(192 days)
The tangoRS™ is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar and sacral spine (T1-S2). When used as a posterior spine thoracic/lumbar system, the tangoRS is intended for the following indications: degenerative disc disease (as defined by back pain of discogenic with degeneration of the disc confirmed by patient history and radiographic studies), spinal stenosis, spondylolisthesis, spinal deformities (i.e., degenerative scoliosis, kyphosis, and/or lordosis), fracture, spinal tumor, pseudarthrosis and failed previous fusion.
The tangoRS™ consists of rods, polyaxial pedicle screws and crosslinks with locking set screws. The components are available in various sizes to accommodate differing patient anatomy. Rods are available in one diameter and a variety of lengths. Polyaxial screws are available in a variety of diameter-length combinations. Crosslinks are offered in four lengths. The purpose of this submission is to add components (a screw size and crosslink options) and two indications for use.
The provided 510(k) summary for the tangoRS™ multifunctional posterior system describes its performance through mechanical testing. Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criterion (Implicit) | Reported Device Performance |
|---|---|
| Biocompatibility (Material safety) | Components manufactured from titanium alloy (Ti-6Al-4V ELI) as described by ASTM F136. |
| Mechanical Strength and Durability (Static Compression Bending) | tangoRS™ performs as well as or better than predicate devices according to ASTM F1717 testing. |
| Mechanical Strength and Durability (Dynamic Compression Bending) | tangoRS™ performs as well as or better than predicate devices according to ASTM F1717 testing. |
| Mechanical Strength and Durability (Static Torsion) | tangoRS™ performs as well as or better than predicate devices according to ASTM F1717 testing. |
| Substantial Equivalence to Predicate Devices (Technological Characteristics & Intended Use) | Same intended use, basic design (rod-based, screw anchors), material (titanium alloy), and comparable sizes to predicate devices. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not specify a "sample size" in terms of individual patients or clinical cases for a test set. The performance evaluation is based on mechanical testing of "worst case tangoRS™ constructs." The exact number of constructs tested is not provided, but it would typically involve multiple samples for each test condition to ensure reproducibility.
- Data Provenance: The data is generated from in-vitro (benchtop) mechanical testing, not from patient-derived data. Therefore, there's no country of origin for clinical data, and it is not retrospective or prospective in the clinical sense.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable to this type of submission. Mechanical testing does not involve human experts establishing "ground truth" on patient data. The "ground truth" for mechanical testing is established by the specified ASTM F1717 standard and the performance of predicate devices.
4. Adjudication Method for the Test Set
This is not applicable as the study is mechanical testing, not a clinical study involving human judgment that would require adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, an MRMC comparative effectiveness study was not done. This submission relies on mechanical bench testing, not clinical studies involving human readers.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
This is not applicable. The device is a physical medical implant (pedicle screw system), not an algorithm or software. Therefore, there's no concept of "standalone performance" in the context of an algorithm. The performance described is inherent to the device's physical and mechanical properties.
7. Type of Ground Truth Used
The "ground truth" used for this type of device and submission is based on:
- ASTM F1717 standards: These provide the accepted methodology and criteria for testing spinal implant constructs.
- Performance of predicate devices: The tangoRS™ system is compared against the mechanical performance of previously cleared predicate devices (flamenco™ {K102853}, CD HORIZON® {K031655/K041460}, Expedium {K041119/K062196}, tangoRS™ {K052385}). The implicit "ground truth" is that if the new device performs "as well as or better than" these established devices under standardized testing, it is considered safe and effective.
8. Sample Size for the Training Set
This is not applicable. There is no "training set" in the context of mechanical testing for a physical implant. The design of the tangoRS™ system is based on established engineering principles and knowledge gained from predicate devices, but there isn't a "training set" of data in the machine learning sense.
9. How the Ground Truth for the Training Set Was Established
This is not applicable as there is no "training set" for this type of device and study. The physical design and materials are chosen based on known biomechanical requirements and historical performance of similar devices.
{0}------------------------------------------------
K120891
page 1 of 2
.
Image /page/0/Picture/1 description: The image shows the logo for Ulrich medical USA. The word "Ulrich" is written in a bold, sans-serif font, with the "U" extending from a black square on the left. Below "Ulrich" is the text "medical USA" in a smaller font.
OCT
1 2012
510(k) Summary
| Date: | 21 March 2012 |
|---|---|
| Sponsor: | ulrich GmbH & Co. KGBuchbrunnenweg 1289081 UlmGermanyPhone: +49 (0) 731-9654-1304Fax: +49 (0) 731-9654-2802 |
| Contact Person: | Hans Stoverulrich medical USA, Inc.612 Trade Center Blvd.Chesterfield, MO 63005(636) 519-0268 Office(636) 519-0271 Fax |
| Proposed Trade Name: | tangoRS™ multifunctional posterior system |
| Device Classifications: | Class II & Class III |
| Classification & Common Name: | Pedicle screw spinal system |
| Regulation: | 888.3070 |
| Device Product Code: | MNI, MNH, NKB |
| Device Description: | The tangoRS™ consists of rods, polyaxial pedicle screws andcrosslinks with locking set screws. The components are available invarious sizes to accommodate differing patient anatomy. Rods areavailable in one diameter and a variety of lengths. Polyaxial screwsare available in a variety of diameter-length combinations. Crosslinksare offered in four lengths.The purpose of this submission is to add components (a screw sizeand crosslink options) and two indications for use. |
| Intended Use: | The tangoRS is intended to provide immobilization and stabilizationof spinal segments as an adjunct to fusion of the thoracic, lumbarand sacral spine (T1-S2). When used as a posterior spinethoracic/lumbar system, the tangoRS is intended for the followingindications: degenerative disc disease (as defined by back pain ofdiscogenic with degeneration of the disc confirmed by patient historyand radiographic studies), spinal stenosis, spondylolisthesis, spinaldeformities (i.e., degenerative scoliosis, kyphosis, and/or lordosis). |
{1}------------------------------------------------
K120891
page 2 of 2
Materials:
The tangoRS™ components are manufactured from titanium alloy (Ti-6Al-4V ELI) as described by ASTM F136. flamenco™ {K102853} CD HORIZON® (K031655/K041460) Expedium (K041119/K062196) tangoRS™ (K052385)
Technological Characteristics:
Predicate Devices:
tangoRS™ possesses the same technological characteristics as one or more of the predicate devices. These include:
- intended use (as described above) .
- basic design (rod-based having screw anchors), .
- . material (titanium alloy),
- sizes (dimensions are comparable to those offered by the . predicate systems) and
The fundamental scientific technology of tangoRS™ is the same as previously cleared devices.
Performance Data:
Mechanical testing of the worst case tangoRS™ constructs included static and dynamic compression bending, and static torsion according to ASTM F1717.
The mechanical test results demonstrate that tangoRS™ performs as well as or better than the predicate devices and therefore that the device is as safe and as effective as the predicates.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
OCT
1 2012
Ulrich GmbH & Company KG % Ulrich Medical USA Mr. Hans Stover President & CEO 612 Trade Center Boulevard Chesterfield, Missouri 63005
Re: K120891
Trade/Device Name: tangoRS™ multifunctional posterior system Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNI, MNH Dated: August 22, 2012 Received: August 23, 2012
Dear Mr. Stover:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
Page 2 - Mr. Hans Stover
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark A. Millikan
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use Statement
510(k) Number: K120891
Device Name: tangoRS™ multifunctional posterior system
Indications for Use:
The tangoRS™ is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar and sacral spine (T1-S2). When used as a posterior spine thoracic/lumbar system, the tangoRS is intended for the following indications: degenerative disc disease (as defined by back pain of discogenic with degeneration of the disc confirmed by patient history and radiographic studies), spinal stenosis, spondylolisthesis, spinal deformities (i.e., degenerative scoliosis, kyphosis, and/or lordosis), fracture, spinal tumor, pseudarthrosis and failed previous fusion.
Prescription Use X
AND/OR
Over-the-Counter Use_
(21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K120891
§ 888.3070 Thoracolumbosacral pedicle screw system.
(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.