K Number
K120435
Manufacturer
Date Cleared
2012-06-04

(112 days)

Product Code
Regulation Number
880.2800
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Browne Aldahol Glutaraldehyde Indicator is a glutaraldehyde concentration monitor dedicated for use with Aldahol V High-Level Disinfectant. The purpose of the Browne Aldahol Glutaraldehyde Indicator is to monitor glutaraldehyde concentration greater than 1.8%.

Device Description

The Browne Aldahol Glutaraldehyde Indicator is a chemical indicator strip consisting of an absorbent paper pad impregnated with the reactive chemicals, which is adhesively bonded to one end of a polymer film. The Browne Aldahol Glutaraldehyde Indicator has been developed to monitor the active glutaraldehyde concentration of Aldahol V High-Level Disinfectant (K113015) solution that has an MRC of 1.8% glutaraldehyde.

AI/ML Overview

Here's an analysis of the provided text, focusing on the acceptance criteria and the study discussing device performance:

Acceptance Criteria and Device Performance Study for Browne Aldahol Glutaraldehyde Indicator

1. Table of Acceptance Criteria and Reported Device Performance

The provided document details various verification activities, their acceptance criteria, and the study results. The device appears to be a chemical indicator strip for monitoring glutaraldehyde concentration.

TestAcceptance Criteria (FAIL @ 1.8% Glutaraldehyde)Acceptance Criteria (PASS @ 2.3% Glutaraldehyde)Study Result
Performance Testing100%≥ 80%Pass
Blind Study Testing100%≥ 80%Pass
Simulated Use (Contaminants) Testing100%≥ 80%Pass
Test Strip Life Outside the Bottle100%≥ 80%Pass
Aggressive Chemical Stability Testing100%≥ 80%Pass
Specificity Testing for glutaraldehyde100%100% FAIL for Tap WaterPass
Stability Testing in three storage environments100%≥ 80%Pass
In-Use Stability Testing in three storage environments100%≥ 80%Pass

Interpretation of Acceptance Criteria:

  • FAIL @ 1.8%: This likely means that when the glutaraldehyde concentration is at or below the minimum recommended concentration (MRC) of 1.8%, the indicator strip must show a "FAIL" result (e.g., no color change, or a specific color indicating failure) 100% of the time. This is critical for patient safety, ensuring used solutions are not reused.
  • PASS @ 2.3%: This indicates that when the glutaraldehyde concentration is above the MRC (e.g., at 2.3% or higher, suggesting a fresh or sufficiently concentrated solution), the indicator strip must show a "PASS" result (e.g., a specific color change indicating usability) at least 80% of the time. This allows for the effective reuse of the solution.
  • Specificity Testing for glutaraldehyde (100% FAIL for Tap Water): This is a critical criterion to ensure the indicator strip is truly reactive to glutaraldehyde and not other common substances. A 'FAIL' for tap water means the strip won't erroneously indicate a usable glutaraldehyde concentration when only water is present.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: The document mentions "Test of 6 Lots" in the table heading. It doesn't explicitly state the number of individual indicator strips or repeated tests performed within each category (e.g., "Performance Testing"). However, the reference to "6 Lots" implies that the tests were conducted across multiple manufacturing batches to account for inter-lot variability. Without further detail, the exact number of test strips used is not specified.
  • Data Provenance: The document is a 510(k) summary submitted by STERIS Corporation, with the contact information indicating the company is based in the United Kingdom and the submission is to the US FDA. The studies described are internal verification activities conducted by the manufacturer to support the safety and effectiveness of the device. The data is prospective in the sense that these tests were designed and executed to evaluate the new device prior to its market release. The country of origin of the data is not explicitly stated beyond the company's dual locations, but the study results are presented as part of a regulatory submission to the FDA.

3. Number of Experts Used and Their Qualifications for Test Set Ground Truth

The document does not provide information on:

  • The number of experts used to establish the ground truth for the test set.
  • The qualifications of any such experts (e.g., specific clinical background or experience).
    Given the nature of the device (a chemical indicator), the "ground truth" for glutaraldehyde concentration would likely be established through quantitative chemical analysis (e.g., titration or spectrophotometry) rather than expert human interpretation of the indicator itself.

4. Adjudication Method for the Test Set

The document does not specify any adjudication method (e.g., 2+1, 3+1, none) for the test set. Given that the indicator produces a chemical reaction result (color change) against a known chemical concentration, there is typically less need for human adjudicators comparing their interpretations, unlike in medical imaging where expert consensus is often required. The "Blind Study Testing" mentioned suggests that the evaluators were blinded to the expected result to prevent bias.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly mentioned or described. This type of study is more common for diagnostic imaging devices where human readers interpret results, and the goal is to assess improvement with AI assistance. For a simple chemical indicator, this study design is not applicable.

6. Standalone Performance Study (Algorithm Only)

Yes, the studies described are essentially standalone performance studies of the device (indicator strip) itself. The device is not an "algorithm" in the typical sense of AI, but rather a chemical reaction-based indicator. The tests (Performance Testing, Blind Study, Simulated Use, Stability, Specificity) all assess the intrinsic performance of the indicator strip without human intervention in its core interpretive function (i.e., it either changes color as expected or it doesn't). The "reading" of the strip (e.g., visually comparing to a color chart) is implied but the performance reported is for the device's ability to react correctly.

7. Type of Ground Truth Used

The ground truth used for evaluating the Browne Aldahol Glutaraldehyde Indicator is based on known chemical concentrations of glutaraldehyde.

  • For the "FAIL @ 1.8%" criteria, the ground truth is a glutaraldehyde solution at or below 1.8% concentration, determined by precise chemical measurement.
  • For the "PASS @ 2.3%" criteria, the ground truth is a glutaraldehyde solution at or above 2.3% concentration, also determined by precise chemical measurement.
  • For Specificity, the ground truth is tap water (absence of glutaraldehyde).

This is a form of analytical ground truth where the actual chemical composition and concentration of the test solutions are precisely known.

8. Sample Size for the Training Set

The document does not mention a "training set." This terminology typically applies to machine learning or AI models. The Browne Aldahol Glutaraldehyde Indicator is a chemical indicator, not an AI-powered device. Therefore, it does not involve a training phase with a dedicated dataset. The development process would have involved formulation and optimization, but not "training" in the AI sense.

9. How the Ground Truth for the Training Set Was Established

As there is no "training set" for this chemical indicator device, this question is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the STERIS logo, along with the code K120436. The STERIS logo is a stylized wave pattern. The image also contains the date JUN - 4 2012.

510(k) Summary For Browne Aldahol Glutaraldehyde Indicator

Albert Browne Ltd., a subsidiary of STERIS Corporation Chancery House 190 Waterside Road Hamilton Industrial Park Leicester LES IQZ United Kingdom Phone: +44 116 276 8636 Fax No: +44 116 276 8639

Contact:

Bill Brodbeck Senior Manager Regulatory Affairs Telephone: (440) 392-7690 Fax No: (440) 357-9198

Summary Date:

February 10, 2012

STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600

.

{1}------------------------------------------------

1. Device Name

Trade Name:Browne Aldahol Glutaraldehyde Indicator
Common/Usual Name:Browne Aldahol Test Strip
Classification Name:Physical/chemical sterilization process indicator (21CFR 880.2800 (b), Product Code JOJ).

2. Predicate Device

  • K924681 3M™ ComplyTM Cold Sterilog™ Glutaraldehyde Monitor 3987MM .

3. Description of Device

The Browne Aldahol Glutaraldehyde Indicator is a chemical indicator strip consisting of an absorbent paper pad impregnated with the reactive chemicals, which is adhesively bonded to one end of a polymer film. The Browne Aldahol Glutaraldehyde Indicator has been developed to monitor the active glutaraldehyde concentration of Aldahol V High-Level Disinfectant (K113015) solution that has an MRC of 1.8% glutaraldehyde.

4. Intended Use

The Browne Aldahol Glutaraldehyde Indicator is a glutaraldehyde concentration monitor dedicated for use with Aldahol V High-Level Disinfectant solution cleared under K113015. The purpose of the Browne Aldahol Glutaraldehyde Indicator is to determine whether the glutaraldehyde concentration of an Aldahol V High-Level Disinfectant solution is above the minimum recommended concentration, allowing the solution to be re-used for reprocessing temperature-sensitive (and other) instruments if the glutaraldehyde concentration is found to be greater than 1.8%.

5. Description of Safety and Substantial Equivalence

The proposed and predicate devices are all single use indicators used to monitor glutaraldehyde concentration in specific solutions. The differences between the proposed Browne Aldahol Glutaraldehyde Indicator and predicate device are limited to differences in the device design and materials. These differences do not raise any new issues of safety and efficacy.

{2}------------------------------------------------

A summary of the technological characteristics of the new device in comparison to those of the predicate device is provided in Section 12 of this premarket notification.

The following table summarizes the verification activities that were performed, with their respective acceptance criteria and results, to demonstrate that the Browne Aldahol Glutaraldehyde Indicator is safe and effective. These studies confirm that .the device's performance meets the requirements of its pre-defined acceptance criteria and intended uses.

Test of 6 LotsAcceptance CriteriaStudyResult
FAIL @ 1.8%PASS @ 2.3%
Performance Testing100%≥ 80%Pass
Blind Study Testing100%≥ 80%Pass
Simulated Use(Contaminants) Testing100%≥ 80%Pass
Test Strip LifeOutside the Bottle100%≥ 80%Pass
Aggressive ChemicalStability Testing100%≥ 80%Pass
Specificity Testingfor glutaraldehyde100%100% FAIL for Tap WaterPass
Stability Testing in threestorage environments100%≥ 80%Pass
In-Use Stability Testing inthree storage environments100%≥ 80%Pass

The Browne Aldahol Glutaraldehyde Indicator is substantially equivalent to its predicate.

r

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized depiction of a caduceus, a symbol often associated with medicine and healthcare. The caduceus features a staff with a snake winding around it, and a pair of wings at the top.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Dr. Bill Brodbeck Senior Manager, Regulatory Affairs STERIS Corporation 5960 Heisley Road Mentor, Ohio 44060

JUN - 4 2012

Re: K120435

Trade/Device Name: Browne Aldahol Glutaraldehyde Indicator Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: JOJ Dated: May 10, 2012 Received: May 11, 2012

Dear Dr. Brodbeck:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Dr. Brodbeck

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting . (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K120435

Device Name:

Browne Aldahol Glutaraldehyde Indicator

Indications for Use:

The Browne Aldahol Glutaraldehyde Indicator is a glutaraldehyde concentration monitor dedicated for use with Aldahol V High-Level Disinfectant. The purpose of the Browne Aldahol Glutaraldehyde Indicator is to monitor glutaraldehyde concentration greater than 1.8%.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Elizabeth F. Clannie-Willin.

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

K120435 510(k) Number:

Page 1 of 1

§ 880.2800 Sterilization process indicator.

(a)
Biological sterilization process indicator —(1)Identification. A biological sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization.(2)
Classification. Class II (performance standards).(b)
Physical/chemical sterilization process indicator —(1)Identification. A physical/chemical sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.(2)
Classification. Class II (performance standards).