K Number
K121320
Manufacturer
Date Cleared
2012-08-10

(100 days)

Product Code
Regulation Number
888.3080
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The C7 Anterior Cervical Intervertebral Fusion Cages are indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level (C2-T1). DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. C7 Anterior Cervical Intervertebral Fusion Cages are used to facilitate intervertebral body fusion in the cervical spine and are placed via an anterior approach and packed with autogenous bone. C7 Anterior Cervical Intervertebral Fusion Cages are to be used with supplemental fixation. Patients should have at least six (6) weeks of non-operative treatment prior to treatment with an intervertebral cage.

Device Description

The Medyssey C7 Anterior Cervical Interbody Fusion Cage system consists of a series of PEEK (polyetheretherketone) cervical interbody spacers of various footprints and thicknesses. The C7 ACIF cages have ridges on their superior and inferior surfaces to prevent migration, and graft windows to allow for bony fusion. Tantalum markers are incorporated into the devices to allow for radiographic visualization of the implants.

AI/ML Overview

The provided document is a 510(k) summary for the Medyssey C7 Anterior Cervical Interbody Fusion Cage. This type of regulatory submission focuses on demonstrating substantial equivalence to a legally marketed predicate device through non-clinical testing. It does not involve studies evaluating the diagnostic accuracy or performance of an AI/ML powered medical device, which would typically involve human readers, ground truth establishment, or training/test sets.

Therefore, many of the requested elements are not applicable to this document. The information provided below reflects what is available in the document.


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Material Equivalence: PEEK with Tantalum markers, trapezoidal shape, central graft containment cavity, ridges to resist expulsion, anterior threaded hole for instrument.The Medyssey C7 Anterior Cervical Interbody Fusion Cage System consists of PEEK (polyetheretherketone) Optima LT1 with tantalum radiographic markers. It is trapezoidal, has a central graft containment cavity, ridges to resist expulsion, and an anterior threaded hole for instrument attachment. The device is manufactured from the same material as the predicate devices and has the same indications for use.
Mechanical Performance (Non-Clinical Tests based on FDA guidance and ASTM F2077 & F2267): - Static Compression - Static Torsion - Static Compressive Shear - Dynamic Compression - Subsidence - ExpulsionNon-clinical tests were performed for: - Static Compression - Static Torsion - Static Compressive Shear - Dynamic Compression - Subsidence - Expulsion Conclusion: Based on the testing and comparison analysis to the predicate devices, Medyssey believes that the subject Medyssey C7 Anterior Cervical Cage System is substantially equivalent to the predicate devices. The minor differences in design were evaluated through testing and do not affect the safety and efficacy of the device for its intended use.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Not applicable. This is a non-clinical testing submission for a physical medical device (interbody fusion cage), not an AI/ML powered device using test sets of data. The "tests" refer to mechanical and material property assessments.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. Ground truth as typically understood for diagnostic AI/ML models is not relevant here. The "ground truth" for a physical device would relate to engineering standards and mechanical properties, assessed by engineers and technicians.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. Adjudication methods are relevant for human interpretation of data, typically in diagnostic studies.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is an interbody fusion cage, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This device is an interbody fusion cage, not an AI algorithm.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

Not applicable. For this device, "ground truth" would be defined by engineering specifications and material science standards.

8. The sample size for the training set

Not applicable. This device is an interbody fusion cage, not an AI algorithm requiring a training set of data.

9. How the ground truth for the training set was established

Not applicable.

{0}------------------------------------------------

K12/320
page 1 of 2

510(k) Summary

AUG 10 2012

Company:Medyssey Co. Ltd.722-3, 4F, Science Tower 4FDongducheon CityGyeonggido, KoreaTel: 82-31-879-0414FAX: 82-31-879-0415
Contact:Michael KvitnitskyChief Operating OfficerMedyssey Co. Ltd8001 N. Lincoln Ave.Suite 401Skokie, IL 60077Tel: 847-982-0100FAX: 888-518-9070
Date Prepared:June 29, 2012
Proprietary Name:Medyssey C7 Anterior Cervical Interbody Fusion Cage
Classification Name:87 ODP- Orthosis, intervertebral body fusion device, cervical 21 CFR888.3080, Class II
Predicate Device:· SpineCraft, LLC ORIO Intervertebral Body Fusion Cages (K090887)· Custom Spine, Inc. Pathway ACIF (K092904)· Biomet Spine C-Thru Anterior Spinal System (K092336)
Product Description:The Medyssey C7 Anterior Cervical Interbody Fusion Cage system consists of aseries of PEEK (polyetheretherketone) cervical interbody spacers of variousfootprints and thicknesses. The C7 ACIF cages have ridges on their superior andinferior surfaces to prevent migration, and graft windows to allow for bony fusion.Tantalum markers are incorporated into the devices to allow for radiographic

Indications for Use:

The C7 Anterior Cervical Intervertebral Fusion Cages are indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level (C2-T1). DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. C7 Anterior Cervical Intervertebral Fusion Cages are used to facilitate intervertebral body fusion in the cervical spine and are placed via an anterior approach and packed with autogenous bone. C7 Anterior Cervical Intervertebral Fusion Cages are to be used with supplemental fixation. Patients should have at least six (6) weeks of non-operative treatment prior to treatment with an intervertebral cage.

visualization of the implants.

{1}------------------------------------------------

K121320
page 2 of 2

Summary of Technological Characteristics:

The Medyssey C7 Anterior Cervical Interbody Fusion Cage System consists of a series of PEEK (polyetheretherketone) Optima LT1 (from Invibio) interbody spacers with tantalum radiographic markers. The C7 ACIF devices are trapezoidal in shape, have a central graft containment cavity, have ridges to resist expulsion and have an anterior threaded hole for instrument. The predicate ORIO ACIF and Pathway ACIF devices are also made from PEEK, with tantalum markers, are trapezoidal in shape, have a central graft containment cavity, ridges to resist expulsion and an anterior threaded hole for instrument attachment.

Medyssey has determined that the minor differences between proposed device and the predicate devices will not impact the safety or effectiveness of the anterior cervical interbody fusion system for its intended use. Analysis has shown that the proposed device is equivalent to the predicate devices.

Identification of Legally Marketed Predicate Device:

Documentation was provided, which demonstrates that the subject Medyssey C7 Anterior Cervical Cage is substantially equivalent to the predicate devices the SpineCraft, LLC ORIO Intervertebral Body Fusion Cages (K090887), Custom Spine, Inc. Pathway ACIF (K092904) and Biomet Spine C-Thru Anterior Spinal System (K092336). The proposed Medyssey device has the same indications for use and is manufactured from the same material. The minor differences in the design were evaluated through testing and do not affect the safety and efficacy of the device for its intended use.

Brief Discussion of Non-Clinical Tests Submitted:

Numerous tests were performed on the Medyssey Anterior Cervical Interbody Fusion System. The tests performed are recommended by the FDA guidance document for intervertebral body fusion devices, including ASTM F2077 and ASTM F2267. List of Tests is below:

  • Static Compression .
  • Static Torsion .
  • Static Compressive Shear .
  • Dynamic Compression �
  • Subsidence .
  • . Expulsion

Conclusions from Non-Clinical Tests:

Based on the testing and comparison analysis to the predicate devices provided in this premarket notification submission, Medyssey believes that the subject Medyssey C7 Anterior Cervical Cage System is substantially equivalent to the SpineCraft, LLC ORIO Intervertebral Body Fusion Cages (K090887), Custom Spine, Inc. Pathway ACIF (K092904) and Biomet Spine C-Thru Anterior Spinal System (K092336)

N

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle. The logo is black and white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

AUG 1 0 2012

Medyssey Company Limited % Mr. Michael Kvitnitsky Chief Operating Officer 8001 North Lincoln Avenue, Suite 401 Skokie, Illinois 60077

Re: K121320

Trade/Device Name: Medyssey C7 Anterior Cervical Interbody Fusion Cages Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: II Product Code: ODP Dated: June 30, 2012 Received: July 23, 2012

Dear Mr. Kvitnitsky:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{3}------------------------------------------------

Page 2 - Mr. Michael Kvitnitsky

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21-GFR Part 820); and if-applicable, the electronicproduct radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson

Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Form

K121320

510(k) Number (if known):

Device Name: Medyssey C7 Anterior Cervical Interbody Fusion Cages

Indications for Use:

The C7 Anterior Cervical Intervertebral Fusion Cages are indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level (C2-T1). DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. C7 Anterior Cervical Intervertebral Fusion Cages are used to facilitate intervertebral body fusion in the cervical spine and are placed via an anterior approach and packed with autogenous bone. C7 Anterior Cervical Intervertebral Fusion Cages are to be used with supplemental fixation. Patients should have at least six (6) weeks of non-operative treatment prior to treatment with an intervertebral cage.

Prescription Use × (Per 21 CFR 801.109). OR

Over-The-Counter Use (Optional Format 1-2-96)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Dision of Surgical, Orthopedic, and Restorative Devices

Page 1 of 1

KI21320 510(k) Number_

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.