(68 days)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
Powder Free Latex Patient Examination Glove, with Protein Content Labeling Claim, Chlorinated (Contains 50 Micrograms per dm2 of glove or Less of Water Extractable Protein) meets all the current specification for ASTM D3578-05 (2010).
The provided text is for a 510(k) Premarket Notification for a medical device, specifically a Powder Free Latex Patient Examination Glove. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, primarily through non-clinical performance data and comparison of technological characteristics, rather than extensive clinical studies as one might find for a novel diagnostic AI device.
Therefore, many of the typical acceptance criteria and study details requested (e.g., sample size for test set, number of experts for ground truth, adjudication method, MRMC studies, standalone algorithm performance, training set details) are not applicable to this type of device and submission. The "study" here refers to the testing performed to meet specific ASTM standards and biocompatibility requirements.
Here's a breakdown of the information that is applicable and how it relates to the acceptance criteria:
1. Acceptance Criteria and Reported Device Performance
The acceptance criteria for this device are defined by existing ASTM standards and biocompatibility testing. The device's performance is reported as meeting these standards.
Characteristic | Acceptance Criteria (Standards) | Reported Device Performance |
---|---|---|
Dimensions | ASTM D 3578-05 (2010) | Meets |
Physical Properties | ASTM D 3578-05 (2010) | Meets |
Freedom from pin-holes | ASTM D 5151-99 (2006), ASTM D 3578-05 (2010) | Meets |
Powder Free Residue | ASTM D 6124-06, ASTM D 3578-05 (2010) | Meets |
Protein Content | ASTM D 5712-10, ASTM D 3578-05 (2010) | Meets |
Biocompatibility (Dermal Sensitization) | ISO 10993-10:2010 | Not a contact skin sensitizer |
Biocompatibility (Primary Skin Irritation) | ISO 10993-10:2010 | Not a primary skin irritant |
Intended Use | Disposable device for medical purposes, worn on examiner's hand to prevent contamination between patient and examiner. | Identical to predicate device |
Regulation Number | 21 CFR Part 880.6250 | Identical to predicate device |
Product Code | 80 LYY | Identical to predicate device |
Design | Ambidextrous, in different sizes per ASTM D3578 dimension requirement. | Identical to predicate device |
Materials | Natural Rubber Latex | Identical to predicate device |
Color | Natural Color | Identical to predicate device |
Sterility | Not Applicable (Non-Sterile) | Identical to predicate device |
Single Use | Yes | Identical to predicate device |
Packaging | Packed in Dispenser Boxes | Identical to predicate device |
Labeling Claim | With Extractable Protein Content Labeling Claim (50 Micrograms per dm² of glove or less of Water Extractable Protein) | Identical to predicate device |
2. Sample size used for the test set and the data provenance
The document specifies that the device "meets all the current specification for ASTM D3578-05 (2010)" and refers to other ASTM and ISO standards for specific characteristics. Clinical data was "not needed for market cleared examination gloves."
- Sample Size for Test Set: The specific sample sizes for each test (e.g., pin-holes, dimensions, physical properties) are not detailed in this 510(k) summary. These would typically be specified within the individual ASTM and ISO standards referenced. For example, ASTM D5151-99 for freedom from pinholes often specifies a sample size based on the acceptable quality level (AQL).
- Data Provenance: The document does not explicitly state the country of origin for the data (e.g., where the tests were physically conducted) nor whether it was retrospective or prospective. Given the nature of performance testing against standards, it would be prospective testing of newly manufactured glove batches. The manufacturer is "Top Calibre Sdn Bhd" in Selangor, Malaysia, so the testing was likely conducted by or for them.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not Applicable: This type of device does not involve a "ground truth" established by experts in a diagnostic or interpretive sense. The ground truth for performance characteristics is defined by the objective measurement criteria within the relevant ASTM and ISO standards (e.g., tensile strength, elongation, water leak rate, protein content measurement). Biocompatibility tests are performed by qualified laboratories and assessed against established toxicology criteria.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not Applicable: Adjudication methods like 2+1 or 3+1 are used for human-reviewed data, often in diagnostic imaging or clinical trials, to resolve discrepancies between readers. For physical and chemical tests of gloves, results are objectively measured against defined thresholds in the standards; there is no subjective interpretation requiring adjudication.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not Applicable: This device is a patient examination glove, not an AI-powered diagnostic tool. Therefore, MRMC studies and assessments of AI assistance are irrelevant.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not Applicable: This device is a physical medical device (glove), not an algorithm or software. Standalone algorithm performance is not relevant.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for this device's performance is established by objective measurements against predefined criteria and thresholds set forth in the referenced ASTM and ISO standards.
- For Dimensions, Physical Properties, Freedom from pin-holes, Powder Free Residue, Protein Content: The ground truth is the quantitative measurement of these properties (e.g., thickness, tensile strength, percentage of gloves passing water leak test, specific protein content in µg/dm²) compared to the specified limits in standards like ASTM D3578-05 (2010), ASTM D5151-99 (2006), ASTM D6124-06, and ASTM D5712-10.
- For Biocompatibility: The ground truth is determined by the results of standardized biological tests (Dermal Sensitization and Primary Skin Irritation Tests as per ISO 10993-10:2010), which assess the device's potential to cause adverse biological reactions. The "ground truth" in this context is the observation of whether a reaction occurs and its severity, interpreted against established toxicity criteria, leading to a "pass" or "fail" determination (e.g., "Not a contact skin sensitizer," "Not a primary skin irritant").
8. The sample size for the training set
Not Applicable: This device does not use machine learning or AI, so there is no "training set."
9. How the ground truth for the training set was established
Not Applicable: As there is no training set, this question is not relevant.
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.