K Number
K122459
Date Cleared
2012-12-13

(122 days)

Product Code
Regulation Number
884.5980
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Pinnacle LITE Pelvic Floor Repair Kit is indicated for tissue reinforcement in women with pelvic organ prolapse, for transvaginal repair of posterior vaginal vault prolapse.

The Uphold LITE Vaginal Support System is indicated for tissue reinforcement in women with pelvic organ prolapse, for the transvaginal repair of anterior and apical vaginal wall prolapse.

Device Description

The proposed devices were submitted for a leg assembly modification to the previously cleared LITE Pelvic Floor Repair Kits: Pinnacle LITE Posterior and Uphold LITE Vaginal Support System. The modification to the leg assembly removes the mesh tack welds, separator weld and adds a second leader loop to maintain the mesh leg location within the sleeve to facilitate mesh leg placement. There are no changes to the mesh design, shape, size, material or indications for use.

The proposed mesh configurations were designed for performing transvaginal vaginal wall repair to facilitate treatment of anterior apical and posterior prolapse repairs. These devices are sterile, single use devices, consisting of one lightweight synthetic mesh assembly and a needle holder.

The mesh assembly consists of a macroporous polypropylene knitted mesh body with integrated legs that are protected by disposable polymer sleeves. At the distal end of the disposable polymer sleeve is a lead with dart designed for use with the Capio™ Suture Capturing Device. The disposable lead was designed to facilitate the passage of the proposed mesh through bodily tissues for placement.

AI/ML Overview

The provided text describes a 510(k) summary for the Boston Scientific Pinnacle LITE Pelvic Floor Repair Kit, Posterior, and Uphold LITE Vaginal Support System. The submission is for a modification to an already cleared device, specifically a change in the leg assembly.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (What was tested)Reported Device Performance (Results)
Leg Assembly FlexibilityDemonstrated equivalence to predicate device.
Mesh/Leg Tensile StrengthDemonstrated equivalence to predicate device.
Leader/Dilator/Sleeve Tensile StrengthDemonstrated equivalence to predicate device.
Sleeve Removal PerformanceDemonstrated equivalence to predicate device.

2. Sample Size Used for the Test Set and Data Provenance:

The document does not specify the sample sizes used for each of the performance tests (Leg Assembly Flexibility, Mesh/Leg Tensile, Leader/Dilator/Sleeve Tensile, Sleeve Removal).

The data provenance is bench evaluation (laboratory testing) conducted by Boston Scientific. The text does not mention any human clinical data, retrospective or prospective studies, or country of origin for such data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

This information is not applicable as the provided study involves bench testing of device components, not a clinical study requiring expert assessment for ground truth.

4. Adjudication Method for the Test Set:

This information is not applicable as the provided study involves bench testing, not an assessment requiring adjudication by experts.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size:

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not reported. The study described is a bench evaluation of physical device characteristics.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

This question is not applicable as the device is a surgical mesh, not an AI algorithm.

7. The Type of Ground Truth Used:

The "ground truth" for the bench tests would be the pre-defined engineering specifications and performance characteristics of the predicate device. The goal was to prove "equivalence" to the predicate, meaning the modified device performs similarly within acceptable tolerances for these physical and mechanical properties.

8. The Sample Size for the Training Set:

This question is not applicable as the study did not involve machine learning or AI, and therefore no training set.

9. How the Ground Truth for the Training Set Was Established:

This question is not applicable for the same reason as above.


Summary of the Study:

The study described is a bench evaluation to demonstrate substantial equivalence of a modified surgical mesh device (Pinnacle LITE Pelvic Floor Repair Kit, Posterior and Uphold LITE Vaginal Support System) to its predicate device. The modification was solely to the "leg assembly" of the device, specifically removing mesh tack welds and a separator weld, and adding a second leader loop to facilitate mesh leg placement. No changes were made to the mesh design, shape, size, material, or indications for use.

The study involved four types of performance testing:

  • Leg Assembly Flexibility
  • Mesh/Leg Tensile Strength
  • Leader/Dilator/Sleeve Tensile Strength
  • Sleeve Removal Performance

The results of these tests indicated that the modified device performed equivalently to the predicate device in these aspects. This bench testing was deemed sufficient to support the claim that the modified device is as safe, as effective, and performs as well as the predicate devices, thereby establishing substantial equivalence for 510(k) clearance.

{0}------------------------------------------------

510(k) Summary for Pinnacle LITE Pelvic Floor Repair Kit, Posterior and Uphold LITE Vaginal Support System

Date Prepared: Sep 30, 2013

A. Sponsor

Boston Scientific Corporation Urology and Gynecology Division 100 Boston Scientific Way Marlborough, MA 01756

B. Contact

Michelle Berry Senior, Regulatory Affairs Specialist 508-683-4941 or Donna Gardner Director, Regulatory Affairs 508-683-4398

C. Device Name

Trade name:Pinnacle LITE Pelvic Floor Repair Kit, Posterior and Uphold LITE Vaginal Support System
Common/usual name:Surgical Mesh
Classification Name:OTP - Mesh, Surgical, Synthetic, Urogynecologic, for Pelvic Organ Prolapse, Transvaginally Placed 21 CFR 878.3300, Class II

D. Predicate Device(s)

Trade name:Pinnacle LITE Pelvic Floor Repair Kit, Posterior andUphold LITE Vaginal Support System
Common/usual name:Surgical Mesh
Classification Name:OTP – Mesh, Surgical, Synthetic, Urogynecologic, forPelvic Organ Prolapse, Transvaginally Placed21 CFR 878.3300, Class II
Premarket Notification:Boston Scientific, K103426

E. Device Description

The proposed devices were submitted for a leg assembly modification to the previously cleared LITE Pelvic Floor Repair Kits: Pinnacle LITE Posterior and Uphold LITE Vaginal Support System. The modification to the leg assembly removes the mesh tack welds, separator weld and adds a second leader loop to maintain the mesh leg location within the sleeve to facilitate mesh leg placement. There are no changes to the mesh design, shape, size, material or indications for use.

{1}------------------------------------------------

The proposed mesh configurations were designed for performing transvaginal vaginal wall repair to facilitate treatment of anterior apical and posterior prolapse repairs. These devices are sterile, single use devices, consisting of one lightweight synthetic mesh assembly and a needle holder.

The mesh assembly consists of a macroporous polypropylene knitted mesh body with integrated legs that are protected by disposable polymer sleeves. At the distal end of the disposable polymer sleeve is a lead with dart designed for use with the Capio™ Suture Capturing Device. The disposable lead was designed to facilitate the passage of the proposed mesh through bodily tissues for placement.

F. Intended Use

The Pinnacle LITE Pelvic Floor Repair Kit is indicated for tissue reinforcement in women with pelvic organ prolapse, for transvaginal repair of posterior vaginal vault prolapse.

The Uphold LITE Vaginal Support System is indicated for tissue reinforcement in women with pelvic organ prolapse, for the transvaginal repair of anterior and apical vaginal wall prolapse.

G. Technological Characteristics

The proposed device has the same identical mesh fiber materials and fiber diameter as the predicate. The leg assembly will include a second leader loop to maintain the mesh leg location within the sleeve to facilitate mesh leg placement.

H. Substantial Equivalence

Utilizing FDA's Guidance for Industry and FDA Staff "Format for Traditional and Abbreviated 510(k)s" and "Guidance for the Preparation of a Premarket Notification Application for a Surgical Mesh", a direct comparison of key characteristics demonstrates that the proposed mesh is substantially equivalent to the predicate mesh in terms of intended use, technological characteristics, and performance characteristics. The proposed device is as safe, as effective, and performs as well as the predicate devices.

I. Performance Testing (Bench Evaluation)

Boston Scientific conducted performance testing in support of the leg assembly change and the following testing was completed:

  • Leg Assembly Flexibility
  • Mesh/Leg Tensile .
  • Leader/Dilator/Sleeve Tensile ●
  • . Sleeve Removal

The results of the performance testing demonstrate equivalence of the proposed and predicate devices. The LITE Pelvic Floor Repair Kits are considered safe and effective for their intended use.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 18, 2013

Boston Scientific Corporation Urology and Women's Health Michelle M. Berry Principal Specialist, Regulatory Affairs 100 Boston Scientific Way Marlborough, MA 01752

Re: K122459

Trade/Device Name: Pinnacle LITE Pelvic Floor Repair Kit, Posterior and Uphold LITE Vaginal Support System Regulation Number: 21 CFR§ 878.3300 Regulation Name: Surgical Mesh Regulatory Class: II Product Code: OTP Dated: (Date on orig SE Itr): November 2, 2012 Received: (Date on orig SE Itr): November 5, 2012

Dear Michelle M. Berry,

This letter corrects our substantially equivalent letter of December 13, 2012.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to 28 May 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Michelle M. Berry

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive. Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure:

{4}------------------------------------------------

Indications for Use Statement

510(k) Number (if Known):

Device Name: Pinnacle LITE Pelvic Floor Repair Kit, Posterior and Uphold LITE Vaginal Support System

Indications For Use:

The Pinnacle LITE Pelvic Floor Repair Kit is indicated for tissue reinforcement in women with pelvic organ prolapse, for transvaginal repair of posterior vaginal vault prolapse.

The Uphold LITE Vaginal Support System is indicated for tissue reinforcement in women with pelvic organ prolapse, for the transvaginal repair of anterior and apical vaginal wall prolapse.

Prescription Use X (21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Benjamin R. Fisher -S 2013.11.18 17:01:46 -05'00'

§ 884.5980 Surgical mesh for transvaginal pelvic organ prolapse repair.

(a)
Identification. Surgical mesh for transvaginal pelvic organ prolapse repair is a prescription device intended to reinforce soft tissue in the pelvic floor. This device is a porous implant that is made of synthetic material, non-synthetic material, or a combination of synthetic and non-synthetic materials. This device does not include surgical mesh for other intended uses (§ 878.3300 of this chapter).(b)
Classification. Class III (premarket approval).(c)
Date premarket application approval or notice of completion of a product development protocol is required. A premarket application approval or notice of completion of a product development protocol for a device is required to be filed with the Food and Drug Administration on or before July 5, 2018, for any surgical mesh described in paragraph (a) of this section that was in commercial distribution before May 28, 1976, or that has, on or before July 5, 2018, been found substantially equivalent to a surgical mesh described in paragraph (a) of this section that was in commercial distribution before May 28, 1976. Any other surgical mesh for transvaginal pelvic organ prolapse repair shall have an approved premarket application or declared completed product development protocol in effect before being placed in commercial distribution.