K Number
K120724
Date Cleared
2012-05-07

(59 days)

Product Code
Regulation Number
888.3060
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Pioneer Lateral Plate System is indicated for use via the lateral or anterolateral surgical approach above the bifurcation of the great vessels in the treatment of thoracic and thoracolumbar (T1-L5) spine instability or via the anterior surgical approach, below the bifurcation of the great vessels in the treatment of lumbar and lumbosacral (L1-S1) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), scoliosis, kyphosis, lordosis, spinal stenosis, or a failed previous spine surgery.

Device Description

The Pioneer Lateral Plate System consists of an assortment of plates and screws manufactured from ASTM F136 Titanium Alloy.

AI/ML Overview

The provided document describes a Pioneer Lateral Plate System, a medical device used for spinal fixation. This is a mechanical device, not an AI/ML powered diagnostic or prognostic tool. Therefore, many of the requested categories related to AI/ML studies (such as sample size for test/training sets, ground truth establishment, expert qualifications, adjudication methods, MRMC studies, standalone performance, etc.) are not applicable and will not be found in the document.

The performance data for this type of device focuses on mechanical testing and demonstrating substantial equivalence to predicate devices based on physical characteristics and intended use.

Here's the information extracted from the provided text:

Acceptance Criteria and Device Performance

Acceptance Criteria (Test Performed)Reported Device Performance
ASTM F1717 Dynamic axial compression testingFunctioned as intended; substantially equivalent to predicate systems
ASTM F1717 Static axial compression testingFunctioned as intended; substantially equivalent to predicate systems
ASTM F1717 Torsional static testingFunctioned as intended; substantially equivalent to predicate systems
ASTM F543 screw pull-out testingFunctioned as intended; substantially equivalent to predicate systems
ASTM F543 screw pull-through testingFunctioned as intended; substantially equivalent to predicate systems

Note: The document states that the test results demonstrate the device "functioned as intended and performed in a manner substantially equivalent to that of predicate systems." It does not provide specific numerical thresholds for acceptance or numerical results for the performance, as is common for substantial equivalence claims based on established ASTM standards for mechanical devices.

Information Not Applicable or Not Found in the Document:

  1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable for mechanical testing of a spinal implant. The "test set" here refers to the physical devices subjected to mechanical stress tests, not a dataset of patient information.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth, in the context of this device, is defined by the physical properties and performance against ASTM standards, not by expert interpretation of data.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable for a mechanical spinal implant.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): For mechanical testing, the "ground truth" is defined by the specifications of the ASTM standards and the physical properties of the materials and design.
  7. The sample size for the training set: Not applicable. There is no training set for a mechanical device.
  8. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

170724

MAY - 7 2012

510(k) Summary Sponsor:

Pioneer Surgical Technology 375 River Park Circle Marquette, MI 49855 (906) 225-5861 Contact: Sarah McIntyre, Emily M. Downs Date: May 3, 2012

Device Name:

. Classification Name: Spinal Intervertebral Body Fixation Orthosis

Classification:

Regulation Number: §888.3060 Spinal intervertebral body fixation orthosis

Product Code: K WQ: Panel Code: 87

· NuVasive Lateral Plate System (K070273) Quantum Spinal System (K080518) Pioneer Lumbar Plate System (K090222)

Pioneer Lateral Plate System

Predicate Devices:

Description:

Intended Use:

Performance Data:

Performance and SE Determination:

The Pioneer Lateral Plate System consists of an assortment of plates and screws manufactured from ASTM F136 Titanium Alloy.

The Pioneer Lateral Plate System is indicated for use via the lateral or anterolateral surgical approach above the bifurcation of the great vessels in the treatment of thoracic and thoracolumbar (T1-L5) spine instability or via the anterior surgical approach, below the bifurcation of the great vessels in the treatment of lumbar and lumbosacral (L1-S1) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), scoliosis, kyphosis, spinal stenosis, or a failed previous spine surgery.

ASTM F1717 Dynamic axial compression, static axial compression and torsional static testing were performed, in addition to ASTM F543 screw pull-out and screw pull-through testing, to establish substantial equivalence. The test results demonstrate that the Pioneer Lateral Plate System functioned as intended and performed in a manner substantially equivalent to that of predicate systems.

Equivalence for the Pioneer Lateral Plate System is based on similarities of intended use, design, and physical characteristics when compared to predicate devices. Therefore. Pioneer Surgical Technology believes that there is sufficient evidence to conclude that the Pioneer Lateral Plate System is substantially equivalent to existing legally marketed devices.

Page 1 of 1

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with its wings spread, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The overall design is simple and professional, reflecting the department's role in public health and welfare.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

MAY - 7 2012

Pioneer Surgical Technology % Ms. Sarah MoIntyre Regulatory Affairs Associate 375 River Park Circle Marquette, Michigan 49855

Re: K120724

Trade/Device Name: Pioneer Lateral Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: March 7, 2012 Received: March 9, 2012

Dear Ms. McIntyre:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable; the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Ms. Sarah McIntyre

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Erin Keith

~ Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

2.0 Indications for Use Statement

K12 0724 510(k) Number (if known):

Device Name:

Pioneer Lateral Plate System

Indications:

The Pioneer Lateral Plate System is indicated for use via the lateral or anterolatera! surgical approach above the bifurcation of the great vessels in the treatment of thoracic and thoracolumbar (T1-L5) spine instability or via the anterior surgical approach, below the bifurcation of the great vessels in the treatment of lumbar and lumbosacral (L1-S1) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), scoliosis, kyphosis, lordosis, spinal stenosis, or a failed previous spine surgery.

Prescription Use (Per 21 CFR 801.109)

Over-the-Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

OR

(Division Sign-Off)

Division of Surgical, Orthopedic, and Restorative Devices

KI20724 510(k) Number_

Pioneer Lateral Plate System Traditional 510(k)

N/A