(78 days)
The device is intended for use by a qualified physician for ultrasound evaluation of Fetal; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transrectal; Transvaginal and Intraoperative (abdominal, thoracic, vascular and neurosurgical).
The LOGIQ S7 Expert AND LOGIQ S7 Pro is a full featured, general purpose diagnostic ultrasound system which consists of a mobile console approximately 62 cm wide, 86 cm deep and 175 cm high that provides digital acquisition, processing and display capability. The user interface includes a computer keyboard, specialized controls, 7-inch LCD touch screen and color 19-inch LCD image display.
This document is a 510(k) Premarket Notification Summary for the GE Healthcare LOGIQ S7 Expert and LOGIQ S7 Pro Ultrasound System. It details the device's intended use and demonstrates its substantial equivalence to previously cleared predicate devices through non-clinical testing. It explicitly states that clinical studies were NOT required to support substantial equivalence.
Therefore, most of the requested information regarding acceptance criteria derived from a clinical study, sample sizes, expert ground truth establishment, adjudication methods, MRMC studies, and standalone algorithm performance cannot be provided from this document.
Here's the information that can be extracted based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Since no clinical studies were performed, there are no specific acceptance criteria for diagnostic accuracy or performance metrics (like sensitivity, specificity, or AUC) that would typically be reported from a clinical study. The acceptance criteria focused on non-clinical aspects and compliance with established standards.
| Acceptance Criteria Category | Reported Device Performance (Compliance) |
|---|---|
| Acoustic Output | Conforms to applicable medical device safety standards and voluntary standards (NEMA UD 3, NEMA UD 2) |
| Biocompatibility | Conforms to applicable medical device safety standards and voluntary standard (ISO10993-1) |
| Cleaning and Disinfection Effectiveness | Conforms to applicable medical device safety standards |
| Thermal Safety | Conforms to applicable medical device safety standards and IEC60601-1, IEC60601-1-2, IEC60601-2-37 |
| Electrical Safety | Conforms to applicable medical device safety standards and IEC60601-1, IEC60601-1-2, IEC60601-2-37 |
| Electromagnetic Safety | Conforms to applicable medical device safety standards and IEC60601-1, IEC60601-1-2, IEC60601-2-37 |
| Mechanical Safety | Conforms to applicable medical device safety standards and IEC60601-1, IEC60601-1-2, IEC60601-2-37 |
| Risk Management | ISO14971 compliant (Application of risk management to medical devices) |
| Digital Imaging and Communications | NEMA, Digital Imaging and Communications in Medicine (DICOM) Set compliant |
| Overall (Substantial Equivalence) | Considered as safe, as effective, and performance is substantially equivalent to predicate devices. |
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: Not applicable, as no clinical performance testing was conducted.
- Data Provenance: Not applicable, as no clinical data was used for performance evaluation in this submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Number of Experts: Not applicable.
- Qualifications of Experts: Not applicable.
4. Adjudication method for the test set
- Adjudication Method: Not applicable.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- MRMC Study: No, an MRMC comparative effectiveness study was not done. The submission explicitly states "The subject of this premarket submission, LOGIQ S7 Expert and LOGIQ S7 Pro, did not require clinical studies to support substantial equivalence." This device is an ultrasound system, not an AI-powered diagnostic tool, so an MRMC study for AI assistance is not relevant to this submission.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- Standalone Performance Study: No, a standalone performance study was not done. This device is an ultrasound system and not an algorithm for standalone diagnosis.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Type of Ground Truth: Not applicable, as no clinical performance testing requiring ground truth was conducted for this submission. The "ground truth" for this submission was compliance with established safety and performance standards for ultrasound devices, assessed through non-clinical means.
8. The sample size for the training set
- Sample Size for Training Set: Not applicable. Training sets are relevant for AI/machine learning models, which are not the subject of this 510(k) submission for an ultrasound system.
9. How the ground truth for the training set was established
- Ground Truth for Training Set Establishment: Not applicable.
{0}------------------------------------------------
Superficial;
Urology
and
Conventional
Image /page/0/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined within a circular border. The letters are stylized and have a flowing, cursive appearance. The logo is black and white.
ОСТ Э 2012
GE Healthcare 510(k) Premarket Notification Submission
510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
Date: July 16, 2012
Submitter: GE Healthcare 9900 Innovation Dr Wauwatosa, WI 53226
| Primary Contact Person: Bryan Behn |
|---|
| Regulatory Affairs Manager |
| GE Healthcare |
| T:(414)721-4214 |
| F:(414)918-8275 |
| Device: Trade Name: | LOGIQ S7 Expert and LOGIQ S7 Pro Ultrasound System |
|---|---|
| Common/Usual Name: | LOGIQ S7 Expert and LOGIQ S7 Pro |
| Classification Names:Product Code: | Class IIUltrasonic Pulsed Doppler Imaging System. 21CFR 892.1550 90-IYNUltrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYODiagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX |
| Predicate Device(s): | K111582 LOGIQ S8 Diagnostic Ultrasound SystemK092271 LOGIQ E9 Diagnostic Ultrasound SystemK101874 LOGIQ P6 Diagnostic Ultrasound SystemK113690 LOGIQ e Diagnostic Ultrasound SystemK112213 Voluson E6/E8/E8 Expert Diagnostic UltrasoundSystem |
| Device Description: | The LOGIQ S7 Expert AND LOGIQ S7 Pro is a full featured,general purpose diagnostic ultrasound system which consists of amobile console approximately 62 cm wide, 86 cm deep and 175cm high that provides digital acquisition, processing and displaycapability. The user interface includes a computer keyboard,specialized controls, 7-inch LCD touch screen and color 19-inchLCD image display. |
| Intended Use: | The device is intended for use by a qualified physician forultrasound evaluation of Fetal; Abdominal; Pediatric; SmallOrgan (breast, testes, thyroid); Neonatal Cephalic; AdultCephalic; Cardiac (adult and pediatric); Peripheral Vascular; |
Musculo-skeletal
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The letters and the border are in black, contrasting with the white background.
(including prostate); Transrectal; Transvaginal and Intraoperative (abdominal, thoracic, vascular and neurosurgical).
The LOGIO S7 Expert and LOGIQ S7 Pro employs the same Technology: fundamental scientific technology as its predicate devices
Summary of Non-Clinical Tests: Determination of
Substantial Equivalence: The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic, and mechanical safety, and has been found to conform to applicable medical device safety standards. LOGIQ S7 Expert and LOGIO S7 Pro and its applications comply with voluntary standards:
- IEC60601-1, Medical Electrical Equipment -1. Part 1: General Requirements for Safety
-
- IEC60601-1-2,Medical Electrical Equipment -Part 1-2:General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility Requirements and Tests
-
- IEC60601-2-37, Medical Electrical Equipment -Part 2-37:Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment
-
- NEMA UD 3, Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
-
- ISO10993-1. Biological Evaluation of Medical Devices- Part 1: Evaluation and Testing- Third Edition
-
- NEMA UD 2, Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment
-
- ISO14971, Application of risk management to medical devices
-
- NEMA, Digital Imaging and Communications in Medicine (DICOM) Set. (Radiology)
{2}------------------------------------------------
The following quality assurance measures were applied to the development of the system:
- . Risk Analysis
- . Requirements Reviews
- . Design Reviews
- . Testing on unit level (Module verification)
- . Integration testing (System verification)
- . Final Acceptance Testing (Validation)
- Performance testing (Verification)
- . Safety testing (Verification)
Transducer materials and other patient contact materials are biocompatible.
Summary of Clinical Tests:
.
The subject of this premarket submission, LOGIQ S7 Expert and LOGIQ S7 Pro, did not require clinical studies to support substantial equivalence.
- Conclusion: GE Healthcare considers the LOGIQ S7 Expert and LOGIQ S7 Pro to be as safe, as effective, and performance is substantially equivalent to the predicate device(s).
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the department's name encircling the perimeter, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion. In the center of the seal is a stylized representation of a caduceus, a symbol often associated with healthcare and medicine.
10903 New Hampshire Avenue Silver Spring, MD 20993
OCT 3 2012
GE Healthcare c/o Brvan Behn 9900 Innovation Dr. Wauwatosa. WI 53226
Re: K122114 Trade Name: LOGIO S7 Expert LOGIO S7 Pro Ultrasound System Regulation Number: 21 CFR 8892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Codes: IYN, IYO, ITX Dated: July 16, 2012 Received: July 17, 2012
Dear Mr. Behn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the LOGIO S7 Expert and LOGIQ S7 Pro, as described in your premarket notification:
Transducer Model Number
| C1-5-D | L8-18i-D | 11L-D |
|---|---|---|
| 9L-D | S4-10-D | 8C |
| ML6-15 | P2D | 3Sp-D |
| IC5-9-D | P6D | |
| 3CRF-D | RAB4-8-D |
{4}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Joshua Nipper at (301) 796-6524.
Sincerely Yours.
Mark D. Okerblom
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure(s)
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular shape. The circle has decorative swirls or flourishes extending outward from the top and bottom, giving it a vintage or classic appearance.
510(k) Number (if known):
K122114
Device Name:
LOGIQ S7 Expert and LOGIQ S7 Pro
Indications for Use:
The device is intended for use by a qualified physician for ultrasound evaluation of Fetal; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transrectal; Transvaginal and Intraoperative (abdominal, thoracic, vascular and neurosurgical).
Prescription Use_X_ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use NA (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Mhhmf D. O
(Division Sign-Off)
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
/2-11.
510(k) Number
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined within a circle. The letters and the circle are black, while the background is white.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert and LOGIQ S7 Pro Ultrasound System
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes | HarmonicImaging | CodedPulse | Other[Notes] |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics[7] | N | N | N | N | N | N | N | N | N | N | 5,6 |
| Abdominal[1] | N | N | N | N | N | N | N | N | N | N | 3,5,6 |
| Pediatric | N | N | N | N | N | N | N | N | N | N | 3,5,6 |
| Small Organ[2] | N | N | N | N | N | N | N | N | N | N | 3,5,6 |
| Neonatal Cephalic | N | N | N | N | N | N | N | N | N | N | 5,6 |
| Adult Cephalic | N | N | N | N | N | N | N | N | N | N | 5,6 |
| Cardiac Adult & Pediatric | N | N | N | N | N | N | N | N | N | N | |
| Peripheral Vascular | N | N | N | N | N | N | N | N | N | N | 3,5,6 |
| Musculo-skeletal Conventional | N | N | N | N | N | N | N | N | N | N | 3,5,6 |
| Musculo-skeletal Superficial | N | N | N | N | N | N | N | N | N | N | 3,5,6 |
| Other[1] | N | N | N | N | N | N | N | N | N | N | 3,5,6 |
| Exam Type, Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal(8) | N | N | N | N | N | N | N | N | N | N | 3,5,6 |
| Transvaginal | N | N | N | N | N | N | N | N | N | N | 3,5,6 |
| Transuretheral | |||||||||||
| Intraoperative(8) | N | N | N | N | N | N | N | N | N | N | 3,5,6 |
| Intraoperative Neurological | N | N | N | N | N | N | N | N | N | N | |
| Intravascular | |||||||||||
| Laparoscopic |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA;
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
Muhi D. Khm
(Division Sign-Off)
Division of Radiological Devices Office of In Vitro Division of Haclological Dovice Evaluation and Safety
610K 122 114
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circular border. The letters and the border are white, set against a black background, creating a high-contrast design. The logo is simple, recognizable, and represents the General Electric brand.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert and LOGIQ S7 Pro with C1-5-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes* | Harmonic Imaging | Coded Pulse | Other (Notes) |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics[7] | P | P | P | P | P | P | P | P | P | P | 5,6 |
| Abdominal[1] | P | P | P | P | P | P | P | P | P | P | 3,5,6 |
| Pediatric | P | P | P | P | P | P | P | P | P | P | 3,5,6 |
| Small Organ[2] | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | |||||||||||
| Cardiac Adult & Pediatric | |||||||||||
| Peripheral Vascular | P | P | P | P | P | P | P | P | P | P | 3,5,6 |
| Musculo-skeletal Conventional | |||||||||||
| Musculo-skeletal Superficial | |||||||||||
| Other[4] | P | P | P | P | P | P | P | P | P | P | 3,5,6 |
| Exam Type, Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal[3] | |||||||||||
| Transvaginal | |||||||||||
| Transuretheral | |||||||||||
| Intraoperative[8] | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA (K111582)
[1] Abdominal includes Renal. GYN/Pelvic. Notes:
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
15| 3D/4D Imaging mode
{6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
(Division Sign Off)
(Division Sign-Off) Division of Radiological Devices Office of In Vitro Division of Radiological Devices
510K 12.2 11.4
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circular frame. The letters are stylized and connected, creating a unique and recognizable design. The logo is in black and white.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert and LOGIQ S7 Pro with 9L-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse | Other(Notes) |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics(7) | P | P | P | P | P | P | P | P | P | 5,6 | |
| Abdominal[1] | P | P | P | P | P | P | P | P | P | 3,5,6 | |
| Pediatric | P | P | P | P | P | P | P | P | P | 3,5,6 | |
| Small Organ(2) | P | P | P | P | P | P | P | P | P | 3,5,6 | |
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | |||||||||||
| Cardiac Adult & Pediatric | |||||||||||
| Peripheral Vascular | P | P | P | P | P | P | P | P | P | 3,5,6 | |
| Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | 3,5,6 | |
| Musculo-skeletal Superficial | P | P | P | P | P | P | P | P | P | 3,5,6 | |
| Other(4) | |||||||||||
| Exam Type. Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal[8] | |||||||||||
| Transvaginal | |||||||||||
| Transuretheral | |||||||||||
| Intraoperative[8] | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication: P = previously cleared by FDA (K111582)
[1] Abdominal includes Renal, GYN/Pelvic. Notes:
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801,109)
(Division Sign-Off)
Devices
Division of Radiological Devices Division of Radiological Devices
Office of in Vitro Diagnostic Device Evaluation and Safety
S'OK. 112 114
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined within a circular border. The letters and the border are black, contrasting with the white background. The logo is a well-known symbol associated with the General Electric company.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert and LOGIQ S7 Pro with ML6-15 Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes* | Harmonic Imaging | Coded Pulse | Other [Notes] | |
| Ophthalmic | ||||||||||||
| Fetal / Obstetrics[7] | ||||||||||||
| Abdominal[1] | ||||||||||||
| Pediatric | P | P | P | P | P | P | P | P | P | 3,5,6 | ||
| Small Organ[2] | P | P | P | P | P | P | P | P | P | 3,5,6 | ||
| Neonatal Cephalic | N | N | N | N | N | N | N | N | N | |||
| Adult Cephalic | ||||||||||||
| Cardiac Adult & Pediatric | ||||||||||||
| Peripheral Vascular | P | P | P | P | P | P | P | P | P | 3,5,6 | ||
| Musculo-skeletal Conventional | N | N | N | N | N | N | N | N | N | 3,5,6 | ||
| Musculo-skeletal Superficial | N | N | N | N | N | N | N | N | N | 3,5,6 | ||
| Other[4] | ||||||||||||
| Exam Type. Means of Access | ||||||||||||
| Transesophageal | ||||||||||||
| Transrectal[8] | ||||||||||||
| Transvaginal | ||||||||||||
| Transuretheral | ||||||||||||
| Intraoperative[8] | ||||||||||||
| Intraoperative Neurological | ||||||||||||
| Intravascular | ||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA (K101874)
[1] Abdominal includes Renal, GYN/Pelvic. Notes:
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
Mhhael D. Khm
(Division signa
Division of Radiological Devices Division of Radiological Device Evaluation and Safety
510K 112.114
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, cursive font. The letters are white and are set against a black circular background. The black circle has a white outline with small white accents around the perimeter.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert and LOGIQ S7 Pro with IC5-9-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | B | M | PW | CW | Color | Color M | Power | Combined Harmonic | Coded | Other | |||
| Anatomy/Region of Interest | Doppler | Doppler Doppler | Doppler | Doppler | Modes | Imaging | Pulse | [Notes) | |||||
| Ophthalmic | |||||||||||||
| Fetal / Obstetrics[7] | P | P | P | P | P | P | b | P | P | 5,6 | |||
| Abdominal(!) | |||||||||||||
| Pediatric | |||||||||||||
| Small Organ[2] | |||||||||||||
| Neonatal Cephalic | |||||||||||||
| Adult Cephalic | |||||||||||||
| Cardiac Adult & Pediatric | |||||||||||||
| Peripheral Vascular | |||||||||||||
| Musculo-skeletal Conventional | |||||||||||||
| Musculo-skeletal Superficial | |||||||||||||
| Other (4) | P | P | P | P | P | P | P | P | b | 3,5,6 | |||
| Exam Type, Means of Access | |||||||||||||
| Transesophageal | |||||||||||||
| Transrectal[8] | P | P | P | P | P | P | P | P | P | 3,5,6 | |||
| Transvaginal | P | P | P | P | P | P | P | P | P | 3.5,6 | |||
| Transuretheral | |||||||||||||
| Intraoperative[8] | |||||||||||||
| Intraoperative Neurological | |||||||||||||
| Intravascular | |||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA (K111582)
[1] Abdominal includes Renal, GYN/Pelvic. Notes:
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Radiological Devices Division of Radiological Device Evaluation and Safety
510K 172 119
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The logo is black and white.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert and LOGIQ S7 Pro with 3CRF-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes | Harmonic Imaging | Coded Pulse | Other [Notes] |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics(7) | P | P | P | P | P | P | P | P | P | 5,6 | |
| Abdominal(1) | P | P | P | P | P | P | P | P | P | 5,6 | |
| Pediatric | P | P | P | P | P | P | P | P | P | 5,6 | |
| Small Organ[2] | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | |||||||||||
| Cardiac Adult & Pediatric | |||||||||||
| Peripheral Vascular | |||||||||||
| Musculo-skeletal Conventional | |||||||||||
| Musculo-skeletal Superficial | |||||||||||
| Other(4) | P | P | P | P | P | P | P | P | P | 5,6 | |
| Exam Type, Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal(8) | |||||||||||
| Transvaginal | |||||||||||
| Transuretheral | |||||||||||
| Intraoperative(8) | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA (K111582)
[1] Abdominal includes Renal, GYN/Pelvic. Notes:
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510K-
ડર્
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circle. The letters and the circle are in black, creating a strong contrast. The logo is simple, recognizable, and represents the General Electric brand.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert and LOGIQ S7 Pro with L8-18i-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | B | M | PW | CW | Color | Color M | Power | Combined | Harmonic | Coded | Other(Notes) | |
| Anatomy/Region of Interest | Doppler | Doppler | Doppler | Doppler | Doppler | Modes | Imaging | Pulse | ||||
| Ophthalmic | ||||||||||||
| Fetal / Obstetrics[7] | ||||||||||||
| Abdominal[1] | ||||||||||||
| Pediatric | P | P | P | P | P | P | P | P | P | 5,6 | ||
| Small Organ[2] | P | P | P | P | P | P | P | P | P | |||
| Neonatal Cephalic | P | P | P | P | P | P | P | P | P | |||
| Adult Cephalic | ||||||||||||
| Cardiac Adult & Pediatric | ||||||||||||
| Peripheral Vascular | P | P | P | P | P | P | P | P | P | 3,5,6 | ||
| Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | 3,5,6 | ||
| Musculo-skeletal Superficial | P | P | P | P | P | P | P | P | P | 3,5,6 | ||
| Other[4] | ||||||||||||
| Exam Type. Means of Access | ||||||||||||
| Transesophageal | ||||||||||||
| Transrectal[8] | ||||||||||||
| Transvaginal | ||||||||||||
| Transuretheral | ||||||||||||
| Intraoperative[8] | P | P | P | P | P | P | P | P | P | 3,5,6 | ||
| Intraoperative Neurological | ||||||||||||
| Intravascular | ||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA(K111582)
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
810K 172114
ﺮ
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined and enclosed within a stylized circular border. The letters and the border are solid black, contrasting with the white background. The logo has a slightly rough or textured appearance, possibly due to the image's source or reproduction method.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert and LOGIQ S7 Pro with S4-10-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes | Harmonic Imaging | Coded Pulse | Other [Notes) | |
| Ophthalmic | ||||||||||||
| Fetal / Obstetrics[7] | P | P | P | P | P | P | P | P | P | P | 5 | |
| Abdominal[1] | P | P | P | P | P | P | P | P | P | P | 5 | |
| Pediatric | P | P | P | P | P | P | P | P | P | P | 5 | |
| Small Organ[2] | P | P | P | P | P | P | P | P | P | P | 5 | |
| Neonatal Cephalic | P | P | P | P | P | P | P | P | P | P | 5 | |
| Adult Cephalic | ||||||||||||
| Cardiac Adult & Pediatric | P | P | P | P | P | P | P | P | P | P | P | |
| Peripheral Vascular | ||||||||||||
| Musculo-skeletal Conventional | ||||||||||||
| Musculo-skeletal Superficial | ||||||||||||
| Other[4] | ||||||||||||
| Exam Type. Means of Access | ||||||||||||
| Transesophageal | ||||||||||||
| Transrectal[8] | ||||||||||||
| Transvaginal | ||||||||||||
| Transuretheral | ||||||||||||
| Intraoperative[8] | ||||||||||||
| Intraoperative Neurological | ||||||||||||
| Intravascular | ||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA (K111582)
(1) Abdominal includes Renal, GYN/Pelvic. Notes:
(2) Small organ includes breast, testes and thyroid
(3) Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
Archi D. Ofil
(Division Sign-Off)
Division of Radiological Devices Office of In Diagnostic Device Evaluation and Safety
510K
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circle. The letters and the circle are black, creating a strong contrast. The logo is simple and recognizable.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert and LOGIQ S7 Pro with P2D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes | HarmonicImaging | CodedPulse | Other[Notes] |
|---|---|---|---|---|---|---|---|---|---|---|---|
| Ophthalmic | |||||||||||
| Fetal / Obstetrics[7] | |||||||||||
| Abdominal[1] | |||||||||||
| Pediatric | |||||||||||
| Small Organ[2] | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | P | P | |||||||||
| Cardiac Adult & Pediatric | P | P | |||||||||
| Peripheral Vascular | P | P | |||||||||
| Musculo-skeletal Conventional | . | ||||||||||
| Musculo-skeletal Superficial | |||||||||||
| Other[4] | |||||||||||
| Exam Type. Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal[8] | |||||||||||
| Transvaginal | |||||||||||
| Transuretheral | |||||||||||
| Intraoperative[8] | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA(K111582)
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
{6} Needle guidance imaging
[7] Includes infertility monitoring of follicle development
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
Marl D. Kirk
(Division Sign-Off) Division of Radiological Devices Office of In Vitro Division of Radiological Dovice Evaluation and Safety
510K. 12-114
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined within a circle. The letters are stylized and connected, creating a recognizable and iconic symbol for the company.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert and LOGIQ S7 Pro with P6D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes* | Harmonic Imaging | Coded Pulse | Other [Notes] | |
| Ophthalmic | ||||||||||||
| Fetal / Obstetrics(7) | ||||||||||||
| Abdominal[1] | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ(2) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | P | P | ||||||||||
| Cardiac Adult & Pediatric | P | P | ||||||||||
| Peripheral Vascular | P | P | ||||||||||
| Musculo-skeletal Conventional | ||||||||||||
| Musculo-skeletal Superficial | ||||||||||||
| Other (4) | ||||||||||||
| Exam Type. Means of Access | ||||||||||||
| Transesophageal | ||||||||||||
| Transrectal[8] | ||||||||||||
| Transvaginal | ||||||||||||
| Transuretheral | ||||||||||||
| Intraoperative(8) | ||||||||||||
| Intraoperative Neurological | ||||||||||||
| Intravascular | ||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA (K111582)
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
Mull D. O'K
(Division Sign-Off)
Division of Radiological Devices Division of Radiological Device Evaiuation and Safety
510K D2-114
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circle. The letters and the circle are in black, contrasting with the white background.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert and LOGIQ S7 Pro with RAB4-8-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes | HarmonicImaging | CodedPulse | Other[Notes] |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics[7] | P | P | P | P | P | P | P | P | P | [5,6] | |
| Abdominal[1] | P | P | P | P | P | P | P | P | P | [5,6] | |
| Pédiatric | P | P | P | P | P | P | P | P | P | [5,6] | |
| Small Organ[2] | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | |||||||||||
| Cardiac Adult & Pediatric | |||||||||||
| Peripheral Vascular | |||||||||||
| Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | [5,6] | |
| Musculo-skeletal Superficial | |||||||||||
| Other[4] | N | N | N | N | N | N | N | N | N | [5,6] | |
| Exam Type. Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal[8] | |||||||||||
| Transvaginal | |||||||||||
| Transuretheral | |||||||||||
| Intraoperative[8] | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA(K092271)
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
Muth D'A
(Division Sign-Off)
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510k
122-114
{17}------------------------------------------------
Image /page/17/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined within a circle. The logo is black and white.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert and LOGIQ S7 Pro with 11L-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes' | HarmonicImaging | CodedPulse | Other(Notes) | |
| Ophthalmic | ||||||||||||
| Fetal / Obstetrics[7] | ||||||||||||
| Abdominal[1] | P | P | P | P | P | P | P | P | P | [3,5,6] | ||
| Pediatric | P | P | P | P | P | P | P | P | P | [3,5,6] | ||
| Small Organ[2] | P | P | P | P | P | P | P | P | P | [3,5,6] | ||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Cardiac Adult & Pediatric | ||||||||||||
| Peripheral Vascular | P | P | P | P | P | P | P | P | P | [3,5,6] | ||
| Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | [3,5,6] | ||
| Musculo-skeletal Superficial | P | P | P | P | P | P | P | P | P | [3,5,6] | ||
| Other[4] | ||||||||||||
| Exam Type. Means of Access | ||||||||||||
| Transesophageal | ||||||||||||
| Transrectal[8] | ||||||||||||
| Transvaginal | ||||||||||||
| Transuretheral | ||||||||||||
| Intraoperative[8] | ||||||||||||
| Intraoperative Neurological | ||||||||||||
| Intravascular | ||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA(K092271)
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801. 109
(Division Sign-Off)
Division Digl-POII)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
Elok
22.114
{18}------------------------------------------------
Image /page/18/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The letters and the border are depicted in black, contrasting with the white background.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert and LOGIQ S7 Pro with 8C Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse | Other[Notes] |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics[7] | |||||||||||
| Abdominal[1] | P | P | P | P | P | P | P | P | P | 5 | |
| Pediatric | P | P | P | P | P | P | P | P | P | 5 | |
| Small Organ[2] | P | P | P | P | P | P | P | P | P | 5 | |
| Neonatal Cephalic | P | P | P | P | P | P | P | P | P | 5 | |
| Adult Cephalic | |||||||||||
| Cardiac Adult & Pediatric | |||||||||||
| Peripheral Vascular | |||||||||||
| Musculo-skeletal Conventional | |||||||||||
| Musculo-skeletal Superficial | |||||||||||
| Other[4] | |||||||||||
| Exam Type. Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal [8] | |||||||||||
| Transvaginal | |||||||||||
| Transuretheral | |||||||||||
| Intraoperative[8] | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA(K101874)
[1] Abdominal includes Renal, GYN/Pelvic. Notes:
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Rediological Devices Division of Habiogroup Of Habitogical Device and Safety
510K 102 114
{19}------------------------------------------------
Image /page/19/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined within a circular border. The letters and the border are in black, creating a high-contrast design. The logo is simple and recognizable, representing the well-known multinational corporation.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert and LOGIQ S7 Pro with 3Sp-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes | Harmonic Imaging | Coded Pulse | Other [Notes] | |
| Ophthalmic | ||||||||||||
| Fetal / Obstetrics(7) | P | P | P | P | P | P | P | P | P | P | 5,6 | |
| Abdominal11 | P | P | P | P | P | P | P | P | P | P | 5,6 | |
| Pediatric | P | P | P | P | P | P | P | P | P | P | 5,6 | |
| Small Organ(2) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | P | P | P | P | P | P | P | P | P | P | ||
| Cardiac Adult & Pediatric | P | P | P | P | P | P | P | P | P | P | ||
| Peripheral Vascular | ||||||||||||
| Musculo-skeletal Conventional | ||||||||||||
| Musculo-skeletal Superficial | ||||||||||||
| Other(4) | ||||||||||||
| Exam Type. Means of Access | ||||||||||||
| Transesophageal | ||||||||||||
| Transrectal(8) | ||||||||||||
| Transvaginal | ||||||||||||
| Transuretheral | ||||||||||||
| Intraoperative(8) | ||||||||||||
| Intraoperative Neurological | ||||||||||||
| Intravascular | ||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA(K112213)
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510K. 122.114
§ 892.1550 Ultrasonic pulsed doppler imaging system.
(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.