K Number
K122501
Date Cleared
2013-03-29

(225 days)

Product Code
Regulation Number
872.4850
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AEI Dr. Terauchi Ultrasonic Tips are intended for use by dental professionals for endodontic root preparation procedures. The Ultrasonic Tips can remove soft and hard tissue from a narrow dentinal structure in order to make space for permanent filling material in the canal. They can also assist in the removal of separated instruments or other intra-canal blockages.

Device Description

The AEI Ultrasonic Tips are an accessory to a Piezo ultrasonic hand piece and unit. These external power supplies are not part of the device submitted for application with the 510(k) submission. The Ultrasonic Tips with a concave spoon shape or straight tip are very small and will not cut significant amounts of dentin during use. The AEI Ultrasonic Tips will be available in M3x0.6 thread with 5 different types of tip design to facilitate reaching specific angles and directions in the root of the tooth.

AI/ML Overview

Here's an analysis of the provided text regarding the AEI Ultrasonic Tips, focusing on the acceptance criteria and the study that proves the device meets those criteria:

Acceptance Criteria and Device Performance

Acceptance CriteriaReported Device Performance
Safety and Effectiveness for Endodontic Root Preparation Procedures: Ability to remove soft and hard tissue from a narrow dentinal structure to create space for permanent filling material, and assist in the removal of separated instruments or other intra-canal blockages.Clinical evaluations conducted outside the US showed that the devices were safe and effective for the proposed indications.
Substantial Equivalence to Predicate Device (DENTSPLY ProUltra® Endo Tips): Based on comparison of indications, intended use, materials, technological characteristics, operating principle, similar design features, and same connection capabilities to piezoelectric ultrasonic generator.The device was determined to be substantially equivalent to the predicate device based on: - Same technological characteristics - Same operating principle - Similar design features - Same connection capabilities to piezoelectric ultrasonic generator - Same material

Study Information:

  1. Sample size used for the test set and the data provenance:

    • The document mentions "Clinical evaluations were conducted outside the US." It does not specify the sample size of these clinical evaluations (test set) nor specific countries of origin. It indicates these were clinical evaluations, implying prospective data collection.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not provided in the document.
  3. Adjudication method for the test set:

    • This information is not provided in the document.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is a medical device (ultrasonic tips), not an AI algorithm. Therefore, an MRMC study related to AI assistance is not applicable and was not conducted.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No. This is a medical device (ultrasonic tips), not an AI algorithm. Therefore, a standalone algorithm performance study is not applicable and was not conducted.
  6. The type of ground truth used:

    • For the safety and effectiveness aspect, the "clinical evaluations" performed likely used clinical outcomes and observations by dental professionals as the ground truth.
    • For substantial equivalence, the "ground truth" was established by direct comparison of the device's technical specifications and intended use against those of the predicate device.
  7. The sample size for the training set:

    • This is not applicable as this isn't an AI/machine learning device. The "training" for such a device would be its design and manufacturing process to meet specifications.
  8. How the ground truth for the training set was established:

    • This is not applicable as this isn't an AI/machine learning device. The design and manufacturing process would be guided by engineering principles and standards relevant to dental instruments.

Summary of the Study:

The document primarily relies on the concept of substantial equivalence to an already cleared predicate device (DENTSPLY ProUltra® Endo Tips, K200880). The "study" proving the device meets the acceptance criteria is detailed as:

  • Non-Clinical Testing: This was not conducted as a separate mechanical test, as substantial equivalence was based on a direct comparison of indications, intended use, and materials. The report explicitly states: "Mechanical testing was, therefore, not conducted."
  • Clinical Testing: "Clinical evaluations were conducted outside the US and showed that the devices were safe and effective for the proposed indications." No details are provided regarding the methodology, sample size, or specific outcomes of these evaluations.

Essentially, the bulk of the "proof" for regulatory approval hinges on the numerous similarities between the AEI Ultrasonic Tips and the predicate device, implying that since the predicate device is safe and effective, and the new device is substantially similar, it too should be considered safe and effective. The clinical evaluations served as a corroborating factor for safety and effectiveness.

{0}------------------------------------------------

1122501

510(k) Summary

MAR 2 9 2013

DATE: August 10, 2012

510(k) SPONSOR: American Eagle Instruments, Inc 6575 Butler Creek Road

· Missoula, MT 59808

CONTACT PERSON:

Kristen Nsytuen Compliance Manager/Engineer (406) 549-7451 knystuen@am-eagle.com

TRADE NAME:

AEI (Dr. Terauchi) Ultrasonic Tips

COMMON NAMES:

CLASSIFICATION, and CLASS:

21 CFR 872.4850, Class II;

Ultrasonic Tips

PRODUCT CODES: ELC

PREDICATE DEVICE:

Company Device DENTSPLY ProUltra® Endo Tips (a.k.a Sapphire Plus® Tips)

510(k) No. K200880

Date Cleared May 10, 1996

DEVICE DESCRIPTION: The AEI Ultrasonic Tips are an accessory to a Piezo ultrasonic hand piece and unit. These external power supplies are not part of the device submitted for application with the 510(k) submission. The Ultrasonic Tips with a concave spoon shape or straight tip are very small and will not cut significant amounts of dentin during use. The AEI Ultrasonic Tips will be available in M3x0.6 thread with 5 different types of tip design to facilitate reaching specific angles and directions in the root of the tooth.

INDICATIONS FOR USE: The AEI Ultrasonic Tips are intended for use by dental professionals for endodontic root preparation procedures. The Ultrasonic Tips can remove soft and hard tissue from a narrow dentinal structure in order to make space for permanent filling material in the canal. They can also assist in the removal of separated instruments or other intra-canal blockages.

{1}------------------------------------------------

SUMMARY OF TECHNOLOGIES/SUBSTANTIAL EQUIVALENCE:

A comparison of the AEI Ultrasonic Tips and the currently marketed ProUltra® Endo Tips indicates the following similarities to the device which received 510(k) clearance:

    • Same technological characteristics
    • Same operating principle
    • Similar design features
    • Same connection capabilities to piezoelectric ultrasonic generator
    • Same material

NON-CLINICAL TESTING:

Substantial equivalence of the AEI Ultrasonic Tips to the predicate device is based on a comparison of indications, intended use and materials. Mechanical testing was, therefore, not conducted.

CLINICAL TESTING:

Clinical evaluations were conducted outside the US and showed that the devices were safe and effective for the proposed indications.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three bars representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 29, 2013

Ms. Kristen Nystuen Compliance Manager/Management Representative American Eagle Instruments, Incorporated 6575 Butler Creek Road MISSOULA MT 59808

Re: K122501

Trade/Device Name: AEI Dr. Terauchi Ultrasonic Tips™ Regulation Number: 21 CFR 872.4850 Regulation Name: Ultrasonic Scaler Regulatory Class: II Product Code: ELC Dated: February 18, 2013 Received: February 26, 2013

Dear Ms. Nystuen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II {Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Nystuen

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to

http://www.fda.gov/MedicalDevices/Safetv/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Kwame O. Ulmer -5

for

Anthony D. Watson. B.S., M.S., M.B.A. Director

Division of Anesthesiology. General Hospital, Respiratory, Infection Control and Dental Devices

Office of Device Evaluation

Center for Devices and

Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Statement

1/2 250) 510(k) Number (if known):

AEI Dr. Terauchi Ultrasonic TipsTM Device Name:

The AEI Dr. Terauchi Ultrasonic Tips are intended for use by dental professionals for endodontic root preparation procedures. The Ultrasonic Tips can remove soft and hard tissue from a narrow dentinal structure in order to make space for permanent filling material in the canal. They can also assist in the removal of separated instruments or other intra-canal blockages.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Mary S. Runner -S
Susan Runner, DDSMA 2013.03.22 12:24:07 -04'00'
(Division Sign-Off)
Do
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices

§ 872.4850 Ultrasonic scaler.

(a)
Identification. An ultrasonic scaler is a device intended for use during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.(b)
Classification. Class II.