K Number
K122242
Manufacturer
Date Cleared
2012-12-05

(131 days)

Product Code
Regulation Number
862.1660
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Elecsys CA 15-3 II CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the assay range established by the Elecsys CA 15-3 II reagent on the indicated Elecsys and cobas e immunoassay analyzers. For in vitro diagnostics use only.

Device Description

The Elecsys CA 15-3 II CalCheck 5 is a lyophilized product consisting of CA 15-3 in equine serum matrix (Level 1) or human serum matrix (Levels 2-5). During manufacture, the analyte is spiked into the matrix at the desired concentration levels.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Device: Elecsys CA 15-3 II CalCheck 5

Intended Use: An assayed control for use in calibration verification and for use in the verification of the assay range established by the Elecsys CA 15-3 II reagent on the indicated Elecsys and cobas e immunoassay analyzers. For in vitro diagnostics use only.


1. Table of Acceptance Criteria and Reported Device Performance

ParameterAcceptance CriteriaReported Device Performance
Value Assignment- Assigned range: ±21% of the assigned value for levels 2-5.- Level 1 assigned value: ≤ 2.5 U/mL.- % CV: 5% for levels 2-5.- Median value on additional analyzers (Elecsys 2010, cobas e 411, cobas e 601, and cobas e 602) must be within 10% of the master platform (MODULAR ANALYTICS E170) assigned value.- Value assignment testing was conducted and passed pre-defined acceptance criteria (specific numerical results for each level are not detailed, but the summary states they passed).- Assigned values from the master platform (MODULAR ANALYTICS E170) were deemed valid for the other platforms (Elecsys 2010, cobas e 411, cobas e 601, and cobas e 602).
Open-Vial Stability (after reconstitution)Reconstituted Elecsys CA 15-3 II CalCheck 5 must be stable for up to 4 hours at 20-25°C.Real-time testing was performed, and the data support the claim of stability for up to 4 hours at 20-25°C.
Shelf-Life Stability- Initial shelf-life claim: 18 months at 2-8°C.- Long-term shelf-life claim: 36 months at 2-8°C (supported by ongoing real-time testing).- Accelerated stability testing at 35°C supports an initial shelf-life claim of 18 months at 2-8°C.- Real-time testing at 2-8°C is ongoing to support a claim of 36 months.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size (Value Assignment): For each CalCheck lot, the CalChecks are run in duplicate on at least three E170 analyzer measuring cells. The assigned value is the median of at least 6 determinations. This process is repeated for each of the five CalCheck levels (levels 1-5). The same procedure is performed on four additional analyzers (Elecsys 2010, cobas e 411, cobas e 601, and cobas e 602) for comparison.
  • Sample Size (Stability): Not explicitly stated, but "real-time and accelerated stability tests were conducted."
  • Data Provenance: Not explicitly stated, but presumably from internal laboratory studies conducted by Roche Diagnostics in the context of their in vitro diagnostic device development. This would be considered prospective data generation for regulatory submission. The country of origin is not specified but Roche Diagnostics is based in Indianapolis, IN, USA.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Not applicable for this type of device. The "ground truth" for in vitro diagnostic control materials is established through extensive analytical validation (value assignment, stability, precision) against reference methods or established procedures, rather than expert interpretation of results. The "truth" is the analytically determined value and its expected performance characteristics.

4. Adjudication Method for the Test Set

  • Not applicable. As this device is an assayed control for calibration verification, there is no subjective adjudication of results by experts. Results are evaluated against predefined analytical precision and accuracy criteria. The median value from multiple determinations is used as the assigned value.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No. This type of study (MRMC) is relevant for diagnostic imaging or interpretation tasks where human readers (interpreters) are involved. The Elecsys CA 15-3 II CalCheck 5 is an in vitro diagnostic control material used to verify instrument performance, not to aid in the interpretation of patient samples by human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Yes, in essence. The performance characteristics (value assignment, stability) are determined for the control material itself, independent of immediate human intervention in the result determination process for patient samples. The evaluation described is of the analytical performance of the control material on specified automated immunoassay analyzers.

7. The Type of Ground Truth Used

  • Analytically Determined/Assigned Values: The ground truth for the CalChecks are their "assigned values," which are determined through a rigorous internal testing process involving multiple runs on multiple analyzer measuring cells. These assigned values are then used as the reference points for subsequent quality control and calibration verification by end-users. The stability data are also based on analytically measured changes in concentration over time.

8. The Sample Size for the Training Set

  • Not applicable. This device is an assayed control material, not a machine learning algorithm that requires a "training set." The development of the product and its performance characteristics rely on analytical validation studies, not on training data in the context of AI/ML.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable, as it's not an AI/ML device requiring a training set.

{0}------------------------------------------------

长122242

510(k) Summary

DEC 5 2012 -

IntroductionAccording to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.
Submitter name, address, contactRoche Diagnostics9115 Hague Road, P.O. Box 50416Indianapolis, IN 46250-0416317-521-3501
Contact Person: Edie EadsPhone: 317-521-4668Fax: 317-521-2324Email: edie.eads@contractors.roche.com
Secondary Contact: Stephanie GreemanPhone: 317-521-2458Fax: 317-521-2324Email: stephanie.greeman@roche.com
Date Prepared: July 24, 2012
Device NameProprietary name: Elecsys CA 15-3 II CalCheck 5Common name: CA 15-3 II CalCheck 5Classification: 21 CFR 862.1660, Single (specified) analyte controls (assayed and unassayed)Product Code: JJX
Predicate deviceThe Elecsys CA 15-3 II CalCheck 5 is substantially equivalent to other products in commercial distribution intended for similar use. We claim equivalency to the currently marketed Elecsys CA 19-9 CalCheck 5 (K101365).
Device DescriptionThe Elecsys CA 15-3 II CalCheck 5 is a lyophilized product consisting of CA 15-3 in equine serum matrix (Level 1) or human serum matrix (Levels 2-5). During manufacture, the analyte is spiked into the matrix at the desired concentration levels.

·

:

Continued on next page

{1}------------------------------------------------

Intended use The Elecsys CA 15-3 II CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the assay range established by the Elecsys CA 15-3 II reagent on the indicated Elecsys and cobas e immunoassay analyzers. For in vitro diagnostics use only. :

ComparisonThe table below compares Elecsys CA 15-3 II CalCheck 5 with the predicate
Tabledevices, Elecsys CA 19-9 CalCheck 5 (K101365),
CharacteristicElecsys CA 15-3 II CalCheck 5(Candidate)Elecsys CA 19-9CalCheck 5 (K101365)Elecsys CA 15-3 CalSet(K001468)
Intended UseThe Elecsys CA 15-3 IICalCheck 5 is an assayed controlfor use in calibration verificationand for use in the verification ofthe assay range established bythe Elecsys CA 15-3 II reagenton the indicated Elecsys andcobas e immunoassay analyzers.For in vitro diagnostics use only.The Elecsys CA 19-9CalCheck 5 is an assayedcontrol for the use incalibration verification andfor use in the verification ofthe assay range establishedby the Elecsys CA 19-9reagent on the indicatedElecsys and cobas eimmunoassay analyzers.Elecsys CA 15-3 CalSet isused for calibrating thequantitative Elecsys CA 15-3assay on Elecsys 1010 or 2010immunoassay systems.
AnalyteCA 15-3CA 19-9CA 15-3
MatrixHuman serum, equine serum(Level 1)Human SerumHuman Serum
LevelsFiveSameTwo
FormatLyophilizedSameNA
HandlingReconstitute Check 1, Check 2,Check 3, Check 4, and Check 5with exactly 1.0 mL distilled ordeionized water. Allow to standclosed for 15 minutes, and thenmix gently by inversion.SameThe calibrators are suppliedready for use in bottles that arecompatible with the system.The calibrators should only beleft on the analyzers duringcalibration at 20-25°C. Afteruse, close the bottles as soonas possible and store at 2-8°C.Because of possibleevaporation effects, not morethan 5 calibration proceduresper bottle set should beperformed.
StabilityUnopened:Store at 2-8°C until expirationdateOpened:20-25°C: 4 hoursSameUnopened:SameOpened:Twelve weeks at 2-8°COn the analyzers:Up to five hours in total

Continued on next page

{2}------------------------------------------------

510(k) Summary, Continued

Performance The Elecsys CA 15-3 II CalCheck 5 was evaluated for value assignment and Characteristics stability. See the following sections for details. Value Value assignment testing was conducted and must pass pre-defined Assignment acceptance criteria. For each Elecsys CA 15-3 II CalCheck 5 lot manufactured, the CalChecks are run in duplicate on at least three E170 analyzer measuring cells. The assigned value of each CalCheck is defined as the median value obtained over at least 6 determinations (duplicate runs on at least 3 analyzer measuring cells) of the respective CalCheck. The target value for each CalCheck is the median value obtained over at least 6 determinations of the respective CalCheck. The assigned range is calculated as ±21% of the assigned value for levels 2 through 5, while level 1 is assigned as a value ≤ 2.5 U/mL. The % CV is 5% for levels 2 through 5. The label states that each laboratory should establish appropriate acceptance criteria when using this product for its intended use. To ensure the values assigned using the master platform are transferrable and valid for the other instrument platforms, the same value assignment procedure is performed on the Elecsys 2010, cobas e 411, cobas e 601, and cobas e 602 analyzers. The assigned values obtained on the additional analyzers are compared to those obtained on the MODULAR ANALYTICES E170. The median value obtained on the four additional analyzers must be within 10% of the master platform assigned value (10% for between analyzer platform tolerances). After this acceptance criterion is met, the assigned values from the master platform are deemed valid for the MODULAR ANALYTICS E170, Elecsys 2010, cobas e 411, cobas e 601, and cobas e 602 immunoassay analyzers. Stability Real-time and accelerated stability tests were conducted to establish the shelflife and open-vial claims. Open-Vial Stability After Reconstitution: Real-time testing was performed and the data support the package insert claim that reconstituted Elecsys CA 15-3 II CalCheck 5 is stable up to 4 hours at 20-25°C. Shelf-Life Stability: The accelerated stability testing performed at 35°C supports an initial shelf-life claim of 18 months at 2-8℃. Real-time testing at 2-8℃ is on-going to support a claim of 36 months.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an emblem featuring a stylized caduceus, a symbol often associated with medicine and healthcare. The caduceus consists of a staff with two snakes coiled around it and a pair of wings at the top.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

Letter Date: December 5, 2012

Roche Diagnostics C/O Ms. Edie Eads Regulatory Affairs Consultant 9115 Hague Road Indianapolis, IN 46250-0416

Re: K122242

Trade/Device Name: Elecsys CA 15-3 II CalCheck 5 Regulation Number: 21 CFR §862.1660 Regulation Name: Quality Control Material (assayed and unassayed) Regulatory Class: Class I Product Code: JJX Dated: November 02, 2012 Received: November 05, 2012

Dear Ms. Eads:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate. commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If vour device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Ms. Edie Eads

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Maria M. Chan

Maria M. Chan, Ph.D.

Director

Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health (OIR)

Center for Devices and Radiological Health

{5}------------------------------------------------

Indication for Use

510(k) Number (if known): K122242

Device Name: Elecsys CA 15-3 II CalCheck 5

Indication for Use:

The Elecsys CA 15-3 II CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the assay range established by the Elecsys CA 15-3 II reagent on the indicated Elecsys and cobas e immunoassay analyzers. For in vitro diagnostics use only.

Prescription Use _ X (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Maria M. Chow

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K122242

§ 862.1660 Quality control material (assayed and unassayed).

(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.