human intervention, not a standalone algorithm. The performance tests assessed the physical and mechanical aspects of the new patient interface.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The "ground truth" for the performance evaluations would be:

  • Intraocular Pressure (IOP): Direct measurement using appropriate medical devices.
  • Holding Force: Direct measurement using mechanical testing methods.
  • Accuracy and Reproducibility of Cataract Treatment Patterns (depths and geometry): Measurement of the laser's physical output (e.g., incision depth, shape) using precision instruments (e.g., optical coherence tomography, microscopy, or other metrology tools).

8. The sample size for the training set:

Not applicable. The device described is a medical device (laser system accessory) and not an AI/ML model that requires a "training set."

9. How the ground truth for the training set was established:

Not applicable, as there is no training set for this type of medical device submission.

{0}------------------------------------------------

4.0 510(K) SUMMARY

b.

DEC 1 2 2012

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(a).

4.1 SUBMITTER'S NAME, ADDRESS, TELEPHONE NUMBER, CONTACT PERSON, AND DATE SUMMARY PREPARED

  • Applicant: Alcon LenSx, Inc. a. 33 Journey Aliso Viejo. CA 92656 Tel: (949) 360-6010 Fax: (949) 360-6028
    Contact Person: Judy Gordon, D.V.M. ClinReg Consulting Services, Inc. 733 Bolsana Drive Laguna Beach, CA 92651 judy@clinregconsulting.com Tel: (949) 715-0609 Fax: (949) 715-0610

4.2 NAME OF DEVICE, INCLUDING TRADE NAME AND CLASSIFICATION NAME

  • Trade/Proprietary Name: LenSx Laser System a. Common/Usual Name: b. LenSx Laser System Classification Name: Laser Instrument, Surgical, Powered C. 21 CFR 886.4390; OOE, HQC, HNO d. Classification Code(s):

PREDICATE DEVICES 4.3

510(K) #TRADE NAMEMANUFACTURER
Laser System
K120732LenSx Laser SystemAlcon LenSx, Inc.
Patient Interface Accessory
K100244Metro Soft, HydrogelSoft Contact LensMetro Optics of Austin, Inc.

{1}------------------------------------------------

DEVICE DESCRIPTION 4.4

The LenSx Laser System uses focused femtosecond laser pulses to create incisions and separate tissue in the lens capsule, crystalline lens, and the cornea. Individual photodisruption locations are controlled by repeatedly repositioning the laser focus. The light pulse is focused into a sufficiently small spot in order to achieve photodisruption of the tissue inside the focus. A tiny volume of tissue, a few microns in diameter, is thereby photodisrupted at the laser focus. The surgical effect is produced by scanning thousands of individual pulses per second to produce a continuous incision or tissue separation.

The location of the tissue photodisruption is controlled by moving the focus of the laser beam to the desired surgical target location. A computer-controlled scanning system directs the laser beam throughout a three-dimensional pattern to produce an incision. The laser pulses are delivered through a Patient Interface accessory that is placed on the surface of the cornea and fixes the eye with respect to the delivery system.

This 510(k) premarket notification is being submitted to seek clearance for an additional patient interface accessory to be offered for use with cataract procedures in addition to the original patient interface accessory to the LenSx Laser System (cleared under K120732 for cataract and lamellar resection procedures). The SoftFit Patient Interface differs from the original Patient Interface (K120732) in that it has an extended suction ring skirt that enables positioning of a soft contact lens against the internal surface of the patient interface glass. The soft contact lens is similar to a standard daily wear contact lens indicated for the correction of ammetropia, with a slightly modified edge to enable fitting into the SoftFit Patient Interface. The materials and manufacturing processes used for the soft contact lens component of the proposed SoftFit Patient Interface accessory are identical to the processes used for the predicate daily wear soft contact lens (K100244).

With the soft contact lens insert in place, the SoftFit Patient Interface is then mounted onto the LenSx laser system for docking onto the eye in a manner that is identical to the original LenSx Laser Patient Interface. As a result, use of the soft contact lens substantially reduces intraocular pressure (IOP) during the laser procedure, enhancing comfort and addressing potential concerns in patients with a history of glaucoma.

For cataract procedures, either a LenSx Laser Patient Interface or a SoftFit Patient Interface may be used. The original LenSx Laser Patient Interface is used for keratoplasty and corneal flaps.

{2}------------------------------------------------

INDICATION FOR USE 4.5

The following indications for use for the LenSx Laser System are unchanged from the previously cleared device (K120732):

  • · In the creation of corneal cuts/incisions, anterior capsulotomy and laser phacofragmentation during cataract surgery. Each of these procedures may be performed either individually or consecutively during the same surgery.
  • · In the creation of a lamellar cut/resection for lamellar keratoplasty, and in the creation of a penetrating cut/incision for penetrating keratoplasty
  • · In the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea

TECHNOLOGICAL CHARACTERISTICS COMPARISON 4.6

The LenSx Laser is unchanged and this 510(k) is intended to introduce an alternative Patient Interface accessory (the SoftFit Patient Interface) for use with cataract surgery procedures. The original Patient Interface (cleared under K120732) will continue to be available for use with cataract, keratoplasty and corneal flap procedures.

The SoftFit Patient Interface consists of a slightly modified Patient Interface and a daily wear soft contact lens insert. The SoftFit PI has an extended suction ring skirt that enables positioning of the contact lens against the internal surface of the patient interface glass. The soft contact lens is identical to standard daily wear contact lens indicated for the correction of ammetropia (K100244), with a slightly modified edge to enable fitting into the SoftFit Patient Interface. The outer curvature of the soft contact lens matches the internal curvature of the patient interface glass. The inner curvature of the soft contact lens approximates the average corneal curvature. With the soft contact lens in place, the SoftFit Patient Interface is then mounted onto the LenSx Laser system for docking onto the eye in a manner that is identical to the original Patient Interface. The SoftFit Patient Interface results in a significantly lower intraocular pressure elevation during the laser procedure than the original Patient Interface.

BRIEF SUMMARY OF PERFORMANCE TEST RESULTS 4.7

The performance data supporting substantial equivalence of the SoftFit Patient Interface accessory include:

  • Evaluation of the holding force and average intraocular pressure increase . generated by original and SoftFit Patient Interfaces.
  • Evaluation of the accuracy and reproducibility of the depths and geometry of each . of the previously cleared cataract treatment patterns for the LenSx Laser when the SoftFit Patient Interface is used.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a black and white logo. The logo is circular in shape and contains text around the perimeter. Inside the circle are three stylized, curved lines that appear to be a symbol or emblem. The text around the perimeter is not clear enough to read.

DEPARTMENT OF HEALTH & HUMAN SERVICES

December 12, 2012

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

Alcon LenSx, Inc. % Ms. Judy F. Gordon, D.V.M. Regulatory Consultant to Alcon LenSx, Inc ClinReg Consulting Services, Inc. 733 Bolsana Drive Laguna Beach, CA 92651

Re: K123120

Trade/Device Name: LenSx® Laser System Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic laser Regulatory Class: Class II Product Code: OOE, HQC, HNO Dated: November 26, 2012 Received: November 27, 2012

Dear Ms. Gordon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Judy F. Gordon, D.V.M.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21-CFR-Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Kesia Y. Ålexander

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known): 上 | 乙 | こ の

LenSx® Laser System Device Name(s):

Indications for Use: -

The LenSx Laser System is indicated for use:

  • · In the creation of corneal cuts/incisions, anterior capsulotomy and laser phacofragmentation during cataract surgery. Each of these procedures may be performed either individually or consecutively during the same surgery.
  • · In the creation of a lamellar cut/resection for lamellar keratoplasty, and in the creation of a penetrating cut/incision for penetrating keratoplasty.
  • · In the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

1 of 1 Page

Daniel P. Abdul

(Division Sign-Off) Division of Ophthalmic and Ear, Nose and Thruat Devices 5 10(k) Number

§ 886.4390 Ophthalmic laser.

(a)
Identification. An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.(b)
Classification. Class II.