K Number
K122928

Validate with FDA (Live)

Manufacturer
Date Cleared
2013-01-30

(128 days)

Product Code
Regulation Number
892.1680
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

1417PGA Digital Flat Panel X-Ray Detector is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. Not to be used for mammography.

Device Description

The 1417PGA is a portable digital X-ray flat panel detector that can generate images of any part of the body. This X-ray imaging system consists of a scintillator directly coupled to an a-Si TFT sensor. It makes high-resolution, high-sensitive digital images. 1417PGA is designed specifically to be integrated with an operating PC and a X-ray generator to digitalize X-ray images into RAW files. The RAW files can be made to DICOM compatible image files for a radiographic diagnosis and analysis by console SW.

AI/ML Overview

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Device: 1417PGA Digital Flat Panel X-ray Detector
Predicate Device: SDX-4336CP by Samsung Mobile Display Co., Ltd. (K102321)

1. Table of Acceptance Criteria and Reported Device Performance

The submission focuses on demonstrating substantial equivalence to the predicate device, rather than defining explicit "acceptance criteria" against a fixed benchmark. The performance is compared directly against the predicate device. The primary performance metrics mentioned are Modulated Transfer Function (MTF) and Detective Quantum Efficiency (DQE).

MetricAcceptance Criteria (Compared to Predicate Device SDX-4336CP)Reported Device Performance (1417PGA)
MTF (Modulated Transfer Function)Equivalent or better resolution performance and sharpnessPerformed better than SDX-4336CP at 0-2.2 lp/mm due to smaller pixel size (127 µm vs 143 µm) and new bonding mechanism.
DQE (Detective Quantum Efficiency)Equivalent or better ability to visualize object details of a certain size and contrast; higher Signal-to-Noise Ratio (SNR) transferDemonstrated better DQE performance than SDX-4336CP at various spatial frequencies. Zero-frequency DQE values: 1417PGA (0.761) > SDX-4336CP (0.740).
NPS (Noise Power Spectrum)(Implied better image quality through reduced noise)Exhibited NPS which has lower performance than SDX-4336CP, indicating reduced noise and improved image accuracy.
Diagnostic Image Quality (Clinical)Equivalent or "better diagnostic image quality" in expert reviewClinical images reviewed by a licensed US radiologist found equivalent or better diagnostic image quality compared to the predicate device.

2. Sample Size Used for the Test Set and Data Provenance

  • Non-clinical (MTF, DQE, NPS) Test Set: The sample size is not explicitly stated. The report mentions "the MTF, DQE and NPS test results of 1417PGA and SDX-4336CP by using the identical test equipment and same analysis method described by IEC 62220-1." This suggests controlled laboratory measurements, not a patient-based dataset.
  • Clinical Test Set: The sample size for clinical images is not explicitly stated. It mentions that "clinical images are taken from both devices."
  • Data Provenance: The submission is from Rayence Co., Ltd. in Korea. The "non-clinical test report and clinical consideration report were prepared and submitted to FDA separately." The clinical images for the expert review were compared "according to age group and anatomical structures." The country of origin for the clinical data is not specified, but the expert reviewer is a "licensed US radiologist."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Non-clinical: Ground truth was established through physical measurements and standard analysis methods (IEC 62220-1). No human experts were involved in establishing ground truth for these objective metrics.
  • Clinical: "a licensed US radiologist" was used for the expert opinion on clinical images. The specific qualifications beyond "licensed US radiologist" (e.g., years of experience, subspecialty) are not provided. The number of experts was one.

4. Adjudication Method for the Test Set

  • Non-clinical: Not applicable, as ground truth was established by objective measurements based on standards like IEC 62220-1.
  • Clinical: Not applicable. The expert review was a single reader's opinion. There is no mention of multiple readers or an adjudication process. The phrasing "a licensed US radiologist to render an expert opinion" suggests a single-reviewer assessment.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not explicitly stated or described. The clinical assessment involved a single radiologist reviewing images from both devices. There is no mention of measuring how much human readers improve with AI vs. without AI assistance, as this device itself is a digital X-ray detector, not an AI-powered diagnostic tool. The comparison is between two different hardware detectors.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

Yes, the non-clinical performance testing (MTF, DQE, NPS) represents standalone performance of the detector hardware without human interpretation. These are objective measurements of the device's image quality characteristics. The device itself is a digital X-ray panel, not an algorithm that outputs a diagnostic decision; its "standalone" performance is its ability to acquire high-quality images.

7. The Type of Ground Truth Used

  • Non-clinical (MTF, DQE, NPS): The ground truth was established through physical measurements and standardized testing procedures (IEC 62220-1).
  • Clinical: The "ground truth" for the clinical comparison was essentially the expert opinion of a single licensed US radiologist comparing the diagnostic image quality of images produced by both the proposed and predicate devices. This is a form of expert consensus, albeit from a single expert.

8. The Sample Size for the Training Set

Not applicable. This device is a digital X-ray detector (hardware), not an AI algorithm that requires a training set. The submission is for a medical imaging device, not a machine learning model.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this hardware device.

{0}------------------------------------------------

510(k) Submission – 1417PGA

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date 510K summary prepared: September 21, 2012

Submitter's Name, address, telephone number, a contact person:

Submitter's Name :Rayence Co., Ltd.
Submitter's Address:(Seogu-dong, 2F/4F) 14, Samsung 1ro 1-gil,
Hwasung-si, Gyeonggi-do, 445-170, Korea
Submitter's Telephone:+82-31-8015-6459
Contact person:Mr. Kee Dock Kim / Manager
Official Correspondent:Dave Kim (davekim@mtech-inc.net)
(U.S. Designated agent)
Address:12946 Kimberley Ln, Houston, TX 77079
Telephone:+713-467-2607
Fax:+713-464-8880

Name of the device, including the trade or proprietary name if applicable, the common or usual name and the classification name, if known:

Trade/proprietary name:1417PGA
Common Name:Digital Flat Panel X-ray Detector
Classification Name :Solid State X-ray Imaging Device, Class2
Regulation Number / Name:21CFR 892.1680 / Stationary X-ray system

Rayence Co., Ltd.

JAN 3 0 2013

{1}------------------------------------------------

510(k) Submission - 1417PGA

Product Code:MQB
Predicate Device :
Manufacturer: Samsung Mobile Display Co., Ltd.
Device: SDX-4336CP
510(k) Number: K102321 (Decision Date - Feb. 11. 2011)

Device Description :

The 1417PGA is a portable digital X-ray flat panel detector that can generate images of any part of the body. This X-ray imaging system consists of a scintillator directly coupled to an a-Si TFT sensor. It makes high-resolution, high-sensitive digital images. 1417PGA is designed specifically to be integrated with an operating PC and a X-ray generator to digitalize X-ray images into RAW files. The RAW files can be made to DICOM compatible image files for a radiographic diagnosis and analysis by console SW.

Indication for use :

1417PGA Digital Flat Panel X-Ray Detector is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. Not to be used for mammography.

Rayence Co., Ltd.

.
.
.
.
.
.

{2}------------------------------------------------

Summary of the technological characteristics of the device compared to the predicate device:

1417PGA SSXI detector described in this 510(k) has the same indications for use and similar technical characteristics as its predicate device, SDX-4336CP, of Samsung Mobile Display Co., Ltd. Table 1 summarizes the technological characteristics of the 1417PGA and SDX-4336CP the predicate device.

CharacteristicProposedRayence Co.,Ltd.1417PGAProposedSamsung Mobile DisplayCo., Ltd.SDX-4336CP
510(k) number-K102321
Intended Use1417PGA Digital Flat PanelX-Ray Detector is indicatedfor digital imaging solutiondesigned for generalradiographic system forhuman anatomy. It isintended to replace film orscreen based radiographicsystems in all generalpurpose diagnosticprocedures. Not to be usedfor mammography.SDX-4336CP Digital Flat PanelX-Ray Detector is indicated fordigital imaging solution designedfor providing generalradiographic diagnosis of humananatomy targeting both adult andchildren. It is intended to be usedby physicians and radiologists.and not to be used formammography.
Detector TypeAmorphous Silicon, TFTAmorphous Silicon, TFT
ScintillatorGadolinium OxysulfideCesium Iodide
Imaging Area14 x 17 inches14 x 17 inches
Pixel matrix3328 x 2816 (9.4 million)2880 x 2400 (6.9 million)
Pixel pitch127 $\mu$ m150 $\mu$ m
Resolution3.9lp/mm3.3lp/mm

Table 1: Comparison of 1417PGA and SDX-4336CP ·

{3}------------------------------------------------

510(k) Submission - 1417PGA

A/D conversion14 bit14 bit
Grayscale16384 (14bit)16384 (14bit)
Preview Image2~3 seconds2~3 seconds
Data outputRAW*The RAW files areconvertible into DICOM 3.0by console S/WRAW*The RAW files areconvertible into DICOM 3.0by console S/W
Dimensions460 x 417 x 15.9 mm491 × 480 × 15 mm
Weight3.6 kg3.8 kg
ApplicationPortable systemAvailable with upright stand,table, universal standPortable systemAvailable with upright stand,table, universal stand
FeatureImage: black box with handleImage: gray surface with handle

Table 2: Comparison of 1417PGA and SDX-4336CP

ItemUnit1417PGASDX-4336CF
Pixel sizeμm127 x 127150 x 150
Total horizontal andvertical sizemm422.7 x 357.6432 x 360
Total horizontal andvertical element countpixels3328 x 28162880 x 2400
Active area horizontal andvertical sizemm416.6 x 351.8423 x 351
Active area horizontal andvertical element countpixels3280 x 27702820 x 2340
Pixel spacingμm127150
Fill factor%65.1468.5
WeightKg3.6 Kg3.8 Kg

{4}------------------------------------------------

Summary of Performance Testing:

Indications for use, material, form factor, performance, and safety characteristics between 1417PGA and the predicate device are very similar. The primary difference is Pixel size, Pixel matrix, Pixel pitch, Resolution and Scintillator materials; GdOS(Gd2O2S:Tb) for 1417PGA and Cesium iodide(Csl) for SDX-4336CP, respectively. The non-clinical test report and clinical consideration report were prepared and submitted to FDA separately to demonstrate the substantial equivalency between two different detectors. The non-clinical test report contains the MTF, DQE and NPS test results of 1417PGA and SDX-4336CP by using the identical test equipment and same analysis method described by IEC 62220-1. The comparisoin of the MTF for 1417PGA and SDX-4336CP detector demonstated that the MTF of the 1417PGA detector performed better than SDX-4336CP at 0~2.2 lp/mm. Also, the new detector 1417PGA utilizes a new bonding mechanism to narrow the gap between the panel and scintillator. Moreover, the pixel size of the new detector

1417PGA is 127 um which is smaller than 143 um of SDX-4336CP. Thereofre, the overall

resolution performance and sharpness of 1417PGA is better than SDX-4336CP which results improvement of the ability of the new detector to represent distinct anatomic features whitin the imaged object. The DOE represents the ability to visualize object details of a certain size and contrast. 1417PGA demonstrated better DQE performance than SDX-4336CP at various spatial frequencies and provides a higher Signal-toNoise Ratio (SNR) transfer from the input to the output of a detector as a function of frequency. At the zero-frequency DQE values for 1417PGA is higher than SDX-4336CP; 0.761 and 0.740 respectively. The reduced noise has imporved the accuracy of image and reduced the degree of artifacts for the new detector. 1417PGA exhibited NPS which has lower performance than SDX-4336CP. Therefore, the image quality of 1417PGA is greater than SDX-4336CP at the same patient exposure.

To further demonstrate the substantial equivalency of two devices, clinical images are taken from both devcies and reviewd by a licensed US radiologist to render an expert opinion. Both test (1417PGA) and control group (SDX-4336CP) are evaluated according to age group and anatomical structures were compared in accordance with the test protocol of diagnostic radiography evaluation procedure.

Based on the non-clinical and clinical consideration and the outcome of an expert review of image comparisions for both devices, we can claim equivalent or better diagnostic image quality for1417PGA compared to the predicate device, SDX-4336CP.

{5}------------------------------------------------

Safety, EMC and Performance Data :

Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1:1998+A1:1999+A2:1995(Medical electrical equipment Part 1: General Requirements for Safety) was performed, and EMC testing were conducted in accordance with standard (Medical electrical equipment - Part 1-2: General Requirements for safety - Collateral Standard : Electromagnetic Compatibility Requirements and tests).

Non-clinical & Clinical considerations according to FDA Guidance for the Submissions of 510(k)'s for Solid State X-ray Imaging Devices" was performed. All test results were satisfactory.

Conclusions :

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Rayence Co., Ltd. concludes that 1417PGA is safe and effective and substantially equivalent in comparison with the predicate device as described herein.

{6}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized representation of human figures.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 30, 2013

Ravence Co., LTD. C/Q Dave Kim Mtech Group 12946 Kimberley Lane HOUSTON TX 77079

Re: K122928

Trade/Device Name: Digital Flat Panel X-ray Detector 1417PGA Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: January 16, 2013 Received: January 26, 2013

Dear Mr. Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{7}------------------------------------------------

Page 2-Mr. Kim

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Sean M. Boyd -S for

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Indications for Use

510(k) Number (if known): K122928

Device Name: Digital Flat Panel X-Ray Detector /1417PGA

Indications for Use:

1417PGA Digital Flat Panel X-Ray Detector is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. Not to be used for mammography.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Sean M. Boyd

(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health

510(k)

Page 1 of 1

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.