K Number
K120761
Date Cleared
2012-05-08

(56 days)

Product Code
Regulation Number
862.3650
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LZI Opiate 2000 Enzyme Immunoassay is intended for the qualitative and semiquantitative determination of morphine in human urine at the cutoff value of 2000 ng/mL. The assay is designed for professional use with a number of automated clinical chemistry analyzers.

The semi-quantitative mode is for purposes of (1) enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as GCMS and LCMS or (2) permitting laboratories to establish quality control procedures.

The LZI Opiate 2000 Drugs of Abuse (DAU) Calibrators are for use as calibrators in the qualitative and semi-quantitative calibration of the LZI Opiate 2000 Enzyme Immunoassay at a cutoff value of 2000 ng/mL.

The LZI Opiate 2000 Drugs of Abuse (DAU) Controls are for use as assayed quality control materials to monitor the precision of the LZI Opiate 2000 Enzyme Immunoassay at a cutoff value of 2000 ng/mL.

The assay provides only a preliminary analytical result: A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas or liquid chromatography/mass spectrometry (GC/MS or LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.

Device Description

The LZI Opiate 2000 assay is a homogeneous enzyme immunoassay with ready-to-use liquid reagents. The assay is based on competition between drug in the sample and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for a fixed amount of antibody in the reagent. Enzyme activity decreases upon binding to the antibody, and the drug concentration in the sample is measured in terms of enzyme activity. In the absence of drug in the sample, morphine-labeled G6PDH conjugate is bound to antibody, and the enzyme activity is inhibited. On the other hand, when free drug is present in the sample, antibody would bind to free drug; the unbound morphinelabeled G6PDH then exhibits its maximal enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that can be measured spectrophotometrically at 340 nm.

The LZI Opiate 2000 Enzyme Immunoassay is a kit comprised of two reagents, an R1 and R2, which are bottled separately but sold together within the kit.

The R1 solution contains mouse monoclonal anti-morphine antibody, glucose-6phosphate (G6P) nicotinamide adenine dinucleotide (NAD), stabilizers, and sodium azide (0.09%) as a preservative. The R2 solution contains glucose-6-phosphate dehydrogenase (G6PDH) labeled with morphine in buffer with sodium azide (0.09%) as preservative.

The LZI Opiate 2000 Enzyme Immunoassay calibrators and controls designated for use at the 2000 ng/mL cutoff contain 0, 1000, 1500, 2000, 2500, 4000, and 6000 ng/mL of morphine in human urine with sodium azide (0.09%) as preservative. These five calibrators and two controls are sold as individual bottles.

AI/ML Overview

Acceptance Criteria and Device Performance for LZI Opiate 2000 Enzyme Immunoassay

This document outlines the acceptance criteria and the study results demonstrating the performance of the LZI Opiate 2000 Enzyme Immunoassay.

1. Table of Acceptance Criteria and Reported Device Performance

The provided document details the precision and qualitative/semi-quantitative performance of the LZI Opiate 2000 Enzyme Immunoassay. While explicit acceptance criteria are not listed as separate entries (e.g., "accept if Positive Predictive Agreement > X%"), the performance characteristics presented are those that would be assessed against internal or regulatory standards for such a device. For this summary, we interpret the reported performance of the predicate device as an implicit benchmark, and the new device aims to demonstrate comparable or superior performance.

Implicit Acceptance Criteria (based on predicate equivalence and established test performance expectations):

  • Precision: Acceptable within-run and total precision (low %CV) across various concentrations, particularly around the cutoff.
  • Qualitative/Semi-Quantitative Agreement: High agreement with confirmed GC/MS or LC/MS results for both positive and negative samples.
  • Limit of Detection: Ability to reliably detect opiate at a specified low concentration.
  • Linearity: Reportable linear range of detection for semi-quantitative use.
  • Specificity/Lack of Interference: Absence of significant cross-reactivity or interference from endogenous compounds.

Reported Device Performance (LZI Opiate 2000 Enzyme Immunoassay for Hitachi 717 Analyzer):

Performance MetricAcceptance Criteria (Implicit)Reported Device Performance
Precision (Semi-Quantitative)Low % CV for within-run and total precision across concentrationsWithin-Run %CV: • 0 ng/mL: 36.6% • 500 ng/mL: 4.1% • 1000 ng/mL: 2.4% • 1500 ng/mL: 1.3% • 2000 ng/mL: 1.6% • 2500 ng/mL: 1.9% • 3000 ng/mL: 2.9% • 3500 ng/mL: 1.8% • 4000 ng/mL: 2.1% Total Precision %CV: • 0 ng/mL: 39.6% • 500 ng/mL: 4.9% • 1000 ng/mL: 2.4% • 1500 ng/mL: 1.4% • 2000 ng/mL: 1.9% • 2500 ng/mL: 1.9% • 3000 ng/mL: 2.9% • 3500 ng/mL: 2.0% • 4000 ng/mL: 2.4%
Qualitative Agreement (Semi-Quant)High agreement with confirmed resultsAt 2000 ng/mL Cutoff: Within Run (N=22 per concentration): • 0, 500, 1000, 1500 ng/mL: 22 Negative • 2000 ng/mL (0% of cutoff): 18 Positive / 4 Negative • 2500, 3000, 3500, 4000 ng/mL: 22 Positive Total Precision (N=88 per concentration): • 0, 500, 1000, 1500 ng/mL: 88 Negative • 2000 ng/mL (0% of cutoff): 59 Positive / 29 Negative • 2500, 3000, 3500, 4000 ng/mL: 88 Positive
Precision (Qualitative - mA/min)Low % CV for within-run and total precision across concentrationsWithin-Run %CV: • 0 ng/mL: 0.7% • 500 ng/mL: 0.6% • 1000 ng/mL: 0.4% • 1500 ng/mL: 0.4% • 2000 ng/mL: 0.7% • 2500 ng/mL: 0.4% • 3000 ng/mL: 0.7% • 3500 ng/mL: 0.6% • 4000 ng/mL: 0.5% Total Precision %CV: • 0 ng/mL: 0.8% • 500 ng/mL: 0.8% • 1000 ng/mL: 0.8% • 1500 ng/mL: 0.7% • 2000 ng/mL: 0.9% • 2500 ng/mL: 0.7% • 3000 ng/mL: 0.9% • 3500 ng/mL: 0.7% • 4000 ng/mL: 0.6%
Qualitative Agreement (Qualitative)High agreement with confirmed resultsAt 2000 ng/mL Cutoff: Within Run (N=22 per concentration): • 0, 500, 1000, 1500 ng/mL: 22 Negative • 2000 ng/mL (0% of cutoff): 9 Positive / 13 Negative • 2500, 3000, 3500, 4000 ng/mL: 22 Positive Total Precision (N=88 per concentration): • 0, 500, 1000, 1500 ng/mL: 88 Negative • 2000 ng/mL (0% of cutoff): 33 Positive / 55 Negative • 2500, 3000, 3500, 4000 ng/mL: 88 Positive
Limit of Detection (LoD)Ability to confidently detect at low concentrations200 ng/mL (Lowest concentration differentiated from negative urine with 95% confidence)
LinearityStrong linear correlation across the relevant rangey = 1.078x - 85.072, r² = 0.993 for range 0 - 6000 ng/mL
Clinical Sample Method ComparisonHigh agreement for positive and negative samplesSemi-Quantitative Data: 96.49% agreement with positive, 93.55% agreement with negative samples Qualitative Data: 96.49% agreement with positive, 96.77% agreement with negative samples
Interference/SpecificityNo significant undesired interferencesNo significant undesired cross-reactants or endogenous substance interference was observed.

2. Sample Size and Data Provenance for the Test Set

The document describes test data from different studies:

  • Precision Studies: A total of 88 determinations were performed for each concentration level (0, 500, 1000, 1500, 2000, 2500, 3000, 3500, 4000 ng/mL) for both semi-quantitative and qualitative modes. These were laboratory-prepared samples. The provenance of the data (country of origin) is not explicitly stated, but it's likely part of the manufacturer's internal validation. Given these are defined spike levels, the data is prospective and controlled.
  • Method Comparison - Clinical Samples: 150 clinical unaltered samples were used. The document does not specify the country of origin but implies they are real-world samples. This data is retrospective, as it compares the device's performance against existing samples confirmed by another method.

3. Number of Experts and Qualifications for Ground Truth

  • Precision Studies: The ground truth for the precision studies was established by known concentrations of morphine spiked into urine. This does not involve human experts in establishing the "ground truth" concentrations themselves but relies on accurate laboratory preparation and measurement of these spiked samples.
  • Method Comparison - Clinical Samples: The ground truth for the 150 clinical samples was established using a "more specific alternative chemical method," specifically Gas or liquid chromatography/mass spectrometry (GC/MS or LC/MS). These are established analytical methods. The document does not mention the number or specific qualifications of human experts (e.g., laboratory personnel, clinical chemists) who performed these confirmatory tests, as the ground truth is derived from the analytical result of the GC/MS or LC/MS.

4. Adjudication Method for the Test Set

  • Precision Studies: No adjudication method is described beyond the direct measurement of instrument responses to samples with known concentrations.
  • Method Comparison - Clinical Samples: For the clinical samples, an adjudication method is not explicitly stated. The comparison is made between the LZI Opiate 2000 Enzyme Immunoassay result and the GC/MS or LC/MS result. Discrepancies would typically be reviewed by laboratory personnel, but a formal multi-reader adjudication process (e.g., 2+1) is not mentioned as it's a comparison to a definitive analytical method, not human interpretation.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done. This device is an in-vitro diagnostic assay, not an imaging AI device that assists human readers. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable here.

6. Standalone Performance Study

Yes, a standalone performance study was done. The entire performance characterization (Precision, Qualitative/Semi-Quantitative Results, Limit of Detection, Linearity, and Method Comparison with Clinical Samples) describes the performance of the LZI Opiate 2000 Enzyme Immunoassay algorithm only (the assay itself) without human interpretation in the workflow, aside from standard laboratory procedures for running the test. The "Immunoassay Result" reported in the tables is the output of the device.

7. Type of Ground Truth Used

  • For controlled precision and linearity studies, the ground truth was based on known concentrations of morphine in spiked urine samples.
  • For the clinical sample method comparison, the ground truth was analytical confirmation by Gas or liquid chromatography/mass spectrometry (GC/MS or LC/MS), which is considered the gold standard for drug confirmation in such contexts.

8. Sample Size for the Training Set

The document does not explicitly mention a "training set" in the context of an algorithm or machine learning model. This device is a homogeneous enzyme immunoassay, a biochemical test, not a software algorithm that undergoes a distinct training phase in the machine learning sense. Its performance is inherent to the chemical reactions and optical detection, which are then validated against known standards and clinical samples.

9. How Ground Truth for the Training Set Was Established

Since there is no distinct "training set" in the machine learning context for this type of immunoassay, the concept of establishing ground truth for a training set is not applicable. The assay's "learning" or optimization would have occurred during its initial development and formulation, using laboratory-created spiked samples and internal validation data, much like the precision data presented. The performance characteristics described are the result of the final, developed assay.

{0}------------------------------------------------

K120761

510(k) Summary of Safety and Effectiveness

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Introduction

According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.

Submitter name, Address, and Contact

Lin-Zhi International, Inc. 670 Almanor Avenue Sunnyvale, CA 94085 Phone: (408) 732-3856 (408) 732-3849 Fax: e-mail: bclin@lin-zhi.com

Bernice Lin, Ph.D. Contact: VP Operations

Device Name and Classification

Classification Name:

Enzyme Immunoassay, Opiate Class II, DJG (91 Toxicology), 21 CFR 862.3650

Drug Specific Calibrators, Class II, DLJ (91 Toxicology), 21 CFR 862.3200

Drug Specific Controls, Class I, LAS (91 Toxicology), 21 CFR 862.3280

Common Name: Proprietary Name: Homogeneous Opiate 2000 Enzyme Immunoassay LZI Opiate 2000 Enzyme Immunoassay, LZI Opiate 2000 Drugs of Abuse (DAU) Calibrators LZI Opiate 2000 Drugs of Abuse (DAU) Controls

{1}------------------------------------------------

Legally Marketed Predicate Device(s)

The LZI Opiate 2000 Enzyme Immunoassay (EIA) at a cutoff of 2000 ng/mL is substantially equivalent to the LZI Opiate Enzyme Immunoassay and Opiate Calibrators and Controls (K020638 & K020769) manufactured by Lin-Zhi International, Inc with a cutoff of 300 ng/mL. The LZI Opiate 2000 Enzyme Immunoassav is identical or similar to its predicate in terms of intended use, method principle, device components, and clinical performance.

Device Description

The LZI Opiate 2000 assay is a homogeneous enzyme immunoassay with ready-to-use liquid reagents. The assay is based on competition between drug in the sample and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for a fixed amount of antibody in the reagent. Enzyme activity decreases upon binding to the antibody, and the drug concentration in the sample is measured in terms of enzyme activity. In the absence of drug in the sample, morphine-labeled G6PDH conjugate is bound to antibody, and the enzyme activity is inhibited. On the other hand, when free drug is present in the sample, antibody would bind to free drug; the unbound morphinelabeled G6PDH then exhibits its maximal enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that can be measured spectrophotometrically at 340 nm.

The LZI Opiate 2000 Enzyme Immunoassay is a kit comprised of two reagents, an R1 and R2, which are bottled separately but sold together within the kit.

The R1 solution contains mouse monoclonal anti-morphine antibody, glucose-6phosphate (G6P) nicotinamide adenine dinucleotide (NAD), stabilizers, and sodium azide (0.09%) as a preservative. The R2 solution contains glucose-6-phosphate dehydrogenase (G6PDH) labeled with morphine in buffer with sodium azide (0.09%) as preservative.

The LZI Opiate 2000 Enzyme Immunoassay calibrators and controls designated for use at the 2000 ng/mL cutoff contain 0, 1000, 1500, 2000, 2500, 4000, and 6000 ng/mL of morphine in human urine with sodium azide (0.09%) as preservative. These five calibrators and two controls are sold as individual bottles.

{2}------------------------------------------------

Intended Use

The LZI Opiate 2000 Enzyme Immunoassay is intended for the qualitative and semiquantitative determination of morphine in human urine at a cutoff value of 2000 ng/mL. The assay is designed for professional use with a number of automated clinical chemistry analyzers.

The semi-quantitative mode is for purposes of (1) enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as GCMS and LCMS or (2) permitting laboratories to establish quality control procedures.

The LZI Opiate 2000 Drugs of Abuse (DAU) Calibrators are for use as calibrators in the qualitative and semi-quantitative calibration of the LZI Opiate 2000 Enzyme Immunoassay at a cutoff value of 2000 ng/mL.

The LZI Opiate 2000 Drugs of Abuse (DAU) Controls are for use as assayed quality control materials to monitor the precision of the LZI Opiate 2000 Enzyme Immunoassay at a cutoff value of 2000 ng/mL.

The assay provides only a preliminary analytical result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas or liquid chromatography/mass spectrometry (GC/MS or LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.

{3}------------------------------------------------

Comparison to Predicate Device

The Lin-Zhi International, Inc. LZI Opiate 2000 Enzyme Immunoassay used at the 2000 ng/mL cutoff is substantially equivalent to the Lin-Zhi International, Inc. Opiate Enzyme Immunoassay, Calibrators and Controls for Hitachi 717 Systems cleared by the FDA under the premarket notification K020638 & K020769 for its stated intended use.

The following table compares LZI's Opiate 2000 Enzyme Immunoassay at the 2000 ng/mL cutoff with the predicate device.

DeviceCharacteristicsSubject DeviceLZI Opiate 2000 Enzyme Immunoassay,Calibrators and ControlsPredicate Device(K020638 & K020769)LZI Opiate Enzyme Immunoassay,Calibrators and Controls
Intended UseThe LZI Opiate 2000 EnzymeImmunoassay, when used in conjunctionwith Hitachi 717 automated clinicalsystem analyzers, is intended for thequalitative and semi-quantitativedetermination of morphine in humanurine at a cutoff value of 2000 ng/mL.The assay is designed for professional usewith a number of automated clinicalchemistry analyzers.This assay provides a rapid screening procedurefor determining the presence of morphine in urine.The assay provides only a preliminary analyticalresult. A more specific alternative chemicalmethod must be used in order to obtain aconfirmed analytical result. Gas or liquidchromatography/mass spectrometry (GC/MS orLC/MS) is the preferred confirmatory method.Clinical consideration and professional judgmentshould be exercised with any drug of abuse testresult, particularly when the preliminary test resultis positive.The Opiate Enzyme Immunoassay fromLin-Zhi International, Inc., when used inconjunction with Hitachi 717 automatedclinical system analyzers, is intended forthe qualitative and semi-quantitativedetermination of morphine in humanurine at a cutoff value of 300 ng/mL. Theassay is designed for professional usewith a number of automated clinicalchemistry analyzers.This assay provides a rapid screening procedurefor determining the presence of morphinein urine.The assay provides only a preliminary analyticalresult. A more specific alternative chemicalmethod must be used in order to obtain aconfirmed analytical result. Gas or liquidchromatography/mass spectrometry (GC/MS orLC/MS) is the preferred confirmatory method.Clinical consideration and professional judgmentshould be exercised with any drug of abuse testresult, particularly when the preliminary test resultis positive.
Analytemorphinemorphine
Cutoff2000 ng/ml300 ng/mL
MatrixUrineUrine
CalibratorsLevel5 Levels(0, 1000, 2000, 4000, 6000 ng/mL)5 Levels(0, 150, 300, 600, 1000 ng/mL)
Controls Level2 Levels(1500 ng/mL, 2500 ng/mL)2 Levels(225 ng/mL, 375 ng/mL)
Storage2-8 °C until expiration date2-8 °C until expiration date

{4}------------------------------------------------

Performance Characteristics Summary:

Hitachi 717 Analyzer

Precision:

Precision: Semi-Quantitative, ng/mL

N=88Within RunTotal Precision
(ng/mL)MeanSD% CVMeanSD% CV
0 ng/mL64.921.3736.6%64.925.739.6%
500 ng/mL448.918.114.1%448.922.14.9%
1000 ng/mL1014.424.222.4%1014.424.82.4%
1500 ng/mL1559.820.071.3%1559.822.61.4%
2000 ng/mL2018.432.221.6%2018.437.61.9%
2500 ng/mL2492.346.41.9%2492.347.01.9%
3000 ng/mL3065.587.912.9%3065.590.12.9%
3500 ng/mL3620.064.061.8%3620.071.42.0%
4000 ng/mL4013.283.962.1%4013.296.92.4%

Semi-Quantitative Precision Analysis Summary: Qualitative Results

N=88(ng/mL)Within RunTotal Precision
MeanQualitative ResponseMeanQualitative Response
0 ng/mL64.9-64.9-
500 ng/mL448.9-448.9-
1000 ng/mL1014.4-1014.4-
1500 ng/mL1559.8-1559.8-
2000 ng/mL2018.4+2018.4+
2500 ng/mL2492.3+2492.3+
3000 ng/mL3065.5+3065.5+
3500 ng/mL3620.0+3620.0+
4000 ng/mL4013.2+4013.2+

Semi-Quantitative Positive/Negative Results:

2000 ng/mL Cutoff Result:Within RunTotal Precision
SampleConcentration% of CutoffNumber ofDeterminationImmunoassayResultNumber ofDeterminationImmunoassayResult
0 ng/mL-100.0%2222 Negative8888 Negative
500 ng/mL-75.0%2222 Negative8888 Negative
1000 ng/mL-50.0%2222 Negative8888 Negative
1500 ng/mL-25.0%2222 Negative8888 Negative
2000 ng/mL0%2218 Pos/ 4 Neg8859 Pos/ 29 Neg
2500 ng/mL+25.0%2222 Positive8888 Positive
3000 ng/mL+50.0%2222 Positive8888 Positive
3500 ng/mL+75.0%2222 Positive8888 Positive
4000 ng/mL+100.0%2222 Positive8888 Positive

{5}------------------------------------------------

Performance Characteristics Summary: (continued) Hitachi 717 Analyzer

Precision: Qualitative, mA/min

N=88(mA/min)Within RunTotal Precision
MeanSD% CVMeanSD% CV
0 ng/mL229.21.670.7%229.21.90.8%
500 ng/mL271.21.740.6%271.22.20.8%
1000 ng/mL335.91.480.4%335.92.70.8%
1500 ng/mL384.71.680.4%384.72.80.7%
2000 ng/mL414.52.700.7%414.53.50.9%
2500 ng/mL441.91.90.4%441.93.10.7%
3000 ng/mL461.93.360.7%461.94.20.9%
3500 ng/mL478.62.970.6%478.63.50.7%
4000 ng/mL489.42.240.5%489.42.90.6%

Qualitative Positive/Negative Results:

2000 ng/mL Cutoff Result:Within RunTotal Precision
SampleConcentration% of CutoffNumber ofDeterminationImmunoassayResultNumber ofDeterminationImmunoassayResult
0 ng/mL-100.0%2222 Negative8888 Negative
500 ng/mL-75.0%2222 Negative8888 Negative
1000 ng/mL-50.0%2222 Negative8888 Negative
1500 ng/mL-25.0%2222 Negative8888 Negative
2000 ng/mL0%229 Pos/ 13 Neg8833 Pos/ 55 Neg
2500 ng/mL+25.0%2222 Positive8888 Positive
3000 ng/mL+50.0%2222 Positive8888 Positive
3500 ng/mL+75.0%2222 Positive8888 Positive
4000 ng/mL+100.0%2222 Positive8888 Positive

Limit of Detection:

The lowest concentration that can be differentiated from the negative urine with 95% confidence is determined as 200 ng/mL.

Linearity:

Hitachi 717 Instrument: 0 - 6000 ng/mL

When comparing the result (y) and target (x) value, using the least squares regression technique, the regression equation and correlation are as follows: y = 1.078x - 85.072, r2 = 0.993

Method Comparison - Clinical Samples:

From a total of one-hundred fifty (150) clinical unaltered samples

Semi-Quantitative Data:

96.49% agreement with positive,

93.55% agreement with negative samples

Qualitative Data:

96.49% agreement with positive,

96.77% agreement with negative samples

{6}------------------------------------------------

Performance Characteristics Summary: (continued)

Hitachi 717 Analyzer

Endogenous Compound Interference & Specificity & Cross-Reactivity:

No significant undesired cross-reactants or endogenous substance interference was observed.

Summary:

The information provided in this pre-market notification demonstrates that the LZI Opiate 2000 Enzyme Immunoassay at a cutoff value of 2000 ng/mL is substantially equivalent to the legally marketed predicate device for its general intended use. Substantial equivalence was demonstrated through comparison of intended use and physical properties to the commercially available predicate device as confirmed by chromatography/mass spectrometry (GC/MS or LC/MS), an independent analytical method. The information supplied in this pre-market notification provides reasonable assurance that the LZI Opiate 2000 Enzyme Immunoassay at a cutoff value of 2000 ng/mL is safe and effective for its stated intended use.

{7}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration

10903 New Hampshire Avenue Silver Spring, MD 20993

Lin-Zhi International, Inc. c/o Bernice Lin, Ph.D. Vice President, Operations 670 Almanor Ave Sunnyvale, CA 94085

MAY - 8 2012

K120761 Re:

Ite: - - R120701 LZI Opiate 2000 Enzyme Controls

Regulation Number: 21CFR 862.3650 Regulation Name: Opiate Test System Regulatory Class: Class II Product Code: DJG, DLJ and LAS Dated: April 26, 2012 Received: April 27, 2012

Dear Dr. Lin:

We have reviewed your Section 510(k) premarket notification of intent to market the device w & nove and have and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that provision in the Federal with the provisions of the Federal Food, Drug, and Cosmetic nave och toe do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general therefore, market the de nee, acreed requirements for annual registration, listing of devices, good controls provisions of the engans and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device to such additional controls. Existing major regulations affecting your device can be finay oc subject to Back as Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or that FDA has made a dolorinination administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (2) CFR Part with an the Ace 3 requirements, moral 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to begin manisang your dever of your device to a legally prematication: "The PDF Imaning sistems for your device and thus, permits your device to proceed to the market.

{8}------------------------------------------------

Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical

Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance ...

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm

Sincerely yours.

Counney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

Premarket Notification

Indications for Use Statement

510(k) Number (if known):

Device Name: LZI Opiate 2000 Enzyme Immunoassay LZI Opiate 2000 Calibrators and Controls

Indications for Use:

The LZI Opiate 2000 Enzyme Immunoassay is intended for the qualitative and semiquantitative determination of morphine in human urine at the cutoff value of 2000 ng/mL. The assay is designed for professional use with a number of automated clinical chemistry analyzers.

The semi-quantitative mode is for purposes of (1) enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as GCMS and LCMS or (2) permitting laboratories to establish quality control procedures.

The LZI Opiate 2000 Drugs of Abuse (DAU) Calibrators are for use as calibrators in the qualitative and semi-quantitative calibration of the LZI Opiate 2000 Enzyme Immunoassay at a cutoff value of 2000 ng/mL.

The LZI Opiate 2000 Drugs of Abuse (DAU) Controls are for use as assayed quality control materials to monitor the precision of the LZI Opiate 2000 Enzyme Immunoassay at a cutoff value of 2000 ng/mL.

The assay provides only a preliminary analytical result: A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas or liquid chromatography/mass spectrometry (GC/MS or LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.

Prescription Use AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) (Per 21 CFR 801.109)

Division Sign-Off

Office of In Vitro Diagnostic Olice Evaluation and Safety

510(k) K120761

Page 9 of 77

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).