(204 days)
The invendo C25 Colonoscopy System is intended to provide visualization and diagnostic / therapeutic access to the adult lower gastrointestinal tract (including but not limited to, the anus, rectum, sigmoid colon, colon, cecum and ileocecal valve) for endoscopy and endoscopic surgery.
The colonoscope component of the invendo C25 Colonoscopy System, the SC20 colonoscope, is a single-use disposable device. The SC20 colonoscope cannot be reprocessed.
The invendo C25 Colonoscopy System is a modified version of the invendo C20 Colonoscopy System.
Like the predicate device, the invendo C25 Colonoscopy System is steered and controlled by the operator. The distal tip of the colonoscope is deflectable and includes a CMOS camera and LEDs for illumination. A working channel is incorporated to enable biopsies and polypectomies. The invendo C25 Colonoscopy System includes insufflation, irrigation and suction functions.
The invendo C25 Colonoscopy System consists of two components: (1) the invendo C25 base system; and (2) a single-use, disposable colonoscope, the SC20 colonoscope.
The invendo C25 base system provides the SC20 colonoscope with the above mentioned functions and transfers user commands to the colonoscope. To control the colonoscope, the operator uses a handheld control unit. All colonoscope functions are solely activated by the user. The invendo C25 base system consists of the Supply and Processing Unit (SPU or base unit), the handheld control and the drive unit.
The provided document is a 510(k) premarket notification for the invendo C25 Colonoscopy System. This type of submission focuses on demonstrating substantial equivalence to a predicate device, rather than proving a device meets specific acceptance criteria through a clinical study for novel performance claims.
Therefore, many of the requested details about acceptance criteria, clinical study methodology, and training/test set specifics are not applicable or available within this document, as it describes a technical modification and equivalence demonstration, not a de novo clinical performance study against pre-defined metrics.
However, I can extract the information that is present:
1. A table of acceptance criteria and the reported device performance
Since this filing is for substantial equivalence to a predicate device (invendo C20 Colonoscopy System), the "acceptance criteria" are primarily related to meeting established safety standards and demonstrating that the new device performs at least as well as the predicate with its modifications.
| Acceptance Criteria / Standard | Reported Device Performance |
|---|---|
| IEC 60601-1 (General safety) | System functioned as intended, met test specifications. |
| IEC 60601-2-18 (Endoscopic equipment safety) | System functioned as intended, met test specifications. |
| IEC 60601-1-2 (EMC) | System functioned as intended, met test specifications. |
| ISO 10993-5 (In vitro cytotoxicity) | System functioned as intended, met test specifications. |
| ISO 10993-10 (Irritation & sensitization) | System functioned as intended, met test specifications. |
| Test of check valve under worst-case conditions | System functioned as intended, met test specifications. |
| Evaluation of decompression line and system tests | System functioned as intended, met test specifications. |
| Evaluation of backflow without check valves | System functioned as intended, met test specifications. |
| Evaluation of reprocessing performance | System functioned as intended, met test specifications. |
| Performance compared to predicate device | Performed as well as or better than the predicate. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: This document does not specify sample sizes for any human clinical test sets, as the testing described is primarily non-clinical (engineering and biocompatibility). The "Evaluation of the performance compared to the predicate device" would have been a non-clinical, comparative engineering/functional test, not a clinical trial.
- Data Provenance: Not applicable, as the document details non-clinical laboratory testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable. This document describes non-clinical testing against engineering standards and a predicate device, not a study requiring expert-established ground truth on medical data.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable for the type of testing described in this 510(k).
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, an MRMC comparative effectiveness study was not done as this is a device modification submission, not an AI-enabled diagnostic device.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- Not applicable, as this is a medical device (colonoscope system), not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The "ground truth" for the non-clinical tests were the specified engineering standards (e.g., IEC, ISO) and the functional and safety performance of the predicate device.
8. The sample size for the training set
- Not applicable, as this is not an AI/algorithm device that requires a training set. The "design validation" mentioned refers to engineering design and testing, not machine learning model training.
9. How the ground truth for the training set was established
- Not applicable, as there is no training set for an AI/algorithm.
{0}------------------------------------------------
1421582
Special 510(k) premarket Notification Invendo C25 Colonoscopy System
Section 7
endo medical
THE FUTURE OF ENDOSCOPY
DEC 2 0 2012
Special 510k Summary 7
Applicant Information
invendo medical GmbH Name:
- Address: Peterhofstr. 3b 86438 Kissing Germany
- Contact:
Oliver v. Ruepprecht Manager QA/RA Phone : +49 ~ 8233 - 744 98 16 Fax : +49-8233-744 98 15 Mail : oliver.ruepprecht@invendo-medical.de
Device Information
- Device Name: invendo C25 Colonoscopy System
Components: invendoscope SC20 (SC20 colonoscope) invendo C25 Base System (base unit, drive, cleaner, handheld)
Common Name: Colonoscope
Class II, FDF 876.1500 Classification:
Predicate invendo C20 colonoscopy system (K100624) device
Intended use The invendo C25 Colonoscopy System is intended to provide visualization and diagnostic / therapeutic access to the adult lower gastrointestinal tract (including but not limited to, the anus, rectum, sigmoid colon, colon, cecum and ileocecal valve) for endoscopy and endoscopic surgery.
The colonoscope component of the invendo C25 Colonoscopy System, the SC20 colonoscope, is a single-use disposable device. The SC20 colonoscope cannot be reprocessed.
{1}------------------------------------------------
Description
The invendo C25 Colonoscopy System is a modified version of the invendo C20 Colonoscopy System.
Like the predicate device, the invendo C25 Colonoscopy System is steered and controlled by the operator. The distal tip of the colonoscope is deflectable and includes a CMOS camera and LEDs for illumination. A working channel is incorporated to enable biopsies and polypectomies. The invendo C25 Colonoscopy System includes insufflation, irrigation and suction functions.
The invendo C25 Colonoscopy System consists of two components: (1) the invendo C25 base system; and (2) a single-use, disposable colonoscope, the SC20 colonoscope.
The invendo C25 base system provides the SC20 colonoscope with the above mentioned functions and transfers user commands to the colonoscope. To control the colonoscope, the operator uses a handheld control unit. All colonoscope functions are solely activated by the user. The invendo C25 base system consists of the Supply and Processing Unit (SPU or base unit), the handheld control and the drive unit.
Description of the differences
The main differences between the C25 Colonoscopy System and the predicate are check valves integrated into the single use colonoscope, an additional decompression line for the rinse tube and the reprocessing regimen for the base unit that eliminates the need to clean and high-level disinfect after each patient. In addition, the C25 colonoscopy system includes minor technology, material, accessory (lubricant), and cosmetic changes compared to the predicate. These minor changes include, for example, an increased deflecting length of the colonoscope; changes to the internal power supply system; additional settings in the graphical user interface that allow the user to change the settings for rinsing function, deflection speed and insufflation gas during, and not only prior to, an examination; a reduction of the safety current of the light source; and the integration of an additional level sensor in the lubrication chambers of the system.
Performance and safety Tests
The following non-clinical testing was performed to validate the design and to assure conformity with design standards:
- IEC 60601-1 "Medical electrical equipment part 1 General requirements for safety" incl. national deviations
- IEC 60601-2-18 "Medical electrical equipment part 2 Particular requirements for the basic safety and essential performance of endoscopic equipment"
7-2
{2}------------------------------------------------
K121582 Page 3 of 3
- IEC 60601-1-2 "Medical electrical equipment part 1-2: General requirements for basic safety and essential performance -Collateral standard: Electromagnetic compatibility -Requirements and tests"
- ISO 10993-5 "Biological evaluation of medical devices -- Part 5: -Tests for in vitro cytotoxicity"
- ISO 10993-10 "Biological evaluation of medical devices -- Part -10: Tests for irritation and skin sensitization"
- -Test of the check valve under worst case conditions
- Evaluation of the decompression line and the corresponding system tests
- -Evaluation of the backflow into the system without any check valves
- -Evaluation of the reprocessing performance
- Evaluation of the performance compared to the predicate device
In all instances, the SC20 colonoscope and the C25 base system functioned as intended, performed as well as or better than the predicate, and met individual test specifications.
Substantial equivalence
The invendo C25 Colonoscopy System has the same intended use, indications for use, functions, operational principles and components as the predicate device.
As a result, the C25 is substantially equivalent to its predicate.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS) in the USA. The logo features a stylized graphic of an eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the graphic.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 20, 2012
invendo medical GmbH % Mr. Oliver v. Ruepprecht Manager OA/RA Peterhofstr. 3b 86438 KISSING GERMANY
Re: K121582
Trade/Device Name: invendo C25 Colonoscopy System Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FDF Dated: November 21, 2012 Received: November 26, 2012
Dear Mr. Ruepprecht:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Oliver v. Ruepprecht
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements, including, but not limited to: registration and 1 sting (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
nvendo medical
THE FUTURE OF ENDOSCOPY
13 Indications for Use Statement
510(k) Number (if known): _ KI ZI 5 82
Device Name: invendo C25Colonoscopy System
Indications for Use:
The invendo C25 Colonoscopy System is intended to provide visualization and diagnostic / therapeutic access to the adult lower gastrointestinal tract (including but not limited to, the anus, rectum, sigmoid colon, colon, cecum and ileocecal valve) for endoscopy and endoscopic surgery.
The colonoscope component of the invendo C25 Colonoscopy System, the SC20 Colonoscope, is a single use disposable device. The SC20 colonoscope cannot be reprocessed.
Prescription Use × (Part 21 CFR 801 Subpart D) AND
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Benjamin R. Hisher -S 2012.12.20 1/3:11:03 -
(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K121582
confidential
13-1
§ 876.1500 Endoscope and accessories.
(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.