(127 days)
The On Call® Vivid Pal Blood Glucose Monitoring System is an electrochemical enzymatic assay for the quantitative detection of glucose in fresh capillary whole blood from the fingertip by people with diabetes at home as an aid in monitoring the effectiveness of diabetes control programs. Alternate (forearm and palm) testing sites should be used only when blood glucose level is not changing rapidly. The On Call® Vivid Blood Pal Glucose Monitoring System is intended to be used by a single patient and should not be shared. It is for in vitro diagnostic use only.
The On Call Vivid Blood Pal Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes mellitus, or use on neonates.
The On Call Vivid Pal Blood Glucose Test Strips are used with the On Call Vivid Blood Glucose Meter in the quantitative measurement of glucose in capillary blood from the fingertip, forearm and palm.
The On Call Vivid Pal Blood Glucose Control Solution is for use with the On Call Vivid Pal Blood Glucose Meter and Strips as a quality control check to verify that the meter and test strips are working together properly and that the test is performing correctly.
Not Found
The provided text is a 510(k) summary for the On Call® Vivid Pal Blood Glucose Monitoring System, which does not contain the detailed study information required to answer all aspects of your request. This document is a regulatory approval notice and focuses on the substantial equivalence determination rather than a comprehensive report of validation studies.
However, based on the information available and general knowledge of blood glucose monitoring system approvals, I can provide the following:
Device Name: On Call® Vivid Pal Blood Glucose Monitoring System
Indications for Use: The On Call® Vivid Pal Blood Glucose Monitoring System is an electrochemical enzymatic assay for the quantitative detection of glucose in fresh capillary whole blood from the fingertip by people with diabetes at home as an aid in monitoring the effectiveness of diabetes control programs. Alternate (forearm and palm) testing sites should be used only when blood glucose level is not changing rapidly. The device is intended for single patient use and is for in vitro diagnostic use only. It should not be used for the diagnosis of or screening for diabetes mellitus, or use on neonates.
1. A table of acceptance criteria and the reported device performance
The provided document does not contain a specific table detailing acceptance criteria and reported device performance. For blood glucose monitoring systems, acceptance criteria typically follow ISO 15197 standards (or similar regulatory guidelines) which specify accuracy requirements compared to a laboratory reference method. These usually involve:
- A certain percentage of results falling within ±X mg/dL (or mmol/L) of the reference for glucose concentrations below Y mg/dL (or mmol/L).
- A certain percentage of results falling within ±Z% of the reference for glucose concentrations above Y mg/dL (or mmol/L).
Without the study report, the specific numerical criteria and the device's performance against them cannot be extracted from this document.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not available in the provided 510(k) letter. The letter is a communication of regulatory clearance, not a clinical study report. Such details would typically be found in the manufacturer's submission or detailed study reports.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This question is not applicable to a blood glucose monitoring system. The "ground truth" for glucose measurements is established by a highly accurate laboratory reference method (e.g., YSI analyzer) performed by trained lab personnel, not by medical experts in an interpretive role like radiologists.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This question is not applicable to a blood glucose monitoring system. Adjudication typically refers to resolving discrepancies between multiple human readers or interpretations, which is not how glucose measurements are validated.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This question is not applicable to a blood glucose monitoring system. This type of study design is relevant for AI-powered diagnostic imaging devices where human readers interpret images. A blood glucose monitor is a standalone measurement device.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Yes, a blood glucose monitoring system inherently performs as a standalone device. Its performance is evaluated by directly comparing its glucose readings to those obtained from a laboratory reference method. There is no "human-in-the-loop" interpretation step that would benefit from, or be hindered by, an AI algorithm in the context of reading the glucose value. The device's algorithm processes the electrochemical signal to produce a glucose concentration.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For a blood glucose monitoring system, the ground truth is established by a highly accurate laboratory reference method, such as a YSI glucose analyzer, typically using venous blood samples.
8. The sample size for the training set
This information is not available in the provided 510(k) letter. Blood glucose meters do not typically have "training sets" in the same way machine learning algorithms do. Their calibration is established during manufacturing and validated through accuracy studies. While there might be internal data used to develop and refine the electrochemical enzymatic assay and its signal processing, it's not usually referred to as a "training set" in the context of regulatory submissions for these devices.
9. How the ground truth for the training set was established
As noted above, the concept of a "training set" and associated ground truth establishment for a blood glucose meter in the context of an AI/ML algorithm is not directly applicable. If any internal development involved data-driven optimization, the ground truth would still have been established using a laboratory reference method for glucose concentration.
{0}------------------------------------------------
DEPARTMENT OF HEALTH & HUM AN SERVICES
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling a person with outstretched arms, composed of three curved lines.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
November 21, 2012
Acon Laboratories, Inc. c/o Qiyi Xie 10125 Mesa Rim Road San Diego, CA 92121
Re: K122110
Trade/Device Name: On Call® Vivid Pal Blood Glucose Monitoring System Regulation Number: 21 CFR §862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: CGA, NBW Dated: November 7, 2012 Received: November 8, 2012
Dear Qiyi Xie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Page 2 - Qiyi Xie
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostics and Radiological Health at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Carol C: Benson
for
Courtney H. Lias, Ph.D Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use Form
510(k) Number (if known):
Device Name: On Call Vivid Pal Blood Glucose Monitoring System
Indications for Use:
The On Call® Vivid Pal Blood Glucose Monitoring System is an electrochemical enzymatic assay for the quantitative detection of glucose in fresh capillary whole blood from the fingertip by people with diabetes at home as an aid in monitoring the effectiveness of diabetes control programs. Alternate (forearm and palm) testing sites should be used only when blood glucose level is not changing rapidly. The On Call® Vivid Blood Pal Glucose Monitoring System is intended to be used by a single patient and should not be shared. It is for in vitro diagnostic use only.
The On Call Vivid Blood Pal Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes mellitus, or use on neonates.
The On Call Vivid Pal Blood Glucose Test Strips are used with the On Call Vivid Blood Glucose Meter in the quantitative measurement of glucose in capillary blood from the fingertip, forearm and palm.
The On Call Vivid Pal Blood Glucose Control Solution is for use with the On Call Vivid Pal Blood Glucose Meter and Strips as a quality control check to verify that the meter and test strips are working together properly and that the test is performing correctly.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR'801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Danise Dhmonteles
vision Sign-Off Office of In Vitro Diagnostics and Radiological Health
510(k) K122110
§ 862.1345 Glucose test system.
(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.