K Number
K120557

Validate with FDA (Live)

Date Cleared
2012-03-19

(24 days)

Product Code
Regulation Number
888.3080
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Solitaire® PEEK Anterior Spinal System is designed for use with autograft and is indicated for stand-alone intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade I spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.

Device Description

This submission is for a line extension to the Solitaire PEEK Anterior Spinal System to add an additional style of spacer with a modified screw angle. This modified device will be marketed as Solitaire-35.

AI/ML Overview

The provided text describes a 510(k) submission for a medical device, the Solitaire PEEK Anterior Spinal System line extension (Solitaire 35°), which is a lumbar stand-alone intervertebral body fusion device. The focus of the submission is to demonstrate substantial equivalence to predicate devices, primarily through mechanical performance testing.

Here's an analysis of the acceptance criteria and study data based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Set by Special Controls Guidance Document: Intervertebral Body Fusion Device)Reported Device PerformanceComments
Static Axial Compression (ASTM F-2077)Mechanical testing shows the strength of the subject device is sufficient for its intended use.Specific values (e.g., failure loads or displacement) are not provided in the summary, only a general statement of sufficiency.
Dynamic Axial Compression (ASTM F-2077)Mechanical testing shows the strength of the subject device is sufficient for its intended use.Specific values are not provided.
Static Compression-Shear (ASTM F-2077)Mechanical testing shows the strength of the subject device is sufficient for its intended use.Specific values are not provided.
Dynamic Compression-Shear (ASTM F-2077)Mechanical testing shows the strength of the subject device is sufficient for its intended use.Specific values are not provided.
Implant Subsidence Analysis (Engineering Analysis)An engineering analysis regarding implant subsidence was presented.No details on the acceptance criteria or the specific findings of the analysis are provided in this summary.

2. Sample Size Used for the Test Set and Data Provenance

The provided text describes mechanical testing of the device prototypes, not a clinical trial involving human subjects or patient data. Therefore, the concepts of "test set" in the context of patient data, data provenance, or retrospective/prospective studies are not applicable here. The sample size for the mechanical tests (e.g., number of devices tested for each condition) is not specified in the summary.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable as the study described is a mechanical performance evaluation of a physical device, not an assessment requiring expert interpretation of clinical data or images to establish ground truth.

4. Adjudication Method for the Test Set

This is not applicable for mechanical testing. Adjudication methods are typically employed in clinical studies or studies involving human readers/interpreters where disagreements in assessment need to be resolved.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

No, an MRMC comparative effectiveness study was not conducted or described in the provided text. The study focuses purely on the mechanical performance of the device itself, not on human reader performance with or without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This question is not applicable as the device is a physical intervertebral body fusion device and not an algorithm or AI system.

7. The Type of Ground Truth Used

For mechanical testing, the "ground truth" would be the physical properties and performance characteristics of the device under specific load conditions, measured by calibrated testing equipment. The adherence to ASTM standards for these tests establishes the methodology for determining this performance.

8. The Sample Size for the Training Set

This question is not applicable as the device is a physical medical implant, not an AI or machine learning model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This question is not applicable for the same reason as above.


Summary of Study and Findings:

The study described is a set of mechanical performance tests conducted on the Solitaire PEEK Anterior Spinal System line extension (Solitaire 35°). The purpose of these tests was to demonstrate that the device meets the special controls guidance document "Class II Special Controls Guidance Document: Intervertebral Body Fusion Device." The tests included:

  • Static Axial Compression (ASTM F-2077)
  • Dynamic Axial Compression (ASTM F-2077)
  • Static Compression-Shear (ASTM F-2077)
  • Dynamic Compression-Shear (ASTM F-2077)
  • An engineering analysis regarding implant subsidence.

The conclusion drawn from these tests is that "Mechanical testing shows the strength of the subject device is sufficient for its intended use." This conclusion, along with the technological characteristics and indications for use, supported the determination of substantial equivalence to predicate devices. No specific quantitative performance metrics or sample sizes for the mechanical tests are provided in this summary.

{0}------------------------------------------------

MAR 1 9 2012

Image /page/0/Picture/2 description: The image shows the logo for Biomet Spine. The word "BIOMET" is in a bold, sans-serif font, with each letter connected to the next. Below the word "BIOMET" is the word "SPINE" in a smaller, sans-serif font.

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR § 807.92.

Preparation Date:February 23, 2012
Applicant/Sponsor:Biomet Spine and Bone Healing Technologies100 Interpace ParkwayParsippany, NJ 07054
Contact Person:Margaret F. CrowePhone: 973-299-9300, ext. 2260Fax: 973-257-0232
Trade name:Solitaire PEEK Anterior Spinal System Line.Extension (Solitaire 35°)
Common Name:Lumbar stand-alone intervertebral body fusiondevice
Classification Name(Product Code):Intervertebral Body Fusion Device with IntegratedFixation/Lumbar (OVD)
Device Panel - Regulation No.:Orthopedic - 21 CFR 888.3080

Device Description:

This submission is for a line extension to the Solitaire PEEK Anterior Spinal System to add an additional style of spacer with a modified screw angle. This modified device will be marketed as Solitaire-35.

Indications for Use:

The Solitaire PEEK Anterior Spinal System is designed for use with autograft and is indicated for stand-alone intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.

There is no change to the Indications for Use as a result of this design modification.

Summary of Technologies:

The technological characteristics of the new components are the same as, or similar to, the predicate devices. Examples of predicate devices include:

  • Solitaire PEEK-Optima Anterior Spinal System (Biomet Spine K081395) .
  • Independence Spacer (Globus Medical K082252) .
  • Expandable PEEK Spacer (Biomet Spine K082406) .

Page 5-2

{1}------------------------------------------------

Performance Data:

Mechanical testing was conducted in accordance with the special controls guidance document entitled "Class II Special Controls Guidance Document: Intervertebral Body Fusion Device". The testing conducted, along with the ASTM standard, are listed below:

  • . Static Axial Compression (ASTM F-2077)
  • Dynamic Axial Compression (ASTM F-2077) .
  • Static Compression-Shear (ASTM F-2077) .
  • Dynamic Compression-Shear (ASTM F-2077) .

An engineering analysis regarding implant subsidence was presented.

Mechanical testing shows the strength of the subject device is sufficient for its intended use.

Substantial Equivalence:

The modified spacers of the Solitaire PEEK Anterior Spinal System are substantially equivalent to the predicate spacers with respect to intended use and indications, technological characteristics, and principles of operation and do not present any new issues of safety or effectiveness. The predicate devices listed above are distributed for similar indications, and have similar design features.

Conclusion:

The subject device is substantially equivalent to its predicates when used as an intervertebral body fixation device. The indications for use and fundamental technology of the device remain unchanged. Furthermore, mechanical testing and other supporting information sufficiently demonstrate the substantial equivalence of the subject device to the other components in the Solitaire PEEK Anterior Spinal System. Based on this information, the subject device does not raise any new issues regarding the safety or efficacy.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a stylized eagle with three tail feathers, representing the three branches of government. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Biomet Spine and Bone Healing Technologies % Ms. Margaret F. Crowe 100 Interpace Parkway Parsippany, New Jersey 07054

MAR 1 9 2012

Re: K120557

Trade/Device Name: Solitaire® PEEK Anterior Spinal System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVD Dated: February 23, 2012 Received: February 24, 2012

Dear Ms. Crowe:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{3}------------------------------------------------

Page 2 - Ms. Margaret F. Crowe

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Attachment 4-1

Indications for Use Statement

510(k) Number (if known): K120557

Device Name: Solitaire® PEEK Anterior Spinal System

Indications for Use:

The Solitaire® PEEK Anterior Spinal System is designed for use with autograft and is indicated for stand-alone intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade I spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Signature

(Division Sign-Off) Division of Surgical, Orthopedic, Jand Restorative Devices

KI20557 510(k) Number.

N/A