K Number
K121157
Date Cleared
2012-08-02

(108 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)that meets all of the requirements of ASTM standard D 5250-06.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the "Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)" based on the provided text, structured according to your requested points:


Acceptance Criteria and Device Performance Study

The submission details the non-clinical testing performed to demonstrate that the Powder Free Vinyl Patient Examination Gloves meet predefined acceptance criteria and are safe and effective.

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicsAcceptance Criteria (Standard)Reported Device Performance
DimensionASTM standard D 5250-06Meets
Physical PropertiesASTM standard D 5250-06Meets
Freedom from pinholes21 CFR 800.20Meets
Powder ResidualASTM standard D 5250-06 and D6124-06Meets (<2mg/glove)
Biocompatibility: Primary Skin IrritationPrimary Skin Irritation in rabbitsPasses (Not a Primary Skin Irritation)
Biocompatibility: Dermal SensitizationDermal sensitization in the guinea pigPasses (Not a Dermal sensitization)

2. Sample size used for the test set and the data provenance

The document does not explicitly state the specific sample sizes used for each individual test (Dimension, Physical Properties, Freedom from pinholes, Powder Residual, Biocompatibility). The tests are described as conforming to established standards (ASTM D 5250-06, ASTM D6124-06, 21 CFR 800.20, and ISO10993-10), which would typically specify sample sizes.

Data Provenance: The document does not specify the country of origin of the data or whether it was retrospective or prospective. It refers to standard tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided in the document. For these types of device performance tests (physical and biocompatibility), "ground truth" is typically established by the results of the standardized tests themselves, performed by qualified laboratory personnel, rather than expert consensus on interpretive data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable and therefore not provided. Adjudication methods like 2+1 or 3+1 are typically used in studies involving subjective interpretation (e.g., radiology image readings) where multiple experts initially assess a case, and discrepancies are resolved by an additional expert or group. The reported tests are objective measurements and standardized assessments.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study was done. This type of study is relevant for AI-powered diagnostic tools where human readers (e.g., radiologists) interpret images or data. These gloves are a Class I medical device and do not involve AI or human interpretation in their intended use.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No standalone algorithm performance study was done. This device is a physical medical glove and does not involve any algorithm or AI component.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for the tests is defined by the objective criteria and methodologies outlined in the referenced ASTM standards, 21 CFR regulations, and ISO standard. For example:

  • Dimension and Physical Properties: Direct measurements and physical testing according to the ASTM D 5250-06 standard.
  • Freedom from pinholes: Water leak test as per 21 CFR 800.20, with results compared against AQL (Acceptable Quality Limit) levels.
  • Powder Residual: Measurement of powder content according to ASTM D 5250-06 and D6124-06.
  • Biocompatibility: Results from standardized animal model tests (rabbits for skin irritation, guinea pigs for dermal sensitization) according to ISO10993-10 protocols.

8. The sample size for the training set

Not applicable. This device is a physical product and does not involve machine learning or AI, and therefore no "training set" is used.

9. How the ground truth for the training set was established

Not applicable. As there is no training set, there is no ground truth established for one.

{0}------------------------------------------------

510(k) Summary

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92(c)."

"The assigned 510(k) number is: K121 | 57 ." (applicant leave blank)

Premarket Notification [510(k)] Summary

((a)(1)). The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared :

Submitter's name :SHIJIAZHUANG JIUYUAN PLASTIC CO., LTD
Submitter's address :XIBAIPO INDUSTRY PARK, PINGSHAN COUNTY,SHIJIAZHUANG, HEBEI, CHINA
Phone number :(86) 311-67699916
Fax number :(86) 311-67699906
Name of contact person:Miss. Gui Suo Guo
Date the summary was prepared:Apr. 1 th, 2012

[(a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known

Device Name:Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)
Proprietary/Trade name:Powder Free Vinyl Patient Examination Gloves
Common Name:Patient examination glove
Classification Name:Patient examination glove
Device Classification:I
Regulation Number:21 CFR 880.6250
Panel:General Hospital (80)
Product Code:LYZ

[(a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence .

Class I* Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) that meets all of the requirements of ASTM standard D 5250-06.

Predicate device : FUGUAN (Brand) Powder-Free Vinyl Patient Examination Gloves, Shijiazhuang Fuguan Plastic Products Co., Ltd., K032908 .

[(a)(4)] A description of the device

Device Description : Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)that meets all of the requirements of ASTM standard D 5250-06.

{1}------------------------------------------------

[(a)(5)] The summary describes the intended use of the device

Device Intended Use: Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

((a)(6)) A.summary of the technological characteristics of new device compared to the predicate device.

The powder free vinyl patient examination gloves, non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

CharacteristicsStandardDevice performance
DimensionASTM standard D 5250-06Meets
Physical PropertiesASTM standard D 5250-06Meets
Freedom from pinholes21 CFR 800.20Meets
Powder ResidualASTM standard D 5250-06Meets
and D6124-06<2mg/glove
BiocompatabilityPrimary Skin Irritation inrabbitsPasses
Not a Primary Skin Irritation
Dermal sensitization in theguinea pigPasses
Not a Dermal sensitization

[(b)(1)] A brief discussion of the non-clinical testing data that was submitted, referenced or relied on to demonstrate that the Subject Device is safe and effective, and whose performance meets the requirements of its pre-defined acceptance criterie and intended uses.

Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)meet requirements per ASTM D5250-06, per ASTM D6124-06, per 21 CFR 800.20 and ISO10993-10. , it is safe and effective and it's performance meets the requirements of its pre-defined acceptance criteria and intended uses.

[(b)(2)] A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

((b)(3)] The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe,as effective, and performed as well or better than the legally marketed device identified in (a)(3).

It can be concluded that the Powder Free Vinyl Patient Examination Gloves meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL, meet labeling claims . It is as safe,as effective, and performed as well the legally marketed device identified in (a)(3).

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

2 2012 AUG

Shijazihuang Jiuyuan Plastic Company, Limited C/O Mr. Chu Xiaoan Room 1606 Building 1 Jianxiang Yuan No. 209 Bei Si Huan Zhong Road Haidian District. Beijing 100083 P.R China

Re: K121157

Trade/Device Name: Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: June 20, 2012 Received: June 27, 2012

Dear Mr. Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Mr. Xiaoan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Tylkman V, An

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE

Applicant: SHIJIAZHUANG JIUYUAN PLASTIC CO., LTD

510(k) Number (if known):_* KI21157

Device Name: Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)

Indications For Use:

Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Hanne Smith

(Division Sign-Off) Division of Anesthesiology, General Hospital ിശ്വാസ Control, Dental Devices

10(k) Number: ﺭ

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.