(265 days)
Not Found
No
The summary describes a standard electrochemical blood glucose monitoring system and does not mention any AI or ML components or capabilities.
No.
The device is an in vitro diagnostic device intended for monitoring blood glucose levels, which aids in managing diabetes, but does not directly treat the condition.
Yes
Explanation: The "Intended Use / Indications for Use" section explicitly states that the device is for "in vitro diagnostic use." Although it specifies "should not be used for the diagnosis of or screening for diabetes," its purpose is to aid in monitoring the effectiveness of diabetes control, which falls under the broader category of diagnostic support. Furthermore, the "Device Description" also labels it as an "in vitro diagnostic device."
No
The device description explicitly states that the system consists of a meter, test strips, control solutions, a lancing device, and sterile lancets, which are all hardware components.
Yes, this device is an IVD (In Vitro Diagnostic).
The text explicitly states this multiple times:
- "The Veri-Q self-testing MGD-1001 Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use)..."
- "The Veri-Q plus MGD-1001 Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use)."
- "The blood glucose test system is an in vitro diagnostic device..."
N/A
Intended Use / Indications for Use
The Veri-Q self-testing MGD-1001 Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from fingertips, palm, or forearm. The Veri-Q self-testing MGD-1001 Blood Glucose Monitoring System is intended to be used by a single patient and should not be shared. The Veri-Q self-testing MGD-1001 Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Veri-O self-testing MGD-1001 Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady state times (when glucose is not changing rapidly).
The Veri-Q self-testing MGS-01 test strips are for use with the Veri-Q self-testing MGD-1001 meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from fingertips. palm. or forcarm.
The Veri-Q self-testing Glucose Control Solutions are for use with the Veri-Q self-testing MGD-1001 meter and the Veri-Q self-testing MGS-01 test strips to check that the meter and test strips are working together properly and providing accurate results.
The Veri-Q plus MGD-1001 Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose (sugar) in fresh capillary whole blood drawn from fingertips, palm, or forearm. The Veri-Q plus MGD-1001 Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use). The Veri-Q plus MGD-1001 Blood Glucose Monitoring System is intended for multi-patient use in a professional healthcare setting, as an aid to monitor the effectiveness of diabetes control. This system is only used with single-use, auto-disabling lancing devices. The Veri-Q plus MGD-1001 Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).
The Veri-Q plus MGS-01 test strips are for use with the Veri-Q plus MGD-1001 meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from fingertips, palm, or forearm. The Veri-Q plus Glucose Control Solutions are for use with the Veri-Q plus MGD-1001 meter and the Veri-Q plus MGS-01 test strips to check that the meter and test strips are working together properly and providing accurate results.
Product codes (comma separated list FDA assigned to the subject device)
NBW, CGA, JJX
Device Description
The proposed Veri-Q self-testing / plus MGD-1001 Blood Glucose Monitoring System consists of a meter, test strips, control solutions (3 levels), a lancing device and sterile lancets. The blood glucose test system is an in vitro diagnostic device designed for measuring the The orded gracess in whole blood sample by means of an electrical current produced in the test strip and sent to the meter for measurement. The Veri-Q self-testing / plus MGD-1001 meter is a no-code meter, so the responsibility of coding has been removed from the user. Each individual test strip contains technology that allows it to automatically calibrate and code the meter once it is inserted.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
fingertips, palm, or forearm
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Home use for self-testing by people with diabetes; Multi-patient use in a professional healthcare setting.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
The testing for the Veri-Q self-testing / plus MGD-1001 Blood Glucose Monitoring System was performed in accordance with ISO 15197:2003. Clinical evaluation included method was performous in arrormance and alternative-site blood glucose measurement. Non-clinical compulsion, asor perform were conducted to establish the performance, functionality and reliability characteristics of the Veri-Q self-testing / plus MGD-1001 Blood Glucose Monitoring System. The device passed all of the tests based on pre-determined Pass/Fail criteria.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 862.1345 Glucose test system.
(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
0
510(k) Summary
JUN 1 9 2013
This summary of 510(k)-safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
05/27/2013 Date: _
1. Submitter:
MiCoBioMed Co., Ltd. 55-13, Singeonji-Dong, Anseong-Si, Gyenggi-Do, 456-781, Republic of Korea Tel : +82-31-8056-5940 Fax : +82-31-8056-5518
Submission Correspondent: 2.
Priscilla Chung LK Consulting Group 1515 E Katella Ave Unit 2115, Anaheim, CA 92805 Phone: 714-202-5789 Fax: 714-409-3357 Email: info@lkconsultinggroup.com
- Device:
.
| Proprietary Name: | Veri-Q self-testing MGD-1001 Blood Glucose Monitoring System
Veri-Q plus MGD-1001 Blood Glucose Monitoring System |
|------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Common Name: | Blood Glucose Test System |
| Classification Name: | System, Test, Blood Glucose, Over The Counter
Glucose Oxidase
Single(specified) analyte controls |
| Classification: | Class II, 21 CFR 862.1345 |
| Classification Product Code: | NBW |
| Subsequent Product Codes | CGA, JJX |
4. Predicate Device:
CareSens N(K083468) by i-SENS, Inc.
1
પં Description:
The proposed Veri-Q self-testing / plus MGD-1001 Blood Glucose Monitoring System consists of a meter, test strips, control solutions (3 levels), a lancing device and sterile lancets. The blood glucose test system is an in vitro diagnostic device designed for measuring the The orded gracess in whole blood sample by means of an electrical current produced in the test strip and sent to the meter for measurement. The Veri-Q self-testing / plus MGD-1001 meter is a no-code meter, so the responsibility of coding has been removed from the user. Each individual test strip contains technology that allows it to automatically calibrate and code the meter once it is inserted.
6. Indications for use:
6.1 Veri-Q self-testing MGD-1001 Blood Glucose Monitoring System
The Veri-Q self-testing MGD-1001 Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from fingertips, palm, or forearm. The Veri-Q self-testing MGD-1001 Blood Glucose Monitoring System is intended to be used by a single patient and should not be shared. The Veri-Q self-testing MGD-1001 Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Veri-Q self-testing MGD-1001 Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).
The Veri-Q self-testing MGS-01 test strips are for use with the Veri-Q self-testing MGD-1001 meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from fingertips, palm, or forearm.
The Veri-O self-testing Glucose Control Solutions are for use with the Veri-Q self-testing MGD-1001 meter and the Veri-Q self-testing MGS-01 test strips to check that the meter and test strips are working together properly and providing accurate results.
6.2 Veri-Q plus MGD-1001 Blood Glucose Monitoring System
The Veri-Q plus MGD-1001 Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose (sugar) in fresh capillary whole blood drawn from fingertips, palm, or forearm. The Veri-Q plus MGD-1001 Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use). The Veri-Q plus MGD-1001 Blood Glucose Monitoring System is intended for multi-patient use in a professional healthcare setting, as an aid to monitor the effectiveness of diabetes control. This system is only used with single-use, auto-disabling lancing devices. The Veri-Q plus MGD-1001 Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).
The Veri-Q plus MGS-01 test strips are for use with the Veri-Q plus MGD-1001 meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from fingertips, palm, or forearm.
510k summary Page 2 of 3
2
The Veri-Q plus Glucose Control Solutions are for use with the Veri-Q plus MGD-1001 meter and the Veri-Q plus MGS-01 test strips to check that the meter and test strips are working together properly and providing accurate results.
7. Technological Characteristics:
The Veri-Q self-testing / plus MGD-1001 Blood Glucose Monitoring System has the same fundamental scientific technology as the predicate device. The indications for use, technology, design of glucose meter, and strips are almost the same. The major difference is that for the Veri-Q meter, non coding is required on user side but the predicate device employs autocoding method. There are also minor difference such as shelf life and memory capacity.
8. Performance Data:
The testing for the Veri-Q self-testing / plus MGD-1001 Blood Glucose Monitoring System was performed in accordance with ISO 15197:2003. Clinical evaluation included method was performous in arrormance and alternative-site blood glucose measurement. Non-clinical compulsion, asor perform were conducted to establish the performance, functionality and reliability characteristics of the Veri-Q self-testing / plus MGD-1001 Blood Glucose Monitoring System. The device passed all of the tests based on pre-determined Pass/Fail criteria.
Conclusions: 9.
MiCoBioMed Co., Ltd. concludes that the Veri-Q self-testing / plus MGD-1001 Blood Glucose Monitoring System is safe and effective also, substantially equivalent to the predicate device.
510k summary Page 3 of 3
3
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three wavy lines, resembling a human figure or a symbol representing health and well-being. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
· June 19, 2013
MiCoBioMed Co., Ltd. C/O Priscilla Chung C/O LK Consulting Group 1515 E. Katella Ave, Unit 2115 ANAHEIM CA 92805
Re: K123007
Trade/Device Name: Veri-Q self-testing MGD-1001 Blood Glucose Monitoring System Veri-Q plus MGD-1001 Blood Glucose Monitoring System
Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW, CGA, JJX Dated: May 28, 2013 Received: May 30, 2013
Dear Ms. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
4
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Carol C. Benson -S for
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
5
Indications for Use
510(k) Number K 123007
Device Name: Veri-Q self-testing MGD-1001 Blood Glucose Monitoring System
Indication for use:
The Veri-Q self-testing MGD-1001 Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from fingertips, palm, or forearm. The Veri-Q self-testing MGD-1001 Blood Glucose Monitoring System is intended to be used by a single patient and should not be shared. The Veri-Q self-testing MGD-1001 Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Veri-O self-testing MGD-1001 Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady state times (when glucose is not changing rapidly).
The Veri-Q self-testing MGS-01 test strips are for use with the Veri-Q self-testing MGD-1001 meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from fingertips. palm. or forcarm.
The Veri-Q self-testing Glucose Control Solutions are for use with the Veri-Q self-testing MGD-1001 meter and the Veri-Q self-testing MGS-01 test strips to check that the meter and test strips are working together properly and providing accurate results.
| Prescription Use
(Part 21CFR801 Subpart D) | AND/OR Over-The-Counter Use
(Part 21CFR807 Subpart C) |
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- | ------------------------------------------------------------------------ |
---|
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro and Radiological Health (OIR)
Katherine Serrano -S
Division Sign-Off Office of In Vitro and Radiological Health Evaluation and Safety
510(k) K123007 Page I of 2
6
Indications for Use
510(k) Number K 123007
Device Name: Veri-Q plus MGD-1001 Blood Glucose Monitoring System
Indication for use:
The Veri-Q plus MGD-1001 Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose (sugar) in fresh capillary whole blood drawn from fingertips, palm, or forearm. The Veri-Q plus MGD-1001 Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use). The Veri-Q plus MGD-1001 Blood Glucose Monitoring System is intended for multi-patient use in a professional healthcare setting, as an aid to monitor the effectiveness of diabetes control. This system is only used with single-use, auto-disabling lancing devices. The Veri-Q plus MGD-1001 Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).
The Veri-Q plus MGS-01 test strips are for use with the Veri-Q plus MGD-1001 meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from fingertips, palm, or forearm. The Veri-Q plus Glucose Control Solutions are for use with the Veri-Q plus MGD-1001 meter and the Veri-Q plus MGS-01 test strips to check that the meter and test strips are working together properly and providing accurate results.
Prescription Use (Part 21CFR801 Subpart D) Over-The-Counter (Part 21CFR807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE OF NEEDED)
AND/OR
Concurrence of CDRH, Office of In Vitro and Radiological Health (OIR)
Katherine Serrano -S
Division Sign-Off Office of In Vitro and Radiological Health Evaluation and Safety
510(k) K123007 Page 2 of 2