K Number
K122284
Device Name
HOFFMANN 3
Manufacturer
Date Cleared
2012-10-24

(86 days)

Product Code
Regulation Number
888.3030
Reference & Predicate Devices
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Hoffmann 3 Modular External Fixation System components are external fixation frame components for use with the components of the Hoffmann II MRI and Hoffmann II Compact MRI External Fixation Systems, in conjunction with Apex Pins. It is intended to provide stabilization of open and/or unstable fractures and where soft tissue injury precludes the use of other fracture treatments such as IM rods, casts or other means of internal fixation.

The indications for use of external fixation devices include:

  • Bone fracture fixation
  • Osteotomy
  • Arthrodesis
  • Correction of deformity
  • Revision procedure where other treatments or devices have been unsuccessful
  • Bone reconstruction procedures
Device Description

This Traditional 510(k) submission is intended to address the addition of new components to the previously cleared Hoffmann 3 Modular External Fixation System. The Hoffmann 3 Modular External Fixation System consists of Rods. Posts: Couplings, Clamps and Pins that can be combined to construct different frame configurations that are MR conditional. The additional components consist of a Multiplanar Rod to Rod Coupling, Multiplanar Pin to Rod Coupling, 30° Rod Coupler, 4/5 Apex Pin, 5/6 Transfixing Pin and a Ø11mm Semi Circular Rod. This external fixation system may also be used with the components in other Howmedica Osteonics external fixation systems such as the Hoffmann II MRI and Hoffmann II Compact MRI External Fixation Systems and in conjunction with other commercially available Apex Pins.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Device: Hoffmann 3 Modular External Fixation System


1. Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criteria/Tests PerformedReported Device Performance
Material PropertiesCorrosion TestingSubstantially equivalent to predicate devices.
Mechanical PerformanceInsertion TestingSubstantially equivalent to predicate devices.
Pullout Strength TestingSubstantially equivalent to predicate devices.
Rotation TestingSubstantially equivalent to predicate devices.
Static and Dynamic Cantilever Bending TestingSubstantially equivalent to predicate devices.
Force and Torque TestingSubstantially equivalent to predicate devices.
MR Environment SafetyMagnetic Resonance Environment TestingSubstantially equivalent to predicate devices.
Radio Frequency Heating TestingSubstantially equivalent to predicate devices.
Artifact TestingSubstantially equivalent to predicate devices.

Overall Conclusion: The non-clinical testing demonstrated that the Hoffmann 3 System components are substantially equivalent to devices currently cleared for marketing. No specific quantitative performance metrics beyond "substantially equivalent" are provided in this summary.


2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: Not explicitly stated. The document refers to "non-clinical laboratory testing and engineering evaluations," which implies various test samples were used for each type of mechanical and material test.
  • Data Provenance: The testing was "non-clinical laboratory testing and engineering evaluations," which suggests the data was generated in a controlled laboratory environment (likely in the US, given the FDA submission). It is considered prospective testing for the purpose of this submission.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Not applicable. This submission relies on engineering and laboratory testing for substantial equivalence, not clinical data or expert interpretation of clinical outcomes. Ground truth in this context refers to established engineering and material science standards.

4. Adjudication Method for the Test Set

  • Not applicable. This was based on objective engineering and material testing against established predicate device performance and standards, not on subjective human assessment requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Clinical Testing: Clinical testing was not required for this submission." This indicates that the FDA deemed non-clinical, bench testing sufficient for this 510(k) submission, likely because it was an addition of new components to an already cleared system.

6. Standalone (Algorithm Only Without Human-in-the Loop Performance) Study

  • Not applicable. This is a medical device (external fixation system), not an AI algorithm. Its performance is inherent in its physical and mechanical properties, not in an interpretive algorithm.

7. Type of Ground Truth Used

  • Engineering and Material Standards: The "ground truth" for the non-clinical testing was based on established engineering principles, material science standards, and performance characteristics demonstrated by the legally marketed predicate devices (K111786 Hoffmann 3 Modular External Fixation System, K051306 Hoffmann II MRI External Fixation System, K053472 Hoffmann II MRI Components, K001886 Apex Fixation Pins).

8. Sample Size for the Training Set

  • Not applicable. This device is not an AI/ML algorithm or software that requires a training set. The "training" for proving substantial equivalence came from the existing knowledge of the predicate devices and established engineering principles.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable, as there is no training set mentioned or implied for this type of device submission.

{0}------------------------------------------------

K12284 (pg 1/2)

510(k) Summary of Safety and Effectiveness

OCT 2 4 2012

510(k) Summary of Safety and Effectiveness

LabelsValues
Proprietary Name:Hoffmann 3 Modular External Fixation System
Common Name:External Fixation System
Classification Name and Reference:Single/multiple component metallic bone fixation appliances and accessories 21 CFR §888.3030
Regulatory Class:Class II
Product Codes:87 KTT: Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Components
For Information contact:Estela Celi, Regulatory Affairs Specialist Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430 Phone: (201) 831-6461 Fax: (201) 831-6038 estela.celi@stryker.com
Date Prepared:July 27, 2012

Description:

This Traditional 510(k) submission is intended to address the addition of new components to the previously cleared Hoffmann 3 Modular External Fixation System. The Hoffmann 3 Modular External Fixation System consists of Rods. Posts: Couplings, Clamps and Pins that can be combined to construct different frame configurations that are MR conditional. The additional components consist of a Multiplanar Rod to Rod Coupling, Multiplanar Pin to Rod Coupling, 30° Rod Coupler, 4/5 Apex Pin, 5/6 Transfixing Pin and a Ø11mm Semi Circular Rod. This external fixation system may also be used with the components in other Howmedica Osteonics external fixation systems such as the Hoffmann II MRI and Hoffmann II Compact MRI External Fixation Systems and in conjunction with other commercially available Apex Pins.

Intended Use:

The Hoffmann 3 Modular External Fixation System is used to provide stabilization of open and/or unstable fractures and where soft tissue injury may preclude the use of other fracture treatments such as IM rods, casts or other means of internal fixation.

Indications:

The Hoffmann 3 Modular External Fixation System components are external fixation frame components for use with the components of the Hoffmann II MRI and Hoffmann II Compact MRI External Fixation Systems, in conjunction with Apex Pins. It is intended to provide stabilization of open and/or unstable fractures and where soft tissue injury precludes the use of other fracture treatments such as IM rods, casts or other means of internal fixation.

ﺩﻧﺎ

{1}------------------------------------------------

The indications for use of external fixation devices include:

  • . Bone fracture fixation
  • Osteotomy .
  • Arthrodesis �
  • Correction of deformity .
  • Revision procedure where other treatments or devices have been unsuccessful .
  • Bone reconstruction procedures �

Summary of Technologies:

Device comparison showed that the proposed device is substantially equivalent in intended use, materials and performance characteristics to the following predicate devices:

  • K111786 Hoffmann 3 Modular External Fixation System .
  • K051306 Hoffmann II MRI External Fixation System �
  • K053472 Hoffmann II MRI Components �
  • K001886 Apex Fixation Pins (reference predicate) .

Non-Clinical Testing:

Non-clinical laboratory testing and engineering evaluations were performed for the Hoffmann 3 System components to determine substantial equivalence. Testing demonstrated that the subject Hoffmann 3 System components are substantially equivalent to devices currently cleared for marketing.

The following testing and evaluations were performed:

  • . Corrosion Testing
  • . Insertion Testing
  • Pullout Strength Testing 0
  • . Rotation Testing
  • Static and Dynamic Cantilever Bending Testing .

Magnetic Resonance Environment Testing

  • Radio Frequency Heating Testing .
  • Force and Torque Testing .
  • Artifact Testing g

Clinical Testing: Clinical testing was not required for this submission.

Conclusion: The Hoffmann 3 System is substantially equivalent to the predicate devices identified in this premarket notification.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird-like figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002

Stryker Corporated % Ms. Estela Celi Regulatory Affairs Specialist 325 Corporate Drive Mahwah, New Jersey 07430

OCT 24 2012

Re: K12284

Trade/Device Name: Hoffman 3 Modular External Fixation System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: KTT Dated: July 27, 2012 Received: July 30, 2012

Dear Ms. Celi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, , and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{3}------------------------------------------------

Page 2 - Ms. Estela Celi

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson

Director · Division of Surgical, Orthoped and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): ال 22284 (موارا)

Device Name: Hoffmann 3 Modular External Fixation System

Indications for Use:

The Hoffmann 3 Modular External Fixation System components are external fixation frame components for use with the components of the Hoffmann II MRI and Hoffmann II Compact MRI External Fixation Systems, in conjunction with Apex Pins. It is intended to provide stabilization of open and/or unstable fractures and where soft tissue injury precludes the use of other fracture treatments such as IM rods, casts or other means of internal fixation.

The indications for use of external fixation devices include:

  • Bone fracture fixation ●
  • . Osteotomy
  • Arthrodesis .
  • Correction of deformity .
  • Revision procedure where other treatments or devices have been unsuccessful �
  • Bone reconstruction procedures ●
Prescription Use(Part 21 CFR 801 Subpart D)X
AND/OROver-The-Counter Use(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF
NEEDED)

(Division Sign-Off)

Concurrence of CDR Division of Surgical Evaluation (ODE)
and Restorative Devices

510(k) Number16122284
-------------------------

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.