(385 days)
The Intellisave AX700 Anesthesia System is intended to provide general inhalational anesthesia and ventilation support to neonatal, pediatric and adult patients. The device is intended to provide volume or pressure controlled ventilation.
The Intellisave AX700 suitable for most types of inhalation anesthesia. O2 and N2O or O2 and Air can be administered, and anesthesia vaporizer(s) can be connected to a back bar. The gasses can be supplied from a centralized supply or cylinders. The Intellisave AX 700 is a prescriptive device restricted to use by, or on the order of, a physician. The Intellisave AX700 includes the following: An Electronic Gas Mixer (EGM), an anesthesia ventilator with various ventilation modes, an Integrated Breathing System (IBS), and an optional multigas module.
The Philips Intellisave AX700 Anesthesia System is an anesthesia breathing machine, which is a medical device and therefore its performance is compared to a similar predicate device already on the market rather than having explicit acceptance criteria set. The study conducted to demonstrate equivalence is a non-clinical performance assessment which shows how the AX700 is substantially equivalent to the GE Avance Anesthesia Breathing Machine (K112722).
1. Table of Acceptance Criteria and Reported Device Performance
Since this is a substantial equivalence submission for a medical device (anesthesia system), direct "acceptance criteria" are not explicitly defined as pass/fail thresholds for an AI algorithm's performance. Instead, the "acceptance criteria" can be interpreted as demonstrating that the Philips Intellisave AX700 Anesthesia System performs equivalently to the predicate device, the GE Avance Anesthesia Breathing Machine (K112722), across various technical and functional specifications. The table below summarizes the comparison presented in the 510(k) summary, effectively acting as the reported device performance against the predicate's established performance.
Category | Predicate Device (GE Avance) Performance | Proposed Device (AX700) Performance | Equivalence Assessment (Comments) |
---|---|---|---|
Ventilation Modes | Volume Control, Pressure Control, SIMV, Pressure Support, PCV-VG, Cardiac Bypass Mode | Volume Controlled Ventilation, Pressure Controlled Ventilation, SIMV, Pressure Supported Ventilation, Volume Supported Ventilation, PRVT, Heart-Lung Mode | Equivalent |
Intended Use | General inhalation anesthesia and ventilatory support to neonatal, pediatric, adult patients; volume or pressure control ventilation. | General inhalational anesthesia and ventilation support to neonatal, pediatric, and adult patients; volume or pressure controlled ventilation. | Equivalent |
Ranges (Ventilation) | |||
Tidal volume | 20 to 1500 mL | 20 to 1500 mL | Equivalent |
Minute volume | 0 to 99.9 L/min | 0.2 to 60.0 L/min | Equivalent |
Pressure, inspired | 5 to 60 cmH2O | 4 to 67 cmH2O | Equivalent |
Pressure limit | 12 to 100 cmH2O | 10 to 80 cmH2O | Equivalent |
Pressure Support | Off, 2 to 40 cmH2O | 4 to 50 cmH2O | Equivalent |
Respiration Rate | 4 - 100 BPM (VCV, PCV, PCV-VG), 2 to 60 BPM (SIMV, PSV) | 4 to 80 BPM | Both within range required for application. |
I:E Ratio | 2:1 to 1:8 | 3:1 to 1:9.9 | Equivalent |
Flow trigger | 1 to 10 L/min | 1 to 10 LPM | Equivalent |
Inspiration termination | 5 to 50% | 10 to 80% (expiratory triggering) | Both within range required for application. |
Inspiratory Pause | 0-60% | 0 to 70% | Equivalent |
PEEP | OFF, 4 to 30 cmH2O | OFF, 4 to 20 cmH2O | Both within range required for application. |
Ventilator gas flow | Continuous: Max. 80 L/min, Peak: 120 L/min | Continuous: Max. 80 L/min, Peak: 120 L/min | Equivalent |
(Fresh gas) Flow compensation | 200 mL/min to 15 L/min | 200 mL/min to 20 L/min | Equivalent |
Accuracy (Ventilation) | |||
Tidal volume delivery | > 210 mL: ±7%, 210 mL = ±9%, 60 cmH2O: 12 cm H2O for 15 seconds (+PEEP-2cmH2O with PEEP on) | PEEP + 5 cmH2O for 15 sec | Equivalent |
Alarm silence | 120 seconds | 120 seconds | Equivalent |
Fresh Gas | |||
Flow range (per gas) | OFF, 100 mL/min-15 L/min | OFF, 100 mL/min to 10 LPM | Both within range required for application. |
Total Flow range | OFF, 150 mL/min-15 L/min | OFF, 200 mL/min to 20 LPM | Both within range required for application. |
O2 flow accuracy | ± 5% or ± 20 mL/min | ± 10% or ± 50 mL/min | Both within range required for application. |
Balance gas flow accuracy | ± 5% or ± 20 mL/min (larger of) Air/N2O | ± 10% or ± 50 mL/min | Both within range required for application. |
Total flow accuracy | ± 10% or ± 40 mL/min | ± 10% or ± 50 mL/min | Equivalent |
O2 concentration range (Air) | 21%, 25 to 100% (Air); 25 to 100% (N2O) | 21 to 100% (balance gas Air); 25 to 100% (balance gas N2O) | Equivalent |
O2 concentration accuracy | ± 5% V/V for flows 1 LPM | ± 5% V/V | Equivalent |
Compensation | to 20℃ and 101.3 kPa | to 20℃ and 101.3 kPa | Equivalent |
Alternate O2 flow | 500 mL/min – 10 L/min | 5 L/min - 15 L/min | Both within range required for application. |
Electrical Specs | |||
Power input | 100-120 Vac, 50/60 Hz | 100-120 Vac, 50/60 Hz | Equivalent |
Rating | 10A@220Vac/15A@120Vac | 10A@220Vac/15A@120 Vac | Equivalent |
Outlets | 3 outlets on back with 2A fuses, and one 3A fuse (120 V), breakers, isolation transformer | 3 outlets on back with 3.15A fuse (120 V) and one 1 outlet in front (vaporizer) with 3.15 A fuse. Isolation transformer | Equivalent |
Pneumatic Specs | |||
Aux. Fresh gas Outlet | ISO 22 mm OD / 15 mm ID | ISO 22 mm OD / 15 mm ID | Equivalent |
Gas supply input | 35 to 100 psi | 58 to 87 psi | Both within range for Operating Room application. |
Adjustable Pressure Limiting valve | 0.8 to 70 cm H2O | Spontaneous, 5 to 75 cmH2O | Equivalent |
Environmental Specs | |||
Operation Temperature | 10° to 40°C | 10° to 40°C | Equivalent |
Operation Humidity | 15 to 95% relative humidity | 10 to 90% relative humidity | Equivalent |
Operation Altitude | -440 to 3565 m | -100 to 3000 m | Both within range required for application. |
Storage Temperature | -25° to 60°C / -13° to 140°F | -20° to 60°C | Equivalent |
Storage Humidity | 10 to 95% RH | 10 to 90% relative humidity | Equivalent |
2. Sample Size Used for the Test Set and Data Provenance
This submission is for a traditional medical device (anesthesia system), not an AI/ML-driven device that relies on a test set of data samples for algorithmic performance evaluation.
Therefore, there is no "test set" sample size in the context of data (e.g., images, patient records) or any mention of data provenance (e.g., country of origin, retrospective/prospective).
The "testing" mentioned refers to engineering verification and validation testing of the physical and software components of the device.
3. Number of Experts Used to Establish Ground Truth for the Test Set and their Qualifications
As there is no "test set" of data samples in the sense of AI/ML, there were no experts used to establish ground truth for such a test set. This device is evaluated based on its functional specifications and compliance with established medical device standards.
4. Adjudication Method for the Test Set
Not applicable, as there is no "test set" in the context of data requiring adjudication. The evaluation is based on engineering tests and comparison to a predicate device.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its Effect Size with AI vs. without AI assistance
No MRMC comparative effectiveness study was done as this is not an AI/ML-enabled diagnostic or assistive device. The submission describes a standalone anesthesia system.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was done
This question is not directly applicable in the context of this device. The device is an anesthesia system, a piece of medical equipment, not an algorithm being evaluated for its standalone performance. The "standalone" performance is the intrinsic technical and functional performance of the system as measured during bench testing and detailed in the comparison table. The device itself is designed for human-in-the-loop operation by a physician.
7. The Type of Ground Truth Used
The "ground truth" for this device's performance is established through engineering specifications, compliance with recognized industry standards (e.g., IEC, ISO), and direct measurement of physical and software outputs during verification and validation testing. The predicate device's established performance serves as the benchmark against which the proposed device's "truth" (its actual performance characteristics) is compared.
8. The Sample Size for the Training Set
Not applicable. This is not an AI/ML device that requires a "training set" of data.
9. How the Ground Truth for the Training Set was Established
Not applicable, as there is no "training set."
§ 868.5160 Gas machine for anesthesia or analgesia.
(a)
Gas machine for anesthesia —(1)Identification. A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The device may include a gas flowmeter, vaporizer, ventilator, breathing circuit with bag, and emergency air supply.(2)
Classification. Class II (performance standards).(b)
Gas machine for analgesia —(1)Identification. A gas machine for analgesia is a device used to administer to a patient an analgesic agent, such as a nitrous oxide-oxygen mixture (maximum concentration of 70 percent nitrous oxide).(2)
Classification. Class II (performance standards).