K Number
K122035
Device Name
HEPAFAT SCAN
Date Cleared
2013-12-13

(520 days)

Product Code
Regulation Number
892.1000
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

HepaFat-Scan is a software device intended for quantitative measurement of the triglyceride fat fraction in magnetic resonance images of the liver. It utilises magnetic resonance images that exploit the difference in resonance frequencies between hydrogen nuclei in water and triglyceride fat. The quantitative triglyceride fat fraction is based on the measurement of a magnetic resonance parameter that reflects the ratio of the proton density signal of triglyceride fat to the total proton density signal in the liver.

When interpreted by a trained physician, the results provide information that can aid in diagnosis.

Device Description

Standalone software application to facilitate the import visualization of multi-slice, gradient-echo MRI encompassing the abdomen, with functionality independent of the MRI equipment, to provide objective and reproducible determination of the triglyceride fat fraction in magnetic resonance images of the liver. It utilises magnetic resonance images that exploit the difference in resonance frequencies between hydrogen nuclei in water and triglyceride fat. The quantitative triglyceride fat fraction is based on the measurement of a magnetic resonance parameter that reflects the ratio of the proton density signal of triglyceride fat to the total proton density signal in the liver.

AI/ML Overview

Here's an analysis of the HepaFat-Scan device, outlining its acceptance criteria and the study used for validation, based on the provided text:

HepaFat-Scan Acceptance Criteria and Validation Study

The provided 510(k) summary for HepaFat-Scan focuses on demonstrating substantial equivalence to existing predicate devices, rather than establishing specific quantitative acceptance criteria in the traditional sense of a performance study with defined thresholds for sensitivity, specificity, etc. The primary acceptance criterion here is "acceptable agreement" with a recognized ground truth method across a clinically relevant range.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implied)Reported Device Performance
Acceptable agreement between HepaFat-Scan output and volumetric fat fraction from liver biopsy samples."Acceptable agreement was attained when comparing the liver biopsy volumetric fat fraction measurements to the HepaFat-Scan device output over a clinically wide range of liver fat fractions."
Equivalence between different MRI scanner models.Demonstrated through a phantom study.

Note: The document does not specify numerical thresholds (e.g., correlation coefficients, mean absolute error limits) for "acceptable agreement." This is common in substantial equivalence claims where the goal is to show the new device performs similarly to established methods, rather than surpassing a specific, pre-defined performance metric.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: Not explicitly stated. The document refers to "in-vivo human clinical studies" where volumetric fat fractions were determined from liver biopsy samples. The exact number of patients or biopsy samples is not provided.
  • Data Provenance: "in-vivo human clinical studies." The country of origin is not specified, but the applicant's address is Australia, suggesting the studies could have been conducted there. The studies were likely retrospective or a combination of retrospective and prospective, as biopsy data is often collected as part of routine patient care.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Number of Experts: Not explicitly stated.
  • Qualifications of Experts: Not explicitly stated, however, the ground truth was established by determining "volumetric fat fractions from liver biopsy samples... using a stereological method based on the Delesse principle." This implies the involvement of pathologists or trained laboratory technicians skilled in histopathology and stereological analysis.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not explicitly mentioned. Since the ground truth involved a quantitative stereological method on biopsy samples, it's less likely to involve a consensus-based adjudication process typical of qualitative image interpretations. The "agreement" was likely a direct comparison of the HepaFat-Scan's quantitative output against the quantitative biopsy-derived fat fraction.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • MRMC Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted in the context of human readers improving with AI assistance. The study focused on the standalone performance of the HepaFat-Scan software compared to biopsy, not on its impact on human interpretation.

6. Standalone Performance (Algorithm Only without Human-in-the-loop Performance)

  • Standalone Performance: Yes, a standalone performance study was conducted. The HepaFat-Scan is described as a "Standalone software application," and its "Accuracy and reproducibility of the quantification of the triglyceride fat fraction has been demonstrated through a combination of bench testing and in-vivo human clinical studies." This indicates the algorithm's output was directly compared to the ground truth.

7. Type of Ground Truth Used

  • Type of Ground Truth: The primary ground truth for the in-vivo human clinical studies was pathology-derived data. Specifically, "Volumetric fat fractions were determined from liver biopsy samples... using a stereological method based on the Delesse principle, which directly measures volumetric fat fraction."

8. Sample Size for the Training Set

  • Training Set Sample Size: Not provided. The document describes the validation study (test set) but does not disclose information about the training data used to develop the HepaFat-Scan algorithm.

9. How the Ground Truth for the Training Set Was Established

  • Ground Truth for Training Set: Not provided. The document focuses exclusively on the validation study and its ground truth. Information on how the algorithm was developed or trained, including the ground truth used for that process, is not included in this summary.

{0}------------------------------------------------

K122035
Page 1 of 2

Image /page/0/Picture/1 description: The image contains the word "Resonance" in a stylized font. To the right of the word, there is a circular logo with a pattern inside. The word "health" is also present, but it is partially obscured and difficult to read clearly.

DEC 13 2013

and

sets

data

triglyceride fat. The quantitative triglyceride fat fraction is based on the measurement of a magnetic resonance parameter that reflects the ratio of the proton density signal of triglyceride fat to the total proton

510(k) Summary

DATE: October 30th, 2013 SUBMITTED BY: Resonance Health Analysis Services Pty Ltd 278 Stirling Highway Claremont WA 6010 Australia Contact in Australia: Mr Andrew Bathgate Quality Assurance and Regulatory Affairs Manager Tel: +61 8 9286 5300 Fax: +61 8 9286 1179 Mr Greg Holland Contact in the US: Regulatory Specialists Inc. 3722 Ave. Sausalito Irvine, CA 92606 Tel: 949-262-0411 Fax: 949 552 2821 NAME OF DEVICE: HepaFat-Scan CLASSIFICATION NAME: Magnetic Resonance Diagnostic Device (21 CFR 892.1000) FRODUCT CODE: 90 LNH HepaFat-Scan TRADE NAME: PREDICATE DEVICE(S): KI03411: IDEAL IQ Software Option R2-MRI Analysis System K043271: DEVICE DESCRIPTION: Standalone software application to facilitate the import visualization of multi-slice, gradient-echo MRI encompassing the abdomen, with functionality independent of the MRI equipment, to provide objective and reproducible determination of the triglyceride fat fraction in magnetic resonance images of the liver. It utilises magnetic resonance images that exploit the difference in resonance frequencies between hydrogen nuclei in water and

density signal in the liver.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Resonancel Health. The logo consists of the word "Resonancel" in a serif font, followed by the word "Health" in a smaller, sans-serif font. To the right of the text is a circular graphic with a pattern of small shapes inside.

INTENDED USE:

HepaFat-Scan is a software device intended for quantitative measurement of the triglyceride fat fraction in magnetic resonance images of the liver. It utilises magnetic resonance images that exploit the difference in resonance frequencies between hydrogen nuclei in water and triglyceride fat. The quantitative triglyceride fat fraction is based on the measurement of a magnetic resonance parameter that reflects the ratio of the proton density signal of triglyceride fat to the total proton density signal in the liver.

When interpreted by a trained physician, the results provide information that can aid in diagnosis.

SUBSTANTIAL EQUIVALENCE INFORMATION:

The device has been shown to be substantially equivalent to the R2-MRI Analysis System (K043271) and the IDEAL IQ Software Option (K103411). Technological similarities between HepaFat-Scan and the two predicates include:

  • . Similar Intended Uses in that the devices utilize MR images to make quantitative measurements of a physiological characteristic of the liver.
  • Similar Intended Purposes in that the devices are used to support clinical diagnoses. .
  • Devices are stand-alone software applications, independent from the MR scanner.
  • . Devices utilize multi-echo and multi slice images.
  • Devices facilitate the import, visualization and analysis of MR images. .

In addition. HepaFat-Scan has the following specific technological similarities with the Ideal IQ Software device (K103411):

  • . Both devices utilize the resonance frequency differences between water and fat protons in the fat measurements.
  • . Both devices incorporate techniques to take into account the effects of T2* on the fat measurement.
  • Both devices provide a fat fraction measurement. .

Accuracy and reproducibility of the quantification of the triglyceride fat fraction has been demonstrated through a combination of bench testing and in-vivo human clinical studies. A phantom was utilized to demonstrate equivalence between different scanner models. Volumetric fat fractions were determined from liver biopsy samples for clinical comparison with HepaFat-Scan using a stereological method based on the Delesse principle, which directly measures volumetric fat fraction. Acceptable agreement was attained when comparing the liver biopsy volumetric fat fraction measurements to the HepaFat-Scan device output over a clinically wide range of liver fat fractions.

CONCLUSION

Resonance Health considers the HepaFat-Scan software device to be substantially equivalent to the predicate devices R2-MRI Analysis System (K043271) and the IDEAL IO Software Option (K103411), in respect to safety and effectiveness.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo of the Department of Health & Human Services. The logo includes the department's seal on the left and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" on the right. The seal features a stylized human figure within a circle, representing the department's focus on health and human well-being. The text is in a bold, sans-serif font, emphasizing the department's name.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 13, 2013

RESONANCE HEALTH ANALYSIS SERVICES PTY LTD % GREG HOLLAND, CONSULTANT REGULATORY SPECIALIST, INC. 3722 AVENUE SAUSALITO IRVINE CA 92606

Re: K122035

Trade/Device Name: Hepafat Scan Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: October 30, 2013 Received: November 04, 2013

Dear Mr. Holland:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2-Mr. Holland

ﺪ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﻮﺿﻮﻋﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/dclault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michael D. O'Hara for

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration . Pat . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Indications for Use

510(k) Number (if known) K 122035

·

Device Name HepaFat-Scan

Indications for Use (Describe)

HepaFat-Scan is a software device intended for quantitative measurement of the triglyceride fat fraction in magnetic resonance images of the liver. It utilises magnetic resonance images that exploit the difference in resonance frequencies between hydrogen nuclei in water and triglyceride fat. The quantitative triglyceride fat fraction is based on the measurement of a magnetic resonance parameter that reflects the ratio of the proton density signal of triglyceride fat to the total proton density signal in the liver.

When interpreted by a trained physician, the results provide information that can aid in diagnosis.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)Michael D. O'Hara

FORM FDA 3881 (9/13)

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

2017년 10월 1일 1월 1일 1월 14시 1월 14시 1 EF

§ 892.1000 Magnetic resonance diagnostic device.

(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.