K Number
K122253
Date Cleared
2013-03-22

(238 days)

Product Code
Regulation Number
870.1025
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PROCARE Monitor B20 is a portable multiparameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport.

The PROCARE Monitor B20 is intended for use under the direct supervision of a licensed health care practitioner.

The PROCARE Monitor B20 is not intended for use during MRI.

The PROCARE Monitor B20 monitors and displays oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), invasive blood pressure, end-tidal carbon dioxide, heart/pulse rate, respiration rate, ECG (including arrhythmia and ST segment analysis), temperature with a reusable or disposable electronic thermometer for continual monitoring.

Esophageal/Nasopharyngeal/Tympanic/Rectal/Bladder/Axillary/Skin/Airway/Room/Myocardial/Core/Surface temperature, and functional oxygen saturation (SpO2) and pulse rate via continuous monitoring, including monitoring during conditions of clinical patient motion or low perfusion.

Device Description

The PROCARE Monitor B20 is a multi-parameter patient monitor. The PROCARE Monitor B20 has a10.4 inch LCD display with integrated keypad and a pre-configuration patient parameter measurement module (Hemo module), the PROCARE Monitor B20 also supports a thermal recorder and Airway gas module (E-MiniC, K052582) with an extension rack.

The PROCARE Monitor B20 includes features and subsystems that are optional or configurable. The PROCARE Monitor B20 interfaces to a variety of existing central station systems via a cabled network interface.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study information for the PROCARE™ Monitor B20:

1. Table of Acceptance Criteria and Reported Device Performance:

Based on the provided text, there are no specific, quantitative acceptance criteria or performance metrics explicitly stated for the PROCARE™ Monitor B20, as it is claiming substantial equivalence to a predicate device. The claim of performance is based on the assertion that its technical specifications remain the same as the predicate, with the exception of the screen size and backlight.

Acceptance Criteria (Not Explicitly Stated)Reported Device Performance
General Performance: Device's ability to obtain patient measurementsUnchanged compared to predicate device (PROCARE Monitor B40) as there are no changes to the parameter measuring hardware.
Risk Profile: Absence of new or altered risksRe-evaluated and found to be unchanged compared to predicate device.
Technical Specifications: (e.g., accuracy of physiological parameters like SpO2, ECG, NIBP)Remain the same as the predicate device (PROCARE Monitor B40), with the exception of screen size.
Safety and Effectiveness:As safe and effective as the predicate device (PROCARE Monitor B40).

2. Sample Size Used for the Test Set and Data Provenance:

  • Test Set Sample Size: Not applicable. The document states, "The subject of this premarket submission, The PROCARE Monitor B20 did not require clinical studies to support substantial equivalence." This means no dedicated test set from clinical trials was used to prove the performance of this specific device.
  • Data Provenance: Not applicable. As no clinical studies were performed for this device, there is no clinical data provenance to report. The justification for equivalence is based on technical and design comparisons to the predicate device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • Number of Experts: Not applicable. Since no clinical studies were performed, no experts were needed to establish ground truth for a test set for this device.
  • Qualifications of Experts: Not applicable.

4. Adjudication Method for the Test Set:

  • Adjudication Method: Not applicable. No clinical test set.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:

  • MRMC Study: No. This device is a multi-parameter patient monitor, not an AI-assisted diagnostic or interpretive system that would typically be evaluated with MRMC studies comparing human readers with and without AI assistance.
  • Effect Size: Not applicable.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done:

  • Standalone Performance: Yes, implicitly. The device itself performs monitoring and analysis of physiological parameters (e.g., arrhythmia and ST segment analysis, SpO2, NIBP, heart rate). The claim of substantial equivalence implies that these internal algorithms perform adequately, as they are stated to be unchanged from the predicate device. However, no specific performance metrics for these algorithms are provided in this summary, nor are they evaluated in a standalone clinical study for this specific submission. The "standalone" here refers to the device's inherent functional performance as a monitor.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

  • Type of Ground Truth: Not applicable for this submission directly. The "ground truth" for the performance of the PROCARE Monitor B20 is established by its direct functional and technical equivalence to the predicate device, the PROCARE Monitor B40, which presumably had performance established through its own clearances. The submission relies on the assertion that "there are no changes to the parameter measuring hardware" and "all technical specification remains the same."

8. The Sample Size for the Training Set:

  • Training Set Sample Size: Not applicable. This device is a patient monitor, and the submission does not describe it as using machine learning or AI models that would require a "training set" in the conventional sense for its physiological parameter measurements. Its functionality is based on established signal processing and measurement algorithms.

9. How the Ground Truth for the Training Set Was Established:

  • Ground Truth for Training Set: Not applicable. As no training set is described for this device, no method for establishing its ground truth is mentioned.

{0}------------------------------------------------

MAR 2 2 2013

8200 West Tower Avenue Milwaukee, Wisconsin, 53223

GE Medical Systems Information Technologies

510(k) Summary (revised Aug 29, 2012)

In accordance with 21 CFR 807.92 the following summary of information is provided: Date: July 27, 2012

Submitter: Sun YanLi

Regulatory Affairs Manager GE MEDICAL SYSTEMS CHINA CO., LTD. No. 19 Changjiang road National Hi-Tech Dev. Zone Wuxi, Jiangsu, China 214028

Primary Contact Person:

Robert Casarsa Regulatory Affairs Leader GE Medical Systems Information Technologies, Inc. 8200 West Tower Avenue Milwaukee. Wisconsin 53223 Telephone: 414-362-3063 Fax at 414-362-2585 E-mail: Robert.casarsa@ge.com

Secondary Contact Person:

Kentz, Douglas Regulatory Affairs Director GE Medical Systems Information Technologies, Inc. 8200 West Tower Avenue Milwaukee, Wisconsin 53223 Phone: 414 362-2038 Fax: 414-262-2585 E-mail: Douglas.kentz@ge.com

Device: Trade Name: Common/Usual Name: Classification Names:

Product Code: Predicate Device(s): Device Description:

PROCARE™ Monitor B20 Multi-parameter patient monitor 21 CFR 870.1025 Arrhythmia detector and alarm (including STsegment measurement and alarm)

MHX

K120598 PROCARE Monitor B40

The PROCARE Monitor B20 is a multi-parameter patient monitor. The PROCARE Monitor B20 has a10.4 inch LCD display with integrated keypad and a pre-configuration patient parameter measurement module (Hemo module), the PROCARE Monitor B20 also supports a thermal recorder and Airway gas

{1}------------------------------------------------

K122253

module (E-MiniC, K052582) with an extension rack.

The PROCARE Monitor B20 includes features and subsystems that are optional or configurable. The PROCARE Monitor B20 interfaces to a variety of existing central station systems via a cabled network interface.

The PROCARE Monitor B20 is a portable multiparameter unit to Intended Use: be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intrahospital transport.

The PROCARE Monitor B20 is intended for use under the direct supervision of a licensed health care practitioner.

The PROCARE Monitor B20 is not intended for use during MRI.

The PROCARE Monitor B20 monitors and displays oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), invasive blood pressure, end-tidal carbon dioxide, heart/pulse rate, respiration rate, ECG (including arrhythmia and ST segment analysis), temperature with a reusable or disposable electronic thermometer for continual monitoring

Esophageal/Nasopharyngeal/Tympanic/Rectal/Bladder/Axillary/S kin/Airway/Room/Myocardial/Core/Surface temperature, and functional oxygen saturation (SpO2) and pulse rate via continuous monitoring, including monitoring during conditions of clinical patient motion or low perfusion.

The PROCARE Monitor B20 employs the same functional Technology: scientific technology as the predicate device the PROCARE Monitor B40 (K120598).

Determination of Substantial Equivalence:

Changes from the predicate PROCARE Monitor B40:

  • Product name New Model name is PROCARE Monitor . B20

  • Display LCD display size was changed from 12 inch to . 10.4 inch to meet customer needs.

  • LCD backlight was changed from CCFL to LED for ROHS compliance.

  • Accessories Expanded the Spo2 Sensors in accessories . list, the additional SpO2 sensors are equivalent to the existing SpO2 sensors used with the predicate B40 and

{2}------------------------------------------------

have been 510(k) cleared under separate submissions. These added accessories are also applicable to the predicate PROCARE B40 Monitor

Summary of Non-Clinical Tests:

The PROCARE Monitor B20 and its applications comply with voluntary standards as detailed in this premarket submission. The following quality assurance measures were applied to the development of the system:

  • Risk Analysis ●
  • . Requirements Reviews
  • Design Reviews
  • . Testing on unit level (Module verification)
  • . Integration testing (System verification)
  • Final acceptance testing (Validation) ●
  • . Performance testing (Verification)
  • Safety testing (Verification) e

Summary of Clinical Tests:

The subject of this premarket submission, The PROCARE Monitor B20 did not require clinical studies to support substantial equivalence.

Conclusion:

The design changes made to create the new model have no effect on the device's ability to obtain patient measurements as there are no changes to the parameter measuring hardware. To assess if the changes had any significant impact to the device, all related risks were re-evaluated and found to be unchanged.

With the exception of the screen size, all technical specification remains the same.

GE Healthcare considers the PROCARE Monitor B20 to be as safe, as effective, and performance is substantially equivalent to the predicate device.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three waves emanating from its body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

March 22, 2013

GE Healthcare c/o Mr. Robert L. Casarsa Regulatory Affairs Leader GE Healthcare 8200 West Tower Avenue Milwaukee, WI 53223

Re: K122253

Trade/Device Names: PROCARE™ Monitor B20 Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (including ST-Segment Measurement and Alarm) Regulatory Class: Class II (two) Product Code: MHX Dated: February 12, 2013 Received: February 20, 2013

Dear Mr. Casarsa:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Robert L. Casarsa

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Owen-PFaris-S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K122253

. Indications for Use

510(k) Number (if known):

Device Name: PROCARE™ Monitor B20

Indications for use:

The PROCARE Monitor B20 is a portable multiparameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport.

The PROCARE Monitor B20 is intended for use under the direct supervision of a licensed health care practitioner.

The PROCARE Monitor B20 is not intended for use during MRI.

The PROCARE Monitor B20 monitors and displays oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), invasive blood pressure, end-tidal carbon dioxide, heart/pulse rate, respiration rate, ECG (including arrhythmia and ST segment analysis), temperature with a reusable or disposable electronic thermometer for continual monitoring.

Esophageal/Nasopharyngeal/Tympanic/Rectal/Bladder/Axillary/Skin/Airway/Room/Myo cardial/Core/Surface temperature, and functional oxygen saturation (SpO2) and pulse rate via continuous monitoring, including monitoring during conditions of clinical patient motion or low perfusion.

Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Owen P. Faris -S
2013.03.22
13:08:51 -04'00'

Page 14

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.