(89 days)
ECAT LSO PET/CT 16, ImagenPRO®
Not Found
No
The summary describes image processing software for correcting scatter and contamination in PET data, but there is no mention of AI, ML, or related concepts like training or test sets.
No
The device is a software application for image processing, specifically correcting for photon scatter and gamma contamination in PET 3D data, and is used to produce reconstructed tomograms. It does not directly provide therapy or affect the body's structure or function for treatment.
No
This device is a software application that corrects for photon scatter and gamma contamination in PET 3D data, which then allows for the reconstruction of PET and/or PET/CT tomograms. It processes existing image data to improve its quality for reconstruction, rather than directly diagnosing.
Yes
The device is explicitly described as a "software application" that processes existing imaging data. It does not include or require any specific hardware components beyond the computer it runs on and the imaging systems it receives data from.
Based on the provided information, the Imagen3D™ system is not an In Vitro Diagnostic (IVD).
Here's why:
- IVDs are used to examine specimens from the human body. The Imagen3D™ system processes imaging data (PET and PET/CT) acquired from a patient, not biological samples like blood, urine, or tissue.
- The intended use is image processing for diagnostic imaging. The system corrects for artifacts in PET data to improve the quality of reconstructed tomograms, which are then used by healthcare providers for diagnosis. This falls under the category of medical imaging software, not IVD.
- The device description focuses on image processing techniques. It describes correcting for photon scatter and prompt gamma contamination in PET data.
Therefore, the Imagen3D™ system is a medical device used in conjunction with diagnostic imaging equipment, but it does not meet the definition of an In Vitro Diagnostic.
N/A
Intended Use / Indications for Use
The Imagen3D™ system is software application that works in conjunction with ImagenPRO® to correct for photon scatter and prompt gamma contamination present in PET 3D data. The resulting 3D datasets can be reconstructed using ImagenPRO® to produce PET and/or PET/CT reconstructed tomograms.
Product codes (comma separated list FDA assigned to the subject device)
KPS, LLZ
Device Description
Imagen3D™ is a Windows application that works in conjunction with ImagenPRO® as an additional processing step for data acquired in 3D (septa retracted mode). Because of the increase count rate of PET scanners in 3D mode, a lower dosage of the radiopharmaceutical is recommended to avoid dead time issues with the scanners. The resulting 3D sinograms can then be processed in ImagenPRO® to produce 3D tomograms. The use of this system is limited to qualified, licensed healthcare providers (radiologists, nuclear cardiologists or nuclear medicine physicians) trained in the use of nuclear medicine imaging devices.
The Imagen3D™ system is designed to take Fourier rebinned sinogram data from commercially available dedicated PET and PET/CT systems acquired in 3D (septa retracted) and scatter correct the data for export into ImagenPRO® for tomographic processing.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
PET, PET/CT
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
qualified, licensed healthcare providers (radiologists, nuclear cardiologists or nuclear medicine physicians)
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Testing and comparison of technological characteristics and intended uses found that all components of the Imagen3D™ system are equivalent to the predicates.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
ECAT LSO PET/CT 16, ImagenPRO®
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 892.1200 Emission computed tomography system.
(a)
Identification. An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.(b)
Classification. Class II.
0
OCT 2 4 2012
6. 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with 21 CFR 807.92.
General Information:
| A. | Submitted By: | Cardiovascular Imaging Technologies
4320 Wornall Road, Suite 114
Kansas City, MO 64111
Tel: 816-531-2842
Fax: 816-531-0643 |
|----|----------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| | Contact Person: | James A. Case |
| | Date Prepared: | July 20, 2012 |
| B. | Device Trade Name: | Imagen3D™ |
| | Classification Name: | System, Emission Computed Tomography
21 CFR 892.1200 (KPS)
System, Image Processing, Radiological
21 CFR 892.2050 (LLZ) |
| C. | Predicate Devices: | ECAT LSO PET/CT 16
ImagenPRO® |
D. Device Description:
Imagen3D™ is a Windows application that works in conjunction with ImagenPRO® as an additional processing step for data acquired in 3D (septa retracted mode). Because of the increase count rate of PET scanners in 3D mode, a lower dosage of the radiopharmaceutical is recommended to avoid dead time issues with the scanners. The resulting 3D sinograms can then be processed in ImagenPRO® to produce 3D tomograms. The use of this system is limited to qualified, licensed healthcare providers (radiologists, nuclear cardiologists or nuclear medicine physicians) trained in the use of nuclear medicine imaging devices.
The Imagen3D™ system is designed to take Fourier rebinned sinogram data from commercially available dedicated PET and PET/CT systems acquired in 3D (septa retracted) and scatter correct the data for export into ImagenPRO® for tomographic processing.
1
-
E. Indications for Use:
The Imagen3D™ system is software application that works in conjunction with ImagenPRO® to correct lower dosage, PET 3D data for photon scatter. The resulting 3D datasets can be reconstructed using ImagenPRO® to produce PET and/or PET/CT reconstructed tomograms. -
Comparison of Technical Characteristics to Predicate Device: F.
The Imagen3D™ system and its predicates, the ECAT LSO PET/CT 16 and the ImagenPRO® utilize the same type of data sets for analysis and calculation of data. -
H. Summary:
Testing and comparison of technological characteristics and intended uses found that all components of the Imagen3D™ system are equivalent to the predicates.
2
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular fashion around the symbol. The text is in all caps and appears to be in a sans-serif font.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Cardiovascular Imaging Technologies % Ms. Melanie Hasek Manager, Regulatory Affairs PRA 9755 Ridge Drive LENEXA KS 66219
Re: K122241
Trade/Device Name: Imagen3DTM Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: KPS and LLZ Dated: July 20, 2012 Received: July 27, 2012
Dear Ms. Hasek:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
OCT 24 2012
3
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to '
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
4
Indications for Use
510(k) Number (if known): K122241
Device Name:
Imagen3D™
Indications For Use:
The Imagen3D™ system is software application that works in conjunction with ImagenPRO® to correct for photon scatter and prompt gamma contamination present in PET 3D data. The resulting 3D datasets can be reconstructed using ImagenPRO® to produce PET and/or PET/CT reconstructed tomograms.
Prescription Use
(Part 21 CFR 801 Subpart D)
✓
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Michael D'Am
Page 1 of
H. Mitchell
Division Of Radiological Health
Office of in Vitr
51000 K1224//