(263 days)
The PLEX-ID Flu assay is a qualitative nucleic acid in vitro diagnostic test intended for the detection and differentiation of influenza A HINI (2009), influenza A H1N1 (seasonal), influenza A H3N2 (seasonal) and influenza B viral nucleic acids in nasopharyngeal swab specimens from patients symptomatic for respiratory tract infection. The PLEX-ID Flu assay is intended for use on the PLEX-ID System (version 1.2) as an aid in the diagnosis of influenza infection in conjunction with clinical and epidemiological information. This assay is not intended to detect influenza C virus.
Negative results do not preclude influenza virus infection and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions,
If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
The PLEX-ID Flu assay consists of the following kits:
- . PLEX-ID Flu Amplification Kit (List No. 05N21-91)
- . PLEX-ID Flu Control Kit (List No. 05N21-80)
The PLEX-ID Flu assay is a qualitative in vitro diagnostic test used for the detection and identification of influenza A and B viruses directly from human samples. First, nucleic acids are extracted from samples. Nucleic acids are then amplified via a reverse transcription polymerase chain reaction (RT-PCR). The PCR products are subsequently desalted and analyzed in a mass spectrometer to determine the base composition of the PCR products. Analysis of the base composition of PCR products is used to determine identity by comparison to a proprietary database. Desalting, mass spectrometry and data analysis are conducted on the PLEX-ID Analyzer. Reported results include an identification of influenza virus species and subtypes of influenza A virus.
Here's a breakdown of the acceptance criteria and the study details for the PLEX-ID Flu assay, based on the provided text:
1. Acceptance Criteria and Reported Device Performance
The document does not explicitly state pre-defined acceptance criteria in terms of specific sensitivity and specificity thresholds. However, the reported performance metrics (Positive Percent Agreement - PPA and Negative Percent Agreement - NPA) serve as de-facto acceptance criteria demonstrated by the device. The study design aims to show a high level of agreement with a comparator device.
| Target | Performance Metric | Reported Performance (Prospective Study) | Reported Performance (Retrospective Study) |
|---|---|---|---|
| Influenza A H1N1 (2009) | Positive % Agreement | 100.0% (95% CI: 89.3%, 100.0%) (32/32) | 98.2% (95% CI: 96.7%, 99.0%) (538/548) |
| Influenza A H1N1 (2009) | Negative % Agreement | 99.6% (95% CI: 99.1%, 99.8%) (1239/1244) | 94.3% (95% CI: 92.5%, 95.7%) (746/791) |
| Influenza A H1N1 (seasonal) | Positive % Agreement | 57.1% (95% CI: 25.0%, 84.2%) (4/7) | 96.4% (95% CI: 82.3%, 99.4%) (27/28) |
| Influenza A H1N1 (seasonal) | Negative % Agreement | 100.0% (95% CI: 99.7%, 100.0%) (1269/1269) | 100.0% (95% CI: 99.7%, 100.0%) (1311/1311) |
| Influenza A H3N2 (seasonal) | Positive % Agreement | 98.0% (95% CI: 89.3%, 99.6%) (48/49) | 100.0% (95% CI: 34.2%, 100.0%) (2/2) |
| Influenza A H3N2 (seasonal) | Negative % Agreement | 99.8% (95% CI: 99.4%, 100.0%) (1225/1227) | 100.0% (95% CI: 99.7%, 100.0%) (1337/1337) |
| Influenza B | Positive % Agreement | 100.0% (95% CI: 85.1%, 100.0%) (22/22) | 92.9% (95% CI: 77.4%, 98.0%) (26/28) |
| Influenza B | Negative % Agreement | 99.8% (95% CI: 99.4%, 100.0%) (1252/1254) | 99.9% (95% CI: 99.6%, 100.0%) (1312/1313) |
2. Sample Size Used for the Test Set and Data Provenance
- Prospective Study:
- Sample Size: 1276 unique nasopharyngeal swab specimens were included in the final analysis (starting from 1287 collected, with exclusions).
- Data Provenance: Prospectively collected from two U.S. clinical sites and three specimen suppliers from September 2009 through June 2010.
- Retrospective Study:
- Sample Size: 1339 unique specimens were included in the final analysis (starting from 1378 collected, with exclusions).
- Data Provenance: Banked specimens collected in February 2008, January 2009 through March 2009, and leftover archived specimens from April 2009 through December 2009. The country of origin for these specimens is not explicitly stated but can be inferred to be largely U.S. given the U.S. clinical sites for prospective collection.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
The ground truth was established using FDA-cleared nucleic acid assays as comparators, not human experts.
- Comparator 1: An FDA-cleared assay for the detection and discrimination of influenza A virus, influenza B virus, and Respiratory Syncytial virus nucleic acids.
- Comparator 2: An FDA-cleared assay for the detection and discrimination of influenza A virus subtypes [influenza A H1N1 (2009), influenza A H1N1 (seasonal), influenza A H3N2 (seasonal)].
Therefore, the concept of "number of experts" and their "qualifications" for ground truth establishment is not applicable in this context, as the ground truth was based on established laboratory methods.
4. Adjudication Method for the Test Set
For discrepant samples between Comparator 1 and Comparator 2, Comparator 1 vs. PLEX-ID Flu, and Comparator 2 vs. PLEX-ID Flu, PCR followed by bi-directional sequencing was used as an adjudication method. However, the results from bi-directional sequence testing were not used to calculate the Positive Percent Agreement and Negative Percent Agreement of the PLEX-ID Flu assay presented in the tables. This implies that the primary agreement calculations were against the Comparator assays, with sequencing used for further investigation of discrepancies but not as the sole ground truth for the primary performance metrics.
5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study
No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not done. This study is for an in vitro diagnostic (IVD) assay designed to detect viral nucleic acids, which does not involve human interpretation in the same way an imaging AI device would.
6. Standalone (i.e. algorithm only without human-in-the loop performance) Study
Yes, this was a standalone performance study. The PLEX-ID Flu assay automatically detects and identifies influenza viral nucleic acids based on the mass spectrometry analysis of PCR products, comparing base composition to a proprietary database. The results are reported directly by the system without human interpretation as part of the core diagnostic function. The "human-in-the-loop" concept, as typically applied to AI in imaging, is not relevant here; the device itself performs the analysis.
7. The Type of Ground Truth Used
The primary ground truth for the clinical performance evaluation was based on results from FDA-cleared comparator nucleic acid assays. For samples with discordant results between the PLEX-ID Flu assay and the comparator assays, PCR followed by bi-directional sequencing was used as an additional, higher-resolution method for further investigation and characterization of the discrepant sample's true status. However, as noted in point 4, the primary agreement metrics were calculated against the comparator assays.
8. The Sample Size for the Training Set
The document does not explicitly state the sample size for a training set. This is common for IVD assays where "training" might refer to internal assay development and optimization rather than a distinct, separate training set for a machine learning algorithm as would be described for AI devices in imaging or other fields. The performance characteristics sections focus on validation against clinical and contrived samples.
9. How the Ground Truth for the Training Set Was Established
Since a specific "training set" with established ground truth is not explicitly described in the context of an AI training paradigm, this information is not applicable in the conventional sense. The assay's "training" or development would involve molecular biology techniques, primer design, and database development, likely drawing upon known influenza sequences and characterized viral samples, rather than a separate dataset for explicit AI algorithm training. The proprietary database against which base compositions are compared would have been developed using known viral sequences and characteristics.
{0}------------------------------------------------
5.0 510(k) Summary
510(k) Number: K121003
Device Name: PLEX-ID Flu
Purpose of the Submission: The purpose of this 510(k) is to gain clearance to market the PLEX-ID Flu (List No. 05N21) assay.
5.1 Official Correspondent to the File
Darren Clarke Name Director Regulatory Affairs Title: Telephone: (760) 476-3224 Fax: (760) 476-3301 Email: darren.clarke@abbott.com
Name Samuel Mathew Title: Associate Director of Regulatory Affairs Telephone: (224) 361-7062 Fax: (847) 775-6777 Email: samuel.mathew@abbott.com
Address: Abbott Molecular Inc. 1300 E. Touhy Avenue Des Plaines, IL 60018
5.2 Date of Preparation
March 30, 2012
5.3 Sponsor
Abbott Molecular Inc. is sponsor of the PLEX-ID Flu (List No. 05N21) 510(k) submission.
5.4 Intended Use
The PLEX-ID Flu assay is a qualitative nucleic acid in vitro diagnostic test intended for the detection and differentiation of influenza A HINI (2009), influenza A H1N1 (seasonal), influenza A H3N2 (seasonal) and influenza B viral nucleic acids in nasopharyngeal swab specimens from patients symptomatic for respiratory tract infection. The PLEX-ID Flu assay is intended for use on the PLEX-ID System (version 1.2) as an aid in the diagnosis of influenza infection in conjunction with clinical and epidemiological information. This assay is not intended to detect influenza C virus.
PLEX-ID Flu 510(k) K121003 December 2012
510(k) Summary Page 1 of 31
DEC 2 1 2012
{1}------------------------------------------------
Negative results do not preclude influenza virus infection and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions,
If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
ર્સ્ડ Trade Name
PLEX-ID Flu (List No. 05N21)
5.6 Common Name
Nucleic acid assay for the detection of influenza virus
5.7 Classification Name and Regulation
- . Regulation Number: 21 CFR 866.3980
- Classification Name: Respiratory viral panel multiplex nucleic acid assay .
5.8 Predicate Devices
{2}------------------------------------------------
5.9 Device Description
The PLEX-ID Flu assay consists of the following kits:
- . PLEX-ID Flu Amplification Kit (List No. 05N21-91)
- . PLEX-ID Flu Control Kit (List No. 05N21-80)
The PLEX-ID Flu assay is a qualitative in vitro diagnostic test used for the detection and identification of influenza A and B viruses directly from human samples. First, nucleic acids are extracted from samples. Nucleic acids are then amplified via a reverse transcription polymerase chain reaction (RT-PCR). The PCR products are subsequently desalted and analyzed in a mass spectrometer to determine the base composition of the PCR products. Analysis of the base composition of PCR products is used to determine identity by comparison to a proprietary database. Desalting, mass spectrometry and data analysis are conducted on the PLEX-ID Analyzer. Reported results include an identification of influenza virus species and subtypes of influenza A virus.
5.10 Background on Influenza
Influenza is a respiratory virus that can cause mild to severe disease, and at times can lead to death. Every year in the United States, on average 5% to 20% of the population gets the flu; more than 200,000 people are hospitalized from flu complications, and about 3,000 to 49,000 people die from flu-related causes. Some people, such as adults 65 years or older, young children, and people with certain health conditions, are at high risk for serious flu complications.1,2
There are two main types of influenza (flu) virus, A and B, which are responsible for seasonal epidemics of flu amongst humans. Influenza A viruses are divided into subtypes based on two proteins on the surface of the virus: hemagglutinin (H) and neuraminidase (N). The common Influenza A virus subtypes found in people are A (HINI) and A (H3N2). In April of 2009, a new strain of Influenza A emerged. On June 11, 2009 WHO classified the 2009 H1N1 Influenza A strain as the cause of a phase 6 pandemic.3
PLEX-ID Flu 510(k) K121003 December 2012
{3}------------------------------------------------
5.11 Technological Characteristics of the Device as Compared to the Predicate
These devices are similar in that they are designed to prepare nucleic acids for amplification, amplify specific influenza sequences, detect the amplified products, and report qualitative results.
The primary similarities and differences between the PLEX-ID Flu assay and the predicate devices are shown in Table 1.
{4}------------------------------------------------
Table 1
Similarities and Differences Between PLEX-ID Flu and Predicate Device
| Current Application | Predicate Devices | Feature | Current Application | Predicate Devices | ||
|---|---|---|---|---|---|---|
| PLEX-ID Flu | Prodesse ProFLU+ | Prodesse ProFAST+ | Assay Type | PLEX-ID Flu | Prodesse ProFLU+ | Prodesse ProFAST+ |
| K121003 | K073029 | K101855 | Analyte Targets | Qualitative influenza genus→ Primer pair 2798 targets polymerase PB1 influenza A → Primer Pair 1266 targets nucleoprotein influenza A → Primer Pair 1279 targets matrix protein influenza A → Primer Pair 1287 polymerase A protein influenza A → Primer Pair 2775 non-structural protein 1, NS1 influenza A → Primer Pair 1259 targets polymerase, PB influenza A H1N1 (2009) → Primer Pair 5101 targets hemagglutinin, HA influenza A H1N1 (2009) → Primer Pair 4998 targets neuraminidase, NA influenza B→ Primer Pair 1261 targets polymerase, PB2 | Qualitative Influenza A Virus → Matrix Respiratory Syncytial Virus A →Polymerase Respiratory Syncytial Virus B →Polymerase Influenza B Virus → Non-structural NS1 and NS2 | Qualitative Seasonal H1 Influenza A → Hemagglutinin Seasonal H3 Influenza A → Hemagglutinin 2009 H1N1 Influenza Virus → Hemagglutinin |
| 21 CFR 866.3980OEP, OCC, OQW, OTA | 21 CFR 866.3980OCC | 21 CFR 866.3332OQW | Input Sample Types | nasopharyngeal swab | nasopharyngeal swab | nasopharyngeal swab |
| Feature | Current Application | Predicate Devices | ||||
| Intended Use | PLEX-ID FluThe PLEX-ID Flu assay is a qualitative nucleic acid in vitro diagnostic test intended for the detection and differentiation of influenza A H1N1 (2009), influenza A H1N1 (seasonal), influenza A H3N2 (seasonal) and influenza B viral nucleic acids in nasopharyngeal swab specimens from patients symptomatic for respiratory tract infection. The PLEX-ID Flu assay is intended for use on the PLEX-ID System (version 1.2) as an aid in the diagnosis of influenza infection in conjunction with clinical and epidemiological information. This assay is not intended to detect influenza C virus. Negative results do not preclude influenza virus infection and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens. | Prodesse ProFLU+The ProFlu™+ Assay is a multiplex Real-Time PCR (RT-PCR) in vitro diagnostic test for the rapid and qualitative detection and discrimination of Influenza A Virus, Influenza B Virus, and Respiratory Syncytial Virus (RSV) nucleic acids isolated and purified from nasopharyngeal (NP) swab specimens obtained from symptomatic patients. This test is intended for use to aid in the differential diagnosis of Influenza A, Influenza B and RSV viral infections in humans and is not intended to detect Influenza C. Negative results do not preclude influenza or RSV virus infection and should not be used as the sole basis for treatment or other management decisions. It is recommended that negative RSV results be confirmed by culture. Performance characteristics for Influenza A Virus were established when Influenza A/H3 and A/H1 were the predominant Influenza A viruses in circulation. When other Influenza A viruses are emerging, performance characteristics may vary. If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens. | Prodesse ProFAST+The ProFAST™+ Assay is a multiplex Real Time RT-PCR in vitro diagnostic test for the qualitative detection and discrimination of seasonal Influenza A/H1, seasonal Influenza A/H3 and 2009 H1N1 Influenza viral nucleic acids isolated and purified from nasopharyngeal (NP) swab specimens from human patients with signs and symptoms of respiratory infection in conjunction with clinical and epidemiological risk factors. This assay targets conserved regions of the Hemagglutinin (HA) gene for seasonal Influenza A/H1, seasonal Influenza A/H3 and 2009 H1N1 Influenza Virus, respectively. This assay is not intended to detect Influenza B or Influenza C Viruses. A negative ProFAST+ Assay result is a presumptive negative result for Influenza A. These results should be confirmed by an FDA cleared nucleic acid-based test (NAT) detecting Influenza A. Negative results do not preclude Influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions. If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens. |
.
510(k) Summary
:
.....
PLEX-ID Flu 510(k) K12
December 2012
{5}------------------------------------------------
.
:
PLEX-ID Flu 510
December 2012
Page 6 of 31
{6}------------------------------------------------
:
PLEX-ID (K) KT21003
December 2012
510(k) Summary
. T
.
.
.
.
{7}------------------------------------------------
| Feature | Current ApplicationPLEX-ID Flu | Prodesse ProFLU+ | Predicate DevicesProdesse ProFAST+ |
|---|---|---|---|
| SampleCollection | Immediately following collection ofthe sample from the patient, place theswab into a tube containing 3 mL ofviral transport media. Place on ice orrefrigerate. Samples for the PLEX-IDFlu assay may be collected in: Micro TestTM M4R ViralTransport Medium Micro TestTM M5R ViralTransport Medium Micro TestTM M6R ViralTransport Medium Micro TestTM M4RTR ViralTransport Medium Copan Universal TransportMedium BD Universal Viral TransportMedium | Following collection, place the swab intoa tube containing 2 to 3 mL of viraltransport medium (Remel M4, M4RT,M5, M6; Copan UTM; or BectonDickenson UVT). | Following collection, place the swab into atube containing 2 to 3 mL of viral transportmedium (Remel M4, M4RT, M5, M6; CopanUTM; or Becton Dickenson UVT). When starting from purified nucleic acidsamples that have been previously processedfor testing with the ProFLU+ Assay, beginwith the RT-PCR reaction. |
| Input SampleVolume | 300 µL | 180 µL | 180 µL |
PLEX-ID December 2012
·
.
510(k) Summary
:
·
.
"
{8}------------------------------------------------
| Feature | Current ApplicationPLEX-ID Flu | Predicate DevicesProdesse ProFLU+ | Predicate DevicesProdesse ProFAST+ | |||
|---|---|---|---|---|---|---|
| Principlesof theProcedure | Automated sample to extraction plate Automated nucleic acid extraction Automated transfer of sample eluate to PCR Plate Automated PCR amplification Automated desalting of PCR products Mass Spectrometry analysis converts mass of PCR products to base composition Base composition signature from multiple PCR products is then used to identify the influenza virus species and subtype | Automated nucleic acid isolation Manual transfer of sample eluate to PCR Plate Automated PCR amplification Optical detection of stimulated fluorescence The fluorescence reader monitors real-time fluorescence during every PCR amplification cycle | Automated nucleic acid isolation Manual transfer of sample eluate to PCR Plate Automated PCR amplification Optical detection of stimulated fluorescence The fluorescence reader monitors real-time fluorescence during every PCR amplification cycle | Nucleic Acid isolation instrument Amplification and detection instrument | Automated nucleic acid isolation instrument (MagNA Pure LC System using the Total Nucleic Acid Isolation (TNAI) Kit or NucliSENS easyMAG System using the Automated Magnetic Extraction Reagents) | Automated PCR amplification and detection by fluorescence analysis (Cepheid SmartCycler II) |
| InstrumentationPrincipleSystemComponents | The PLEX-ID System integrates sample preparation (PLEX-ID SP, PLEX-ID FH) with PCR amplification (PLEX-ID TC) followed by mass spectrometry analysis to generate assay results (PLEX-ID Analyzer). | Nucleic Acid isolation instrument Amplification and detection instrument | Automated nucleic acid isolation instrument (MagNA Pure LC System using the Total Nucleic Acid Isolation (TNAI) Kit or NucliSENS easyMAG System using the Automated Magnetic Extraction Reagents) | Automated PCR amplification and detection by fluorescence analysis (Cepheid SmartCycler II) | ||
| SamplePreparationInstrumentComponents | Liquid handling and robotic manipulation platform (PLEX-ID SP, PLEX-ID FH) using the PLEX-ID Viral RNA Isolation Kit | |||||
| AmplificationandDetectionInstrument | Automated PCR amplification (PLEX-ID TC) is followed by mass spectrometry analysis and influenza virus identification (PLEX-ID |
510(k) Summary
.
{9}------------------------------------------------
| Feature | Current Application | Predicate Devices | |
|---|---|---|---|
| AmplificationControls | PLEX-ID Flu Each PCR reaction well contains an Amplification Control that consists of a synthetic RNA molecule specific to the primer pairs used, but distinguishable from known target sequences. The Amplification Control serves as the internal PCR control and provides scaling information for the level of influenza RNA present in the sample. It also detects PCR inhibition, reagent failure, and process errors. | Prodesse ProFLU+ Internal Control to detect PCR inhibition in individual samples and Reagent failure or process error Prodesse ProFAST+ Internal Control to detect PCR inhibition in individual samples and Reagent failure or process error | |
| AmplificationEnzymeType(s) | DNA Polymerase Reverse Transcriptase | Taq DNA Polymerase Reverse Transcriptase Taq DNA Polymerase Reverse Transcriptase | |
| DetectionProcedure | Mass Spectrometry analysis converts mass of PCR products to base composition Base composition signature from multiple PCR products is then used to identify the influenza virus species and subtype | Optical detection of stimulated fluorescence. The fluorescence reader monitors real-time fluorescence during every PCR amplification cycle. Optical detection of stimulated fluorescence. The fluorescence reader monitors real-time fluorescence during every PCR amplification cycle. | |
| Feature | Current ApplicationPLEX-ID Flu | Predicate DevicesProdesse ProFLU+ | Predicate DevicesProdesse ProFAST+ |
| Limit ofDetection(LOD) | Influenza A H1N1 (2009) (Strain A/SwineNY/02/2009 H1N1) – $1.3 × 10^{-1}$ TCID50/mL Influenza A H1N1 (seasonal) (Strain A/New Caledonia/20/99 H1N1) – 4.6 TCID50/mL Influenza A H3N2 (seasonal) (Strain A/Wisconsin/67/05 H3N2) – $3.3 × 10^{2}$ TCID50/mL Seasonal Influenza B (Strain B/Panama/45/90) – 15 TCID50/mL | Influenza A/Port Chalmers/1/73 (H3N2) – $10^{1}$ TCID50/mL Influenza A/CA/7/04 (H3N2) – $10^{0}$ TCID50/mL Influenza A/New Caledonia/12/99 (H1N1) – $10^{2}$ TCID50/mL Influenza A/WS/33 (H1N1) – $10^{0}$ TCID50/mL Influenza B/Lee/40 – $10^{1}$ TCID50/mL Influenza B/Wisconsin/2/06 – $10^{0}$ TCID50/mL | H1N1 A/Virginia/1/06 – $5 × 10^{-1}$ TCID50/mL H1N1 A/Hong Kong/2652/06 – $5 × 10^{-2}$ TCID50/mL H3N2 A/Anhui/1239/05 – $1 × 10^{1}$ TCID50/mL H3N2 A/California/07/04 – $5 × 10^{1}$ TCID50/mL 2009 H1N1 Clinical Isolate #1 – $1 × 10^{2}$ TCID50/mL 2009 H1N1 Clinical Isolate #5 – $1 × 10^{2}$ TCID50/mL |
| AssayControls | Negative Control Positive Control | Negative Control Positive Control Extraction Control | Negative Control Positive Control Extraction Control |
| ResultsReporting | The PLEX-ID Flu assay reports results for both influenza A and influenza B in separate sections of the report. The influenza A report can show the following under Detected Microbe: 2009 (Pandemic) H1N1 Influenza A Seasonal H1N1 Influenza A Seasonal H3N2 Influenza A Other influenza A Not Detected The influenza B report can show the following under Detected Microbe: Seasonal Influenza B Not detected | Influenza A result of either "POS" for positive or "NEG" for negative for influenza A Influenza B result of either "POS" for positive or "NEG" for negative for influenza B "Unresolved" – PCR inhibition or reagent failure. | H1 result of either "POS" for positive or "NEG" for negative for Seasonal Influenza A/H1 H3 result of either "POS" for positive or "NEG" for negative for Seasonal Influenza A/H3 2009 H1N1 result of either "POS" for positive or "NEG" for negative for 2009 H1N1 Influenza RNA "Unresolved" – PCR inhibition or reagent failure. |
·
510(k) Summary
Page 10 of 31
.
·
:
·
・
. T
·
:
:
{10}------------------------------------------------
・
:
:
:
・
510(k) Summary
Page 11 of 31
1
{11}------------------------------------------------
| Feature | Current ApplicationPLEX-ID Flu | Prodesse ProFLU+ | Predicate DevicesProdesse ProFAST+ |
|---|---|---|---|
| Drug /CompoundInterference | Samples containing FluMist® may cause false positive results in the assay. FluMist is an intranasal vaccine that contains live attenuated influenza virus that is detected by the PLEX-ID Flu assay. No interference was detected from other common cold medicines or mucin | An interference study evaluating potentially interfering common cold medications was not performed. | The ProFAST+ assay was not affected by the presence of a panel of endogenous and exogenous potential RT-PCR inhibitors with the exception of mucin. Results indicate that mucin at a concentration of 5% and 1% w/v added to NP swab matrix, resulting in a highly viscous sample, had an inhibitory effect on detection of influenza A subtypes and the Internal RNA Control (IC) near the LoD using the ProFAST+ Assay. |
| Reagents andControlsStorageConditions | Frozen (-10°C or colder) | - 70°C or colder | - 70°C or colder |
510(k) Summary
Page 12 of 31
:
·
. ·
.
·
{12}------------------------------------------------
5.12 Performance Characteristics
5.12.1 Expected Values
The clinical performance of the PLEX-ID Flu assay on the PLEX-ID System was established using nasopharyngeal (NP) swab specimens prospectively collected from two U.S. clinical sites and three specimen suppliers from September 2009 through June 2010. A total of 1287 prospective NP swab specimens were collected for the PLEX-ID Flu clinical study. Of the 1287, five specimens were excluded: duplicate specimens were collected from two subjects (the second specimen from each subject was excluded from the analysis), one specimen from a subject who did not meet the inclusion criteria, and two specimens had an invalid PLEX-ID Flu result, for a total of 1282 included specimens. The prevalence of influenza A 2009 H1N1 decreased significantly during the collection period, consequently, there were only 37 positive influenza A 2009 H1N1, 4 positive influenza A HINI (seasonal), 50 positive influenza A H3N2 (seasonal), and 24 positive influenza B specimens among the 1282 specimens. Prevalence rates and demographics are presented in Table 2 and Table 3, respectively, for four of the five sites that provided prospectively collected specimens. No demographic information was available from supplier Site 3 (N=199) and one subject from clinical Site 1,
PLEX-ID Flu 510(k) K121003 December 2012
510(k) Summary Page 13 of 31
{13}------------------------------------------------
Table 2
Influenza Prevalence Rate - Prospectively Collected Specimens
| Number of Positives by the PLEX-ID Flu Assay(Observed Prevalence Rate) | |||||
|---|---|---|---|---|---|
| Age Group | N | 2009 H1N1 | SeasonalH1N1 | SeasonalH3N2 | Influenza B |
| < 2 years | 202 | 3 (1.5%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) |
| 2 to 5 years | 46 | 1 (2.2%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) |
| 6 to 11 years | 20 | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) |
| 12 to 18 years | 49 | 1 (2.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) |
| 19 to 64 years | 730 | 18 (2.5%) | 0 (0.0%) | 0 (0.0%) | 1 (0.1%) |
| > 65 years | 35 | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) |
| Unknown* | 200 | 14 (7.0%) | 4 (2.0%) | 50 (25.0%) | 23 (11.5%) |
| Total | 1282 | 37 (2.9%) | 4 (0.3%) | 50 (3.9%) | 24 (1.9%) |
- Demographic information was unavailable.
Table 3
Subject Demographics - Prospectively Collected Specimens
| Gender/Age | Number of Subjects (%) |
|---|---|
| Female | 603 (47.0%) |
| Male | 479 (37.4%) |
| Unknown* | 200 (15.6%) |
| Total | 1282 (100.0%) |
- Demographic information was unavailable.
5.12.2 Clinical Performance
5.12.2.1 Prospective Clinical Study
Clinical performance of the PLEX-ID Flu assay was assessed by testing nasopharyngeal (NP) swab specimens from subjects presenting with symptoms of influenza-like illness.
Prospectively collected clinical specimens were tested at three sites. All specimens were
tested with an assay FDA cleared for the detection and discrimination of influenza A
virus, influenza B virus, and Respiratory Syncytial virus nucleic acids isolated and PLEX-ID Flu 510(k) K121003 510(k) Summary December 2012 Page 14 of 31
{14}------------------------------------------------
purified from NP swab specimens obtained from symptomatic patients (Comparator 1). Specimens with a positive influenza A test result went on for further testing with Comparator 2, an assay FDA cleared for the detection and discrimination of influenza A virus subtypes [influenza A HINI (2009), influenza A H1NI (seasonal), influenza A H3N2 (seasonal)] from nucleic acids isolated and purified from NP swab specimens. Aliquots from the same specimens were tested with the PLEX-ID Flu assay. For each subtype of influenza A [H]N] (2009), H1N1 (seasonal), H3N2 (seasonal)], the point estimate for positive percent agreement (PPA) and negative percent agreement (NPA) was calculated. PCR followed by bi-directional sequencing was used to test discrepant samples between Comparator 1 and Comparator 2, Comparator 1 vs. PLEX-ID Flu. and Comparator 2 vs. PLEX-ID Flu. The results from bi-directional sequence testing were not used to calculate the positive percent agreement and negative percent agreement of the PLEX-ID Flu assay.
Of the 1282 specimens included in the study, 6 were unresolved by Comparator 1 and were excluded, resulting in 1276 specimens included in the final analysis.
A total of 90 unique specimens were influenza A positive by Comparator 1 and 1186 were negative. All 90 specimens were tested by Comparator 2 to determine influenza A subtypes [influenza A HINI (2009), influenza A H1N1 (seasonal), influenza A H3N2 (seasonal)]. From these 90 unique specimens, 6 were negative for all three subtypes by Comparator 2. The remaining 84 specimens generated a total of 88 positive results. Four specimens generated two different subtype results each by Comparator 2. For one of these four specimens, Comparator 2 assay reported both HIN1 and H3N2 while the PLEX-ID Flu assay reported an HIN1 (seasonal) result. For the remaining 3 specimens, Comparator 2 reported both H3N2 and H1N1, whereas the PLEX-ID Flu assay reported H3N2 (seasonal). These 88 positive results comprised 32 influenza A H1N1 (2009), 7 influenza A HINI (seasonal) positive, and 49 influenza A H3N2 (seasonal) positive results.
From the 1287 prospective NP swab specimens collected for this study and tested, 98.7% (1285/1302) of PLEX-ID assays results generated were valid.
PLEX-ID Flu 510(k) K121003 December 2012
510(k) Summary Page 15 of 31
{15}------------------------------------------------
The comparison of the PLEX-ID Flu assay and Comparator 2 results for influenza A HINI (2009) are shown in Table 4, influenza A HINI (seasonal) in Table 5, and influenza A H3N2 (seasonal) in Table 6.
Table 4
PLEX-ID Flu
Influenza A H1N1 (2009) Specimen Testing
| Comparator 2Positive | ComparatoraNegative | Total | |
|---|---|---|---|
| PLEX-ID FluPositive | 32 | 5b | 37 |
| PLEX-ID FluNegative | 0 | 1239 | 1239 |
| Total | 32 | 1244 | 1276 |
| PercentAgreement(95% CI) | Positive PercentAgreement100.0% (32/32)(89.3%, 100.0%) | Negative PercentAgreement99.6% (1239/1244)(99.1%, 99.8%) |
4 Negative for influenza A by Comparator 1 (and therefore untested by Comparator 2) or positive for influenza A by Comparator 1 and negative by Comparator 2 reflex testing for influenza A HINI (2009).
brive samples were negative for influenza A 2009 H1N1 by Comparator 2; three samples were positive by bi-directional sequence analysis for 2009 HINI, one had insufficient sample volume for bidirectional sequence analysis, and 2009 HIN1 nucleic acid was not detected by bi-directional sequence analysis in one sample.
{16}------------------------------------------------
Table 5
PLEX-ID Flu
Influenza A H1N1 (Seasonal) Specimen Testing
| Comparator 2Positive | ComparatoraNegative | Total | |
|---|---|---|---|
| PLEX-ID FluPositive | 4 | 0 | 4 |
| PLEX-ID FluNegative | 3b | 1269 | 1272 |
| Total | 7 | 1269 | 1276 |
| PercentAgreement(95% CI) | Positive PercentAgreement57.1% (4/7)(25.0%, 84.2%) | Negative PercentAgreement100.0% (1269/1269)(99.7%, 100.0%) |
"Negative for influenza A by Comparator 1 (and therefore untested by Comparator 2) or positive for influenza A by Comparator 1 and negative by Comparator 2 reflex testing for influenza A H1N1 (Seasonal).
6/H1 nucleic acids were not detected by bi-directional sequence analysis in these three samples.
Table 6
PLEX-ID Flu
Influenza A H3N2 (Seasonal) Specimen Testing
| Comparator 2Positive | ComparatoraNegative | Total | |
|---|---|---|---|
| PLEX-ID FluPositive | 48 | 2b | 50 |
| PLEX-ID FluNegative | 1c | 1225 | 1226 |
| Total | 49 | 1227 | 1276 |
| PercentAgreement(95% CI) | Positive PercentAgreement98.0% (48/49)(89.3%, 99.6%) | Negative PercentAgreement99.8% (1225/1227)(99.4%, 100.0%) |
"Negative for influenza A by Comparator 1 (and therefore untested by Comparator 2) or positive for influenza A by Comparator 1 and negative by Comparator 2 reflex testing for influenza A H3N2 (Seasonal).
,
o Two samples were negative for influenza A by Comparator 1 and were not reflexed to Comparator 2 for subtype determination, but the samples were positive by bi-directional sequence analysis for A/H3. 6 A/H3 nucleic acid was not detected by bi-directional sequence analysis in this sample.
PLEX-ID Flu 510(k) K121003 December 2012
510(k) Summary Page 17 of 31
{17}------------------------------------------------
A total of 22 positive specimens were obtained for influenza B. The positive percent agreement and negative percent agreement were calculated. The results of the PLEX-ID Flu compared to Comparator 1 for the detection of influenza B are shown in Table 7. Bidirectional sequence analysis was not performed for discordant influenza B specimens.
Table 7
PLEX-ID Flu
Influenza B Specimen Testing
| Comparator 1Positive | Comparator 1Negative | Total | |
|---|---|---|---|
| PLEX-ID FluPositive | 22 | 2 | 24 |
| PLEX-ID FluNegative | 0 | 1252 | 1252 |
| Total | 22 | 1254 | 1276 |
| PercentAgreement(95% CI) | Positive PercentAgreement100.0% (22/22)(85.1%, 100.0%) | Negative PercentAgreement99.8% (1252/1254)(99.4%, 100.0%) |
5.12.2.2 Retrospective Specimen Testing
Due to the decreased prevalence of influenza A 2009 H1N1 and the absence of influenza A seasonal H1N1 and H3N2, the prospectively collected specimen population was supplemented with pre-selected banked specimens that were collected in February 2008, and from January 2009 through March 2009, and left over archived specimens collected from April 2009 through December 2009.
A total of 1378 retrospective specimens were tested. Of these specimens, 1341 were included in the analysis (the remaining 37 were excluded either due to unavailable results for the PLEX-ID Flu assay or Comparator 1 assay, duplicate subjects, or failure to meet subject inclusion criteria).
Of the 1341 retrospective specimens, 634 were influenza A positive and 707 were negative by Comparator 1. All 634 positive specimens were tested to determine influenza A subtype (influenza A 2009 H1N1, influenza A (seasonal) HINI, influenza A
{18}------------------------------------------------
(seasonal) H3N2). Two specimens were unresolved by Comparator 2 and, therefore, excluded, resulting in 1339 total specimens in the analysis. Fifty-four of the 634 Comparator 1 influenza A positive specimens were negative for all three subtypes by Comparator 2 and were included as Comparator negative specimens. Consequently, 578 positive results were included in the analysis: 548 influenza A HINI (2009), 28 influenza A HINI (seasonal), and 2 influenza A H3N2 (seasonal).
From the 1378 retrospective NP swab specimens tested in this study, 93.4% (1368/1464) of PLEX-ID assays results generated were valid.
The comparison of the PLEX-ID Flu and Comparator 2 results for influenza A H1N1 (2009) are shown in Table 8, influenza A H1N1 (seasonal) in Table 9, and influenza A H3N2 (seasonal) in Table 10.
The following analysis was performed on the leftover archived and pre-selected banked specimens.
Table 8
PLEX-ID Flu
| Comparator 2Positive | ComparatoraNegative | Total | |
|---|---|---|---|
| PLEX-ID FluPositive | 538 | 45b | 583 |
| PLEX-ID FluNegative | 10c | 746 | 756 |
| Total | 548 | 791 | 1339 |
| PercentAgreement(95% CI) | Positive PercentAgreement98.2% (538/548)(96.7%, 99.0%) | Negative PercentAgreement94.3% (746/791)(92.5%, 95.7%) |
Influenza A H1N1 (2009) Specimen Testing
"Negative for influenza A by Comparator 1 (and therefore untested by Comparator 2) or positive for influenza A by Comparator 1 and negative by Comparator 2 reflex testing for influenza A HINI (2009). 841 samples were positive by bi-directional sequence analysis for 2009 H1N1, 1 was unresolved, 2 had insufficient sample volume for bi-directional sequence analysis, and HINI(2009) nucleic acid was not detected in one sample.
CA/HINI (2009) was detected in 8 samples by bi-directional sequence analysis.
PLEX-ID Flu 510(k) K121003 December 2012
510(k) Summary Page 19 of 31 .
{19}------------------------------------------------
Table 9
PLEX-ID Flu
Influenza A H1N1 (Seasonal) Specimen Testing
| Comparator 2Positive | ComparatoraNegative | Total | |
|---|---|---|---|
| PLEX-ID FluPositive | 27 | 0 | 27 |
| PLEX-ID FluNegative | 1b | 1311 | 1312 |
| Total | 28 | 1311 | 1339 |
| PercentAgreement(95% CI) | Positive PercentAgreement96.4% (27/28)(82.3%, 99.4%) | Negative PercentAgreement100.0% (1311/1311)(99.7%, 100.0%) |
"Negative for influenza A by Comparator 1 (and therefore untested by Comparator 2) or positive for influenza A by Comparator 1 and negative by Comparator 2 reflex testing for influenza A HINI (Seasonal).
b A/H I nucleic acid was not detected by bi-directional sequence analysis in this sample.
Table 10
PLEX-ID Flu
Influenza A H3N2 (Seasonal) Specimen Testing
| Comparator 2Positive | ComparatorᵃNegative | Total | |
|---|---|---|---|
| PLEX-ID FluPositive | 2 | 0 | 2 |
| PLEX-ID FluNegative | 0 | 1337 | 1337 |
| Total | 2 | 1337 | 1339 |
| PercentAgreement(95% CI) | Positive PercentAgreement100.0% (2/2)(34.2%, 100.0%) | Negative PercentAgreement100.0% (1337/1337)(99.7%, 100.0%) |
"Negative for influenza A by Comparator 1 (and therefore untested by Comparator 2) or positive for influenza A by Comparator 1 and negative by Comparator 2 reflex testing for influenza A H3N2 (Seasonal).
A total of 28 positive specimens were obtained for influenza B. The positive percent agreement and negative percent agreement were calculated. The comparison results of the PLEX-ID Flu and Comparator 1 to detect influenza B are shown in Table 11. Bidirectional sequence analysis was not performed for discordant influenza B specimens.
{20}------------------------------------------------
Table 11
PLEX-ID Flu
Influenza B Specimen Testing
| Comparator 1Positive | Comparator 1Negative | Total | |
|---|---|---|---|
| PLEX-ID FluPositive | 26 | 1 | 27 |
| PLEX-ID FluNegative | 2 | 1312 | 1314 |
| Total | 28 | 1313 | 1341 |
| PercentAgreement(95% CI) | Positive PercentAgreement92.9% (26/28)(77.4%, 98.0%) | Negative PercentAgreement99.9% (1312/1313)(99.6%, 100.0%) |
5.12.2.3 Other Clinical Supportive Data
Not applicable.
5.12.2.4 Clinical Cut-Off
Not applicable.
5.12.3 Summary of Nonclinical Studies
5.12.4 Limit of Detection
The limit of detection (LOD) for the PLEX-ID Flu assay was determined by testing quantitated influenza strains diluted in negative nasopharyngeal swab matrix. Twenty replicates of the following were tested: influenza A HINI (2009), influenza A H1N1 (seasonal), influenza A H3N2 (seasonal), and influenza B. The LOD was defined as the lowest influenza concentration at which ≥ 95% of replicates tested positive. The results are presented in Table 12.
ない
{21}------------------------------------------------
| Influenza | Limit of Detection (TCID50/mL) |
|---|---|
| influenza A H1N1 (2009) | $1.3 \times 10^{1}$ |
| influenza A H1N1 (seasonal) | $4.6 \times 10^{0}$ |
| influenza A H3N2 (seasonal) | $3.3 \times 10^{2}$ |
| influenza B | $1.5 \times 10^{1}$ |
Table 12 PLEX-ID Flu Limit of Detection (LOD)
5.12.4.1 Reproducibility
The reproducibility of the PLEX-ID Flu assay was evaluated at 3 clinical sites. Reproducibility was assessed using a nine member panel of contrived samples that included a negative member, low positive panel members targeted near the assay LOD and moderate positive panel members targeted at approximately 2 to 3 times the LOD for influenza B and the 3 influenza A subtypes detected by the assay. Samples were tested in replicates of 4 with one replicate each of the positive and negative control per batch. Two (2) batches were run per day and each was run by a different operator with 2 operators per site. The study was designed to allow for a total of 120 results for each panel member over 5 days of testing. The results are presented in Table 13.
PLEX-ID Flu 510(k) K121003 December 2012
510(k) Summary Page 22 of 31
{22}------------------------------------------------
| Table 13 | ||
|---|---|---|
| PLEX-ID Flu | ||
| Reproducibility Study |
:
| Panel Member Description | Correct Results/Number Tested(% Agreement) | Total of All Sites(95% CI) | ||
|---|---|---|---|---|
| Site 1 | Site 2 | Site 3 | ||
| Negative | 40/40(100.0%) | 40/40(100.0%) | 40/40(100.0%) | 120/120 (100.0%)(96.9%, 100.0%) |
| influenza A H1N1 (2009) Low Positive | 40/40(100.0%) | 39/39a(100.0%) | 38/40(95.0%) | 117/119a (98.3%)(94.1%, 99.5%) |
| influenza A H1N1 (2009) Moderate Positive | 40/40(100.0%) | 40/40(100.0%) | 40/40(100.0%) | 120/120 (100.0%)(96.9%, 100.0%) |
| influenza A H1N1 (seasonal) Low Positive | 39/40(97.5%) | 40/40(100.0%) | 39/40(97.5%) | 118/120 (98.3%)(94.1%, 99.5%) |
| influenza A H1N1 (seasonal) Moderate Positive | 40/40(100.0%) | 40/40(100.0%) | 40/40(100.0%) | 120/120 (100.0%)(96.9%, 100.0%) |
| influenza A H3N2 (seasonal) Low Positive | 40/40(100.0%) | 40/40(100.0%) | 39/39a(100.0%) | 119/119a (100.0%)(96.9%, 100.0%) |
| influenza A H3N2 (seasonal) Moderate Positive | 39/40(97.5%) | 40/40(100.0%) | 40/40(100.0%) | 119/120 (99.2%)(95.4%, 99.9%) |
| influenza B Low Positive | 40/40(100.0%) | 39/39a(100.0%) | 39/40(97.5%) | 118/119a (99.2%)(95.4%, 99.9%) |
| influenza B Moderate Positive | 40/40(100.0%) | 40/40(100.0%) | 39/40(97.5%) | 119/120 (99.2%)(95.4%, 99.9%) |
ªMissing replicate due to instrument error.
PLEX-ID Flu 510(k) K121003 December 2012
:
:
{23}------------------------------------------------
5.12.4.2 ·Within Laboratory Precision
Within laboratory precision was determined by testing all nine members of the reproducibility panel in duplicate in two runs per day for a total of twelve days. Each run contained the assay positive and negative controls for a total of 20 samples per run. Multiple PLEX-ID Flu assay lots were used in the study. The results are presented in Table 14.
| Table11 | 0 |
|---|---|
| ----------------- | --- |
PLEX-ID Flu Within Laboratory Precision
| Panel Member Description | Correct Results/Number Tested(% Agreement) | 95%ConfidenceInterval |
|---|---|---|
| Negative | 48/48 (100%) | (93%, 100%) |
| influenza A HINI (2009)Low Positive | 48/48 (100%) | (93%, 100%) |
| influenza A H1N1 (2009)Moderate Positive | 48/48 (100%) | (93%, 100%) |
| influenza A H1N1 (seasonal)Low Positive | 48/48 (100%) | (93%, 100%) |
| influenza A HIN1 (seasonal)Moderate Positive | 48/48 (100%) | (93%, 100%) |
| influenza A H3N2 (seasonal)Low Positive | 48/48 (100%) | (93%, 100%) |
| influenza A H3N2 (seasonal)Moderate Positive | 48/48 (100%) | (93%, 100%) |
| influenza BLow Positive | 48/48 (100%) | (93%, 100%) |
| influenza BModerate Positive | 48/48 (100%) | (93%, 100%) |
PLEX-ID Flu 510(k) K121003 December 2012
510(k) Summary Page 24 of 31
{24}------------------------------------------------
5.12.4.3 Cross-Reactivity
The effect of 34 common respiratory microbes on the PLEX-ID Flu assay was evaluated. Negative samples and samples containing influenza A HINI (2009), influenza A HINI (seasonal), influenza A H3N2 (seasonal), and influenza B were tested. No interference in the performance of the PLEX-ID Flu assay was observed in the presence of the potential cross-reactants shown in Table 15 for all positive and negative samples tested.
Table 15 PLEX-ID Flu Cross-Reactivity
| Microbe | Target SpikeConcentration |
|---|---|
| Bordetella pertussis | 106 CFU/mL |
| Chlamydia pneumoniae | 106 CFU/mL |
| Corynebacterium sp. | 106 CFU/mL |
| Escherichia coli | 106 CFU/mL |
| Haemophilus influenzae | 106 CFU/mL |
| Lactobacillus sp. | 106 CFU/mL |
| Legionella pneumophilia | 106 CFU/mL |
| Moraxella cartarrhalis | 106 CFU/mL |
| Mycobacterium tuberculosis avirulent | 106 genomes/mL |
| Mycoplasma pneumoniae | 106 CFU/mL |
| Neisseria meningitides | 106 CFU/mL |
| Neisseria sp. | 105 CFU/mL |
| Pseudomonas aeruginosa | 106 CFU/mL |
| Staphylococcus aureus | 106 CFU/mL |
| Staphylococcus epidermidis | 106 CFU/mL |
| Streptococcus pneumoniae | 106 CFU/mL |
| Streptococcus pyogenes | 106 CFU/mL |
PLEX-ID Flu 510(k) K121003 December 2012
510(k) Summary Page 25 of 31
{25}------------------------------------------------
| Microbe | Target SpikeConcentration |
|---|---|
| Streptococcus salivarius | $10^5$ CFU/mL |
| Adenovirus Type 1 | $10^5$ TCID50/mL |
| Adenovirus Type 7 | $10^5$ TCID50/mL |
| Human coronavirus OC 43 | $10^4$ TCID50/mL |
| Human coronavirus OC229E | $10^5$ TCID50/mL |
| Cytomegalovirus | $10^5$ TCID50/mL |
| Enterovirus | $10^5$ TCID50/mL |
| Epstein Barr virus | $10^5$ TCID50/mL |
| Measles | $10^5$ TCID50/mL |
| Human metapneumovirus | $10^5$ TCID50/mL |
| Mumps virus | $10^5$ TCID50/mL |
| Respiratory syncytial virus Type B | $10^5$ TCID50/mL |
| Rhinovirus Type 1A | $10^4$ TCID50/mL |
| Human parainfluenza Type 1 | $10^5$ TCID50/mL |
| Human parainfluenza Type 2 | $10^5$ TCID50/mL |
| Human parainfluenza Type 3 | $10^5$ TCID50/mL |
| Herpes simplex virus Type 1 | $10^5$ TCID50/mL |
.
・
· :
510(k) Summary Page 26 of 31
・
.
{26}------------------------------------------------
5.12.4.4 Potentially Interfering Substances
The effect of substances that may be present in nasopharyngeal swab samples for potential interference with the PLEX-ID Flu assay was evaluated. Negative samples and samples containing influenza A HINI (2009), influenza A HIN1 (seasonal), influenza A H3N2 (seasonal), and influenza B were tested at concentrations approximately 2× LOD. No interference in the performance of the PLEX-ID Flu assay was observed in the presence of the potentially interfering substances shown in Table 16 for all positive and negative samples tested.
FluMist®, an intranasal vaccine that contains live attenuated influenza virus, was detected by the PLEX-ID Flu assay. Samples containing FluMist may cause false positive results in the assay.
| Mucin: bovine submaxillary gland, type I-S | Boiron Sulphur iodatum |
|---|---|
| Blood (human) | Cepacol Dual Relief Sore Throat Spray |
| Neo-Synephrine | CVS Pharmacy Sore Throat Spray |
| Equaline 12-hour Nasal Spray | Cool Mint Listerine Antiseptic |
| Equaline Premium Saline Nasal Spray | Halls Mentho-Lyptus |
| Beconase | Tamiflu |
| Decadron | Flumadine |
| Nasarel | SYMMETREL |
| Allernaze | Relenza |
| Rhinocort Aqua Nasal Spray | BACTROBAN NASAL |
| Nasonex | Tobi |
| Oxymetazonine HCl | Adult Robitussin Chest Congestion |
| Flonase | CVS Children's Allergy LiquidMedication |
| Zicam | Delsym Cough Suppressant |
Table 16 PLEX-ID Flu Potentially Interfering Substances
PLEX-ID Flu 510(k) K121003 December 2012
{27}------------------------------------------------
| Boiron Galphimia glauca | CVS Non-Drowsy Nasal Decongestant |
|---|---|
| Boiron Histaminum hydrochloricum |
5.12.4.5 Carryover
Sample carryover on the PLEX-ID System was evaluated using high target concentrations (1 × 105 viral genomes per mL) of influenza A and influenza B alternately loaded with negative sample rack, representing a worst-case, checkerboard configuration. Following 5 independent isolations, no positive results were reported among the 120 negative samples tested, resulting in an observed carryover rate of 0% (0/120).
5.12.4.6 Inclusivity
Multiple viral strains representing temporal and geographical diversity for each influenza type and subtype were tested with the PLEX-ID Flu assay. The samples were targeted at concentrations at approximately 2× LOD. The results reported for each strain tested matched the expected result in 3 out of 3 replicates, as shown in Table 17.
Table 17 PLEX-ID Flu Inclusivity
| Viral Strain | Concentration | Influenza A H1N1 (2009) | Influenza A H1N1 (seasonal) | Influenza A H3N2 (seasonal) | Influenza A Other | Influenza B |
|---|---|---|---|---|---|---|
| influenza A Swine H1N1 NY01 (2009 H1N1) | 6.57 TCID50/mL | + | - | - | - | - |
| influenza A Swine H1N1 NY02 (2009 H1N1) | $1.06x10^1$ TCID50/mL | + | - | - | - | - |
| influenza A Swine H1N1 NY03 (2009 H1N1) | 9.37 TCID50/mL | + | - | - | - | - |
| influenza A/Brisbane/59/07 | $1.37\times10^1$ TCID50/mL | - | + | - | - | - |
| Viral Strain | Concentration | Influenza A H1N1(2009) | Influenza A H1N1(seasonal) | Influenza A H3N2(seasonal) | Influenza A Other | Influenza B |
| (Seasonal H1N1) | ||||||
| influenzaA/NewCal/20/1999(Seasonal H1N1) | $1.29×10^3$ TCID50/mL | - | + | - | - | - |
| influenza A/Taiwan/42/06(Seasonal H1N1) | $4.45×10^3$ TCID50/mL | - | + | - | - | - |
| influenza A/SolomonIslands/03/06(Seasonal H1N1) | $1.40×10^1$ TCID50/mL | - | + | - | - | - |
| influenza A/PR/8/34(Seasonal H1N1) | $4.02×10^4$ CEID50/mL | - | + | - | - | - |
| influenza A/WS/33(Seasonal H1N1) | $4.14×10^2$ CEID50/mL | - | + | - | - | - |
| influenza A/NewJersey/8/76 (SeasonalH1N1) | $1.24×10^2$ CEID50/mL | - | + | - | - | - |
| influenza A/Weiss/43(Seasonal H1N1) | $3.41×10^6$ CEID50/mL | - | + | - | - | - |
| influenza A/FM/1/47(Seasonal H1N1) | $6.52×10^3$ CEID50/mL | - | + | - | - | - |
| influenza A/Mal/302/54(Seasonal H1N1) | $7.41×10^2$ CEID50/mL | - | + | - | - | - |
| influenza A/PortChalmers/1/73 (SeasonalH3N2) | $6.15×10^3$ CEID50/mL | - | - | + | - | - |
| influenza A/HongKong/8/68 (SeasonalH3N2) | $4.12×10^4$ CEID50/mL | - | - | + | - | - |
| influenza A/Victoria/3/75(Seasonal H3N2) | $2.46×10^3$ CEID50/mL | - | - | + | - | - |
| Viral Strain | Concentration | Influenza A H1N1 (2009) | Influenza A H1N1 (seasonal) | Influenza A H3N2 (seasonal) | Influenza A Other | Influenza B |
| influenza A/Aichi/2/68(Seasonal H3N2) | $4.72\times10^3$ CEID50/mL | - | - | + | - | - |
| influenza B/Lee/40 | $3.28\times10^4$ CEID50/mL | - | - | - | - | + |
| influenza B/Mass/3/66 | $9.41\times10^3$ CEID50/mL | - | - | - | - | + |
| influenza B/Taiwan/2/62 | $1.65\times10^2$ CEID50/mL | - | - | - | - | + |
| influenza B/Hong Kong/5/72 | $2.63\times10^4$ CEID50/mL | - | - | - | - | + |
| influenza B/GL/1739/54 | $1.56\times10^3$ CEID50/mL | - | - | - | - | + |
| influenza B/Allen/45 | $1.30\times10^2$ CEID50/mL | - | - | - | - | + |
| influenza B/Maryland/1/59 | 8.50 CEID50/mL | - | - | - | - | + |
| influenza B/Russia/69 | $1.24\times10^3$ CEID50/mL | - | - | - | - | + |
| influenza B/Florida/02/06 | $8.11\times10^1$ TCID50/mL | - | - | - | - | + |
| influenza B/Malaysia/2506/04 | $2.07\times10^1$ TCID50/mL | - | - | - | - | + |
PLEX-ID FIL December 2012
Page 28 of 31
{28}------------------------------------------------
.
:
510(k) Summary
Page 29 of 31
.
{29}------------------------------------------------
.
ﺘﺮﺍ ﻓﻲ ﺍﻟﻤﺘﻮﺳﻂ ﺍﻟﺘﻲ ﺗﻌﺘﺒﺮ ﺍﻟﻤﺘﻮﺳﻂ ﺍﻟﺘﻲ ﺗﻌﺘﺒﺮ ﺍﻟﻤﺘﻮﺳﻂ ﺍﻟﻤﺘﻮﺳﻂ ﺍﻟﻤﺘﻮﺳﻂ ﺍﻟﻤﺘﻮﺳﻂ ﺍﻟﻤﺘﻮﺳﻂ ﺍﻟﻤﺘﻮﺳﻂ ﺍﻟﻤﺘﻮﺳﻂ ﺍﻟﻤﺘﻮﺳﻂ ﺍﻟﻤﺘﻮﺳﻂ ﺍﻟﻤﺘﻮﺳﻂ ﺍﻟﻤﺘﻮﺳﻂ ﺍﻟﻤﺘﻮﺳﻂ ﺍﻟﻤﺘﻮﺳﻂ ﺍﻟﻤﺘﻮﺳﻂ ﺍﻟﻤﺘﻮﺳﻂ ﺍﻟﻤﺘﻮﺳﻂ
·
·
{30}------------------------------------------------
5.13 References
- CDC Influenza (Flu), What Everyone Should Know About Flu and the Flu l . Vaccine: www.cdc.gov/flu/keyfacts.htm [8/17/2009]
-
- U.S. Department of Health & Human Services: www.flu.gov/about the flu/seasonal/index.html [9/17/2012]
- Chan, Margaret. World now at the start of 2009 influenza pandemic, Statement to 3. the press by WHO Director-General Dr Margaret Chan 11 June 2009: http://www.who.int/mediacentre/news/statements/2009/h1n1 pandemic phase6 20 090611/en/index.html. September 2012.
PLEX-ID Flu 510(k) K121003 December 2012
510(k) Summary Page 31 of 31
{31}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Abbott Laboratories c/o Darren Clarke 1300 East Touhy Avenue Des Plaines, Illinois, 60018
December 21, 2012
Re: K121003
Trade/Device Name: Abbott Plex-ID Flu Assay Regulation Number: 21 CFR 866.3980 Regulation Name: Respiratory viral panel multiplex nucleic acid assay Regulatory Class: Class II Product Code: OCC, OEP, OQW, OTA Dated: December 15, 2012 Received: December 17, 2012
Dear Mr. Clarke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{32}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostics and Radiological Health at (301) 796-( 5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Uwe Scherf for
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{33}------------------------------------------------
Indications of Use Statement
510(k) Number: K121003
4.0
Device Name: PLEX-ID Flu
Indications for Use for the PLEX-ID Flu:
The PLEX-ID Flu assay is a qualitative nucleic acid in vitro diagnostic test intended for the detection and differentiation of influenza A H1N1 (2009), influenza A H1N1 (seasonal), influenza A H3N2 (seasonal) and influenza B viral nucleic acids in nasopharyngeal swab specimens from patients symptomatic for respiratory tract infection. The PLEX-ID Flu assay is intended for use on the PLEX-ID System (version 1.2) as an aid in the diagnosis of influenza infection in conjunction with clinical and epidemiological information. This assay is not intended to detect influenza C virus.
Negative results do not preclude influenza virus infection and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.
If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
Prescription Use X (Per 21 CFR 801.109) AND/OR
Over-The-Counter Use (Per 21 CFR Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Tausael Feldberg
Division Sign-Off
Office of In Vitro Dlagnostic Device Evaluation and Safety
510(k) K121003
§ 866.3980 Respiratory viral panel multiplex nucleic acid assay.
(a)
Identification. A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B;
(2) Influenza A subtype H1 and Influenza A subtype H3;
(3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B;
(4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus;
(5) Human Metapneumovirus;
(6) Rhinovirus; and
(7) Adenovirus.
(b)
Classification. Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.