K Number
K122509
Manufacturer
Date Cleared
2012-12-17

(122 days)

Product Code
Regulation Number
888.3050
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VertiFlex® Spinous Process Fixation Plate is a posterior, non-pedicle supplemental fixation device intended for use at a single level in the non-cervical spine (T1 - S1). It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fusion in the following conditions: Degenerative disc disease (defined as back pain of discogenic origin with degenerationof the disc confirmed by history and radiographic studies); spondylolisthesis; Irauma (i.e., fracture or dislocation); and/or tumor. The VertiFlex® Spinous Process Fixation Plate is intended for use with bone graft, and is not intended for stand-alone use.

Device Description

The VertiFlex® Spinous Process Fixation (SPF) Plate is a one-piece bilateral locking plate device which attaches to the posterior non-cervical spine by securely grasping two adjacent spinous processes. The SPF Plate is available in multiple sizes to accommodate different anatomical requirements, and it is composed entirely of titanium 6AL-4V allov. The SPF Plate may be implanted by either conventional surgical methods, or via minimally-invasive techniques. Proprietary manual instrumentation for implantation of the SPF Plate is available for both conventional and minimally-invasive surgical procedures.

AI/ML Overview

The provided text describes the VertiFlex® Spinous Process Fixation Plate, a medical device for spinal fixation. It details the device's indications for use, comparison to predicate devices, and a summary of non-clinical tests. However, it explicitly states, "No clinical testing was conducted to support this submission."

Therefore, the information requested regarding acceptance criteria and a study proving the device meets those criteria, specifically concerning performance metrics from clinical trials, a test set, ground truth, expert involvement, or AI-related data, is not available in the provided text.

Based on the document, here's what can be extracted and what cannot:

1. A table of acceptance criteria and the reported device performance

  • Acceptance Criteria: Not explicitly stated as pass/fail criteria from a specific clinical study for device performance. The "acceptance criteria" for regulatory submission seems to be substantial equivalence to predicate devices, based on design, materials, indications, and non-clinical performance.
  • Reported Device Performance:
    • Non-Clinical Tests: The device demonstrated "substantial equivalence... in terms of performance" to a commercially available predicate through tests conducted in accordance with ASTM F1717-12 (Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model), static pull-off testing in synthetic bone, and static dissociation testing. Specific performance values (e.g., load limits, displacement) are not provided.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not applicable. No clinical test set data is provided. The non-clinical tests likely used test samples of the device and synthetic bone models.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable. No clinical test set and thus no ground truth established by medical experts for device performance are mentioned.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable. No clinical test set is mentioned.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No. This device is a mechanical implant, not an AI-powered diagnostic tool. MRMC studies are not relevant to its regulatory pathway as described.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • No. This device is a mechanical implant. The concept of "standalone algorithm performance" is not applicable.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Not applicable for clinical ground truth. For non-clinical testing, the "ground truth" would be the engineering specifications and established performance characteristics of the predicate devices against which the VertiFlex® plate was compared for substantial equivalence.

8. The sample size for the training set

  • Not applicable. No machine learning or AI training set is mentioned as this is a mechanical device.

9. How the ground truth for the training set was established

  • Not applicable. No machine learning or AI training set is mentioned.

{0}------------------------------------------------

August 16, 2012

5.0 510(k) SUMMARY

In accordance with Title 21 Code of Federal Regulations (21 CFR), Part 807, and in particular, §807.92, the following 510(k) summary is provided for the VertiFlex® Spinous Process Fixation Plate:

Submitted By: 5.1

VertiFlex®, Incorporated 1351 Calle Avanzado San Clemente, California 92673

Contact: Steve Reitzler, Vice President, Clinical & Regulatory Affairs

Date Prepared: August 16, 2012

5.2 Device Name

Trade or Proprietary Name:VertiFlex® Spinous Process Fixation Plate
Common or Usual Name:Spinous Process Fixation Plate
Classification Name:Spinal Interlaminal Fixation Orthosis

5.3 Predicate Devices

The subject device is substantially equivalent, in whole or in part, to the following commercially available predicate devices:

AffixTM Spinous Process Plate System (K073278; NuVasive, Inc.) Aspen™ Spinal Fixation System (K071877, K100935; Lanx, LLC) Axle™ Interspinous Fusion System (K101471; X-spine Systems, Inc.) PrimaLOK™ Interspinous Fusion System (K100354; OsteoMed Spine, Inc.) SP-Fix™ Spinous Process Fixation Plate (K102195; Globus Medical, Inc.) Spire™ Spinous Process Plate (K032037, K043053, K102866; Medtronic, Inc.)

Device Description 5.4

The VertiFlex® Spinous Process Fixation (SPF) Plate is a one-piece bilateral locking plate device which attaches to the posterior non-cervical spine by securely grasping two adjacent spinous processes. The SPF Plate is available in multiple sizes to accommodate different anatomical requirements, and it is composed entirely of titanium 6AL-4V allov. The SPF Plate may be implanted by either conventional surgical methods, or via minimally-invasive techniques. Proprietary manual instrumentation for implantation of the SPF Plate is available for both conventional and minimally-invasive surgical procedures.

{1}------------------------------------------------

ર.5 Intended Use

The subject device is indicated for use as follows:

The VertiFlex® Spinous Process Fixation Plate is a posterior, non-pedicle supplemental fixation device intended for use at a single level in the non-cervical spine (TI - SI). It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fusion in the following conditions: Degenerative disc disease (defined as back pain of discogenic origin with degenerationof the disc confirmed by history and radiographic studies); spondylolisthesis; Irauma (i.e., fracture or dislocation); and/or tumor. The VertiFlex® Spinous Process Fixation Plate is intended for use with bone graft, and is not intended for stand-alone use.

5.6 Comparison to Predicate Devices

Testing and comparisons of design characteristics and features have established that the subject SPF Plate is substantially equivalent in design, materials of composition, indications, performance, and other features, to other predicate spinous process fixation devices commercially available in the U.S.

5.7 Summary of Non-Clinical Tests

Non-clinical tests conducted in accordance with such recognized standards as ASTM F1717-12, Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model, static pull-off testing in synthetic bone, and static dissociation testing, have demonstrated the substantial equivalence of the subject SPF Plate to a commerciallyavailable predicate in terms of performance.

. 5.8 Summary of Clinical Tests

No clinical testing was conducted to support this submission.

રું તે તે જે જે જે જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં મુખ્યત્વે ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં મ Conclusions

The results of all testing and comparison demonstrated the substantial equivalence of the subject SPF Plate to the identified predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the eagle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-00002

December 17, 2012

VertiFlex®. Incorporated % Mr. Steve Reitzler · Vice President, Clinical & Regulatory Affairs 1351 Calle Avanzado San Clemente, California 92673

Re: K122509

Trade/Device Name: VertiFlex® Spinous Process Fixation Plate Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: KWP Dated: December 11, 2012 Received: December 12, 2012

Dear Mr. Reitzler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical

{3}------------------------------------------------

Page 2 - Mr. Steve Reitzler

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Image /page/3/Picture/6 description: The image shows the name "Erin FD Keith" in a stylized font. The first name, "Erin," is written in a bold, sans-serif font. The middle initials, "FD," are designed with a layered, geometric pattern. The last name, "Keith," is written in a similar style to the middle initials, with a layered effect.

Mark N. Meikerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

August 16, 2012

Indications for Use

510(k) Number (if known): _ K122509

VertiFlex® Spinous Process Fixation Plate Device Name:

Indications for Use:

The VertiFlex® Spinous Process Fixation Plate is a posterior, non-pedicle supplemental fixation device intended for use at a single level in the non-cervical spine (T1 - S1). It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fusion in the following conditions: Degenerative disc disease (defined as back pain of discogenic origin with degenerationof the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); and/or tumor. The VertiFlex® Spinous Process Fixation Plate is intended for use with bone graft, and is not intended for stand-alone use.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Ronald P. Jean -S

(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K122509

| 4

§ 888.3050 Spinal interlaminal fixation orthosis.

(a)
Identification. A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.(b)
Classification. Class II.